510(k) Data Aggregation

    K Number
    K173363

    Validate with FDA (Live)

    Date Cleared
    2018-01-24

    (90 days)

    Product Code
    Regulation Number
    876.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cycler Drain Bag Set is indicated for use by patients with acute and stage renal disease undergoing peritoneal dialysis (PD) in a healthcare facility or at home. The Cycler Drain Bag Set is an optional receptacle that connects to the Luer-lock end of a Cycler Set drain line in order to collect patient effluent during PD treatments. This device is to be used only with Fresenius Medical Care (FMCNA) Cyclers and Cycler Sets.

    Device Description

    The Drain Bag Set (tubings and flexible bags) is a passive, closed drainage system used as an optional receptacle during an APD treatment. The Drain Bag Set is used to collect effluent in bags rather than letting the effluent flow directly to a drain. The Drain Bag Set connects to the drain line of a cycler set to collect patient effluent by means of gravity and the pumping action of a Peritoneal Dialysis (PD) cycler. The Drain Bag Set connects to the drain line using a standard Luer lock connection. The Drain Bag Set consists of three (3) interconnected 7-liter drain bags, each with a sampling port and an occluding clamp. The tubing lines that connect each drain bag to the drain line have a snap-disconnect tubing segment. This tubing segment allows the user to disconnect individual bags by snapping the tubing apart.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for the Cycler Drain Bag Set (K173363), a medical device used in peritoneal dialysis.

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ConductedAcceptance CriteriaReported Device Performance
    Drop TestThe drain bag shall not leak (visual inspection). Acceptance criteria from ISO 15747:2010 were applied.Pass, results within acceptance criteria
    Internal Pressure TestThe drain bag shall not leak (visual inspection). Acceptance criteria from ISO 15747:2010 were applied.Pass, results within acceptance criteria
    Hanger Holes Tensile StrengthThe drain bag hanger holes shall withstand a tensile load of 15 N for 60 minutes. Acceptance criteria from ISO 15747:2010 were applied.Pass, results within acceptance criteria
    Film Roughness TestThe drain bag's film shall have a roughness ≤ 0.75 µm (Roughness Average).Pass, results within acceptance criteria
    Tensile Strength of BondingAll bonded engagements between plastic composites shall require >10 lbf to detach.Pass, results within acceptance criteria
    Maintenance of Sterility testingAcceptance criteria from ISO 11607-1:2006/(R)2010 - A1:2014 were applied.Pass, results within acceptance criteria
    Male Luer lock connector testingAcceptance criteria from ISO 594-2:1998 were applied.Pass, results within acceptance criteria
    Drain Bag Set CapacityThe Drain Bag Set shall hold at least 24 L of water (volume ≥ 24 L).Pass, results within acceptance criteria
    Sampling Port Leakage TestThe sampling port shall not leak after being punctured with a 23 gauge needle one time for 15 seconds and subjected to a pressure of 20 kPa under water for 15 seconds. Acceptance criteria from ISO 15747:2010 were applied.Pass, results within acceptance criteria
    Snap Connector Breakage TestThe snap connector shall have a break point between 60 oz-in and 120 oz-in.Pass, results within acceptance criteria
    Clamp TestNo leaks past the clamp. Acceptance criteria from ISO 8638:2010 were applied.Pass, results within acceptance criteria
    Vibratory (ship) TestingNo tubing kinks or loose or damaged components.Pass, results within acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each individual test. It states "The samples" were used for various tests (e.g., sampling port leakage test) but doesn't quantify how many samples.

    The data provenance is from in-house laboratory testing performed by Fresenius Medical Care. It is not externally sourced and is not described as retrospective or prospective clinical data. The tests performed are primarily engineering and material science tests according to established ISO and ASTM standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and therefore not provided in the document. The tests performed are objective, quantitative engineering and material assessments against pre-defined industry standards (e.g., ISO, ASTM). They do not involve expert interpretation or clinical ground truth determination in the way a diagnostic AI device would.

    4. Adjudication Method for the Test Set

    This information is not applicable. Since the tests are objective physical and performance assessments against documented standards, there is no need for an adjudication method by experts. The results are "Pass" or "Fail" based on whether the measured values meet the specified criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance

    This information is not applicable. The device is a physical medical accessory (drain bag set) and does not involve AI or human reader interpretation for its function. Therefore, no MRMC comparative effectiveness study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. As mentioned, the device is a physical accessory and does not contain any algorithms or software.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests conducted is based on established international and industry standards (ISO 15747:2010, ISO 11607-1:2006/(R)2010 - A1:2014, ISO 594-2:1998, ISO 8638:2010, ASTM D4169-14) and pre-defined engineering specifications (e.g., film roughness ≤ 0.75 µm, tensile strength >10 lbf, break point between 60 oz-in and 120 oz-in, capacity ≥ 24 L). These are objective, measurable criteria.

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical device, and the testing described here is for performance verification. There is no "training set" in the context of machine learning or AI algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as point 8.