Found 1 results

510(k) Data Aggregation

    K Number
    K192855

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-24

    (143 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DELTA XTEND Prosthesis is indicated for use in treatment of a grossly deficient rotator cuff joint with:
    • severe arthropathy and/or;
    • a previous failed joint replacement and/or;
    • Fracture-dislocations of the proximal humerus where the articular surface is severely communited, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory
    The patient's joint must be anatomically suited to receive the selected implant(s), and a functional deltoid muscle is necessary to use the device.
    DELTA XTEND hemi-shoulder replacement is also indicated for hemi-arthroplasty if the glenoid is fractured intraoperatively or for the revision of a previously failed DELTA XTEND Reverse Shoulder. Porous-coated epiphysis are indicated for use in total shoulder replacement only.
    The metaglene component is HA coated and is intended for cementless use with the addition of screws for fixation. The modular humeral stem is HA coated and is intended for cementless use. The HA coated humeral epiphysis is intended for cementless use. The porous-coated epiphysis is intended for cemented or cementless use.
    All other metallic components are intended for cemented use only.

    Device Description

    The DELTA XTEND Reverse Shoulder System consists of humeral stem, modular epiphysis, humeral spacer, humeral cup, glenosphere, metaglene and metaglene screws used for reverse shoulder arthroplasty. The humeral spacer can be added between the epiphysis and the humeral cup if necessary. Humeral head can be used in hemi-shoulder arthroplasty in place of the humeral cup and glenoid components.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the DELTA XTEND™ Reverse Shoulder System, a medical device (shoulder prosthesis). It describes the non-clinical performance testing conducted to demonstrate substantial equivalence to legally marketed predicate devices.

    Therefore, the requested information about "acceptance criteria and the study that proves the device meets the acceptance criteria" refers to engineering and material performance tests, not a study involving human-in-the-loop AI or diagnostic performance.

    Here's an analysis of the provided text in relation to your questions, framed for a medical device rather than an AI/diagnostic system:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of numerical acceptance criteria with corresponding performance values for the device. Instead, it lists the types of non-clinical tests performed to demonstrate safety and efficacy and asserts that the device was "determined as substantially equivalent."

    Here's what can be inferred/extracted:

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Fatigue AnalysisDevice withstands fatigue loads representative of its intended use and service life (comparable to predicate devices).Met requirements, supported substantial equivalence.
    Tolerance AnalysisManufacturing tolerances ensure proper fit, function, and interchangeability of components within design specifications.Met requirements, supported substantial equivalence.
    Epiphysis Comparison andDesign and material characteristics suitable for intended use and comparable to predicate devices.Met requirements, supported substantial equivalence.
    Design Justification
    Range of Motion AnalysisDevice allows for a functional range of motion, comparable to predicate devices and physiological limits.Met requirements, supported substantial equivalence.
    Biocompatibility StudyMaterials are biocompatible and do not elicit adverse biological responses (e.g., cytotoxicity, sensitization, irritation).Materials are biocompatible, supported substantial equivalence.
    Bacterial Endotoxin TestingMeets specified limits for bacterial endotoxin (ANSI AAMI ST-72:2011).The proposed devices meet the requirement of bacterial endotoxin testing.
    MRI Compatibility TestingDevice is safe for use in an MRI environment, indicating specific field strengths, SAR limits, and artifact characteristics.Testing conducted in support of adding MRI compatibility language to labeling.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Sizes: The document does not specify the sample sizes (e.g., number of implants or test specimens) used for each non-clinical test (Fatigue Analysis, Biocompatibility, etc.).
    • Data Provenance: The data provenance is from non-clinical laboratory testing of the device components. The specific country of origin of these tests is not stated, but the manufacturer "DePuy (Ireland)" is based in Ireland. These are prospective tests performed on newly manufactured devices or components to demonstrate performance characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this type of device submission.

    • Ground Truth: For a shoulder prosthesis, "ground truth" for performance is established through engineering principles, material science, and regulatory standards (e.g., ISO, ASTM standards for implant testing). It's based on objective measurements rather than expert consensus on individual cases.
    • Experts: While engineers, material scientists, and regulatory experts would have designed and overseen these tests, the document does not specify their number or qualifications as it would for, say, a clinical diagnostic study.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable to this type of device. Adjudication methods like 2+1 or 3+1 are used in clinical studies where human readers (e.g., radiologists) interpret images and their interpretations need to be reconciled for ground truth establishment. For mechanical device testing, results are objective measurements from laboratory equipment and are compared against predetermined engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is relevant for AI-powered diagnostic devices where the performance of human readers, potentially aided by AI, is being evaluated. This submission is for a mechanical medical implant, not a diagnostic imaging device or AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. There is no algorithm in this medical device. The "standalone performance" for this device would be its mechanical performance in a laboratory setting, which is what the non-clinical tests assessed.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this medical device's performance is based on:

    • Engineering and material science principles: Adherence to established mechanical, physical, and chemical properties required for an implantable device.
    • Regulatory standards: Compliance with international and national standards for implantable medical devices (e.g., for fatigue, biocompatibility, endotoxin levels).
    • Comparison to predicate devices: The "ground truth" of performance is largely the established safe and effective performance of previously cleared, substantially equivalent devices.

    8. The sample size for the training set

    This question is not applicable. This device is a mechanical implant; there is no "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no "training set."