Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K222827

    Validate with FDA (Live)

    Date Cleared
    2022-10-20

    (31 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upon intravenous administration and use of an ICG drug product consistent with its approved label, the da Vinci Firefly Imaging System is intended to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The da Vinci Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near infrared imaging.

    Fluorescence imaging of biliary ducts with the da Vinci Firefly Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

    Upon interstitial administration and use of an ICG drug product consistent with its approved label, the da Vinci Firefly Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    The da Vinci Firefly Imaging System provides visualization in two modes: standard, visible-light imaging mode and a near-infrared fluorescence imaging mode, consisting of either a black-and-white surgical image or completely black background with the nearinfrared fluorescence depicted in a color overlay. For near-infrared fluorescence imaging, indocyanine green (ICG) fluorescence imaging agent is required.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    785nm Fluorescence System PerformanceMet the acceptance criteria for all design requirements and specifications, specifically for Firefly Clinical Applications test cases.
    Safety and Effectiveness (Change Impact)No issues of safety or effectiveness and no new risks were identified with the change to the 785nm Fluorescence System.
    Illumination Requirements (Firefly Mode)Verification testing (limited to illumination requirements in Firefly Mode) demonstrated compliance.
    Compatibility with EndoscopesThe 785nm Fluorescence Vision System and applicable endoscopes met compatibility acceptance criteria. No issues of safety or effectiveness and no new risks were identified regarding compatibility.
    White Light PerformanceNot applicable; white light LED and components are unchanged from the previously cleared 805nm system, so no new testing was required or reported for white light performance in this particular submission.
    Substantial Equivalence (Near-Infrared)The modified device, using an excitation wavelength of 785 nm, maintains substantially equivalent near-infrared imaging compared to the predicate device (805nm laser).
    Substantial Equivalence (Imaging Modes/Energy)In terms of imaging modes and energy used and/or delivered, the subject device is substantially equivalent to the predicate devices.
    LabelingLabeling was updated to reflect the 785nm laser change and is substantially equivalent to the predicate device's labeling.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample size for the test set used in the "Delta validation testing" or "Delta design verification testing."

    The data provenance is also not explicitly detailed. It can be inferred that the testing was conducted internally by Intuitive Surgical, Inc. as part of their design verification and validation processes. There is no mention of country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    The provided text does not provide information regarding the number of experts used or their specific qualifications for establishing ground truth within these tests. The tests are described as "validation testing" and "design verification testing" against "design requirements and specifications" and "clinical applications test cases," suggesting an internal product-centric validation rather than a clinical ground truth established by external experts in an MRMC study.

    4. Adjudication Method for the Test Set

    The provided text does not mention any adjudication method like 2+1 or 3+1. The description of testing ("met the acceptance criteria for all of the design requirements and specifications") implies a direct assessment against predefined engineering or performance metrics rather than a consensus-based review of clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or reported in this document. The device is an imaging system (da Vinci Firefly Imaging System), not an AI-assisted diagnostic tool that directly aids human readers in interpretation. The filing describes a modification to an existing imaging system (change in excitation wavelength from 805nm to 785nm), and the testing focused on verifying the performance of this modified system against its design specifications and maintaining substantial equivalence. Therefore, there is no discussion of human reader improvement with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The device is an imaging system used during surgery, and its functionality involves providing real-time visible and near-infrared fluorescence imaging to surgeons. It is inherently a human-in-the-loop device as it provides visual information for a surgeon to interpret and act upon. The text does not describe any standalone algorithmic performance without human interpretation.

    7. The Type of Ground Truth Used

    The "ground truth" for the testing described appears to be based on engineering design requirements and specifications for the imaging system's performance, particularly related to the 785nm fluorescence system, illumination, and compatibility. It also references "Firefly Clinical Applications test cases," which suggests scenarios designed to evaluate the system's performance in clinical contexts, likely validated against anticipated imaging results or known technical benchmarks. It does not mention expert consensus, pathology, or outcomes data as the primary ground truth for this particular submission.

    8. The Sample Size for the Training Set

    The provided text does not mention a training set because this submission describes a modification to a medical device (an imaging system, specifically a change in laser wavelength) and not an AI/ML algorithm that would typically require a training set. The testing performed was for verification and validation of the hardware change.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a training set, the method for establishing its ground truth is not applicable to this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141077

    Validate with FDA (Live)

    Date Cleared
    2014-08-12

    (109 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The da Vincil Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci Firefly Imaging System enables surgeous to perform minimally invasive suggedy using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion. and at least one of the major extra-hepatic bile duct. common bile duct or common hepatic duct), wing near infrared imaging.

    Fluorescence imaging of billiary ducts with the da Vinci Firefly Imaging System is intended of care white light and when indicated intraoperady. The device is not intended for standalone use for bliany duct visualization

    Device Description

    The Intuitive Surgical da Vinci® Firefly™ Imaging System uses the existing endoscopic imaging system as submitted in K131861 (cleared March 28, 2014) for high definition (HD) visible light and near-infrared fluorescence imaging during minimally invasive surgery. The da Vincie Firefly™ Imaging System utilizes the following existing components of the da Vinci Xi Surgical System:

    • 8 mm 0° and 30° endoscopes (PNs 470026 and 470027) .
    • Endoscope Controller (PN 372601) .

    For near-infrared fluorescence imaging, the Fluorescence Imaging Kit is also required. The kits will be provided to the end user in an identical manner to the current supply and distribution chain for the predicate device. The kit is unchanged from K101077 (February 4, 2011)/K124031 (September 13, 2013).

    • Fluorescence Imaging Kit (PN 950156) [Includes IndoCyanine Green (ICG) fluorescence . imaging agent, aqueous solvent, and syringe trays]
    AI/ML Overview

    This document (K141077) describes the da Vinci® Firefly™ Imaging System, which is intended for real-time endoscopic visible and near-infrared fluorescence imaging during minimally invasive surgery. It allows for visual assessment of vessels, blood flow, related tissue perfusion, and major extra-hepatic bile ducts using near-infrared imaging. It is intended for use with standard white light and intraoperative cholangiography, but not for standalone use for biliary duct visualization.

    Here's a breakdown of the acceptance criteria and supporting studies as presented in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics like sensitivity/specificity/accuracy for the imaging system's clinical performance. Instead, it focuses on demonstrating that the device meets safety and effectiveness requirements through various tests. The acceptance criteria are implicitly that the device performs as intended, meets design requirements, and is substantially equivalent to predicate devices.

    Test CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Bench TestingDevice meets dimensional, mechanical, functional, and electrical requirements and specifications. Objective pass/fail criteria were defined.All final tests PASSED. Verified optical, illumination, mechanical, environmental, and labeling aspects.
    Animal Testing (Design Validation)Device meets user needs and intended use as documented in product requirements. Objective pass/fail criteria were defined.All final tests PASSED. Evaluated Firefly mode performance characteristics in various surgical tasks.
    Animal Testing (Device Comparison)Performance is comparable to the predicate device (IS3000 Firefly mode) in basic clinical function and fluorescence image quality.All final tests PASSED. Side-by-side comparison showed comparable fluorescence image quality.
    Animal Testing (Surgeon Evaluation)Clinically acceptable performance allowing for safe and effective surgical use, as assessed by independent external surgeons.All final tests PASSED. Evaluators assessed various Firefly vision parameters in at least one application.
    Human Factors and Usability TestingSafe and effective use by intended users in the intended use environment, addressing critical tasks and use-related risks.Summative usability validation study conducted with 16 teams. Results provided evidence of safe and effective use.

    2. Sample size used for the test set and the data provenance:

    • Bench Testing: Sample sizes "up to 5 units" were used for design verification tests (Optical, Illumination, Mechanical, Environmental, Labeling). Data provenance is not explicitly stated but implies internal laboratory testing.
    • Animal Testing:
      • Design Validation: Two labs were used, with one canine and one porcine.
      • Device Comparison: Two porcines were used.
      • Surgeon Evaluation: Four labs were used, utilizing canines or porcines.
      • Data provenance is from animal models (canine and porcine) in a simulated clinical setting. This is prospective testing.
    • Human Factors and Usability Testing: 16 teams of users (surgeons and operating room staff) were involved. The study was conducted in a simulated operating room, which is a prospective test setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document mentions "independent, external surgeon evaluators" were used for the Animal Testing (Surgeon Evaluation) section. Four such surgeons served as evaluators.
    • Their specific qualifications (e.g., years of experience) are not detailed other than being "independent, external surgeons".
    • For bench and animal testing, "objective pass/fail criteria were defined in the protocol and used," implying internal expert definition, but specific numbers and qualifications are not provided.
    • For human factors, 16 "teams of users (surgeons and operating room staff)" participated. Their specific qualifications are not detailed beyond their roles.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document does not explicitly describe an adjudication method like 2+1 or 3+1 for any of the testing.
    • For the surgeon evaluations, it states "All evaluators completed at least one Firefly application to assess the various Firefly vision parameters." This suggests individual assessments, but no process for resolving discrepancies among the four evaluators is detailed. For other tests, "objective pass/fail criteria" were defined, implying a predetermined standard rather than expert adjudication of subjective outcomes.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving human readers with AI assistance versus without AI assistance was not mentioned or performed.
    • The device described is an imaging system, not an AI diagnostic tool. The "Firefly" system provides near-infrared fluorescence imaging. The "Device Comparison" in animal testing was a side-by-side comparison of image quality between the new device and a predicate device, not a comparison of human reader performance with and without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The device is an imaging system used by surgeons in real-time. It is not an algorithm-only device that operates without human interaction.
    • The product description explicitly states, "The device is not intended for standalone use for biliary duct visualization," meaning it's always used with human surgical intervention and standard white light. Therefore, no standalone (algorithm-only) performance testing would be relevant or performed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Bench Testing: Ground truth was based on pre-defined technical specifications and objective pass/fail criteria for dimensional, mechanical, functional, and electrical requirements.
    • Animal Testing (Design Validation, Device Comparison, Surgeon Evaluation): Ground truth was based on:
      • Pre-defined test procedures and objective pass/fail criteria.
      • User needs and intended use documented in product requirements.
      • Comparison to the predicate device (IS3000 Firefly mode) for image quality.
      • Assessments by independent, external surgeon evaluators regarding clinically acceptable performance.
      • While not explicitly stated as "pathology" for, e.g., identifying bile ducts, the clinical assessment by surgeons in an animal model serves as the practical ground truth for the imaging system's utility in visualizing these structures in a surgical context.
    • Human Factors and Usability Testing: Ground truth was based on whether the system allowed for "safe and effective use" during "typical workflow scenarios" and "troubleshooting scenarios related to safety-critical tasks" in a simulated environment, as assessed by observing user teams.

    8. The sample size for the training set:

    • The document describes performance testing for a medical device (an imaging system), not an AI/ML algorithm that requires a "training set." Therefore, the concept of a training set sample size is not applicable to this submission. The tests performed are for design verification and validation of hardware and integrated software.

    9. How the ground truth for the training set was established:

    • As noted above, this device does not involve a "training set" in the context of AI/ML. All ground truth establishment described (e.g., objective pass/fail criteria, surgeon assessments) pertains to testing and validation, not to the training of a machine learning model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1