Found 4 results

510(k) Data Aggregation

    K Number
    K243466

    Validate with FDA (Live)

    Date Cleared
    2025-01-06

    (59 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in hip arthroplasty in skeletally mature patients with the following conditions:

    1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
    2. Avascular necrosis of the femoral head.
    3. Acute traumatic fracture of the femoral head or neck.
    4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement or total hip replacement.
    5. Certain cases of ankylosis.
      Conformity stem is for cementless use only.
    Device Description

    The subject device is an extension line of 510(K) cleared device Conformity stem, cementless (K183312, K242249), which introduces four new variations: (1) Conformity stem, #0; (2) Conformity stem, collared, #0; (3) Conformity stem, coxa vara, 125° STD, collared, #0; and (4) Conformity stem, short neck, collared #0. The indications, design rationales, materials, manufacturing process, and sterilization method of the subject devices are identical to the 510(k) cleared Conformity Stem, cementless (K183312, K242249).
    Conformity Stem can collocate with “United” metallic femoral heads (K994078, K022520, K111546, K122504, K152439, K162957, K221675) or ceramic femoral heads (K103497, K112463, K122185).

    • Conformity stem, #0
      Conformity stem, #0 is an extension in terms of sizes to 510 (k) cleared device Conformity stem, #1-11 (K183312). The material, coating thickness, neck angle (135°), and neck length (35.5mm) design are identical to 510 (k) cleared device Conformity Stem, #1, #11 (K183312)The only differences in the items are the offset specifications and stem length.
    • Conformity stem, collared, #0
    • Conformity stem, collared, #0 is an extension in terms of sizes to 510 (k) cleared device Conformity stem, collared, #1-11 (K183312). The material, coating thickness, neck angle (135°), and neck length (35.5mm) design are identical to 510 (k) cleared device Conformity stem, collared, #1, #11 (K183312).The only differences in the items are the offset specifications and stem length.
    • Conformity Stem, coxa vara, 125° STD, collared, #0
    • Conformity Stem, coxa vara, 125° STD, collared, #0 is an extension in terms of sizes to 510 (k) cleared device Conformity stem, coxa vara 125° STD, collared, #1-7 (K242249). The material, coating thickness, neck angle (125°), and neck length (32.5mm) design are identical to 510 (k) cleared device Conformity stem, 125° STD, collared, #1, #7 (K242249). The only differences in the items are the offset specifications and stem length.
    • Conformity Stem, short neck, collared #0
    • Conformity Stem, short neck, collared #0 is an extension in terms of sizes to 510 (k) cleared device Conformity Stem, short neck, collared, #1-7 (K183312 and K242249). The material, coating thickness, neck angle (135°), and neck length (28.5mm) design are identical to 510 (k) cleared device Conformity Stem, short neck, collared, #1, #7 (K183312 and K242249). The only differences in the items are the offset specifications and stem length.
    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (hip stem extension line). It details the device's identity, intended use, and comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets those criteria, especially not in the context of an AI/ML-driven device.

    The document primarily focuses on demonstrating substantial equivalence to existing legally marketed devices based on:

    1. Identical Indications for Use: The new device has the same indications as the predicate devices.
    2. Similar Technological Characteristics: The material, principle of operation, regulation number, product code, risk class, intended users, fixation method, surface coating, locking mechanism, and sterilization method are the same as the predicate device.
    3. Non-Clinical Performance Data: The submission includes tests like Range of Motion, Neck Fatigue Assessment, Stem Fatigue Test, Characterization of HA Plasma Spray Coating, and Usability Evaluation. These are standard engineering and material tests for orthopedic implants, not AI/ML performance metrics.

    Therefore, I cannot extract the requested information regarding:

    • A table of acceptance criteria and reported device performance (for AI/ML).
    • Sample size and data provenance for a test set (for AI/ML).
    • Number and qualifications of experts for ground truth (for AI/ML).
    • Adjudication method (for AI/ML).
    • MRMC comparative effectiveness study or human reader improvement with AI.
    • Standalone algorithm performance.
    • Type of ground truth (expert consensus, pathology, outcomes data, etc.) for AI/ML validation.
    • Sample size for the training set (for AI/ML).
    • How ground truth for the training set was established (for AI/ML).

    The document is about a physical orthopedic implant and its substantial equivalence based on mechanical and material testing, not an AI/ML diagnostic or assistive device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242249

    Validate with FDA (Live)

    Date Cleared
    2024-08-30

    (30 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in hip arthroplasty in skeletally mature patients with the following conditions:

    1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia.
    2. Avascular necrosis of the femoral head.
    3. Acute traumatic fracture of the femoral head or neck.
    4. Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement or total hip replacement.
    5. Certain cases of ankylosis.
      Conformity stem is for cementless use only.
    Device Description

    The "United" Conformity Stem cementless type (K183312) is designed for cementless use in hip arthroplasty. The Conformity Stem cementless type is a proximal trapezoid and distal quadrangular stem with 12/14 neck taper, 135° and 125° neck angle. It is manufactured from Ti-6Al-4V alloy, which conforms to ASTM F136-13 and the distal part of each femoral stem is coated with hydroxyapatite (HA) (ASTM F1185-03) to provide biological fixation. The "United" Conformity Stem cementless is available in various sizes with collarless and collared types to accommodate various hip surgical requirements.
    For the subject device, it's an extension line of the 510(K) cleared device Conformity Stem cementless type (K183312), which introduces two new variations: (1) Conformity stem, coxa vara, 125° STD, collared, #1-7; and (2) Conformity stem, short neck, #4-7. The indications, design rationales, materials, major manufacturing process, and sterilization method of the subject devices are identical to the 510(k) cleared Conformity Stem, cementless (K183312).
    Conformity Stem can collocate with "United" metallic femoral heads (K994078, K022520, K111546, K122504, K152439, K162957, K221675) or ceramic femoral heads (K103497, K112463, K122185).

    • Conformity stem, coxa vara, 125° STD, collared
      Conformity stem, coxa vara, 125° STD, collared is a new coxa vara type variation of 510(k) cleared Conformity stem, coxa vara (1110-52XX, K183312). The same as this 510(k) cleared device, it's also made of Ti-6Al-4V alloy (ASTM F136), and the distal part is coated with hydroxyapatite(HA) (ASTM F1185). In addition, the subject device is also a collared stem with a 125° neck angle. The only difference is the offset design. The subject device is standard offset, while the 510(k) cleared coxa vara collared stem (110-52XX) is high offset
    • Conformity Stem, short neck, #4-7
      Conformity stem, short neck, #4-7 is an extension in terms of sizes to 510 (k) cleared device Conformity stem, short neck, #1-3 (K183312). The design, material, coating thickness, geometrical characteristics, neck angle (135°), and neck length (28.5mm) design are identical to 510 (k) cleared device Conformity Stem, short neck, #1-3 (K183312). The only different is the stem length sizes.
    AI/ML Overview

    This is an FDA 510(k) clearance letter for a medical device called "Conformity Stem Extension Line." This document does not describe acceptance criteria for an AI/ML device or a study proving that an AI/ML device meets those criteria. Instead, it describes mechanical analyses performed on a physical orthopedic implant.

    Therefore, I cannot extract the requested information regarding acceptance criteria and study details for an AI/ML device from this document. The document pertains to a hip joint prosthesis, which is a physical implant, not a software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191056

    Validate with FDA (Live)

    Date Cleared
    2020-01-03

    (259 days)

    Product Code
    Regulation Number
    888.3350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in hip arthroplasty in skeletally mature patients with the following conditions:

    1. A severely painful and/or disabled joint from osteoarthritis, rheumatoid arthritis, or congenital hip dyplasia.
    2. Avascular necrosis of the femoral head.
    3. Acute traumatic fracture of the femoral head or neck.
    4. Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacment or total hip replacement.
    5. Certain cases of ankylosis.
      Conformity stem, cemented is for cemented use only.
    Device Description

    The "United" Conformity stem, cemented is a triple tapered, polished, collarless stem. It is available in various sizes with standard and high offset types to accommodate various hip surgical requirements. Conformity stem, cemented is manufactured from Co-Cr-Mo alloy which conform to ASTM F799-11 (Raw material: ASTM F1537-11/ISO 5832-12:2007). It is intended to be fixed only with the use of PMMA bone cement and should be used with "United" Cement restrictor.
    For total hip arthroplasty, Conformity stem, cemented can be used with "United"acetabular liner, cup and femoral head. For bipolar hip replacement, Conformity stem, cemented can be used with "United" bipolar prosthesis.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Conformity stem, cemented) and does not describe a study involving an AI/algorithm. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data (mechanical properties, FEM analysis, ROM simulation, bacterial endotoxin testing).

    Therefore, I cannot provide information on acceptance criteria and study details related to an AI/algorithm according to your request. The document explicitly states "Clinical Performance Data/Information: None provided as a basis for substantial equivalence."

    If this was a misunderstanding and you were looking for the acceptance criteria and performance of the hip stem itself, and not an AI or algorithm, please let me know.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183312

    Validate with FDA (Live)

    Device Name
    Conformity Stem
    Date Cleared
    2019-08-08

    (252 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in hip arthroplasty in skeletally mature patients with the following conditions:

    1. A severely painful and/or disabled joint from osteoarthritis, rheumatoid arthritis, or congenital hip dyplasia.
    2. Avascular necrosis of the femoral head.
    3. Acute traumatic fracture of the femoral head or neck.
    4. Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacment or total hip replacement.
    5. Certain cases of ankylosis.
      Conformity stem is for cementless use only.
    Device Description

    The "United" Conformity Stem cementless type is design for cementless use in hip arthroplasty. The Conformity Stem cementless type is a proximal trapezoid and distal quadrangular stem with 12/14 neck taper, 135° and 125° neck angle. It is manufactured from Ti-6A1-4V alloy which conform to ASTM F136-13 and the distal part of each femoral stem is coated with hydroxyapatite (HA) (ASTM F1185-03) to provide biological fixation. The subject device is available in various sizes with collarless and collared types to accommodate various hip surgical requirements.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "Conformity Stem," a hip stem for hip arthroplasty. The document asserts the device's substantial equivalence to a legally marketed predicate device.

    However, the content does not contain any information related to an AI/ML-driven medical device, nor does it discuss acceptance criteria, device performance metrics (like accuracy, sensitivity, specificity, AUC), sample sizes for test or training sets, ground truth establishment by experts, adjudication methods, or MRMC studies.

    The document focuses on the mechanical and material performance of the hip stem, including tests like stem fatigue, neck fatigue, range of motion, and fretting corrosion, all of which are standard engineering and biocompatibility evaluations for an implantable orthopedic device. It explicitly states "Clinical Performance Data/Information: None provided as a basis for substantial equivalence."

    Therefore, I cannot fulfill the request as the provided text does not contain the necessary information about an AI/ML device study.

    To answer your request, the input text would need to describe an AI/ML device, its intended use for diagnostic or prognostic purposes, and the clinical study conducted to validate its performance according to the specified criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1