510(k) Data Aggregation

    K Number
    K183601

    Validate with FDA (Live)

    Device Name
    CryoTouch
    Manufacturer
    Date Cleared
    2019-04-18

    (113 days)

    Product Code
    Regulation Number
    878.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CryoTouch™ is intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures.

    The list below shows examples of the types of lesions that may be treated:

    • Verruca (warts) ●
    • Plantar warts
    • Seborrheic Keratoses
    • Actinic Keratoses
    • Achrochordon ●
    • Molluscum Contagiosum ●
    • Age Spots
    • Dermafibroma
    • Small Keloids ●
    • Granuloma Annulare ●
    • Porokeratosis Plantaris ●
    • Angiomas
    • Lentigo Maligna
    • Keratoacanthoma
    • Basal Cell Carcinomas ●
    • Bowen's Disease
    • Lentigo Discrete
    • Chondrodermatitis
    • Epithelial Nevus
    • Leukoplakia ●
    • Granuloma Pyogenicum ●
    • Kaposi's Sarcoma ●
    • o Pyogenic Granuloma
    Device Description

    The CryoTouch™ device gently dispenses cryogen gas into an applicator or a uniquely designed cone. The applicator or cone is then used to apply the extreme cold to the lesion being treated to effect cryosurgical treatment.

    AI/ML Overview

    The CryoTouch™ device is a cryosurgical unit intended for the destruction of various lesions by applying cryogenic gases at low temperatures. The submission aims to demonstrate substantial equivalence to the predicate device, Verruca Freeze K982506.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state quantitative "acceptance criteria" in the traditional sense (e.g., a specific temperature range to be achieved with a defined tolerance). Instead, the performance objective is to demonstrate equivalency of temperature and cryosurgical effect compared to the predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalency of Temperature Attained"Bench testing in this submission focused on demonstrating the equivalency of temperatures over their treatment times while using cone or applicators from the CryoTouch™ and Verruca Freeze." "The comparison bench testing of the CryoTouch TM to Verruca Freeze shows that the temperatures attained are sufficient for cryosurgery." The CryoTouch™ cryogen gas boiling point is -52°C, compared to -48°C for the predicate, indicating a similar or slightly colder temperature.
    Sufficient Cryosurgical Effect (Tissue Destruction)"When sufficient heat has been removed from the treated spot, the tissue is destroyed through cell lysis and vascular stasis." "A second bench test also used the CryoTouch™ and Verruca Freeze, comparing each device and its ability to destroy red blood cells immobilized in agar." "The data, therefore, supports the substantial equivalence between CryoTouch™ and the predicate."
    Similar Mode of Operation and Safety CharacteristicsThe comparison table shows remarkably similar characteristics for intended use, materials, mode of use, mechanism of action, storage conditions, safety, treatment procedure, operation, disposal, and defined operators. The only significant difference is the specific cryogen mixture and its boiling point, which are handled in the temperature equivalency testing.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The study involved "bench testing" and "a second bench test." The quantity of devices, cones/applicators, or agar samples used is not detailed.
    • Data Provenance: The data is generated from bench testing, meaning it was conducted in a laboratory setting. There is no information about country of origin of the data as it's a technical bench study. It is inherently prospective as the tests were designed and executed to compare the devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to the bench testing described. The "ground truth" here is objective physical measurements (temperature, red blood cell destruction) rather than expert interpretation of medical images or clinical outcomes. There is no mention of experts being used in this capacity for the bench tests.

    4. Adjudication Method for the Test Set

    Not applicable. The bench tests involved objective measurements, not subjective evaluations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC study is relevant for diagnostic devices where human readers interpret medical images. This submission describes a cryosurgical device, and its performance evaluation focused on its physical characteristics and direct effect on biological material in a lab setting. It does not involve human readers interpreting cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is not an AI/algorithm-driven device. Its performance is assessed physically.

    7. Type of Ground Truth Used

    The ground truth used in the bench testing was:

    • Physical Measurements: Primarily temperature profiles attained by the device.
    • Direct Biological Effect: The device's ability to destroy red blood cells immobilized in agar, serving as a simplified model for tissue destruction.

    8. Sample Size for the Training Set

    Not applicable. This is a medical device, not an AI/ML model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.