510(k) Data Aggregation

    K Number
    K970241

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-08-14

    (204 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DePuy Proximal Femoral Replacement Prosthesis is indicated for uncemented or cemented use as the femoral component in total hip arthroplasty for replacing the hip joint in the following indications, as appropriate: 1) a severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, 2) avascular necrosis of the femoral head, 3) acute traumatic fracture of the femoral head or neck, 4) failed previous surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or other total hip replacement, and 5) certain cases of ankylosis.

    Device Description

    The DePuy Proximal Femoral Replacement Prosthesis is indicated for uncemented and cemented use as the femoral component in total hip arthroplasty for replacing the hip joint in the following indications, as appropriate: 1) a severely painful and/or a severely disabled joint resulting from osteoarthritis, or theumatoid arthritis, 2) avascular necrosis of the femoral head, 3) acute traumatic fracture of the femoral head or neck, 4) failed previous surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or other total hip replacement, and 5) certain cases of ankylosis. It is a porous coated (except for the tapered distal tip), straight, cast Co-Cr-Mo stem available in three femoral sizes. The stem has a driving platform, a self-locking 12/14, 0540" taper for attachment of interchangeable modular femoral heads.

    AI/ML Overview

    This 510(k) premarket notification is for the DePuy Proximal Femoral Replacement Prosthesis, a hip prosthesis.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    It's important to note that this 510(k) summary does not present acceptance criteria or quantitative performance data in the way you might expect for a modern AI/device study. The basis of clearance for this device in 1997 is substantial equivalence to previously cleared predicate devices, not performance against predefined acceptance criteria. Therefore, the "acceptance criteria" here are implicitly the characteristics and intended use of the legally marketed predicate devices.

    Acceptance Criteria (Implicit from Predicate Devices)Reported Device Performance (Characteristics of DePuy PFR Prosthesis)
    Intended Use: - Femoral component in total hip arthroplasty - Replacing hip joint for: osteoarthritis, rheumatoid arthritis, avascular necrosis, traumatic fracture, failed previous surgery, ankylosisIntended Use: - Femoral component in total hip arthroplasty - Replacing hip joint for: osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis of femoral head, acute traumatic fracture of femoral head or neck, failed previous surgery (joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, other total hip replacement), certain cases of ankylosis. - Uncemented and cemented use (similar to Solution Hip Prosthesis).
    Material: - Co-Cr-Mo alloy (for Porocoat Proximal Femoral Third Prosthesis and Solution Hip Prosthesis) or Ti-6Al-4V (for Proximal Femur Replacement Stem).Material: - Cast Co-Cr-Mo stem (identical to Porocoat Proximal Femoral Third Prosthesis and Solution Hip Prosthesis).
    Porous Coating Placement: - Various placements on predicate devices (e.g., total stem except distal tip, distal stem, entire stem).Porous Coating Placement: - Porous coated (except for the tapered distal tip) - identical to Solution Hip Prosthesis.
    Distal Tip Design: - "Bullet" tapered (Porocoat Proximal Femoral Third, Solution Hip) or "rounded" non-tapered (Proximal Femur Replacement Stem).Distal Tip Design: - "Bullet" tapered distal tip (similar to Porocoat Proximal Femoral Third and Solution Hip Prosthesis).
    Collar/Stem Feature: - Collarless, mid-stem "step", or collar.Collar/Stem Feature: - Collarless (similar to Proximal Femur Replacement Stem).
    Modularity: - Designed to accept modular femoral heads or non-modular.Modularity: - Designed to accept modular femoral heads (similar to Proximal Femur Replacement Stem and Solution Hip Prosthesis). Specifically, self-locking 12/14, 0540" taper for attachment of interchangeable modular femoral heads.
    General Design: - Basic design (e.g., 12/14 taper) (of Proximal Femur Replacement Stem).General Design: - Basic design with a 12/14 taper (unchanged from Proximal Femur Replacement Stem cleared in 1990).
    Manufacturing Process:Manufacturing Process: - Unchanged from Proximal Femur Replacement Stem cleared by FDA in 1990.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as this is a 510(k) submission based on substantial equivalence, primarily comparing design and material specifications to predicate devices. There is no mention of a "test set" in the context of clinical performance data in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Substantial equivalence claims typically do not involve establishing ground truth from a test set with expert adjudicators. The "ground truth" for the clearance is essentially the regulatory status and established safety/effectiveness of the predicate devices.

    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring adjudication is mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical hip prosthesis, not an AI or imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic or AI device.

    7. The Type of Ground Truth Used

    The "ground truth" for this 510(k) clearance is the established safety and effectiveness of the legally marketed predicate devices, as determined by the FDA. The submission demonstrates that the new device is "substantially equivalent" to these predicates.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set involved as this is a physical medical device clearance based on substantial equivalence to existing products, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set is mentioned in this submission.