Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241387

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-13

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CoLink® NeoFuse MIS Plating System is a fusion plate indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia and talus. The addition of a posterior compression screw and a fully threaded screw through the tibiotalar joint (example CoLag 6.7mm screw) are required for ankle fusion procedure.

    Device Description

    The CoLink® NeoFuse MIS Plating System includes sterile fusion plates and screws for anterior fixation of ankle arthrodesis and fractures, including the distal tibia and talus. The addition of a medial-lateral compression screw through the tibiotalar joint is required. These subject devices are a line extension of the CoLink® NeoFuse Plating System.

    AI/ML Overview

    The provided text describes a medical device, the "CoLink® NeoFuse MIS Plating System," which is a fusion plate used for anterior fixation of ankle arthrodesis and fractures. This document is a 510(k) Pre-market Notification from the FDA, indicating a review of the device's substantial equivalence to existing legally marketed devices.

    Based on the provided text, the following information can be extracted regarding the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of numerical acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as would be typical for an AI-powered diagnostic device. Instead, the "acceptance criteria" are implied by demonstrating substantial equivalence to predicate devices through various performance tests and comparisons.

    The primary acceptance criteria for this type of medical device (a bone fixation system) revolve around:

    • Mechanical Performance: The device must meet sufficient mechanical strength and fatigue properties to perform its intended function.
    • Biocompatibility: The materials used must be safe for implantation in the human body.
    • Sterility: The device must be sterile and maintain its sterility.
    • Packaging Integrity: The packaging must protect the device and maintain sterility until use.
    • Dimensional and Material Equivalence: The subject device's geometry, dimensions, and materials should be comparable to predicate devices.

    Table of Performance (Based on provided text):

    Performance AspectAcceptance Criteria (Implied by Substantial Equivalence to Predicates)Reported Device Performance
    MechanicalThe mechanical performance should be comparable to or better than predicate devices for bone fixation. Demonstrated adequacy through testing simulating in vivo loading conditions. (ASTM F382, modified). The device should withstand anticipated forces without failure and maintain stability for fixation."Testing was conducted per a modified ASTM F382 set up for the plates for the CoLink NeoFuse MIS Plating System and the plates were found to be substantially equivalent." "The test setup was modified from ASTM F382 to more accurately simulate anticipated mechanical loading on the plate in vivo." "No additional mechanical testing was required for the screws." (This implies the screws either meet requirements via predicate comparison or prior testing, which isn't detailed for the subject device.)
    BiocompatibilityThe device materials must be biologically safe for human implantation. (ISO 10993-1)."The CoLink NeoFuse MIS Plating System was validated per ISO 10993-1 for biocompatibility..."
    SterilizationThe device must be terminally sterilized and maintain sterility. Its sterilization method (gamma) must be validated to achieve a specified sterility assurance level (SAL). (ISO 11137-2)."...ISO 11137-2 for gamma sterilization..."
    PackagingThe packaging must maintain product sterility and integrity until the point of use. (ISO 11607-1)."...and ISO 11607-1 for packaging."
    EndotoxinsThe device must have endotoxin levels below specified limits for implantable devices. (ANSI/AAMI ST72)."Endotoxin testing was completed per ANSI/AAMI ST72."
    MaterialMust use medical-grade materials suitable for implantation, comparable to predicate devices."Titanium Alloy (ASTM F136)" - Same as predicates.
    Intended Use & Indications for Use (IFU)The subject device's intended use and Indications for Use should be substantially equivalent to the predicate devices, allowing for minor differences that do not raise new questions of safety or effectiveness. Minimal differences from predicate IFU regarding anatomical locations and required additional screws.Subject device: "The CoLink® NeoFuse MIS Plating System is a fusion plate indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia, talus, and calcaneus. The addition of a posterior compression screw and a fully threaded screw through the tibiotalar joint (example CoLag 6.7mm screw) is required." Predicate K213069: "...including the distal tibia, talus, and calcaneus. The addition of a compression screw through the tibiotalar joint (example CoLag 6.7mm screw) is required." (Minor difference: subject specifies "posterior compression screw and a fully threaded screw" vs. just "compression screw").

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide numerical sample sizes for mechanical testing (e.g., number of plates tested). It states "Testing was conducted per a modified ASTM F382 set up for the plates...". For biocompatibility, sterilization, and packaging, it refers to validation according to international standards (ISO, ANSI/AAMI), which implies that appropriate sample sizes were used as dictated by those standards for validation studies, but these specific numbers are not detailed in this summary.

    There is no mention of data provenance (e.g., country of origin or retrospective/prospective nature) as this is not a clinical study involving human patient data, but rather performance testing of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The device is a physical bone fixation system, not a diagnostic AI device requiring expert ground truth for interpretation of medical images or patient outcomes. The "ground truth" for the mechanical performance testing would be the physical properties and failure points observed during the standardized tests, compared against established engineering and regulatory benchmarks.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods are relevant for studies involving human interpretation (e.g., of medical images) where discrepancies need to be resolved. This document details engineering and performance testing, not clinical interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. An MRMC study is relevant for AI-powered diagnostic devices that assist human readers. The CoLink® NeoFuse MIS Plating System is a surgical implant, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance testing is based on:

    • Standardized Test Methods: Adherence to and performance within established engineering standards (e.g., ASTM F382, ISO 10993-1, ISO 11137-2, ISO 11607-1, ANSI/AAMI ST72).
    • Predicate Device Performance: The underlying assumption of substantial equivalence is that the predicate devices are safe and effective, and thus the subject device, performing comparably, is also considered safe and effective.

    8. The sample size for the training set

    This is not applicable. There is no "training set" as this device is not an AI algorithm requiring machine learning.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1