510(k) Data Aggregation

    K Number
    K242499

    Validate with FDA (Live)

    Date Cleared
    2025-01-06

    (137 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital radiography X-ray system is used in hospitals, clinics, and medical practices.

    The Digital radiography X-ray system enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. It is not intended for mammographic applications.

    The Digital radiography X-ray system uses digital detectors for generating diagnostic images by converting X-rays into image signals.

    Device Description

    The Digital Radiography X-ray System is a radiography X-ray system. It is designed as a modular system with components such as a ceiling suspension with an X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless, and fixed integrated detectors that may be combined into different configurations to meet specific customer needs. Software is of Basic level of concern, which is also based on a predicate device.

    AI/ML Overview

    The provided documents describe the Digital Radiography X-ray System (Sontu100-Rad(E), Sontu300-Mars(E)) and its substantial equivalence to predicate devices (K220919 and K213700). However, the document does not contain specific acceptance criteria, reported device performance metrics in numerical form, details about a clinical study for comparative effectiveness, or information regarding ground truth establishment beyond comparison to predicate device characteristics and phantom/patient images.

    Based on the information provided, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance

    The document mentions that "all the performance parameters of flat panel detector meet the requirements, and the performance parameters of proposed detector and predicate detector are the same." It also states that a "concurrence evaluation on a certain number of clinical images from the same phantom or patients" was performed, and the results "show the ability of the device to provide images with equivalent diagnostic capability to those of the cleared predicate devices."
    However, specific quantitative acceptance criteria and their corresponding reported performance values are not provided in the document. The document primarily focuses on demonstrating substantial equivalence through component and technical characteristic comparisons with predicate devices and compliance with relevant IEC and ASTM standards.

    2. Sample size used for the test set and the data provenance

    The document mentions "a certain number of clinical images from the same phantom or patients" for the concurrence evaluation.

    • Sample size: "a certain number" (specific number not provided).
    • Data provenance: "clinical images from the same phantom or patients". The country of origin is not specified, nor is whether the data is retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document.

    4. Adjudication method for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study.
    • AI assistance: The device described is a Digital Radiography X-ray System, which is hardware for generating images. There is no mention of AI assistance or an AI component in this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Since there is no mention of an AI algorithm or software component for image analysis (beyond the basic level of concern software managing the X-ray system), a standalone algorithm-only performance study like this is not applicable and not mentioned. The "performance data" refers to the overall X-ray system and its detector's image quality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document indicates that the performance of the flat panel detectors was evaluated based on standard image quality metrics (Effect Image Area, Linear Dose Range, Linear Dynamic Range, Spatial Resolution, Low Contrast Resolution, Flat Uniformity, Modulation Transfer Function, Detective Quantum Efficiency, Artifact, Erasure Thoroughness) and comparison to predicate devices. For clinical images, a "concurrence evaluation" was performed to show "equivalent diagnostic capability to those of the cleared predicate devices." This suggests the ground truth was implied to be equivalent to the diagnostic output of the predicate devices, likely through comparison of images by human readers, but the specifics of how this diagnostic capability was established as ground truth (e.g., expert consensus) are not detailed.

    8. The sample size for the training set

    The document describes an X-ray imaging system, not a device that uses a training set for machine learning. Therefore, a "training set" in this context is not applicable and not mentioned.

    9. How the ground truth for the training set was established

    As there is no training set for this type of device, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240841

    Validate with FDA (Live)

    Date Cleared
    2024-12-09

    (257 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities).

    This device is not intended for mammography and pediatric patients.

    Device Description

    This Digital Radiography System (ManntiX K) is a portable digital device developed, designed and manufactured by Shenzhen Browiner Tech Co., Ltd. A detailed comparison table with an equivalent device are in 510(k) summary. The device consists of the following major components: portable X-ray Assembly, X-ray Protective Device, Mobile Stand, Digital Detector, image processing system. The difference between ManntiX B and ManntiX K are as below.

    ModelManntiX BManntiX K
    Portable X-ray Assembly (including collimator and tube)PX10PX10 Pro
    X-ray Protective Device (optional)BA-01
    Mobile StandMS-05PMS-10P
    Digital Detector (optional)CareView 1500Cwe
    Mars1717X
    Luna 1012X
    Image Processing System (optional)MOC (V03)
    Image Acquisition Workstation (optional)BWS-10

    Note:

      1. ManntiX B and ManntiX K are delivered with one of CareView1500Cwe, Mars1717X and Luna 1012X, or two of them, or three of them.
      1. A computer system, necessary for image viewing and manipulation, is not part of the device.

    The major components of Digital Radiography System include: portable x-ray assembly, x-ray protective device, mobile stand, digital detector, image processing system, image acquisition workstation.

    Portable X-ray assembly including collimator and tube, produces X-rays at the clinically required dose; X-ray protective device is used to protect the human body during diagnostic radiography; mobile stand is used to support or hang X-ray source components; digital detector receives X-ray signals and converts them into visible light, which is converted into electrical signals by photoelectric elements and then converted into digital signals by analog-to-digital conversion before being transmitted to the computer; image processing system has multiple functions. including new patient registration, image acquisition and processing, data transmitting, etc; image acquisition workstation is used in addition to the image processing system to adjust exposure parameters, check the status of the entire machine, and view logs.

    AI/ML Overview

    The provided document is a 510(k) summary for a Digital Radiography System (ManntiX B, ManntiX K). It describes the device, its intended use, and compares it to a predicate device. However, it does not contain information about acceptance criteria and a study proving that the device meets those criteria, especially in the context of an AI/ML-driven device's performance.

    The document primarily focuses on:

    • Regulatory information: FDA review, regulations, and contact details.
    • Device description: Components, technical specifications (generator, detector, image processing), and intended use.
    • Comparison to a predicate device: Highlighting similarities in intended use, technology, and functionality to establish substantial equivalence.
    • Non-clinical performance tests: Listing compliance with various IEC and ISO standards related to electrical safety, electromagnetic compatibility, radiation protection, software lifecycle, and usability.
    • Biocompatibility: Statement of evaluation according to ISO 10993-1.

    There is no mention of:

    • Specific acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy) beyond general safety and electrical performance standards.
    • A study proving the device meets such acceptance criteria.
    • Sample sizes for test sets, data provenance, expert ground truth adjudication (for AI/ML performance evaluation), MRMC studies, standalone performance, type of ground truth used (pathology, outcomes), or training set details.

    This type of 510(k) submission primarily demonstrates that the proposed device is as safe and effective as a legally marketed predicate device, often through a comparison of technical specifications and adherence to recognized standards. It does not typically involve the detailed clinical or AI/ML performance evaluation aspects you've asked about unless the device incorporates novel AI/ML components that significantly alter its diagnostic output or workflow. In this case, the device is a Digital Radiography System, which is an imaging modality, and the details provided relate to the hardware and image acquisition/processing capabilities rather than an AI/ML diagnostic aid.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232298

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-26

    (269 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DIGITAL RADIOGRAPHY CXDI-RF Wireless B1 is indicated for use in generating radiographic images of human anatomy to replace the radiographic film/screen systems in all general purpose diagnostic procedures. It is also used for generating fluoroscopic images, when integrated into the X-ray diagnostic systems, to replace the spot-film devices and the X-ray image intensifiers.

    Not intended for mammography applications.

    Device Description

    The proposed CXDI-RF Wireless B1 is a solid state X-ray imager, which is capable of both fluoroscopic and spot radiographic imaging and has a common Thallium-doped Cesium Iodide (CsI:TI) scintillator. The CXDI-RF Wireless B1 will be integrated as a component into the Canon fluoroscopy x-ray system for fluoroscopic and radiographic imaging. The CXDI-RF Wireless B1 intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values. The digital values are sent to the PC via a wired or wireless connection, converted to images with the CXDI Control Software (CXDI Controller RF version 3.11), and then displayed on the PC/monitors. The PC/monitors used with the CXDI-RF Wireless B1 are not a part of this submission.

    AI/ML Overview

    The provided document from the FDA 510(k) premarket notification for the Canon Inc. DIGITAL RADIOGRAPHY CXDI-RF Wireless B1 does not contain information about acceptance criteria and a study proving that the device meets specific acceptance criteria related to AI/algorithm performance.

    The document states that the device is a digital radiography and fluoroscopy system, which is hardware (a solid-state X-ray imager) with associated control software. It is not an AI/CADe (Computer-Aided Detection/Diagnosis) device with performance metrics like sensitivity, specificity, or AUC, which are typically evaluated against AI acceptance criteria.

    The "Summary of Non-Clinical/Test Data" section primarily references adherence to technical standards, internal specifications, and FDA guidance documents for medical devices (e.g., biological evaluation, risk management, software life cycle, cybersecurity, electrical safety, radiation protection, usability, etc.). It emphasizes that "No clinical testing was required as non-clinical testing should be sufficient to demonstrate that the DIGITAL RADIOGRAPHY CXDI-RF Wireless B1 works as intended."

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets AI/algorithm-specific acceptance criteria, as the document does not describe such a study or performance metrics for AI.

    If you have a document that specifically details the AI/CADe component of a different device, I would be happy to analyze it for the information you've requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240284

    Validate with FDA (Live)

    Date Cleared
    2024-03-28

    (56 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technician on adult subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

    Device Description

    The Digital Radiography System (MobileApex 60E, MobileApex 60F, MobileApex 60G, MobileApex 60H) is comprised of a High Voltage Generator with a maximum power output of 63kW, the Digital Radiography System (MobileApex 60E, MobileApex 60F, MobileApex 60C, MobileApex 60H) can meet different exposure needs for varying positions and body mass. With the Digital Radiography System's (MobileApex 60E, MobileApex 60F, MobileApex 60G, MobileApex 60H) state-of-the-art design and powerful 63kW generator coupled with the Digital Detector (CareView 1500Cwe, CareView 750Cw, Mars1717X, Mars1417X, Luna 1012X) and Digital Radiography Operator Console, users can obtain clear images quickly and easily.

    AI/ML Overview

    Based on the provided text, the device in question is a Digital Radiography System (MobileApex 60E, MobileApex 60F, MobileApex 60G, MobileApex 60H). The acceptance criteria and the study proving the device meets these criteria are focused on demonstrating substantial equivalence to a predicate device, the MX40 Mobile Digital X-ray System (K181874), rather than proving a specific, quantitative performance metric for a clinical application (like detection of a disease).

    The documentation primarily discusses non-clinical performance tests related to electrical safety, electromagnetic compatibility, radiation protection, software validation, and usability. There is no mention of a clinical study to assess diagnostic performance (e.g., accuracy in detecting specific conditions).

    Therefore, the following information is extracted and where information is not provided in the text, it is explicitly stated as "Not provided in the text."


    Acceptance Criteria and Device Performance for Digital Radiography System

    The acceptance criteria for this device are based on demonstrating substantial equivalence to a predicate device in terms of intended use, technological characteristics, and safety and effectiveness, primarily through compliance with recognized standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (based on substantial equivalence)Reported Device Performance (based on non-clinical testing)
    1. Similar Intended Use: Intended for diagnostic radiographic exposures of various body parts on adult subjects (skull, spinal column, chest, abdomen, extremities). Not for mammography.Device confirmed to have the "similar intended use" as the predicate.
    2. Similar Technological Characteristics: (e.g., Mobile X-ray System, Manual Operation, Radiology Panel, Class II, 21 CFR 892.1720 Regulation, IZL/MQB Product Code, operates on battery/line, 100-240V, uses specific X-ray tubes, specific imaging panels (already cleared), Windows OS, DICOM, Wi-Fi).The comparison table (Section 8) explicitly states that the "Proposed Device has the similar intended use, similar technological characteristics as the predicate device." Key specifications are provided and compared to the predicate, with minor differences acknowledged (e.g., generator power level is 63 kW for proposed vs. 50 kW for predicate, but this is not identified as raising new safety/effectiveness issues).
    3. Safety and Effectiveness: Compliance with relevant international and FDA standards for medical electrical equipment, radiation protection, software life cycle, usability, and diagnostic X-ray systems.The device has been tested and found compliant with:IEC 60601-1 (General requirements for basic safety and essential performance)IEC 60601-1-2 (Electromagnetic disturbances)IEC 60601-1-3 (Radiation protection in diagnostic X-ray equipment)IEC 60601-2-54 (Particular requirements for X-ray equipment for radiography and radioscopy)IEC 62304 (Medical device software - Software life cycle processes)IEC 62366-1 (Usability engineering)21 CFR 1020.30 (Diagnostic X-Ray systems)21 CFR 1020.31 (Radiographic equipment)All non-clinical testing results indicate the device is "as safe and effective as the predicate device."

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not applicable/Not provided. The summary describes non-clinical engineering and performance testing against standards, not a clinical test set with patient data for diagnostic performance.
    • Data provenance: Not applicable/Not provided. The testing described is type testing and verification against standards, not clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not provided. Ground truth in a diagnostic sense (e.g., for disease detection) was not established as part of this submission's described testing, which is focused on technical performance and safety.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided. No clinical test set requiring adjudication of ground truth is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study was not done. This device is a mobile X-ray system, not an AI-powered diagnostic software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a hardware device (X-ray system), not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable for diagnostic performance. For the described non-clinical performance tests (e.g., electrical safety, radiation output), the "ground truth" is defined by the requirements and specifications of the relevant international and FDA performance standards (e.g., IEC 60601 series, 21 CFR 1020.30/31).

    8. The sample size for the training set

    • Not applicable/Not provided. This document describes a traditional medical device (X-ray system) and its non-clinical testing, not an AI/ML product that would require a training set.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. (See point 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K222855

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-17

    (26 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-Pro / D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-Pro, also called the DIGITAL RADIOGRAPHY D1, (hereinafter referred to as CXDI-Pro) is a solid-state x-ray imager. The CXDI-Pro is a series of detectors, and in the predicate submission (K221876) consists of the CXDI-703C Wireless and CXDI-403C Wireless detectors, also called the AR-D3543W and AR-D4343W detectors respectively. The detectors intercept x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values. The digital values are sent to the PC via a wired or wireless connection, converted to images with the CXDI Control Software, and then displayed on the PC/monitors. The PC/monitors used with the CXDI-Pro are not a part of this submission. The proposed changes to the predicate device, CXDI-Pro, includes the addition of the new detector, CXDI-803C Wireless (also called the AR-D2735W) to the CXDI-Pro series; a firmware update from 01.01.03.00 to 01.02.00.01; and a CXDI Control Software version update from 3.10.2.2 to 3.10.2.6. The new detector, CXDI-803C Wireless, which differs in pixel count, imaging area, external dimensions, and weight, has the same image performance as the predicate detectors. None of the CXDI-Pro detectors have any dynamic functions (such as fluoroscopy).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical "acceptance criteria" in the format of a table with thresholds and corresponding performance values. Instead, it focuses on demonstrating substantial equivalence to a predicate device by comparing technological characteristics and ensuring continued conformance with safety and performance standards.

    The closest to "reported device performance" are the shared technical specifications and the statement that the new detector has the "same image performance as the predicate detectors."

    Characteristic / StandardAcceptance/Conformance Statement (Implied Criteria)Reported Device Performance
    Intended UseIdentical to predicate device: Digital image capture for conventional film/screen radiographic examinations, displaying radiographic images of human anatomy, replacing film/screen systems in all general diagnostic procedures (excluding mammography).The proposed device's Indications for Use are identical to the predicate device.
    Functional DesignSubstantially equivalent to predicate device.The flat panel detector units are functionally the same, using the same components. The fundamental scientific technology has not been modified.
    Device ComponentsPrimarily identical, with specific modifications in detector model additions and firmware/software updates not impacting safety or effectiveness.Proposed Device: CXDI-703C, CXDI-803C (NEW), CXDI-403C Wireless detectors. CXDI Control Software V3.10.2.6. Detector Firmware V01.02.00.01. Predicate Device: CXDI-703C, CXDI-403C Wireless detectors. CXDI Control Software V3.10.2.2. Detector Firmware V01.01.03.00. (The table lists the predicate as also having CXDI-803C, implying it was part of the earlier submission, but the text states the addition of 803C.)
    Image PerformanceSame as predicate detectors.The new detector, CXDI-803C Wireless, has the same image performance as the predicate detectors.
    Safety StandardsConformance with U.S. Performance Standard for radiographic equipment and relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility testing (specifically IEC 60601-1, 60601-1-2, 60601-1-6, and 60601-2-54). Changes did not impact conformance and raised no new questions regarding safety or effectiveness.Evaluation and verification/validation activities successfully demonstrated that the device continues to meet the standards for areas impacted by modifications.
    CybersecurityConformance with "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" guidance.Not explicitly detailed, but stated as followed for applicable guidance documents.
    Wireless TechnologyConformance with "Radio Frequency Wireless Technology in Medical Devices" guidance.Not explicitly detailed, but stated as followed for applicable guidance documents.
    Mechanical/ElectricalConformance with "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" guidance.Not explicitly detailed, but stated as followed for applicable guidance documents.

    2. Sample size used for the test set and the data provenance

    The document states: "Adequate detector bench testing should be sufficient to demonstrate that the subject detector. CXDI-Pro, works as intended."
    It also mentions "verification/validation activities" which included "detector bench testing."

    • Sample Size for Test Set: Not explicitly stated in terms of number of images or cases. The testing appears to be primarily at the component level (detector, firmware, software) through bench testing and conformance to standards, rather than a clinical study with a specific number of patient cases.
    • Data Provenance: Not specified, but given the nature of "bench testing" and "conformance with U.S. Performance Standard," it would likely involve laboratory test data and manufactured test images, rather than patient data from specific countries. This is a "Special 510(k) Submission" for modifications, so extensive new clinical data is often not required if substantial equivalence can be demonstrated through non-clinical means.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was a non-clinical submission focused on technical equivalency, safety, and performance standards. There was no clinical ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was a non-clinical submission; no adjudication method for a clinical test set was required or mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The device described is a digital radiography detector system, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study regarding human reader improvement with AI assistance is not relevant or reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device is a hardware component (detector) and associated software/firmware for image acquisition and display, not an algorithm providing a standalone diagnostic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For this specific submission, the "ground truth" was established against technical specifications, engineering performance metrics (e.g., MTF, DQE), and conformance to internationally recognized safety and performance standards (e.g., IEC 60601 series). There was no clinical ground truth (expert consensus, pathology, outcomes data) as this was a non-clinical submission for device modifications.

    8. The sample size for the training set

    Not applicable. This device is a diagnostic imaging system component, not a machine learning algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. There was no training set for a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222687

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-09-29

    (23 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-Elite / E1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-Elite, also called the DIGITAL RADIOGRAPHY E1, (hereinafter referred to as CXDI-Elite) is a solid-state x-ray imager. The CXDI-Elite is a series of detectors, and in the predicate device consists of the CXDI-720C Wireless detector (also called the AR-E3543W detector), the CXDI- 420C Wireless detector (also called the AR-E4343W detector), and the CXDI-420C Fixed detector (also called the AR-E4343S detector). The detector intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values. The digital values are sent to the PC via a wired or wireless connection, converted to images with the CXDI Control Software, and then displayed on the PC/monitors. The PC/monitors used with the CXDI-Elite are not a part of this submission. The proposed changes to the predicate device, DIGITAL RADIOGRAPHY CXDI-Elite (K221620), include the addition of a new detector, the CXDI-820C Wireless (also called the AR-E2735W), to the CXDI-Elite series; the standardization of the thickness of CXDI-720C Wireless and CXDI-420C Wireless; the firmware update from F2.00.03.00 to 02.01.04.00; the CXDI Control Software version update from 3.10.2.2 to 3.10.2.6; and the addition of a 25 m of the CXDI-420C Fixed detector cable and a 10 m of Wiring Cable (WC-01). Together, these changes make up the CXDI-Elite.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Digital Radiography CXDI-Elite / E1, based on the provided text:

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive performance study of a novel device. Therefore, a traditional "study that proves the device meets the acceptance criteria" in terms of clinical outcomes or diagnostic accuracy with human-in-the-loop is expressly stated as not necessary for this submission. The "acceptance criteria" here are primarily about conformance to safety and performance standards after design changes, and the "study" is the verification/validation activities.


    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Conformance with U.S. Performance Standard for radiographic equipmentThe evaluation confirmed that the changes (addition of new detector, firmware, software updates) did not impact CXDI-Elite conformance with the U.S. Performance Standard for radiographic equipment.
    Conformance with relevant voluntary safety standards for Electrical safetyThe evaluation confirmed that the changes did not impact CXDI-Elite conformance with relevant voluntary safety standards for Electrical safety, specifically IEC 60601-1, 60601-1-3, and 60601-2-54.
    Conformance with relevant voluntary safety standards for Electromagnetic Compatibility (EMC)The evaluation confirmed that the changes did not impact CXDI-Elite conformance with relevant voluntary safety standards for Electromagnetic Compatibility, specifically IEC 60601-1-2.
    Conformance with relevant voluntary safety standards for UsabilityThe evaluation confirmed that the changes did not impact CXDI-Elite conformance with relevant voluntary safety standards for Usability, specifically IEC 60601-1-6.
    Device works as intendedAdequate detector bench testing successfully demonstrated that the subject detector CXDI-Elite / E1 works as intended. The verification/validation activities conducted support a determination of substantial equivalence, indicating the device performs as expected for its intended use (digital image capture for conventional film/screen radiographic examinations, replacing film/screen systems in general purpose diagnostic procedures, not for mammography). No new questions regarding safety or effectiveness were raised compared to the predicate device.
    Substantial Equivalence to Predicate Device (K221620)Achieved, based on similarities in primary intended use, principles of operation, functional design, and established medical use.

    Study Details:

    This submission describes verification/validation activities rather than a clinical study.

    1. Sample size used for the test set and the data provenance: Not applicable in the context of a clinical test set. The submission refers to "adequate detector bench testing" and "verification/validation activities" which would involve testing the device itself against various parameters and standards. No specific number of images or patient data is mentioned for testing in the summary provided. The focus is on the device's technical performance and safety, not diagnostic accuracy based on a dataset.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic decision-making from expert interpretation is not relevant for this type of submission, as it's not a diagnostic AI device requiring human reader comparison. The "ground truth" here is compliance with engineering specifications and regulatory standards, established through technical testing.

    3. Adjudication method for the test set: Not applicable. As there were no human readers establishing ground truth for diagnostic accuracy, no adjudication method was used.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a general-purpose digital radiography system, not an AI-assisted diagnostic tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, in terms of product performance and safety, the device itself was evaluated in a standalone manner (through bench testing and compliance checks). The "device works as intended" conclusion stems from this.

    6. The type of ground truth used:

      • Engineering specifications and performance standards: For parameters like pixel pitch, scintillator type, spatial resolution, IP level, wireless communication, and physical dimensions.
      • Regulatory standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-1-6, and the U.S. Performance Standard for radiographic equipment.
    7. The sample size for the training set: Not applicable. This document does not describe the development of an AI algorithm, so there is no "training set."

    8. How the ground truth for the training set was established: Not applicable, as there is no training set for an AI algorithm mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222661

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-09-27

    (25 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-CS01 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-CS01 (hereinafter referred to as CXDI-CS01) is a solid-state x-ray imager. The CXDI-CS01 is a series of detectors, and in the predicate device (K203849) consists of the CXDI-702C Wireless, CXDI-402C Wireless, CXDI-710C Wireless, CXDI-810C Wireless, and CXDI-410C Wireless.

    The CXDI-CS01 intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values. The digital values are sent to the PC via a wired or wireless connection, converted to images with the CXDI Control Software, and then displayed on the PC/monitors. The PC/monitors used with the CXDI-CS01 are not a part of this submission.

    The subject of this Special 510(k) submission is a change to the CXDI-CS01 to add the DIGITAL RADIOGRAPHY CXDI-Pro (hereinafter referred to as CXDI-Pro) series and the DIGITAL RADIOGRAPHY CXDI-Elite (hereinafter referred to as CXDI-Elite) series detectors to the CXDI-CS01 series of detectors. In addition, the CXDI-Pro and CXDI-Elite components and features (Bluetooth connection and the Built-in AEC Assistance) have been added to the CXDI-CS01. The CXDI Control Software has been updated from V2.19.0.7 to V3.10.2.2 to add the Intelligent NR function (cleared under K212269), update the GPU driver, add several minor functional changes, and fix bugs. Together, these changes make up the CXDI-CS01.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "DIGITAL RADIOGRAPHY CXDI-CS01". It describes the device, its intended use, and its similarities to a predicate device. However, it explicitly states that clinical testing was not necessary for this submission due to minor differences from the predicate device. Therefore, the document does not contain any information about acceptance criteria, device performance from a clinical study, sample sizes, ground truth establishment, or expert involvement for a clinical study.

    The document only discusses bench testing to demonstrate conformance with performance standards and safety standards.

    Here's a breakdown of what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance

    This information is not available in the provided text. The document refers to conformance with various standards (e.g., IEC 60601-1) but does not provide specific acceptance criteria or reported performance data for these standards in a tabular format. The table provided in the 510(k) summary compares technological characteristics with the predicate, such as DQE, pixel pitch, and spatial resolution, but these are characteristics of the device, not performance benchmarks against acceptance criteria from a study demonstrating clinical effectiveness.

    Technological Characteristics Comparison (from the provided text, not acceptance criteria/performance from a clinical study):

    CharacteristicProposed Device PerformancePredicate Device Performance
    ScintillatorCsI(Tl)CsI(Tl)
    Pixel PitchCXDI-702C Wireless, CXDI-402C Wireless, CXDI-710C Wireless, CXDI-810C Wireless, CXDI-410C Wireless, and CXDI-Elite: 125μm. CXDI-Pro: 140 μm.CXDI-702C Wireless, CXDI-402C Wireless, CXDI-710C Wireless, CXDI-810C Wireless, CXDI-410C Wireless: 125μm.
    Spatial ResolutionCXDI-702C Wireless, CXDI-402C Wireless, CXDI-710C Wireless, CXDI-810C Wireless, and CXDI-410C Wireless: 35% [MTF@2lp/mm]. CXDI-Pro and CXDI-Elite: 45% [MTF@2lp/mm].CXDI-702C Wireless, CXDI-402C Wireless, CXDI-710C Wireless, CXDI-810C Wireless, CXDI-410C Wireless: 35% [MTF@2lp/mm].
    DQECXDI-702C Wireless: 58% [@0.5 lp/mm, 1 mR]. CXDI-402C Wireless: 58% [@0.5 lp/mm, 3.5 uGy]. CXDI-710C Wireless, CXDI-810C Wireless, and CXDI-410C Wireless: 60% [@0 lp/mm, 4 uGy]. CXDI-Pro: 58% [@0.5 lp/mm, 3.5 uGy]. CXDI-Elite: 67% [@0.5 lp/mm, 3.5 uGy].CXDI-702C Wireless: 58% [@0.5 lp/mm, 1 mR]. CXDI-402C Wireless: 58% [@0.5 lp/mm, 3.5 uGy]. CXDI-710C Wireless, CXDI-810C Wireless, and CXDI-410C Wireless: 60% [@0 lp/mm, 4 uGy].

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not available. The document explicitly states: "Clinical testing is not necessary for the current submission, based on the minor differences from the predicate device. Adequate detector bench testing should be sufficient to demonstrate that the subject detector CXDI-CS01 works as intended."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not available due to the absence of a clinical study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not available due to the absence of a clinical study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not available. The device is a "Solid State X-Ray Imager" and its primary function is digital image capture. There's no mention of AI assistance for human readers in the document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not available. The device is a digital radiography system, not an AI algorithm, and no clinical performance studies are mentioned.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not available due to the absence of a clinical study.

    8. The sample size for the training set

    This information is not available. There is no mention of an AI algorithm requiring a training set in this submission.

    9. How the ground truth for the training set was established

    This information is not available due to the absence of a training set as no AI algorithm is detailed in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221876

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-15

    (17 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-Pro / D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-Pro, also called the DIGITAL RADIOGRAPHY D1, (hereinafter referred to as CXDI-Pro) is a solid-state x-ray imager. The CXDI-Pro is a series of detectors, and in the predicate device (K220098) consists of the CXDI-703C Wireless and CXDI-403C Wireless detectors, also called the AR-D3543W and AR-D4343W detectors respectively. The detectors intercept x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values The digital values are sent to the PC via a wired or wireless connection, converted to images with the CXDI Control Software, and then displayed on the PC/monitors The PC/monitors used with the CXDI-Pro are not a part of this submission.

    The subject of this Special 510(k) submission is a change to the predicate device, including compatibility with the Multi Box and Status Indicator, DIGITAL RADIOGRAPHY CXDI-Elite components that were cleared under K213780. The Status Indicator can be used when the proposed device is used in combination with the Multi Box. In addition, the CXDI Control Software has been updated from V3.10.0.8 to V3.10.2.2 to fix some bugs. Together, these changes make up the CXDI-Pro.

    AI/ML Overview

    Based on the provided text, the device in question, DIGITAL RADIOGRAPHY CXDI-Pro / D1, is a digital radiography system, and the submission (K221876) is a Special 510(k). This means the submission focuses on modifications to an already cleared predicate device (K220098) rather than a completely new device. Therefore, the clinical testing section is deemed "not necessary" due to "minor differences from the predicate device."

    This significantly limits the type of performance study and acceptance criteria information available. Special 510(k) submissions typically rely heavily on non-clinical bench testing and verification/validation activities to demonstrate that changes do not adversely affect safety or effectiveness, rather than large-scale clinical trials measuring diagnostic accuracy.

    Here's an analysis based on the constraint that "clinical testing is not necessary" for this specific submission:

    1. A table of acceptance criteria and the reported device performance:

    Since clinical testing was not performed for this submission, there are no acceptance criteria related to diagnostic accuracy or human-in-the-loop performance. The acceptance criteria and performance reported are focused on demonstrating that the modifications (addition of Multi Box and Status Indicator, software update, and firmware update) do not negatively impact the fundamental scientific technology and existing safety/performance standards of the predicate device.

    Acceptance Criteria CategorySpecific Acceptance Criteria (from text)Reported Device Performance / Evaluation (from text)
    Safety & Electrical PerformanceConformance with U.S. Performance Standard for radiographic equipment.Evaluation confirmed changes did not impact conformance.
    Conformance with relevant voluntary safety standards for Electrical safety (IEC 60601-1).Evaluation confirmed changes did not impact conformance.
    Electromagnetic Compatibility (EMC)Conformance with relevant voluntary safety standards for Electromagnetic Compatibility (IEC 60601-1-2).Evaluation confirmed changes did not impact conformance.
    Usability / Software LifecycleConformance with IEC 60601-1-6 (Usability). (Implied by software update context)Verification/validation activities demonstrated the device continues to meet standards.
    X-ray Imaging EquipmentConformance with IEC 60601-2-54 (X-ray equipment for radiography and radioscopy).Evaluation confirmed changes did not impact conformance.
    Software FunctionalityCXDI Control Software V3.10.2.2 fixes previous bugs (from V3.10.0.8).Verification/validation activities demonstrated the CXDI-Pro works as intended with the updated software.
    Bio-compatibilityConformance with ISO 10993-1 for biological evaluation.Follows applicable elements of ISO 10993-1.
    Overall EquivalenceNo new questions regarding safety or effectiveness compared to the predicate device.Verification/validation activities support a determination of substantial equivalence.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document explicitly states: "Clinical testing is not necessary for the current submission, based on the minor differences from the predicate device. Adequate verification and validation activities should be sufficient to demonstrate that the CXDI-Pro works as intended."

    Therefore, there is no human patient "test set" or related sample size or data provenance information. The "test set" here refers to the bench tests and engineering validations performed on the device and its components to ensure compliance with the mentioned technical standards and functionality.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as human clinical evaluation or ground truth adjudication by experts (e.g., radiologists) for diagnostic performance was not part of this "Special 510(k)" submission. The "ground truth" for the non-clinical tests would be established through engineering specifications, calibration standards, and regulatory requirements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable, as there was no test set requiring human expert adjudication for diagnostic performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was done. This device is a digital radiography system; the submission focuses on hardware and software updates to the imaging system itself, not an AI feature that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No standalone performance evaluation of a diagnostic algorithm was done. This is a review of a general-purpose digital X-ray system, not an AI diagnostic algorithm. The software update mentioned is for the CXDI Control Software, which governs the imaging process and display, not for an AI interpretation algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this submission's evaluation is based on engineering specifications, regulatory standards (e.g., IEC 60601 series, U.S. Performance Standard for radiographic equipment), and the performance characteristics of the predicate device. It's not clinical ground truth (like pathology or patient outcomes) because clinical testing was not required.

    8. The sample size for the training set:

    Not applicable. This submission is for modifications to a general radiography system, not for an AI algorithm. There is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221620

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-06-14

    (11 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    5 - 18
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-Elite / E1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-Elite, also called the DIGITAL RADIOGRAPHY E1, (hereinafter referred to as CXDI-Elite) is a solid-state x-ray imager. The CXDI-Elite is a series of detectors, and in the predicate device consists of the CXDI-720C Wireless detector unit, also called the AR-E3543W detector. The detector intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values, and the images will be displayed on monitors. The digital value can be communicated to the operator console via a wired or wireless connection.

    The subject of this Special 510(k) submission is a change to the CXDI-Elite to add new detectors, the CXDI-420C Wireless (also called AR-E4343W) and the CXDI-420C Fixed (also called AR-E4343S), to the CXDI-Elite series of detectors. In addition, a Detector Stand (DS-01) has also been added as a component to be used with both the CXDI-720C Wireless and CXDI-420C Wireless. Compatibility with the Power Box (PB-01), a CXDI-Pro component that was cleared under K220098 has also been added. The CXDI Control Software has been updated from V3.10.0.4 to V3.10.2.2 to fix some bugs, update the GPU driver, and add several minor functional changes. Together, these changes make up the CXDI-Elite.

    AI/ML Overview

    Upon review of the provided document, it is important to note that the submission is for an imaging device (Digital Radiography CXDI-Elite / E1), not an AI/Software as a Medical Device (SaMD) that performs diagnostic or assistive functions requiring performance studies with acceptance criteria based on accuracy metrics.

    The document primarily addresses a Special 510(k) submission for changes to an existing, cleared device. These changes involve adding new detectors, an optional accessory, and software bug fixes/minor functional changes. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and safety performance, rather than clinical efficacy measured by AI model output and human reader performance.

    The provided text does not contain information about:

    • A table of acceptance criteria and reported device performance related to diagnostic accuracy.
    • Sample sizes for test sets (in the context of clinical performance evaluation).
    • Data provenance (country of origin, retrospective/prospective).
    • Number of experts for ground truth establishment or their qualifications.
    • Adjudication methods.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance studies.
    • Types of ground truth (expert consensus, pathology, outcomes data).
    • Sample size for training sets.
    • How ground truth for training sets was established.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device by confirming that the modifications do not raise new questions of safety or effectiveness. This is primarily achieved through non-clinical/bench testing and adherence to relevant standards.

    Here's a summary of what the document does provide regarding acceptance criteria and performance, as it pertains to the hardware and basic software functionality of a digital radiography system:

    Acceptance Criteria and Study for Digital Radiography CXDI-Elite / E1 (Hardware/Software Updates)

    The acceptance criteria are implicitly met by demonstrating continued compliance with relevant international and FDA standards, and by showing that the changes (new detectors, accessory, software update) do not negatively impact the fundamental performance or safety characteristics of the device.

    1. Table of Acceptance Criteria (Implicit) and Reported Device Performance (as stated in the document):

    Acceptance Criteria (Implicit/Standard Compliance)Reported Device Performance (as per "Summary of Non-Clinical / Test Data")
    Fundamental Scientific Technology Unchanged"The fundamental scientific technology of the CXDI-Elite has not been modified."
    Conformance with U.S. Performance Standard for Radiographic Equipment"Evaluation of the addition of the new detector and optional accessories confirmed that the changes did not impact CXDI-Elite conformance with the U.S. Performance Standard for radiographic equipment..."
    Compliance with Voluntary Safety Standards (Electrical Safety, EMC)"...and with relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-2-54, and 60601-1-6."
    Device Works as Intended (Functional Performance)"Adequate detector bench testing should be sufficient to demonstrate that the subject detector CXDI-Elite / E1 works as intended." (This indicates bench testing was done to verify the new components and software perform correctly within the system.)
    No New Questions Regarding Safety or Effectiveness Compared to Predicate"These verification/validation activities successfully demonstrated that the device continues to meet the standards for the areas impacted by the modifications to the predicate device and raises no new questions regarding either safety or effectiveness when compared to the predicate device."
    Pixel Pitch125μm (Identical to predicate)
    ScintillatorCsI(Tl) [Cesium Iodide doped with Thallium] (Identical to predicate)
    Spatial Resolution45% [MTF@2lp/mm] (Identical to predicate)
    DQE (Detective Quantum Efficiency)67% [@0.5 lp/mm, 3.5 uGy] (Identical to predicate)
    IP Level (Ingress Protection)CXDI-720C Wireless: IP57, CXDI-420C Wireless: IP57, CXDI-420C Fixed: N/A (Modified from predicate, detailed for each new component)
    Wireless CommunicationIEEE 802.11a/b/g/n/ac, 2.4GHz/5GHz, Bluetooth Low Energy (only for CXDI-720C Wireless and CXDI-420C Wireless) (Identical to predicate's wireless capabilities)
    External DimensionsCXDI-720C Wireless: 384 x 460 x 15.5 mm, CXDI-420C Wireless: 460 x 460 x 15.5 mm, CXDI-420C Fixed: 460 x 460 x 15.3 mm (Modified from predicate, detailed for each new component)
    Photographing Mode (Functional Modes of Operation)Standard Synchronization Mode, Standard Synchronization Mode with Built in AEC Assistance, Non Generator Connection Mode, Standalone Mode (only for CXDI-720C Wireless and CXDI-420C Wireless) (Modified from predicate to reflect modes available with new detectors; one mode removed for fixed detector)
    Compliance with Applicable FDA Guidance DocumentsThe proposed device follows applicable elements of several FDA guidance documents (e.g., Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, etc.)

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document states "Adequate detector bench testing should be sufficient to demonstrate that the subject detector CXDI-Elite / E1 works as intended." This implies physical testing of the devices, but does not specify sample sizes for "test sets" in the clinical performance sense, as this was not a clinical performance study. The testing appears to be primarily engineering verification and validation of hardware and software components.
    • Data provenance: Not applicable in the context of clinical images or patient data. The testing is a validation of the device's technical specifications and compliance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts:

    • Not applicable. This was not a study requiring expert readers to establish ground truth on clinical images, as it was not a diagnostic performance study for an AI algorithm.

    4. Adjudication Method for the Test Set:

    • Not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This was not an AI-enabled diagnostic device or a comparative effectiveness study involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This device is a digital radiography system, not a standalone AI algorithm. Its function is to capture and display radiographic images.

    7. The Type of Ground Truth Used:

    • Not applicable in the clinical diagnostic sense. The "ground truth" for this submission are the established engineering specifications, safety standards (IEC 60601 series), and FDA guidance documents, against which the device's technical performance (e.g., MTF, DQE, IP level, functional modes) is verified via bench testing.

    8. The Sample Size for the Training Set:

    • Not applicable. This device does not involve machine learning or AI models with training sets.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable for the reasons stated above.

    In conclusion, the provided document is a 510(k) summary for a digital radiography hardware system with minor modifications. The "acceptance criteria" and "study" described pertain to engineering validation, safety testing, and demonstration of substantial equivalence by showing that the updated device continues to meet established performance and safety standards, and does not pose new risks. It is not an AI/SaMD performance study seeking to establish diagnostic accuracy against clinical ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220098

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-02-04

    (23 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DIGITAL RADIOGRAPHY CXDI-Pro / D1 provides digital image capture for conventional film/screen radiographic examinations. This device is intended to capture, for display, radiographic images of human anatomy, and to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

    Device Description

    The DIGITAL RADIOGRAPHY CXDI-Pro, also called the DIGITAL RADIOGRAPHY D1, (hereinafter referred to as CXDI-Pro) is a solid-state x-ray imager. The CXDI-Pro is a series of detectors, and in the predicate device consists of the CXDI-703C Wireless detector unit, also called the AR-D3543W detector. The detector intercepts x-ray photons, and the scintillator emits visible spectrum photons that illuminate an array of photodetectors that create electrical signals. After the electrical signals are generated, the signals are converted to digital values, and the images will be displayed on monitors. The digital image can be communicated to the operator console via a wired or wireless connection.
    The subject of this Special 510(k) submission is a change to the CXDI-Pro add a detector, the CXDI-403C Wireless (also called AR-D4343W) to the CXDI-Pro series of detectors. In addition, a Detector Stand (DS-01) and a new Battery Charger (BC-01) have been added as optional components to be used with the proposed CXDI-Pro detectors (CXDI-703C Wireless and CXDI-403C Wireless), the sleep to ready cycle time for the proposed CXDI-Pro detectors has been decreased, and the CXDI Contro Software has been updated from V3.10.0.3 to V3.10.0.8 to add the Intelligent NR function (cleared under K212269) and apply some bug fixes. Together, these changes make up the CXDI-Pro.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Canon Inc. Digital Radiography CXDI-Pro / D1, based on the provided FDA 510(k) summary:

    This submission (K220098) is a Special 510(k) submission, which indicates that the changes made to the device (new detector, accessories, reduced cycle time, software update) did not fundamentally alter its intended use or fundamental scientific technology. Therefore, the primary focus of the performance evaluation is to ensure that these changes do not negatively impact the device's conformance with relevant standards and do not raise new questions of safety or effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a Special 510(k) for modifications to an already cleared device, the "acceptance criteria" are primarily related to maintaining conformance with existing performance and safety standards, rather than establishing new clinical performance targets. The reported device performance confirms that these standards are still met.

    Acceptance Criteria (related to modified aspects)Reported Device Performance
    Maintain conformance with U.S. Performance Standard for radiographic equipment.Evaluation of the addition of the new detector (CXDI-403C Wireless) and optional accessories (Detector Stand, Battery Charger) confirmed that the changes did not impact CXDI-Pro conformance with this standard.
    Maintain conformance with relevant voluntary safety standards for Electrical safety and Electromagnetic Compatibility (EMC), specifically IEC 60601-1, 60601-1-2, 60601-2-54, 60601-1-6, and IEC 60529.Verification/validation activities successfully demonstrated that the device continues to meet these standards.
    Ensure the reduction in "Sleep to Ready Cycle Time" (from 7 sec to less than 5 sec) does not compromise safety or effectiveness.Implied: The successful verification/validation activities encompass this change, indicating it does not raise new questions of safety or effectiveness. (Specific performance data for this change beyond the stated time is not detailed, but its successful integration into the system meeting overall standards implies acceptance.)
    Ensure the software update (V3.10.0.3 to V3.10.0.8, including Intelligent NR function K212269) and firmware update (01.00.00.03 to 01.00.02.00) do not compromise safety or effectiveness.Implied: The software update to incorporate an already cleared feature (Intelligent NR, K212269) and bug fixes, along with the firmware update, were part of the verification/validation activities that demonstrated the device continues to meet standards and raises no new questions regarding safety or effectiveness. (Specific performance data for the software/firmware changes beyond their successful integration and bug fixes are not detailed in this summary.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set sample size" in terms of patient images or clinical cases. The performance evaluation is described as "Verification/validation activities" which focused on technical performance and safety standards. This suggests that the testing involved:

    • Engineering tests on the physical components (new detector, accessories).
    • Software testing for the updated control software and firmware.
    • Electromagnetic compatibility (EMC) testing.
    • Electrical safety testing.

    The data provenance is industrial and engineering testing, rather than clinical data from specific countries or retrospective/prospective studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable in the context of this Special 510(k). Ground truth for clinical diagnostic performance (e.g., presence/absence of disease) is not the focus of this submission. The "ground truth" here is adherence to engineering specifications and safety standards, which would be assessed by engineers and quality control personnel.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study requiring reader adjudication of clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed to demonstrate the clinical performance of a diagnostic aid, which is not the scope of this Special 510(k) submission focused on hardware/software modifications and maintaining safety/performance standards.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The document primarily describes the performance of the overall digital radiography system, which includes the detector and control software. While the software update includes an "Intelligent NR function" (cleared under K212269), this submission's performance evaluation focuses on the integrated system's safety and continued conformance, not a standalone algorithm's diagnostic performance. Its effectiveness would have been established in the K212269 submission.

    7. The Type of Ground Truth Used

    The ground truth used for these verification and validation activities is based on:

    • Engineering specifications (for the new detector, cycle time, accessories).
    • International standards (IEC 60601-1, 60601-1-2, 60601-2-54, 60601-1-6, IEC 60529).
    • U.S. Performance Standard for radiographic equipment.
    • Software testing protocols to verify bug fixes and the integration of the Intelligent NR function without introducing new issues.

    8. The Sample Size for the Training Set

    Not applicable. This submission does not describe a machine learning model being developed or trained on a dataset. The software update includes an "Intelligent NR function" which would have involved a training set in its original K212269 submission, but that information is not provided here.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set described in this submission.