Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K973757

    Validate with FDA (Live)

    Date Cleared
    1997-11-07

    (36 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.O.R.C. SOLUX Light Source is indicated for intraocular illumination in vitreoretinal surgery.

    Device Description

    The D.O.R.C. SOLUX Light Source uses a halogen lamp and an internal focusing system to focus the light into the end of the optical fiber. Accessories to the Illumination Unit include adapters and color filters.

    AI/ML Overview

    This 510(k) summary (K973757) describes the D.O.R.C. SOLUX Light Source, an ophthalmic light source intended for intraocular illumination in vitreoretinal surgery. The submission establishes substantial equivalence to existing predicate devices (Grieshaber Light Source and Escalon VitLite I) primarily based on operational and technological characteristics, rather than clinical performance studies with acceptance criteria in the typical sense of AI/software devices.

    Here's an analysis based on the provided document, addressing your specific points:

    #Information RequestedDetails from K973757
    1Table of acceptance criteria and reported device performanceAcceptance Criteria (Implied by Predicate Comparison): The device (SOLUX Light Source) must have technological and operational characteristics similar to legally marketed predicate devices, particularly in areas like intended use (intraocular illumination for vitreoretinal surgery), lamp type (Halogen), variable intensity, UV filtration, and ability to adapt to multiple accessories. While specific numerical "acceptance criteria" are not explicitly stated for each characteristic in the way one would for a diagnostic accuracy study, the implicitly accepted performance is that which is "substantially equivalent" to the predicate devices. Reported Device Performance (SOLUX Light Source):- Indication: Endoillumination for vitreoretinal surgery (YES)- Lamp Type: Halogen- Lamp Rating: 100 watts- Light Output (lumens): High: 3800, Low: 3200- Color Temperature (degrees K): High: 3220, Low: 3160- Variable Intensity: Yes- Wavelength Range (nm): 400-800- UV filtration: Yes- Additional color filtration: Green, Yellow, Red, Blue, Daylight (all Yes)- Adapts to multiple accessories: Yes
    2Sample size for the test set and data provenanceThis is a conventional medical device (light source), not a software/AI device. Therefore, there is no "test set" or "data provenance" in the context of diagnostic performance evaluation. The substantial equivalence is based on engineering specifications and comparison to predicate devices, not on data from patient cases.
    3Number of experts to establish ground truth for test set and qualificationsNot applicable. No test set for ground truth establishment.
    4Adjudication method for the test setNot applicable. No test set.
    5MRMC comparative effectiveness studyNot applicable. This is a light source, not a diagnostic AI system where human readers would improve with AI assistance.
    6Standalone (algorithm only) performance studyNot applicable. This is a hardware device, not an algorithm. Standalone performance for such a device would refer to its electrical, optical, and mechanical specifications as an engineering product, which were implicitly demonstrated or compared to predicates to establish substantial equivalence.
    7Type of ground truth usedNot applicable. The "ground truth" for this type of device is its physical specifications and functionality, which are compared against known capabilities of predicate devices. There is no diagnostic "ground truth" from pathology, expert consensus, or outcomes data in this context.
    8Sample size for the training setNot applicable. This is not an AI/ML device. There is no "training set."
    9How the ground truth for the training set was establishedNot applicable. No training set.

    Summary of the Study and Acceptance Criteria:

    The "study" in this context is a 510(k) premarket notification which aims to demonstrate substantial equivalence to legally marketed predicate devices. The primary method is a comparison of technological and operational characteristics.

    • Acceptance Criteria (Implied): The D.O.R.C. SOLUX Light Source is accepted if its technical specifications and intended use are similar enough to the predicate devices (Grieshaber Light Source and Escalon VitLite I) that it can be considered equally safe and effective. The comparison table (page 1) serves as the core of this demonstration. For instance, having "YES" for "Indication: Endoillumination for vitreoretinal surgery" for all devices shows alignment in intended use. The fact that its lamp type is Halogen, like the Grieshaber, and its wavelength range (400-800 nm) encompasses that of the Escalon VitLite (450-700 nm), along with features like UV and color filtration, contribute to this claim of equivalence.

    • Proof of Meeting Criteria: The provided table titled "Comparison of SOLUX Light Source to Predicate Devices" directly shows how the D.O.R.C. SOLUX Light Source's characteristics align with or are similar to the predicate devices across various attributes like lamp type, variable intensity, filtration, and intended use. The FDA's letter (page 2-3) confirming "substantial equivalence" is the official recognition that the device meets these implied acceptance criteria based on the information provided in the 510(k) submission.

    This 510(k) is characteristic of traditional medical device clearances based on engineering comparisons rather than clinical performance data typical of AI/ML-powered diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1