Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112142

    Validate with FDA (Live)

    Date Cleared
    2011-11-04

    (101 days)

    Product Code
    Regulation Number
    862.3560
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LITH method is an in vitro diagnostic test for the quantitative measurement of lithium in human serum and plasma on the Dimension Vista® System. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).

    The Drug 4 CAL is an in vitro diagnostic product for the calibration of the LOCI Digoxin (DIGXN) and Lithium (LITH) methods on the Dimension Vista® System.

    Device Description

    The LITH method employs a lithium-specific chromoionophore that forms a complex with the Li+ ion in an alkaline solution. The concentration of lithium in the sample is proportional to the increase in absorbance, which is due to the formation of the dye-lithium complex. The reaction is measured using a Bichromatic (510 and 700 mm) endpoint technique.

    The Drug 4 Calibrator is a five (5) level, liquid calibrator. It is packaged as a kit of ten vials per level (A, B, C, D and E), 2.5 mL per vial. The product matrix is liquid human serum and contains digoxin and lithium. This product is sold separately from the Flex® reagent cartridge.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dimension Vista® Lithium (LITH) Flex® Reagent Cartridge and Dimension Vista® Drug 4 Calibrator, based on the provided text:

    Overview:

    The submission focuses on demonstrating "substantial equivalence" to predicate devices. This means the new devices (Dimension Vista® Lithium Flex® Reagent Cartridge and Dimension Vista® Drug 4 Calibrator) perform similarly enough to existing, legally marketed devices (Dimension® Lithium Flex® Reagent Cartridge K011033 and Dimension® Drug Calibrator K011035) that they don't raise new questions of safety or effectiveness. The primary study type is comparative testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence submission, the "acceptance criteria" are implicitly met if the new device performs comparably to the predicate device, as demonstrated through comparative testing. The document states "Comparative testing described in the submission report demonstrates substantial equivalent performance."

    Here’s a table summarizing the features compared, which serve as the basis for demonstrating substantial equivalence:

    Dimension Vista® Lithium (LITH) Flex® Reagent Cartridge vs. Predicate (K011033)

    FeatureAcceptance Criteria (Predicate: Dimension® Lithium (LI) K011033)Reported Device Performance (Dimension Vista® Lithium (LITH) K4150)
    Intended UseQuantitative measurement of lithium in human serum and plasma on Dimension® systems for drug dosage assurance in mental disturbances.Quantitative measurement of lithium in human serum and plasma on the Dimension Vista® System for drug dosage assurance in mental disturbances. (Identical)
    Sample TypeSerum and Sodium Heparin PlasmaSerum and Sodium Heparin Plasma
    Measuring Range0.20 - 5.00 mmol/L0.20 - 3.00 mmol/L
    Sample Size10 µL2 µL
    MeasurementBichromatic endpoint (540 and 700 nm)Bichromatic endpoint (510 and 700 nm)

    Dimension Vista® Drug 4 Calibrator vs. Predicate (K011035)

    FeatureAcceptance Criteria (Predicate: Dimension® Drug Calibrator K011035)Reported Device Performance (Dimension Vista® DRUG 4 CAL KC460A)
    Intended UseCalibration of DIG, LI, PHNO, PTN, and THEO methods on Dimension Vista® System.Calibration of LOCI Digoxin (DIGXN) and Lithium (LITH) methods on the Dimension Vista® System.
    MatrixHuman serum basedHuman serum based
    PreparationLiquid: Provided ready to use.Liquid: Provided ready to use.
    Target ConcentrationsLevel 1: ≤ 0.20 mmol/LLevel 2: 0.80 - 1.00 mmol/LLevel 3: 1.71 - 1.89 mmol/LLevel 4: 3.42 - 3.78 mmol/LLevel 5: 5.22 - 5.78 mmol/LLevel 1: ≤ 0.20 mmol/LLevel 2: 0.77 - 1.15 mmol/LLevel 3: 1.67 - 2.00 mmol/LLevel 4: 3.35 - 3.85 mmol/LLevel 5: 5.12 - 5.89 mmol/L
    StorageStore at 2 to 8 °C.Store at 2 - 8 °C.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Comparative testing described in the submission report demonstrates substantial equivalent performance." However, the exact sample sizes for the test set (e.g., number of patient samples, controls) are not specified within the provided text.

    The data provenance is also not explicitly stated (e.g., country of origin, retrospective/prospective). As an in vitro diagnostic device, it's highly likely that samples would be human serum and plasma, but the source specifics are missing.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable and not provided in the context of this device. This is a quantitative chemical assay, not an imaging device or diagnostic interpretation requiring expert human adjudication. The "ground truth" for calibrators and reagent cartridges like these is typically established through analytical methods, reference materials, and specified concentrations, not expert consensus.


    4. Adjudication Method for the Test Set

    Not applicable and not provided. As explained in point 3, this is a quantitative chemical assay, not a diagnostic interpretation that would involve adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically for evaluating a diagnostic aid that assists human readers (e.g., AI in radiology). The devices in question are reagents and calibrators for a laboratory instrument, not tools for human interpretation.


    6. If a Standalone Performance Study was Done

    Yes, in essence, standalone performance was assessed as part of the "comparative testing" mentioned. For an in vitro diagnostic device, demonstrating performance against a predicate device (rather than a human reader) is the standard. The device's performance properties (e.g., precision, measuring range, analytical specificity) would be evaluated when operating on its own as designed, and then these results would be compared to the predicate. The document states "Comparative testing described in the submission report demonstrates substantial equivalent performance," indicating the new device's performance was measured.


    7. The Type of Ground Truth Used

    The ground truth for these types of devices is based on analytical validation and reference methods.

    For the Lithium Flex® Reagent Cartridge: The "ground truth" for lithium concentration in samples would be established by reference methods or validated comparative methods, against which the device's quantitative measurements are compared to assess accuracy, precision, and linearity.

    For the Drug 4 Calibrator: The "ground truth" for the calibrator's target concentrations (Levels 1-5) is established by precise formulation and measurement against recognized standards. The matrix itself (human serum) is a specified component.


    8. The Sample Size for the Training Set

    Not applicable and not provided. These are chemical reagents and calibrators for an IVD assay, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The development and optimization of such reagents involve experimental design, formulation studies, and analytical verification, rather than machine learning training.


    9. How the Ground Truth for the Training Set was Established

    Not applicable and not provided. As explained in point 8, there is no "training set" in the context of these devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1