Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K965071

    Validate with FDA (Live)

    Date Cleared
    1998-04-21

    (488 days)

    Product Code
    Regulation Number
    882.5950
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Detachable Silicone Balloon (DSB) is intended for artificial embolization of the internal carotid artery in the presence of certain intracranial aneurysms in patients for whom, in the judgment of the neurosurgical management team, other medical or neurosurgical means would not be indicated.

    Device Description

    The DSB consists of a silicone balloon which is inflated with contrast agent and detached via catheter at the desired site of embolization.

    AI/ML Overview

    This is not an AI/ML device, therefore, the mentioned acceptance criteria and studies are not applicable.

    The Detachable Silicone Balloon (DSB) is a medical device for artificial embolization. Device testing included biocompatibility testing and functional bench testing. Functional bench testing included cycle testing, rupture testing, pressure-volume testing, detachment force testing, inflation-deflation cycle rate testing, and catheter insertion valve testing. The results of these tests demonstrated the DSB to be safe for its intended use.

    This device was determined to be substantially equivalent to predicate devices based on intended use, materials, manufacturing, design parameters, method of delivery, and method of deployment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1