Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K214009

    Validate with FDA (Live)

    Date Cleared
    2023-04-05

    (516 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The D.Kaptain™ PTA High Pressure Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, tibial, peroneal, subclavian, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in the coronary vasculature.

    Device Description

    The D·Kaptain™ PTA High Pressure Balloon Dilatation Catheter is an over-the-wire balloon catheter with a dual lumen shaft design. One lumen is used to pass the catheter over 0.035" guidewires. The second lumen communicates with the balloon and is used to inflate and deflate the balloon during the procedure. The guidewire lumen and the balloon lumen terminate at the proximal end of the catheter in a Y-connector manifold with luer lock fillings. There are radiopaque marker bands located under the balloon to aid in positioning the system during the procedure. A PVP coating is applied to the tip to enhance insertion and withdrawal performance. The D.Kaptain™ PTA High Pressure Balloon Dilatation Catheter will be available with balloon diameters 3.0 mm to 12.0 mm, balloon lengths 20 mm to 150 mm and with shaft lengths of 50 cm, 75 cm, and 130 cm.

    AI/ML Overview

    This is a medical device submission, specifically for a balloon dilatation catheter, not an AI/ML powered device. As such, the typical acceptance criteria and study designs for AI/ML devices (e.g., test set sample size, data provenance, expert adjudication, MRMC studies, standalone performance) are not applicable here.

    The document describes the performance criteria for the physical medical device and the studies conducted to demonstrate that the device meets those criteria, ensuring its substantial equivalence to a legally marketed predicate device.

    Here's the information extracted based on your request, adapted for a medical device:

    1. A table of acceptance criteria and the reported device performance

    The document lists various in-vitro performance tests that were conducted and states that "The test results met all acceptance criteria" and "ensure the design and construction are suitable for its intended use." Specific numerical acceptance criteria are not detailed in this summary, but the successful outcome is reported. Some of these performance characteristics are implied comparisons to the predicate device, where "SE" (Substantial Equivalence) indicates the performance is considered comparable.

    Acceptance Criteria (Implied / Tested)Reported Device Performance
    In vitro performance tests:All met acceptance criteria and ensure design suitability.
    Visual inspectionMet
    Dimensional verificationMet
    Crossing profileMet
    Guidewire compatibilityMet
    Simulated useMet
    Air/liquid leakMet
    Flexibility and kinkMet
    Torque strengthMet
    Particulate releaseMet
    Coating integrityMet
    LubricityMet
    Balloon rated burst pressureMet
    Balloon fatigueMet
    Balloon complianceMet
    Balloon inflation and deflation timeMet
    Catheter bond strengthMet
    Tip pull strengthMet
    TrackabilityMet
    RadiopacityMet
    Biocompatibility testing:Met all acceptance criteria; Device is biocompatible.
    CytotoxicityMet
    Sensitization testMet
    Intracutaneous reactivity testMet
    Systemic Toxicity test (Acute)Met
    Material-mediated pyrogenicityMet
    Hemocompatibility (hemolysis, complement activation, PTT, in vivo thrombo resistance)Met
    Sterility:Sterile with SAL of 10-6 via EO sterilization.
    Packaging/Shelf Life:Shelf life of 2 years.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses in-vitro testing for the device. It does not mention a "test set" in the context of clinical data for AI/ML. The performance data is derived from laboratory tests on the physical device prototypes. No mention of country of origin or retrospective/prospective nature is given for these laboratory tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This concept is not applicable to the evaluation of a physical medical device like a balloon catheter. "Ground truth" established by experts is relevant for diagnostic AI/ML models.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This concept is not applicable to the evaluation of a physical medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or mentioned, as this is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to AI/ML algorithm performance. The D.Kaptain™ PTA High Pressure Balloon Dilatation Catheter is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the physical performance tests, the "ground truth" would be established engineering specifications, industry standards (like ISO 10993-1), and FDA guidance for Class II Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submission. Device performance against these established criteria forms the basis of proving suitability.

    8. The sample size for the training set

    This is not applicable to a physical medical device.

    9. How the ground truth for the training set was established

    This is not applicable to a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1