Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K161408

    Validate with FDA (Live)

    Date Cleared
    2016-11-22

    (186 days)

    Product Code
    Regulation Number
    870.2850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Compass Cast and MAP disposable pressure transducer system with integrated digital display and wireless transmission capability is intended for direct measurement and monitoring of physiologic pressure, including during the infusion of fluids and therapeutic and diagnostic agents.

    Device Description

    The Compass Cast and MAP disposable pressure transducer system with integrated digital display and wireless transmission capability is intended for direct measurement and monitoring of physiologic pressure, including during the infusion of fluids and therapeutic and diagnostic agents.

    The Compass CT and Compass CT Port are disposable, point-of-use pressure measurement and monitoring devices that incorporate a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The devices have a distal male luer fitting to connect to a needle or catheter, and a proximal female luer fitting that can be connected to accessory devices (e.g. syringes, caps, or infusion tubing). The devices measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD. The Compass CT Port has an additional, sealed proximal port through which commercially available guidewires can be inserted during pressure measurement.

    The Compass CT Port device is being modified to add wireless Bluetooth capability to the point-of-use Compass Cast device, in order to allow for optional display of acquired pressure data and information to a separate display monitor. The modified product will consist of two components that are packaged separately, the compass Cast (Subject Device) and MAP (Monitor Accessory Plug - Accessory).

    The Compass Cast (Subject Device) is a single-use, sterile pressure transducer that is physically identical to the predicate device Compass CT Port (510(k) K133624) with the exception of a slight modification to the main CT Port circuit board to accommodate the secondary Bluetooth radio board. All components and functionality of the main circuit board remain identical.

    The MAP (Accessory) is a non-sterile, reusable monitor accessory plug that receives a digital pressure signal from the Compass Cast and converts the digital pressure signal to an analog output that is identical to the analog output of a traditional wired pressure transducer. There are two versions of the MAP device:

    1. Wireless Pressure Receiver - GE Monitor (CWMG001-5)
    2. Wireless Pressure Receiver Philips Monitor (CWMP001-5)
    AI/ML Overview

    The provided text describes information about the submission of the "Compass Cast and MAP System" for FDA clearance. However, it does not contain explicit acceptance criteria and device performance data in a tabular format, nor does it detail a study that proves the device meets specific performance criteria related to its core function of physiological pressure measurement beyond stating adherence to existing standards.

    The document focuses heavily on demonstrating substantial equivalence to a predicate device (Compass CT Port) by highlighting the technological characteristics and the modifications made (addition of Bluetooth capability and a reusable Monitor Accessory Plug - MAP). Performance data mentioned primarily concerns safety aspects due to these modifications, such as sterility, EO residuals, and electromagnetic compatibility.

    Therefore, many of the requested items cannot be fully answered from the provided text.

    Based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state a table of new acceptance criteria established for the Compass Cast and MAP system itself in terms of pressure measurement performance. Instead, it states that the device's fundamental pressure accuracy and functional performance are "identical" to the predicate device and "Meet or exceed ANSI/AAMI BP22:1994(R)2006".

    The only "acceptance criteria" and "reported performance" directly stated are for safety and compatibility tests related to the new wireless functionality and sterilization.

    Acceptance Criteria (Implied / Standard Adherence)Reported Device Performance
    Sterilization: Sterility Assurance Level (SAL) of 1 x 10^-6Validated (in accordance with ANSI/AAMI/ISO 11135-1:2014)
    EO Residuals: < 4 mg/device for EO, < 9 mg/device for ECTesting performed; compliant with ANSI/AAMI/ISO 10993-7:2008 (implied passes, as "testing was performed" on worst-case model and is "in Compliance")
    BET Testing: Endotoxin limit: 2.15 EU/DeviceKinetic Chromogenic Test Method validated; "in Compliance" with AAMI ANSI ST72:2011
    EMC/Electrical Safety: Compliance with IEC 60601-1-2:2007"Confirmed to be compliant" and "in Compliance"
    Wireless Communication: Compliance with relevant standards (e.g., CFR 47, Part 15, subpart B; RSS-247)"In Compliance"
    Pressure Accuracy and Functional Performance: Meet or exceed ANSI/AAMI BP22:1994(R)2006"Identical" to predicate, which meets/exceeds this standard. Explicit mention: -199 to -51 mm Hg: ±3% of reading -50 to +50 mm Hg: ±2 mm Hg 51 to 999 mm Hg: ±3% of reading
    Zero Drift: ±1 mm Hg per 4 hours"Identical" to predicate
    Light Sensitivity: < 1 mm Hg"Identical" to predicate

    2. Sample size used for the test set and the data provenance

    For the safety tests (sterilization, EO residuals, BET, EMC/Wireless), specific sample sizes are mentioned:

    • EO Residuals: "Three samples were processed through two routine (Cycle 20) sterilization cycles".
    • Other tests: Sample sizes are not explicitly stated for EMC/Wireless, Sterilization validation, or BET, beyond stating that testing was performed.
    • Pressure performance: No specific test set sample size is described for evaluating the pressure accuracy of the Compass Cast itself, as its performance is stated to be "identical" to the predicate. Therefore, any data provenance for such a study would refer to the predicate device development.

    The provenance for the safety tests is internal company testing ("Centurion Protocol# 17205-16"). No country of origin is specified for these tests. The nature of these tests is prospective for this device modification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The regulatory submission relies on adherence to established consensus standards (e.g., ANSI/AAMI BP22, ISO 11135, IEC 60601-1-2), rather than a new study requiring expert ground truth for diagnostic accuracy.

    4. Adjudication method for the test set

    Not applicable, as no diagnostic accuracy study requiring expert adjudication is described in the provided text for the modified device. The compliance is against engineering and safety standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a pressure transducer system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation would be the stated compliance to standards like ANSI/AAMI BP22 for pressure accuracy, and the safety/EMC standards. There is no algorithm in the AI sense for a standalone "algorithm only" performance. The device itself is the "standalone" component measuring pressure.

    7. The type of ground truth used

    For pressure accuracy and functional performance, the ground truth is based on established metrology reference standards and methods outlined in ANSI/AAMI BP22:1994(R)2006.
    For sterilization, EO residuals, BET, and EMC, the ground truth is defined by the specific requirements and test methods detailed in the respective international and national standards (e.g., ANSI/AAMI/ISO 11135, ISO 10993-7, AAMI ANSI ST72, IEC 60601-1-2).

    8. The sample size for the training set

    Not applicable. This is not a machine learning/AI device, so there is no training set mentioned.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set involved.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1