Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K223491

    Validate with FDA (Live)

    Date Cleared
    2023-05-25

    (185 days)

    Product Code
    Regulation Number
    892.2090
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Critical Care Suite with Pneumothorax Detection AI Algorithm is a computer-aided triage, notification, and diagnostic device that analyzes frontal chest X-ray images for the presence of a pneumothorax. Critical Care Suite identifies and highlights images with a pneumothorax to enable case prioritization or triage and assist as a concurrent reading aide during interpretation of radiographs.

    Intended users include qualified independently licensed healthcare professionals (HCPs) trained to independently assess the presence of pneumothoraxes in radiographic images and radiologists.

    Critical Care Suite should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. It is not intended to replace the review of the X-ray image by a qualified physician. Critical Care Suite is indicated for adults and Transitional Adolescents (18 to < 22 years old but treated like adults).

    Device Description

    Critical Care Suite is a suite of Al algorithms for the automated image analysis of frontal chest X-rays acquired on a digital x-ray system for the presence of critical findings. Critical Care Suite with Pneumothorax Detection Al Algorithm is indicated for adults and transitional adolescents (18 to <22 years old but treated like adults) and is intended to be used by licensed qualified healthcare professionals (HCPs) trained to independently assess the presence of pneumothoraxes in radiographic images and radiologists. Critical Care Suite is a software module that can be deployed on several computing platforms such as PACS, On Premise, On Cloud or X-ray Imaging Systems.

    Today's clinical workflow, hospitals are overburdened by large volume of orders and long turnaround times for radiologist reports. Critical Care Suite with the Pneumothorax Detection Al Algorithm enables effective prioritization and assists in the detection / diagnosis of pneumothoraxes for radiologists and HCPs that have been trained to independently assess the presence of pneumothoraxes in radiographic images. It performs this task by flagging images with a suspicious finding and providing a localization overlay of the suspected pneumothorax as well as a graphical representation of the algorithm's confidence in the resultant finding. These outputs can be displayed wherever the reviewing physician normally conducts their reads per their standard of care, including PACS, On Premise, On Cloud and Digital Projection Radiographic Systems.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the GE Medical Systems, LLC Critical Care Suite with Pneumothorax Detection AI Algorithm, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on reporting the device's performance against its own established criteria rather than explicitly listing pre-defined "acceptance criteria" tables. However, we can infer the acceptance criteria from the reported performance goals.

    MetricAcceptance Criteria (Implied from Performance)Reported Device Performance (Standalone)Reported Device Performance (MRMC with AI Assistance vs. Non-Aided)
    Pneumothorax Detection (Standalone Algorithm)Detect pneumothorax in frontal chest X-ray images, with high diagnostic accuracy.AUC of 96.1% (94.9%, 97.2%)Not Applicable
    Sensitivity (Overall)High sensitivity for overall pneumothorax detection.84.3% (80.6%, 88.0%)Not Applicable
    Specificity (Overall)High specificity for overall pneumothorax detection.93.2% (90.8%, 95.6%)Not Applicable
    Sensitivity (Large Pneumothorax)High sensitivity for large pneumothoraxes.96.3% (93.1%, 99.2%)Not Applicable
    Sensitivity (Small Pneumothorax)High sensitivity for small pneumothoraxes.75.0% (69.2%, 80.8%)Not Applicable
    Pneumothorax Localization (Standalone Algorithm)Localize suspected pneumothoraxes effectively.Partially localized 98.1% (96.6%, 99.6%) of actual pneumothorax within an image (apical, lateral, inferior regions).Not Applicable
    Full agreement between regions.67.8% (62.7%, 73.0%)Not Applicable
    Overlap with true pneumothorax area.DICE Similarity Coefficient of 0.705 (0.683, 0.724)Not Applicable
    Reader Performance Improvement (MRMC Study)Improve reader performance for pneumothorax detection.Mean AUC improved by 14.5% (7.0%, 22.0%; p=.002) from 76.8% (non-aided) to 91.3% (aided).14.5% improvement in mean AUC
    Reader Sensitivity ImprovementIncrease reader sensitivity.Reader sensitivity increased by 16.3% (13.1%, 19.5%; p<.001) from 67.4% (non-aided) to 83.7% (aided).16.3% improvement in sensitivity
    Reader Specificity ImprovementIncrease reader specificity.Reader specificity increased by 12.4% (9.6%, 15.1%; p<.001) from 76.6% (non-aided) to 89.0% (aided).12.4% improvement in specificity
    Reader Performance Improvement (Large Pneumothorax)Improve reader performance for large pneumothoraxes.Mean AUC improved by 10.5% (3.2%, 17.8%, p=0.009). Sensitivity improved by 13.4% (10.0%, 16.9%, p<.001).10.5% improvement in mean AUC (large); 13.4% improvement in sensitivity (large)
    Reader Performance Improvement (Small Pneumothorax)Improve reader performance for small pneumothoraxes.Mean AUC improved by 17.6% (9.3%, 25.9%, p<0.001). Sensitivity improved by 18.7% (13.8%, 23.6%, p<.001).17.6% improvement in mean AUC (small); 18.7% improvement in sensitivity (small)
    Improvement Across User GroupsDemonstrate improvement across different clinical user types.All physicians (Rad, IM, ER) improved 10.4% (2.8%, 17.9%, p=0.015). Nurse practitioners improved 24.1% (1.2%, 47.0%, p=0.045). Non-radiologists (ER, IM, NP) improved 17.5% (9.6%, 25.4%, p<0.001).Varied improvements across user groups as noted.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 804 images
    • Data Provenance: The test set included images from two North American sites.
    • Retrospective/Prospective: The document does not explicitly state if the test set was retrospective or prospective. However, given it's a "final validation ground truth dataset" that was not used in training, it's highly likely to be retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Three blinded radiologists.
    • Qualifications of Experts: Radiologists (no specific experience level mentioned, but "blinded radiologists" implies qualified professionals).

    4. Adjudication Method for the Test Set

    • Adjudication Method: The ground truth was established by "three blinded radiologists." This implies a consensus method, likely majority rule or a process where discrepancies were resolved to arrive at a single ground truth label. The specific phrase "consensus" or "adjudication" is not used, but the description points to this approach.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • MRMC Study Done: Yes, a multi-reader multi-case study was conducted.
    • Effect Size of Human Reader Improvement with AI vs. Without AI Assistance:
      • Mean AUC: Improved by 14.5% (from 76.8% non-aided to 91.3% aided; p=0.002).
      • Sensitivity: Increased by 16.3% (from 67.4% non-aided to 83.7% aided; p<0.001).
      • Specificity: Increased by 12.4% (from 76.6% non-aided to 89.0% aided; p<001).
      • Large Pneumothorax (Mean AUC): Improved by 10.5% (p=0.009).
      • Large Pneumothorax (Sensitivity): Improved by 13.4% (p<0.001).
      • Small Pneumothorax (Mean AUC): Improved by 17.6% (p<0.001).
      • Small Pneumothorax (Sensitivity): Improved by 18.7% (p<0.001).

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Study Done: Yes, the "standalone performance of the Pneumothorax Detection AI Algorithm was tested against this dataset."

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus by three blinded radiologists.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: The algorithm was developed using "over 12,000 images." This number includes images used for training, verification, and validation, but the specific breakdown for the training set alone is not provided. It's implied that the majority would be for training.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: The document states that the "Pneumothorax Detection AI Algorithm was developed using over 12,000 images from six sources, including the National Institute of Health and sites within the United States, Canada, and India." It then clarifies this data was "segregated into training, verification, and validation datasets." While it doesn't explicitly detail the methodology for establishing ground truth for the training set, it's standard practice that such large datasets for deep learning and medical imaging are meticulously annotated by medical experts (e.g., radiologists) or derived from existing clinical reports and pathology, which would then be reviewed or confirmed by experts. Given the rigor for the validation set, it's reasonable to infer a similar expert-driven process for the training data, although the specifics are not provided in this excerpt.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211161

    Validate with FDA (Live)

    Date Cleared
    2021-10-29

    (193 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Critical Care Suite is a suite of AI algorithms for the automated image analysis of frontal chest X-rays acquired on a digital x-ray system.

    Critical Care Suite with the Endotracheal Tube Position produces an on-screen image overlay that detects and localizes an endotracheal tube, locates the endotracheal tube tip, locates the carina, and automatically calculates the vertical distance between the endoracheal tube tip and carina. This information is also transmitted to the radiologist for review.

    Intended users include licensed qualified healthcare professionals (HCPs) trained to independently place and/or assess endotracheal tube placement and radiologists.

    Critical Care Suite with the Endotracheal Tube Positioning AI Algorithm should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. It is not intended to review of the X-ray image by a qualified healthcare professional. Critical Care Suite with the Positioning AI Algorithm is indicated for adult-sized patients.

    Device Description

    Critical Care Suite with Endotracheal Tube Positioning Al Algorithm is an additional AI Algorithm incorporated into the Critical Care Suite software previously cleared under K183182. It introduces the Endotracheal Tube Positioning Al Algorithm which is a quantification tool that analyzes frontal chest x-ray images and based on the data in the image determines the location of the tip of an intubated patient's endotracheal tube, determines the location of the carina, and then calculates and displays the vertical distance between them. The distance provided is within the x-ray detector imaging plane and does not take into account the geometric magnification resultant from the geometry of the x-ray acquisition based on source to image distance (SID), patient size, or any impacts due to patient rotation or tube rotation. This information can aide clinical care teams and radiologists to determine the proper placement of the endotracheal tube in an intubated patient. All algorithms previously cleared under K183182 are still available with Critical Care Suite, including the Pneumothorax Detection Algorithm for triage and notification. The benefit of the proposed modification is not specific to the platform on which it is deployed. This benefit applies to all previously cleared computational platforms for Critical Care Suite, including PACS, On Premise, On Cloud and Digital Projection Radiographic Systems. The Optima XR240amx was chosen as the initial platform for deployment because endotracheal tube placement images are almost exclusively acquired on mobile X-ray systems due to the immobilization of the patients being intubated with an endotracheal tube.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Critical Care Suite with Endotracheal Tube Positioning AI Algorithm, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Implicit)Reported Device Performance (95% CI)
    ETT DetectionHigh accuracy for detection of endotracheal tubes.AUC: 0.9999 (0.9998, 1.0000)
    High sensitivity for detection of endotracheal tubes.Sensitivity: 0.9941 (0.9859, 1.0000)
    High specificity for detection of endotracheal tubes.Specificity: 1.0000 (1.0000, 1.0000)
    ETT Tip to Carina Distance MeasurementHigh success rate for accurate distance measurement.Success Rate: 0.9851 (0.9722, 0.9981)
    Carina LocalizationHigh success rate for accurate carina localization.Success Rate: 0.9851 (0.9722, 0.9981)
    ETT Tip LocalizationHigh success rate for accurate ETT tip localization.Success Rate: 0.9524 (0.9296, 0.9752)
    ETT Localization (DICE Score)High accuracy for overall ETT localization (segmentation fidelity).DICE: 0.9881 (0.9765, 0.9997)

    Note: The document states that "the results met the defined passing criteria." While specific numerical acceptance thresholds are not explicitly listed in the text, the reported high performance metrics imply that these values exceeded the internal acceptance criteria set by the manufacturer.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document states that the ground truth dataset contained a "sufficient number of images to adequately analyze all the primary and secondary endpoints." However, the exact sample size for the test set is not explicitly provided in the given text.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It only mentions the use of a "ground truth dataset."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts used to establish the ground truth for the test set, nor does it provide their specific qualifications (e.g., radiologist with X years of experience).

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method (e.g., 2+1, 3+1, none) used for establishing the ground truth of the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly conducted or described in the provided document. The clinical tests focused on the standalone performance of the AI algorithm against a ground truth dataset, not on comparing human reader performance with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone study was done. The "Summary of Clinical Tests" section explicitly describes the performance of the Endotracheal Tube Positioning AI Algorithm tested against a ground truth dataset, reporting metrics like AUC, sensitivity, specificity, and success rates for localization and measurement. This indicates a standalone evaluation of the algorithm's performance without direct human-in-the-loop comparison for these specific metrics.

    7. The Type of Ground Truth Used

    The type of ground truth used is expert consensus. The document refers to the algorithm's performance being "tested against a ground truth dataset" without specifying the exact method of ground truth establishment (e.g., pathology, outcomes data). However, for image analysis tasks like ETT positioning and carina localization, ground truth is typically established by multiple experts (e.g., radiologists) providing annotations or measurements, often followed by an adjudication process to reach a consensus.

    8. The Sample Size for the Training Set

    The document does not explicitly provide the sample size for the training set. It mentions the algorithms being "trained with clinical and/or artificial data" but no specific numbers.

    9. How the Ground Truth for the Training Set Was Established

    The document states that the algorithms are "trained with clinical and/or artificial data." It does not explicitly detail how the ground truth for the training set was established. It refers to "nonadaptive machine learning algorithms trained with clinical and/or artificial data," but the process of creating the ground truth annotations for this training data is not described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183182

    Validate with FDA (Live)

    Date Cleared
    2019-08-12

    (266 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Critical Care Suite is a computer aided triage and notification device that analyzes frontal chest x-ray images for the presence of prespecified critical findings (pneumothorax). Critical Care Suite identifies images with critical findings to enable case prioritization or triage in the PACS/workstation.

    Critical Care Suite is intended for notification only and does not provide diagnostic information beyond the notification. Critical Care Suite should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. It is not intended to replace the review of the x-ray image by a qualified physician.

    Critical Care Suite is indicated for adult-size patients.

    Device Description

    Critical Care Suite is a software module that employs Al-based image analysis algorithms to identify pre-specified critical findings (pneumothorax) in frontal chest X-ray images and flag the images in the PACS/workstation to enable prioritized review by the radiologist.

    Critical Care Suite employs a sequence of vendor and system agnostic AI algorithms to ensure that the input images are suitable for the pneumothorax detection algorithm and to detect the presence of pneumothorax in frontal chest X-rays:

    • The Quality Care Suite algorithms conduct an automated check to confirm that the input image is compatible with the pneumothorax detection algorithm and that the lung field coverage is adequate;

    • the PTX Classifier determines whether a pneumothorax is present in the image.

    If a pneumothorax is detected, Critical Care Suite enables case prioritization or triage through direct communication of the Critical Care Suite notification during image transfer to the PACS. It can also produce a Secondary Capture DICOM Image that presents the Al results to the radiologist.

    When deployed on a Digital Projection Radiographic Systems such as Optima XR240amx, Critical Care Suite is automatically run after image acquisition. Quality Care Suite algorithms produce an on-device notification if the lung field has atypical positioning to give the technologist the opportunity to make correction before sending the image to the PACS. The Critical Care Suite output is then sent directly to PACS upon exam closure where images with a suspicious finding are flagged for prioritized review by the Radiologist.

    In parallel, an on-device, technologist notification is generated 15 minutes after exam closure, indicating which cases were prioritized by Critical Care Suite in PACS. The technologist notification is contextual and does not provide any diagnostic information. The on-device, technologist notification is not intended to inform any clinical decision, prioritization, or action.

    The Digital Projection Radiographic System intended use remains unchanged in that the system is used for general purpose diagnostic radiographic imaging.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance:

    MetricAcceptance Criteria (Predicate Device HealthPNX - K190362)Critical Care Suite Reported Performance
    ROC AUC> 0.950.9607 (95% CI [0.9491, 0.9724])
    Specificity93%93.5% (95% CI [91.1%, 95.8%])
    Sensitivity93%84.3% (95% CI [80.6%, 88.0%])
    AUC on large pneumothoraxNot assessed0.9888 (95% CI [0.9810, 0.9965])
    Sensitivity on large pneumothoraxNot assessed96.3% (95% CI [93.3%, 99.2%])
    AUC on small pneumothoraxNot assessed0.9389 (95% CI [0.9209, 0.9570])
    Sensitivity on small pneumothoraxNot assessed75% (95% CI [69.2%, 80.8%])
    Timing of notification (delay in PACS worklist)22 seconds (HealthPNX)No delay (immediately on PACS receipt)

    2. Sample size and Data Provenance for the Test Set:

    • Sample Size: 804 frontal chest X-ray images (N=376 for pneumothorax present; N=428 for pneumothorax absent).
    • Data Provenance: Collected in North America, representative of the intended population. The text does not explicitly state if it was retrospective or prospective, but the nature of a "collected dataset" for evaluation typically implies retrospective analysis of existing images.

    3. Number of Experts and Qualifications for Ground Truth of the Test Set:

    • Number of Experts: 3 independent US-board certified radiologists.
    • Qualifications: "US-board certified radiologists." No specific years of experience or subspecialty are mentioned beyond board certification.

    4. Adjudication Method for the Test Set:

    • The text states the ground truth was "established by 3 independent US-board certified radiologists." It does not explicitly detail a specific adjudication method like 2+1 or 3+1. This implies a consensus-based approach where the radiologists independently reviewed images to establish the ground truth, likely resolving discrepancies through discussion to reach a final determination.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study directly comparing human readers with AI assistance vs. without AI assistance was not reported in this summary. The study focused on the standalone diagnostic performance of the AI algorithm.

    6. Standalone (Algorithm Only) Performance Study:

    • Yes, a standalone performance study of the algorithm without human-in-the-loop was done. The reported metrics (ROC AUC, Sensitivity, Specificity) are direct measurements of the algorithm's performance against the established ground truth.

    7. Type of Ground Truth Used:

    • Expert Consensus: The ground truth was established by "3 independent US-board certified radiologists." This indicates an expert consensus approach.

    8. Sample Size for the Training Set:

    • The document does not explicitly state the sample size for the training set. It mentions the algorithm was "trained on annotated medical images" but provides no further details on the quantity of images used for training.

    9. How the Ground Truth for the Training Set Was Established:

    • The document states the device utilizes a "deep learning algorithm trained on annotated medical images." While it doesn't explicitly describe the method for establishing ground truth for the training set, it is implied that these "annotated medical images" had pre-existing labels or were labeled by experts for the purpose of training the AI.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1