Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K150942

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-11-25

    (231 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Contour Next USB blood glucose monitoring system is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only and should not be shared.

    The system consists of a Contour Next USB blood glucose meter, Contour Next test strips and Contour Next control solutions.

    The Contour Next USB blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

    The Contour Next USB blood glucose monitoring system is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

    The Contour Next Test Strips are for use with the Contour Next USB blood glucose monitoring system for the quantitative measurement of glucose in whole blood.

    Device Description

    The Contour Next USB Blood Glucose Meter consists of a small handheld blood glucose meter that utilizes dry reagent test strips for the measurement of glucose in capillary whole blood by persons with diabetes. Liquid control solution is used to check the performance of the system. The meter, together with the test strips and control solutions, is referred to as the Contour Next USB Blood Glucose Monitoring System.

    The chemical principle utilized for both the predicate and modified devices is based on measurement of electrical current caused by the reaction of glucose in the blood with chemicals on the reagent strip. The blood sample is drawn into the tip of the reagent strip through capillary action. Glucose in the sample reacts with FAD glucose dehydrogenase (FAD-GDH) enzyme on the reagent strip. The electrons generated by this reaction are shuttled to an electrode by a mediator chemical, producing a current that is proportional to the glucose in the sample. After a fixed reaction time, the glucose concentration in the sample is calculated and displayed.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Contour Next USB Blood Glucose Monitoring System, which primarily describes modifications to an existing predicate device (K121087) rather than providing a detailed study of the modified device's performance against specific acceptance criteria. The modifications focus on improved error detection capabilities.

    Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes, expert involvement, and ground truth for a full performance study is not explicitly detailed in the provided text. The submission focuses on demonstrating substantial equivalence to the predicate device due to specific enhancements.

    Here's an attempt to extract the available information based on your request, with noted limitations where data is not provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria in a table format for blood glucose measurement accuracy for the modified device. Instead, it describes functional improvements related to error detection. The "reported device performance" focuses on the functionality of these new error checks and equivalency to the predicate device.

    Acceptance Criteria (Inferred from improvements)Reported Device Performance (Modified Device)
    Detect test strips exposed to degrading chemicalError message given for exposed strips
    Detect un-mixed control solutionError message given for un-mixed control solution
    Detect sample 'perturbation' during a testError message given for disturbed samples
    For overall glucose measurement accuracy:
    Not statistically different from predicate device (K121087)Equivalency testing confirmed performance not statistically different.
    No good results classified as errors by modificationsSoftware verification ensured no good results were misclassified.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: Not specified. The document mentions "bench testing" and an "internal user study" but does not give sample sizes for either.
    • Data Provenance: Not specified, but likely proprietary internal data from Bayer Healthcare. It doesn't specify country of origin or whether it's retrospective or prospective. Given the context of 510(k) submissions, such studies are typically prospective tests conducted by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable or not specified. This type of submission for a blood glucose meter does not typically involve expert consensus to establish ground truth in the same way an imaging or diagnostic AI device would. The "ground truth" for a blood glucose meter is usually established by a laboratory reference method. The document mentions "internal user study," suggesting lay users may have been involved in some testing, but no details on expert involvement for ground truth are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable or not specified. Adjudication methods like 2+1 are typically used for subjective assessments (e.g., image interpretation) where multiple experts review and reconcile differences. For a quantitative device like a blood glucose meter, ground truth is established by a reference method, not by expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are relevant for devices that assist human readers (e.g., AI for medical imaging). This is a standalone diagnostic device. The study mentions an "internal user study" but not in the context of an MRMC design or comparing human performance with/without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in essence. The bench testing and software verification of the error detection mechanisms would fall under "standalone" algorithm testing to ensure the meter correctly identifies errors and does not misclassify good results. The equivalency testing would also implicitly assess the algorithm's accuracy compared to the predicate.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For blood glucose meters, the ground truth for accuracy studies is typically established by a laboratory reference method (e.g., YSI glucose analyzer) for whole blood glucose concentration. While not explicitly stated for this specific submission, this is the standard for blood glucose meter validation. The text focuses on the functional correctness of its new error checks rather than primary accuracy data against a reference method, as it's demonstrating equivalence to an already cleared device.

    8. The sample size for the training set

    Not applicable. As a medical device (blood glucose meter) and not an AI/ML algorithm that requires a "training set" in the conventional sense, this information is not relevant or provided. The "training" for such a device is its design and manufacturing process, and its performance is validated through testing.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the context of AI/ML for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121087

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-06-28

    (79 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CONTOUR® NEXT USB Blood Glucose Monitoring System is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only and should not be shared. The CONTOUR® NEXT USB Blood Glucose Monitoring System is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. The CONTOUR® NEXT USB Blood Glucose Monitoring System may be used as an aid to monitor the effectiveness of a diabetes control program and is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. The Contour® NEXT Test Strips are for use with the CONTOUR® NEXT USB Blood Glucose Monitoring System for the quantitative measurement of glucose in whole blood. The Contour® NEXT Controls are for use with the CONTOUR® NEXT USB Blood Glucose Monitoring System to check that the meter and test strips are working properly.

    Glucofacts® Deluxe Diabetes Management Software is an over-the-counter software program intended for use by health care professionals and patients with diabetes for viewing and printing reports that display blood sugar readings from Baver's Contour and Breeze families of meters.

    Device Description

    The Contour NEXT USB Blood Glucose Monitoring System consists of a small handheld blood glucose meter that is substantially equivalent in look and feel to the predicate system. Contour Next Wireless Blood Glucose Meter, (K110894). The system also consists of dry reagent test strips used for the measurement of glucose in capillary whole blood and includes liquid controls to check the performance of the system.

    The chemical principle of the system is based on measurement of electrical current caused by the reaction of glucose in the blood with chemicals on the reagent strip. The blood sample is drawn into the tip of the reagent strip through capillary action. Glucose in the sample reacts with FAD glucose dehydrogenase (FAD-GDH) enzyme on the reagent strip. The electrons generated by this reaction are shuttled to an electrode by a mediator chemical, producing a current that is proportional to the glucose in the sample.

    AI/ML Overview

    The Bayer Contour NEXT USB Blood Glucose Monitoring System was tested and found to meet the system accuracy requirements of ISO 15197:2003 through analytical performance testing and a user performance evaluation.


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are derived from ISO 15197:2003 for system accuracy.

    System Accuracy - Glucose Concentrations < 75 mg/dL

    Acceptance Criteria (ISO 15197:2003)Reported Device Performance (Contour NEXT USB)
    At least 95% of results must be within ± 15 mg/dL of the reference method.78 of 78 (100%) of results were within ± 15 mg/dL.
    (Not explicitly stated as an ISO requirement in the document, but for context) 95% within ± 10 mg/dL78 of 78 (100%) of results were within ± 10 mg/dL.
    (Not explicitly stated as an ISO requirement in the document, but for context) 95% within ± 5 mg/dL65 of 78 (83.3%) of results were within ± 5 mg/dL.

    System Accuracy - Glucose Concentrations ≥ 75 mg/dL

    Acceptance Criteria (ISO 15197:2003)Reported Device Performance (Contour NEXT USB)
    At least 95% of results must be within ± 20% of the reference method.522 of 522 (100%) of results were within ± 20%.
    (Not explicitly stated as an ISO requirement in the document, but for context) 95% within ± 15%522 of 522 (100%) of results were within ± 15%.
    (Not explicitly stated as an ISO requirement in the document, but for context) 95% within ± 10%512 of 522 (98.1%) of results were within ± 10%.
    (Not explicitly stated as an ISO requirement in the document, but for context) 95% within ± 5%390 of 522 (74.7%) of results were within ± 5%.

    Repeatability (Analytical Testing)

    Target Glucose (mg/dL)Mean Glucose (mg/dL)Standard Deviation (mg/dL)Coefficient of Variation (%)
    40470.81.8
    80871.31.5
    1301311.91.5
    2102112.81.3
    3303426.41.9

    Intermediate Precision (Analytical Testing with Control Solution)

    Control LevelMean Glucose (mg/dL)Standard Deviation (mg/dL)Coefficient of Variation (%)
    Level 1450.92.0
    Level 21332.01.5
    Level 33976.61.7

    User Performance Evaluation - Glucose Concentrations < 75 mg/dL (Subject & Professional Fingertip Results)

    TesterWithin ± 5 mg/dLWithin ± 10 mg/dLWithin ± 15 mg/dL
    Subject8 of 8 (100%)8 of 8 (100%)8 of 8 (100%)
    Professional8 of 8 (100%)8 of 8 (100%)8 of 8 (100%)

    User Performance Evaluation - Glucose Concentrations ≥ 75 mg/dL (Subject & Professional Fingertip Results)

    TesterWithin ± 5%Within ± 10%Within ± 15%Within ± 20%
    Subject152 of 196(77.6%)188 of 196(95.9%)193 of 196(98.5%)195 of 196(99.5%)
    Professional151 of 196(77.0%)194 of 196(99.0%)196 of 196(100%)196 of 196(100%)

    2. Sample Size and Data Provenance for Test Set

    • Analytical System Accuracy Test Set: The "N" for the regression statistics was 600, indicating 600 samples for the primary system accuracy evaluation. The results are further broken down into <75 mg/dL (78 samples) and ≥75 mg/dL (522 samples). The document does not specify the country of origin or whether the data was retrospective or prospective.
    • Repeatability Test Set: 300 venous blood tests per glucose level (40, 80, 130, 210, 330 mg/dL), with 100 tests from each of three lots of test strips.
    • Intermediate Precision Test Set: 300 control solution tests per control level (Level 1, Level 2, Level 3), with 100 tests from each of three lots of test strips.
    • User Performance Evaluation Test Set: For glucose concentrations <75 mg/dL, 8 samples were tested by both subjects (intended users) and professionals. For glucose concentrations ≥75 mg/dL, 196 samples were tested by both subjects and professionals. The data provenance is not explicitly stated regarding country or retrospective/prospective nature.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not provide information about the number or qualifications of experts used to establish the ground truth. It refers to a "reference method" for comparison but does not detail how this reference method was implemented or who performed it.


    4. Adjudication Method for Test Set

    The document does not describe any adjudication method. The comparisons are made against a "reference method."


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is reported. The tests focus on device accuracy against a reference method and user performance without comparison to human readers with and without AI assistance.


    6. Standalone Algorithm Performance

    Yes, the study primarily describes the standalone performance of the device (Contour NEXT USB Blood Glucose Monitoring System). The "System Accuracy" and "Repeatability" sections detail the analytical performance of the device without human-in-the-loop assistance in the measurement process itself, although the "User Performance Evaluation" section involves human users operating the device. The core measurement algorithm's performance is reflected in the accuracy and precision data.


    7. Type of Ground Truth Used

    The ground truth used is a "reference method" for glucose measurement. The specific nature of this reference method (e.g., a laboratory gold-standard analyzer, a specific clinical assay) is not detailed in the provided text.


    8. Sample Size for Training Set

    The document does not provide information on the sample size used for a training set. This is typical for a medical device (blood glucose meter) where the fundamental chemical and electrical principles are fixed, and performance is characterized through analytical and clinical validation, rather than a machine learning model that requires a "training set."


    9. How Ground Truth for Training Set Was Established

    Since no training set is described for a machine learning algorithm, the establishment of ground truth for a training set is not applicable here. The device's underlying principles are based on the reaction of glucose with FAD-GDH enzyme and electrical current measurement.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1