Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K171517

    Validate with FDA (Live)

    Date Cleared
    2018-01-11

    (232 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.

    Device Description

    The proposed device can acquire ECG signal via twelve leads simultaneously, display or print waveform of ECG signal. The ECG electrical signals from patient body surface are acquired by disposable ECG electrodes; after been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper via an external printer.

    AI/ML Overview

    The provided text is a 510(k) Summary for a Digital Electrocardiograph. It contains information about the device's substantial equivalence to a predicate device, focusing on non-clinical tests and technical specifications. However, it does not include any details about clinical studies, acceptance criteria for an algorithm's performance, or the specific information requested about human reader performance, ground truth establishment for a test set or training set, or sample sizes related to clinical efficacy evaluation.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and the reported device performance: This information is not present. The document focuses on compliance with electrical safety and performance standards for the hardware, not the diagnostic performance of an interpretive algorithm against specific clinical benchmarks.
    • Sample sized used for the test set and the data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set sample size, and ground truth establishment for the training set: None of this information is available in the provided text, as it describes a non-clinical assessment for substantial equivalence, primarily focusing on hardware characteristics and safety standards.

    The document mentions "automated analysis" and "interpretive statements offered to clinician on an advisory basis only," but it does not provide any data or studies related to the performance or validation of this interpretive function from a clinical accuracy perspective. It primarily verifies that the proposed device met design specifications and complied with relevant safety standards (IEC 60601-1, IEC 60601-2-25, IEC 60601-1-2) which are non-clinical requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160092

    Validate with FDA (Live)

    Date Cleared
    2016-02-12

    (28 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. The proposed device has analysis feature, however the ECC with measurements and interpretive statements is offered to clinician on an advisory basis only.

    Device Description

    The proposed digital electrocardiograph, iE 101 and iE 300, are designed based on the cleared digital electrocardiographs K141946 (iE 3). The proposed device has same intended use, functional modules design, working modes, analysis function and accessories. the proposed devices are different than the predicate device in that they has new printing channel for iE 101); new layout of user-machine interface; three different physical specification, such as screen size, device dimension, device weight; three different specification, such as battery capacity, noise level and CMRR; and a modified software which is modified to applicable to the proposed devices.

    The proposed electrocardiographs, iE 101 and iE 300, follow the same design principle, main components, accessories and software; they acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.

    They consist of four modules, which are power supply module, signal collection module, amplification module, and control module.

    They can acquire ECG signal via twelve leads simultaneously, display or print waveform of ECG signal via one channel (iE 101) or three channel (iE 300).

    AI/ML Overview

    The provided text describes a 510(k) submission for a Digital Electrocardiograph, models iE101 and iE300. The submission argues for substantial equivalence to a predicate device (K141946).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission focuses on establishing substantial equivalence to a predicate device, rather than defining specific performance acceptance criteria for diagnostic accuracy. The "acceptance criteria" here are essentially the compliance with relevant safety and performance standards and the demonstration that the proposed device performs similarly to the predicate.

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with IEC 60601-1:2005 + A1: 2012 (General safety)Complies with IEC 60601-1:2005 + A1: 2012
    Compliance with IEC 60601-1-2: 2007 (EMC)Complies with IEC 60601-1-2: 2007
    Compliance with IEC 60601-2-25:2011 (Electrocardiographs particular requirements)Complies with IEC 60601-2-25:2011
    Biocompatibility (ISO 10993)Complies with ISO 10993
    Intended UseSame as predicate device
    Intended PopulationSame as predicate device (Adult and pediatric patients)
    Intended EnvironmentSame as predicate device (Hospital, clinics)
    ConfigurationSame as predicate device
    LeadSame as predicate device (Standard 12-lead)
    Acquisition modeSame as predicate device (Simultaneous 12-lead acquisition)
    Recording formatSame as predicate device (Automatic / Manual / Rhythm)
    Analysis modeYes (Same as predicate, advisory basis only)
    Physical/Technical Specifications (e.g., screen size, channels, battery capacity, noise level, CMRR)Differences noted but deemed not to affect substantial equivalence (e.g., iE101: 5.0" screen, 1 channel; iE300: 5.0" screen, 3 channel vs. predicate's 6.0" screen, 3 channel). Software modified to be applicable to proposed devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no test set in the context of clinical performance evaluation based on patient data. The "tests" conducted were non-clinical, likely bench testing and engineering verification to demonstrate compliance with standards.

    • Sample Size for Test Set: Not applicable as no clinical test set was used.
    • Data Provenance: Not applicable as no clinical data was used for a test set. The non-clinical tests were conducted by the manufacturer (Shenzhen Biocare Bio-medical Equipment Co., Ltd.) in China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable, as no clinical study or test set requiring expert-established ground truth was performed for this submission. The device's "analysis feature" is offered on an advisory basis only, meaning its interpretive statements are not intended as definitive diagnoses.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical study or test set requiring adjudication was performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The submission explicitly states "No clinical study is included in this submission." Therefore, no MRMC study was conducted, and no effect size calculation for human readers with or without AI assistance is available.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    While the device has an "analysis feature," the submission does not include any standalone performance study of this algorithm. The analysis feature provides "measurements and interpretive statements" on an "advisory basis only," indicating it's not cleared for diagnostic standalone use.

    7. The Type of Ground Truth Used

    Not applicable, as no clinical performance study requiring ground truth was conducted. For non-clinical tests (electrical safety, EMC, etc.), the "ground truth" would be the specifications and requirements of the relevant IEC standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is an electrocardiograph hardware with an accompanying analysis feature. The submission does not describe a machine learning algorithm that underwent training with a specific dataset. The analysis function is likely based on pre-programmed algorithms rather than a trained AI model in the modern sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no described training set or AI model training in the submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152384

    Validate with FDA (Live)

    Date Cleared
    2015-12-18

    (116 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and disease. Digital Electrocardiograph shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiograph, ECG-2000, is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease.

    It consists of the acquisition box, optional accessories, installation CD, and USB dongle. The acquisition box is connected with ECG electrodes, and intends to acquire ECG signals from patient. The acquired ECG signal is filtered and amplified, then transferred to the working station. The acquisition box is powered by the working station via a USB line. The working station is a Personal Computer installed with ECG-2000 Software, it intends to receive the processed ECG signal from the acquisition box, and further display and record the signals. Optional accessories include lead wires, limb electrodes and chest electrodes, and general-purpose printer (with USB interface, support PCL language, such as HP1010, P2035, and P2055d series). The CD contains Digital Electrocardiograph installer and dongle installation program module.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a Digital Electrocardiograph, Model: ECG-2000. It focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study to prove the device meets acceptance criteria derived from a novel clinical trial.

    Here's an analysis based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly define "acceptance criteria" in the context of a clinical performance study with specific metrics like sensitivity, specificity, accuracy, etc. Instead, it demonstrates compliance with technical specifications and safety standards to establish substantial equivalence. The "performance" reported is primarily in terms of meeting these engineering and safety benchmarks.

    FeaturePredicate Device 1 (K092010)Description / ValueProposed Device (ECG-2000)Reported Performance
    Product CodeDPSDPS (Same)
    Regulation Number21 CFR 870.234021 CFR 870.2340 (Same)
    ClassClass IIClass II (Same)
    Intended UseAcquire ECG signals from adult patients to help analyze and diagnose heart disease. Used by healthcare professionals.Acquire ECG signals from adult and pediatric patients to help analyze and diagnose heart disease. Used by healthcare professionals. (Similar: extended to pediatric patients)
    ConfigurationChest Electrode and Limb ElectrodeChest Electrode and Limb Electrode (Same)
    ECG LeadStandard 12-leadStandard 12-lead (Same)
    ECG Gain(Implied range from "Similar" to proposed device)1.25, 2.5, 5, 10, 20, 10-5, 20-10 (mm/mV) (More options than Predicate 1, Same as Predicate 2)
    Sampling rate1000Hz1000Hz (Same)
    Input circuit current(Implied value from "Similar" to proposed device, complies with IEC 60601-1)$\leq$ 0.1 μA (Complies with IEC 60601-1)
    Noise level(Implied value from "Similar" to proposed device, complies with IEC 60601-1)$<15\mu$ Vp-p (Complies with IEC 60601-1)
    Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1 (Same)
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2 (Same)
    Patient-contact MaterialChest Electrode: Metal, Limb Electrode: ABSSame

    Note on "Intended Use": The proposed device extends its intended use to include pediatric patients, whereas Predicate Device 1 explicitly mentions "adult patients" in the comparison table. Predicate Device 2, however, is the exact same device, and its previous clearance (K133985) likely covered adult patients, and this submission seeks to add pediatric. The document explicitly states: "This submission is filed for extension of intended use...therefore, only the following standard was tested to verify that the device can meet its additional intended use: IEC 60601-2-25: 2011, Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs."

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. The document states "No clinical study is included in this submission." The testing primarily involved product verification to standards (e.g., IEC 60601-2-25) rather than a clinical evaluation with a test set of patient data.
    • Data Provenance: Not applicable, as no clinical data set was used for performance evaluation in this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. No clinical test set requiring ground truth established by experts was used.

    4. Adjudication method for the test set

    • Not applicable. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The device is a "Digital Electrocardiograph," which primarily acquires and displays ECG signals for analysis by healthcare professionals. It does not appear to incorporate AI for interpretation or act as an AI-assistive tool for human readers, nor was an MRMC study conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The document implies the device performs basic signal acquisition, filtering, and amplification, displaying these signals for human interpretation. There is no mention of an "algorithm" performing diagnostic interpretations in a standalone capacity that would require specific performance testing against a ground truth. The device "can help users to analyze and diagnose heart disease," which suggests it provides raw or processed signals for expert review, not an automated diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. No clinical performance evaluation requiring a ground truth was conducted. The "ground truth" here is implied by the technical specifications and adherence to international standards for electrocardiographs.

    8. The sample size for the training set

    • Not applicable. This device is an acquisition and display system for ECG signals. There's no indication it employs machine learning or AI that would require a "training set."

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for an AI/ML algorithm mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141946

    Validate with FDA (Live)

    Date Cleared
    2015-02-13

    (211 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.

    Device Description

    Digital Electrocardiograph, iE 3/ iE 6/ iE 12/ iE 12A/ iE 15/ ECG-1210/ ECG-3010/ ECG-6010, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.

    All the proposed models, iE 3/ iE 6/ iE 12/ iE 12A/ iE 15/ ECG1210/ ECG3010/ ECG6010, follow the same design principle, main components, accessories and software; they acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.

    They consist of four modules, which are power supply module, signal collection module, amplification module, and control module.

    They can acquire ECG signal via twelve leads simultaneously, display or print waveform of ECG signal via three channel, six channel, twelve channel or fifteen channel.

    The CardioPro analysis program is the analysis program used in this proposed device.

    AI/ML Overview

    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative manner for specific performance metrics like accuracy, sensitivity, or specificity regarding the analysis and diagnosis of heart disease. Instead, the submission focuses on demonstrating substantial equivalence to predicate devices and compliance with international safety and performance standards.

    The table below summarizes the technical specifications and standards compliance, which implicitly serve as performance requirements for the device's functionality.

    Feature / StandardAcceptance Criteria (Implied by Predicate & Standards)Reported Device Performance
    Product CodeDPSDPS
    Regulation Number21 CFR 870.234021 CFR 870.2340
    Lead AcquisitionStandard 12-lead (or 15-lead for proposed device)Standard 12-lead or 15 lead
    Acquisition modeSimultaneous 12-lead acquisitionSimultaneous 12-lead or 15 lead acquisition
    Recording formatAutomatic / Manual / RhythmAutomatic / Manual / Rhythm
    Analysis modeYes (presence of analysis program)Yes (CardioPro analysis program)
    CMRR>60dB>60dB, >100 with AC filter
    Paper SpeedMultiple levels (e.g., 6.25, 12.5, 25, 50mm/s)4-6 levels (e.g., 6.25, 12.5, 25, 50mm/s)
    Input CIR current≤0.1μA≤0.1μA
    Input impedance>50MΩ>50MΩ
    Patient leak current<10μA<10μA
    Frequency response0.05~150Hz0.05~150Hz
    Noise level<15μVp-p<15μVp-p
    Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2
    Particular requirementsComply with IEC 60601-2-25Comply with IEC 60601-2-25
    BiocompatibilityComply with ISO 10993Comply with ISO 10993

    2. Sample size used for the test set and the data provenance

    The document does not mention a specific "test set" in the context of clinical performance for the interpretation of ECGs or diagnosis of heart disease. The provided information is focused on non-clinical testing to demonstrate compliance with general safety and performance standards (IEC standards) and an assessment of technical characteristics against predicate devices. Therefore, details like data provenance (country of origin, retrospective/prospective) are not applicable or provided for a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As there is no mention of a clinical test set with ground truth related to ECG interpretation or diagnosis in this document, there is no information about the number or qualifications of experts used for establishing ground truth. The device's interpretive statements are explicitly stated to be "on an advisory basis only," suggesting that human experts remain responsible for final diagnosis.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set for diagnostic performance is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, the document does not indicate that a multi-reader multi-case (MRMC) comparative effectiveness study was performed. The non-clinical tests relate to hardware compliance and functional specifications.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe a standalone performance study for the algorithm's diagnostic capabilities. The non-clinical tests performed were to verify that the device met design specifications and complied with safety and performance standards. The "CardioPro analysis program" is mentioned, and its interpretive statements are stated to be for "advisory basis only," indicating it's not intended for standalone diagnosis.

    7. The type of ground truth used

    Not applicable, as no clinical test set for diagnostic performance is described. The "ground truth" implicitly referred to in the non-clinical tests would be the established benchmarks and requirements defined by the IEC standards for technical performance (e.g., signal-to-noise ratio, frequency response).

    8. The sample size for the training set

    The document does not provide any information about a "training set" or explicit details on how the "CardioPro analysis program" was developed or trained.

    9. How the ground truth for the training set was established

    Not applicable, as no information on a training set or its ground truth establishment is provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133985

    Validate with FDA (Live)

    Date Cleared
    2015-01-22

    (392 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiograph is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiograph shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiograph, ECG-2000, is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease.

    It consists of the acquisition box, optional accessories, installation CD, and USB dongle. The acquisition box is connected with ECG electrodes, and intends to acquire ECG signals from patient. The acquired ECG signal is filtered and amplified, then transferred to the working station. The acquisition box is powered by the working station via a USB line. The working station is a Personal Computer installed with ECG-2000 Software, it intends to receive the processed ECG signal from the acquisition box, and further display and record the signals. Optional accessories include lead wires, limb electrodes and chest electrodes, and general-purpose printer (with USB interface, support PCL language, such as HP1010, P2035, and P2055d series). The CD contains Digital Electrocardiograph installer and dongle installation program module.

    AI/ML Overview

    The provided text is a 510(k) summary for a Digital Electrocardiograph (ECG-2000) and primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing and comparison of technical specifications. It does not contain information about acceptance criteria for device performance based on clinical studies, nor does it describe such a study.

    Therefore, I cannot provide the requested information regarding acceptance criteria, reported device performance from a clinical study, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or how ground truth was established for a training set.

    The document states that non-clinical tests were conducted to verify that the proposed device met all design specifications and complied with relevant IEC and EN standards for electrical safety, EMC, and particular requirements for electrocardiographs. It also mentions "Software Verification Test was performed to verify the software functions against its intended use." However, these are engineering verification tests, not clinical performance studies for diagnostic accuracy.

    The closest to "acceptance criteria" presented in the document are the technical specifications compared with predicate devices, such as:

    ItemAcceptance Criteria (from predicate device comparison)Reported Device Performance (Proposed Device)
    Product CodeDPSDPS
    Regulation Number21 CFR 870.234021 CFR 870.2340
    ClassClass IIClass II
    ECG LeadStandard 12-leadStandard 12-lead
    Sampling rate1000Hz (at least for one predicate)1000Hz
    Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2
    Input circuit current≤0.1 μA (for one predicate)≤0.1 μA
    Noise level<15μ Vp-p (for one predicate)<15μ Vp-p

    Summary of missing information:

    1. A table of acceptance criteria and the reported device performance from a clinical study: Not available. The document refers to engineering specifications and compliance with standards, not diagnostic performance metrics from clinical data.
    2. Sample sized used for the test set and the data provenance: Not available. No clinical test set is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available. No clinical ground truth establishment is described.
    4. Adjudication method for the test set: Not available.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not available. This device is a Digital Electrocardiograph, not an AI-assisted diagnostic tool as described by the question.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not available. The device acquires ECG signals; it's not an algorithm that performs diagnosis independently for performance evaluation in this context.
    7. The type of ground truth used: Not available.
    8. The sample size for the training set: Not available.
    9. How the ground truth for the training set was established: Not available.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131858

    Validate with FDA (Live)

    Date Cleared
    2014-06-11

    (352 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiographs, E30, E65, E70 and E80, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. There is no specific contraindication identified for the proposed devices.

    The proposed devices acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.

    The components of the proposed devices, Digital Electrocardiograph, include: (1) chest electrodes (2) limb electrodes, (3) cable and (4) Batteries;

    AI/ML Overview

    The provided text is a 510(k) Summary for a Digital Electrocardiograph. It describes the device's intended use, comparison to a predicate device, and non-clinical tests conducted to establish substantial equivalence. However, it does not contain any information regarding specific acceptance criteria for device performance, a study proving the device meets these criteria, or details about sample sizes, ground truth establishment, or expert involvement as requested.

    The summary primarily focuses on demonstrating substantial equivalence to a predicate device by comparing intended use, technological characteristics, and compliance with general safety and performance standards (like IEC 60601-1 and AAMI EC11).

    Therefore, based only on the provided text, I cannot fulfill most of your request.

    Here's what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance

    • Information in text: The document states the device complies with certain standards (IEC 60601-1, IEC 60601-1-2, AAMI / ANSI EC11:1991/(R)2007, ISO 10993-5, ISO 10993-10). These standards define performance requirements, but the specific acceptance criteria and reported device performance against these criteria are not detailed in the summary. The text only mentions "Non clinical tests were conducted to verify that the proposed device met all design specifications" and that it "complies with the following standards."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Information in text: Not available.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Information in text: Not available. The device "can help users to analyze and diagnose heart disease," implying human interpretation, but no details of an expert-adjudicated test set are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Information in text: Not available.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Information in text: Not available. This device is an electrocardiograph for acquiring ECG signals, not an AI-assisted diagnostic tool discussed in the context of MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Information in text: Not available. The device is described as acquiring signals
      and assisting users in analysis and diagnosis, not a standalone automated diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Information in text: Not available.

    8. The sample size for the training set

    • Information in text: Not available. (No mention of a training set for an algorithm).

    9. How the ground truth for the training set was established

    • Information in text: Not available. (No mention of a training set for an algorithm).

    Summary of available information related to performance/acceptance criteria:

    Acceptance Criteria (Standards Met)Reported Device Performance
    IEC 60601-1 (Safety)Complies w/ standard
    IEC 60601-1-2 (EMC)Complies w/ standard
    AAMI / ANSI EC11 (Diagnostic ECG Devices)Complies w/ standard
    ISO 10993-5 (Cytotoxicity)Non-Cytotoxicity
    ISO 10993-10 (Irritation/Hypersensitivity)Non-Irritation, Non-Sensitization

    Further Details:

    • Test Set Sample Size & Provenance: Not provided.
    • Experts for Ground Truth & Qualifications: Not provided.
    • Adjudication Method: Not provided.
    • MRMC Comparative Effectiveness Study: Not conducted/reported.
    • Standalone Performance Study: Not conducted/reported.
    • Type of Ground Truth Used: Not provided.
    • Training Set Sample Size: Not applicable/provided (device is for signal acquisition, not an AI diagnostic algorithm with a training set).
    • Training Set Ground Truth Establishment: Not applicable/provided.

    The document focuses on demonstrating substantial equivalence to an existing legally marketed device (predicate device) by showing similar intended use, technological characteristics, and compliance with recognized performance and safety standards for electrocardiographs. It does not provide the kind of detailed clinical study data often found for AI/CADe devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132758

    Validate with FDA (Live)

    Date Cleared
    2014-01-30

    (148 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiographs, iE 3, iE 6, are intended to acquire, display and record ECG signals from adult and pediatric patients through body surface by ECG electrodes. ECG data result and patient information could be stored in the memory of the device. The obtained ECG records can help users to analyze and diagnose heart disease.

    AI/ML Overview

    The provided text is a 510(k) summary for Digital Electrocardiographs (models iE 3, iE 6) and a letter from the FDA determining their substantial equivalence to a predicate device (iE 12). This document does not contain information about specific acceptance criteria, device performance metrics, or details of a study demonstrating the device meets such criteria.

    The core of this document is a comparison between the proposed devices and a predicate device to establish "substantial equivalence," meaning that the new devices are as safe and effective as a legally marketed device. This is a regulatory pathway, not a clinical study demonstrating performance against specific benchmarks.

    Therefore, many of the requested details cannot be extracted from this document.

    Here's what can be inferred or directly stated from the provided text, and what is explicitly missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Acceptance Criteria: Not explicitly stated in terms of performance metrics (e.g., sensitivity, specificity, accuracy for disease detection). The acceptance criteria, in the context of this 510(k), are primarily related to meeting design specifications and demonstrating substantial equivalence to the predicate device, especially regarding safety and intended use.
    • Reported Device Performance: No specific numerical performance metrics are reported in this document. The document states a "Non-Clinical Test Conclusion" that "Non clinical tests were conducted to verify that the proposed device met all design specifications as vas Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1, IEC 60601-1-2." This refers to electrical safety and electromagnetic compatibility standards, not clinical performance for diagnosing heart disease.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not mentioned.
    • Data Provenance: Not mentioned.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not mentioned, as a clinical study with ground truth established by experts is not detailed here.

    4. Adjudication Method

    • Not mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • If done: No, an MRMC comparative effectiveness study is not described or referenced in this document. This submission focuses on establishing substantial equivalence to a predicate device based on technical characteristics and intended use, not on demonstrating improved human reader performance with AI assistance.
    • Effect size of human readers improvement: Not applicable, as no such study is presented.

    6. Standalone (Algorithm Only) Performance Study

    • If done: No, a standalone performance study of an algorithm is not described or referenced. The device is a "Digital Electrocardiograph" which acquires, displays, and records ECG signals for users (doctors/trained professionals) to analyze and diagnose. While it "can help users to analyze and diagnose heart disease," the document does not detail an inherent diagnostic algorithm's performance.

    7. Type of Ground Truth Used

    • Not mentioned, as no clinical study requiring a ground truth is detailed.

    8. Sample Size for the Training Set

    • Not mentioned, as this document does not describe the development or training of an AI algorithm.

    9. How Ground Truth for the Training Set Was Established

    • Not mentioned, as this document does not describe the development or training of an AI algorithm.

    Summary of what the document does provide:

    • Device Models: iE 3, iE 6
    • Intended Use: To acquire ECG signals from adult and pediatric patients through body surface ECG electrodes, with obtained records helping users to analyze and diagnose heart disease. To be used in healthcare facilities by doctors and/or trained healthcare professionals.
    • Predicate Device: Digital Electrocardiographs, iE 12 (K122712)
    • Regulatory Basis for Acceptance: Substantial Equivalence (SE) to the predicate device, demonstrated through non-clinical tests verifying compliance with design specifications and international safety and EMC standards (IEC 60601-1, IEC 60601-1-2).
    • Key Comparison Points for SE: Intended use, technological characteristics, principles of operation, patient contact material, sterile/single use status, energy source, safety standards, and accessories are deemed equivalent to the predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131262

    Validate with FDA (Live)

    Date Cleared
    2013-11-14

    (195 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiographs, IE3/ IE12/ IE12P/ IE15, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiographs, IE3/ IE12/ IE12/ IE15, are designed to acquire, analyze, display and record ECG signals from patient body surface by ECG electrodes. After been amplified, filtered and analyzed, the ECG signals waveforms and analysis results are displayed in the LCD and recorded in the paper through thermal printer. ECG data, analysis result and patient information could be stored in the memory of the device.
    The device consists of three modules, which are power supply module, amplification module, and control module.
    The device has three recording mode, AUTO mode, MAN mode and RHY mode,
    They are standard twelve leads, including bipolar limb leads, augmented unipolar limb leads and unipolar chest leads.

    AI/ML Overview

    The provided 510(k) summary for K131262 does not contain information related to a clinical study or performance data against specific acceptance criteria for the device's diagnostic capabilities. Instead, it focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing and technical specifications.

    Therefore, many of the requested sections regarding acceptance criteria, study details, and performance metrics cannot be directly addressed from the provided document.

    Here's a breakdown of what can be extracted and what is missing:


    Description of Acceptance Criteria and Device Performance

    The submission focuses on establishing substantial equivalence to predicate devices (K122712 and K123816) based on technical specifications and compliance with international standards, rather than defining specific diagnostic acceptance criteria for the device's analytical function.

    The non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated compliance with the following standards:

    • IEC 60601-1:1990+ A1:1993+ A11:1993+ A12:1993+ A13:1996, Medical electrical equipment, Part 1: General requirements for safety.
    • IEC 60601-2-25:1993+A1:1999. Medical electrical equipment, Part 2-25: Particular requirements for the safety of electrocardiographs.
    • IEC 60601-1-2: 2007, Medical electrical equipment, Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.

    Table of Acceptance Criteria and Reported Device Performance (Based on Technical Specifications and Compliance to Standards):

    Acceptance Criteria (from Predicate/Standard Compliance)Reported Device Performance (K131262 Device)
    Intended Use (Similar to predicates)Acquires ECG signals, helps users analyze/diagnose heart disease
    Lead (Standard 12-lead)Standard 12-lead
    Acquisition mode (Simultaneous 12-lead acquisition)Simultaneous 12-lead acquisition
    Recording format (Automatic / Manual / Rhythm)Automatic / Manual / Rhythm
    Analysis mode (No integrated analysis)No (similar to predicates)
    CMRR (>60dB, >100 with AC filter)>60dB, >100 with AC filter
    Paper Speed (4 or 6 levels)4 levels (6.25, 12.5, 25, 50mm/s) OR 6 levels (5, 6.25, 10, 12.5, 25, 50mm/s)
    Input CIR current (<0.1μA)<0.1μA
    Input impedance (>50MΩ)>50MΩ
    Patient leak current (<10μA)<10μA
    Frequency response (0.05~150Hz)0.05~150Hz
    Noise level (<15μVpp)<15μVpp
    Electrical Safety (Comply with IEC 60601-1)Comply with IEC 60601-1
    EMC (Comply with IEC 60601-1-2)Comply with IEC 60601-1-2
    Particular requirements (Comply with IEC 60601-2-25)Comply with IEC 60601-2-25
    Biocompatibility (Comply with ISO 10993)Comply with ISO 10993

    Missing Information from the Provided Document:

    1. Sample size used for the test set and the data provenance: Not mentioned. The submission relies on non-clinical testing for compliance with standards and technical equivalency, not clinical diagnostic performance data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no diagnostic performance study with specific ground truth from experts is described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an electrocardiograph, which acquires and displays ECG signals for human interpretation, it does not include AI for interpretation or an automated analysis mode (as per "Analysis mode: No" in the comparison table).
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. The device's "Analysis mode" is listed as "No," indicating it does not perform automated algorithmic diagnosis. It is intended for healthcare professionals to analyze and diagnose heart disease from the acquired ECG records.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, as no diagnostic performance study with ground truth is presented. The "ground truth" here is compliance with established safety and performance standards for an ECG device.
    7. The sample size for the training set: Not applicable, as there is no mention of an algorithm or AI component that would require a training set.
    8. How the ground truth for the training set was established: Not applicable.

    Conclusion:

    This 510(k) summary is for a traditional Digital Electrocardiograph, which is a signal acquisition and display device, not an AI-powered diagnostic system. Its approval is based on demonstrating substantial equivalence to existing legally marketed devices through compliance with industry standards and shared technical specifications. The document does not include data from a clinical performance study measuring diagnostic accuracy against specific acceptance criteria for disease detection or classification by an algorithm. The device "can help users to analyze and diagnose heart disease," implying the diagnostic responsibility remains with the human healthcare professional.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123816

    Validate with FDA (Live)

    Date Cleared
    2013-02-27

    (77 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiographs, iE 12A / iE 15S, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiographs, iE 12A / iE 15 / iE 15S, are designed to acquire, display and record ECG signals from patient body surface by ECG electrodes. After been amplified and filtered, the ECG signals waveforms are displayed in the LCD and recorded in the paper through thermal printer. ECG data result and patient information could be stored in the memory of the device.

    All the models. iE 12A / iE 15 / iE 15S, of the proposed device. Digital Electrocardiographs, follow · the same design principle and similar technical specifications.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Digital Electrocardiographs (models iE 12A, iE 15, iE 15S). The submission focuses on demonstrating substantial equivalence to a predicate device (iE 12) rather than presenting a study with specific acceptance criteria related to clinical performance metrics like sensitivity, specificity, or reader improvement.

    Therefore, many of the requested details, such as specific acceptance criteria for diagnostic performance, sample sizes for test and training sets for an AI algorithm, expert qualifications, adjudication methods, or MRMC study results, are not available in this document. This document primarily details the technical specifications and safety/EMC standards met by the device.

    Here's a breakdown of what is available and what is not:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the sense of diagnostic performance metrics (e.g., sensitivity, specificity, AUC). Instead, it refers to verifiable technical specifications and compliance with international standards for safety and electrical performance. The "reported device performance" in this context refers to the device meeting these technical specifications and standards.

    ITEMAcceptance Criteria (Predicate iE 12)Proposed Device Performance (iE 12A / iE 15 / iE 15S)
    LeadStandard 12-leadStandard 12-lead (iE 12A), Standard 12-lead/15-lead (iE 15, iE 15S)
    Acquisition modeSimultaneous 12-lead acquisitionSimultaneous 12-lead acquisition (iE 12A), Simultaneous 12-lead/15-lead acquisition (iE 15, iE 15S)
    Recording formatAutomatic / Manual / RhythmAutomatic / Manual / Rhythm
    Frequency response0.05~150Hz0.05 Hz150 Hz (iE 12A), 0.05 Hz250 Hz (iE 15, iE 15S)
    Noise level<15μVp-p<15μVp-p
    CMRR>60dB, >100dB with AC filter>60dB, >100dB with AC filter
    Recording SpeedSix levels as 5, 6.25, 10, 12.5, 25, 50mm/sSix levels as 5, 6.25, 10, 12.5, 25, 50mm/s
    Input CIR current≤0.1μΑ≤0.1μΑ
    Input impedance>50ΜΩ>50ΜΩ
    External Input Impedance≥100k Ω≥100k Ω
    External Input Sensitivity10mm/V±5%10mm/V±5%
    External Output Impedance≤100 Ω≤100 Ω
    External Output Sensitivity1V/mV±5% (at 10mm/mV)1V/mV±5% (at 10mm/mV)

    Study Proving Acceptance Criteria Met:
    The document states: "Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: EN 60601-1, EN 60601-2-25, EN 60601-1-2, IEC60601-2-51."

    The primary "acceptance criterion" for this 510(k) submission is demonstrating substantial equivalence to the predicate device, which is verified by meeting technical specifications and compliance with relevant safety and performance standards. The differences in frequency response and lead type for the proposed devices compared to the predicate were addressed by conducting relative standard tests to demonstrate compliance.

    2. Sample size used for the test set and the data provenance

    Not applicable/Not provided within the scope of this document. This submission focuses on technical specifications and substantial equivalence, not a clinical study with a patient test set for diagnostic algorithm performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not provided. No diagnostic algorithm performance study is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable/Not provided. No diagnostic algorithm performance study is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document pertains to the hardware and basic signal acquisition capabilities of an electrocardiograph, not an AI-assisted diagnostic system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This document describes the ECG acquisition device itself, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable/Not provided. The "ground truth" here is compliance with engineering standards and technical specifications shown via bench testing.

    8. The sample size for the training set

    Not applicable/Not provided. No machine learning algorithm is described.

    9. How the ground truth for the training set was established

    Not applicable/Not provided. No machine learning algorithm is described.

    In summary: This 510(k) submission is for a traditional Digital Electrocardiograph hardware device. The "acceptance criteria" and "study" described are focused on engineering specifications, electrical safety, and electromagnetic compatibility (EMC) through bench testing, to demonstrate substantial equivalence to a predicate device. It does not involve a clinical study to evaluate diagnostic performance of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122712

    Validate with FDA (Live)

    Date Cleared
    2012-10-25

    (51 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Digital Electrocardiographs, ECG-8080/ iE 3S/ iE 12/ iE 12P, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.

    Device Description

    Digital Electrocardiographs, ECG-8080/ IE 3S/ iE 12/ iE 12P, are designed to acquire, display and record ECG signals from patient body surface by ECG electrodes. After been amplified and filtered, the ECG signals waveforms are displayed in the LCD and recorded in the paperthrough thermal printer. ECG data result and patient information could be stored in the memory of the device.

    All the models, ECG-8080/ iE 3S/ iE 12/ iE 12P, of the proposed device, Digital Electrocardiographs, follow the same design principle and similar technical specifications.

    AI/ML Overview

    The provided document is a 510(k) summary for Digital Electrocardiographs. It focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria, study details, sample sizes for test and training sets, expert qualifications, or ground truth establishment methods.

    The document indicates "Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device." However, it does not provide specific performance criteria or results from these bench tests beyond compliance with general safety and EMC standards (IEC 60601-1, IEC 60601-2-25, IEC 60601-1-2).

    Therefore, I cannot populate the table or answer the specific questions about the study that proves the device meets acceptance criteria based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2