    The Crius PEEK Suture Anchor System is intended for the fixation of soft tissue to bone in the foot,ankle,hand,wrist,elbow,knee,hip and shoulder.

    Device Description

    The Crius PEEK Suture Anchor System is a soft-tissue to bone fixation device, provided preloaded on a disposable inserter. Crius PEEK Suture Anchor System includes four types Thread-Fix Anchor, Thread-Fix Ultra Anchor,Helicoil Anchor and Pile Anchor.The device is available in varying profile diameters. The system consists of a suture anchor, suture(except for Pile Anchor ), and insertion handle and insertion shaft) and is a sterile,single use device. The anchor body is manufactured from Polyetheretherketone(PEEK).The suture is made from ultra high molecular weight polyethylene(UHMWPE) available in white and blue/white colors. The blue dye used in the suture is Chromium-cobalt-aluminum oxide in accordance with 21 CFR 73.1015 regulations. Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture. The insertion shaft is made from 304 stainless steel compliant with ISO 7153-1 and 630 stainless steel compliant with ISO 16061. The surgical instruments are provided non-sterile and shall be sterilized by the end user.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "Crius PEEK Suture Anchor System." This document primarily focuses on establishing substantial equivalence to predicate devices and does not contain information about the performance or acceptance criteria of an AI/ML device.

    Therefore, I cannot provide a detailed answer to your request. The document describes a traditional medical device (suture anchors) and its physical and mechanical performance testing, not the performance of an AI/ML system.

    However, I can extract the information that is present about the traditional device's performance testing:

    Performance Data for Crius PEEK Suture Anchor System (Traditional Medical Device)

    The performance data included in the document relates to mechanical testing of the suture anchors, not an AI/ML model.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states: "Acceptance criteria were based on similar sized predicate suture anchor devices and the results of the performance testing showed the subject device to have substantially equivalent performance compared to the predicate device."
    Specific numerical acceptance criteria are not provided in this summary.

    Test TypeAcceptance Criteria (General Statement)Reported Device Performance (General Statement)
    Insertion Force TestingBased on similar sized predicate suture anchor devices.Showed substantially equivalent performance compared to the predicate device.
    Insertion Torque TestingBased on similar sized predicate suture anchor devices.Showed substantially equivalent performance compared to the predicate device.
    Torsional Strength TestingBased on similar sized predicate suture anchor devices.Showed substantially equivalent performance compared to the predicate device.
    Pullout Strength TestingBased on similar sized predicate suture anchor devices.Showed substantially equivalent performance compared to the predicate device.
    Fatigue Strength TestingBased on similar sized predicate suture anchor devices.Showed substantially equivalent performance compared to the predicate device.
    Pyrogen Limit<20 EU/device (Implicitly, this is the acceptance criterion)Meets the pyrogen limit specification (<20 EU/device).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided text.
    • Data Provenance: The tests were conducted according to ASTM F543-17, ASTM F1839-08, and FDA Guidance, implying laboratory testing on the physical devices. This is not clinical data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is mechanical testing of a physical device, not AI/ML performance evaluation based on expert-labeled data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the mechanical tests, the "ground truth" would be the measured physical properties (e.g., force, torque, ultimate strength) of the device itself and the comparative performance against predicate devices as per the mentioned ASTM standards and FDA guidance.

    8. The sample size for the training set:

    • Not applicable, as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    • Not applicable.