510(k) Data Aggregation

    K Number
    K251117

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-25

    (14 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the cervical spine and are placed via an anterior approach at the C2 to T1 disc levels using autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should have received 6 weeks of non-operative treatment prior to treatment with the devices. If the device is being used without the CSTS Integrated Plate, supplemental fixation must be used.

    Device Description

    The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone through growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy. The device is available in a variety of heights, footprints, and lordotic angles. The CSTS Integrated Plates come in multiple heights and orientations to accommodate the patient's anatomy. Screws are inserted through the anterior portion of the Integrated Plates into adjacent vertebral bodies for bony fixation. The Integrated Plates have a rotating locking tab to prevent back-out of the screws.

    AI/ML Overview

    This FDA 510(k) clearance letter for the Cervical Spine Truss System (CSTS) is for a physical medical device (an intervertebral body fusion device), not an AI/software device. Therefore, the information typically required to describe the acceptance criteria and study proving an AI/software device's performance (as outlined in the prompt's requested information points) is not present in this document.

    The document discusses the mechanical and safety performance of the physical device and its substantial equivalence to a predicate device, but it does not contain details about:

    • Acceptance criteria for an algorithm's performance
    • Sample sizes for data used in AI model testing
    • Expert involvement in ground truth establishment for AI
    • Adjudication methods for AI test sets
    • MRMC studies for AI assistance
    • Standalone AI performance
    • Training set details for AI

    The 510(k) summary explains that the purpose of this specific submission (K251117) is to "add additional configurations to the integrated plate product offering and to provide the integrated plate as sterile packaged." It then states that "The modifications made to the CSTS Integrated Plate for this submission did not create a new worst-case condition for any mechanical or MR safety tests. The results of the predicate testing are applicable to the subject devices and no further testing was required."

    In summary, based solely on the provided FDA 510(k) clearance letter for the Cervical Spine Truss System (CSTS), it is not possible to answer the detailed questions about acceptance criteria and study design for an AI/software device's performance because the device itself is a physical implant, not an AI or software product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231739

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-25

    (103 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the cervical spine and are placed via an anterior approach at the C2 to T 1 disc levels using autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should have received 6 weeks of non-operative treatment prior to treatment with the devices. If the device is being used without the CSTS Integrated Plate. supplemental fixation must be used.

    Device Description

    The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone through growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy. The device is available in a variety of heights, footprints, and lordotic angles. The CSTS Integrated Plates come in multiple heights to accommodate the patient's anatomy. Screws are inserted through the anterior portion of the Integrated Plates into adjacent vertebral bodies for bony fixation. The Integrated Plates have a rotating locking tab to prevent back-out of the screws.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Cervical Spine Truss System - CSTS). This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily through non-clinical performance testing (mechanical and material testing, and finite element analysis).

    It's crucial to understand that 510(k) submissions for devices like the CSTS (intervertebral body fusion devices) typically do not involve clinical studies with human participants in the same way that software-as-a-medical-device (SaMD) or drug trials do. Therefore, the concepts of "acceptance criteria for device performance studies," "sample sizes for test sets (in a clinical context)," "expert adjudication of ground truth," "MRMC studies," "standalone algorithms," and "training sets for AI models" as described in your prompt, are not applicable to this type of regulatory submission or the information contained within this document.

    The "study that proves the device meets the acceptance criteria" in this context refers to the non-clinical performance testing that verifies the mechanical integrity and safety of the implant.

    Here's how to address your prompt based only on the provided text, while clarifying the differences:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical "acceptance criteria" for the mechanical tests in a tabular format, nor does it provide specific numerical "reported device performance" values. Instead, it states that the tests were completed "per the following standards" (ASTM F2077, ASTM F2119, ASTM F2182, ASTM F2213, ASTM F2051, and "accepted industry standard" for Expulsion Testing) and that the results "show that the strength of the CSTS interbody and CSTS Interbody with an Integrated Plate is sufficient for its use and is substantially equivalent to legally marketed predicate devices."

    Table: Non-Clinical Performance Testing and Outcomes

    Acceptance Criteria (Implied by Standards)Reported Device Performance (Summary)
    Meets ASTM F2077 for Static Axial CompressionStrength sufficient for use; substantially equivalent to predicates
    Meets ASTM F2077 for Static Compression ShearStrength sufficient for use; substantially equivalent to predicates
    Meets ASTM F2077 for Static TorsionStrength sufficient for use; substantially equivalent to predicates
    Meets ASTM F2077 for Dynamic Axial Compression (With and Without Integrated Plate)Strength sufficient for use; substantially equivalent to predicates
    Meets ASTM F2077 for Dynamic Compression Shear (With and Without Integrated Plate)Strength sufficient for use; substantially equivalent to predicates
    Meets ASTM F2077 for Dynamic Torsion (With and Without Integrated Plate)Strength sufficient for use; substantially equivalent to predicates
    Meets Screw Pushout Testing (per accepted industry standard)Strength sufficient for use; substantially equivalent to predicates
    Meets Expulsion Testing (per accepted industry standard)Strength sufficient for use; substantially equivalent to predicates
    Validated Finite Element Analysis (FEA) demonstrating non-inferiority to existing devicesWorst-case CSTS device for each endplate geometry is not a new worst-case compared to previously tested CSTS devices.
    Meets ASTM F2119 for MR Image ArtifactResults support substantial equivalence to predicates for MR compatibility.
    Meets ASTM F2182 for MR Induced HeatingResults support substantial equivalence to predicates for MR compatibility.
    Meets ASTM F2213 for MR Induced TorqueResults support substantial equivalence to predicates for MR compatibility.
    Meets ASTM F2051 for MR Induced Displacement ForceResults support substantial equivalence to predicates for MR compatibility.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified in terms of number of surgical cases or patient data, as this is not a clinical study. For mechanical testing, "sample size" refers to the number of physical devices or components tested, which is not detailed in this summary.
    • Data Provenance: Not applicable in the context of clinical data. The "data" refers to results from laboratory mechanical and MR compatibility testing, and finite element analysis.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This concept is not applicable as the "ground truth" for a mechanical device like this is established by adherence to engineering standards and validated models, not by expert consensus on clinical images or outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This concept is not applicable in this context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • An MRMC study is not applicable as this is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This concept is not applicable as this is not an AI/software device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" (or verification of performance) for this device relies on:
      • Adherence to established ASTM (American Society for Testing and Materials) standards for intervertebral body fusion device mechanical testing and MR compatibility.
      • Validated Finite Element Analysis (FEA), which is a computational method used to predict how a product reacts to real-world forces, heat, vibration, etc.
      • Demonstration of substantial equivalence to previously cleared predicate devices through direct comparison of technological characteristics (indications for use, materials, function, sizes) and performance testing.

    8. The sample size for the training set

    • This concept is not applicable as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established

    • This concept is not applicable as this is not an AI/machine learning device.

    In summary: The provided document is a regulatory submission for a physical medical implant. Its "acceptance criteria" and "proof" relate to meeting established engineering and material standards, and demonstrating substantial equivalence to similar devices already on the market, rather than clinical efficacy studies involving human participants or AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221266

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-05

    (248 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cervical Spine Truss System (CSTS) Interbody Fusion Device is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the cervical spine at one level or two contiguous disc levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. CSTS Interbody Fusion Devices are used as an adjunct to fusion in the cervical spine and are placed via an anterior approach at the C2 to T1 disc levels using autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Patients should have received 6 weeks of non-operative treatment prior to treatment with the devices. The device must be used with supplement fixation.

    Device Description

    The device is an open architecture truss design mathematically formulated to provide structural support with open space throughout the implant for bone through growth and fusion. The 4WEB additive manufacturing process provides a hierarchical surface roughness. The implant is made from Ti6Al4V alloy. The device is available in a variety of sizes to accommodate the patient's anatomy.

    AI/ML Overview

    The provided text describes a medical device, the Cervical Spine Truss System (CSTS) Interbody Fusion Device, and its 510(k) premarket notification to the FDA. This submission is for an updated version of an already cleared device, intended to introduce design changes to the interbody devices.

    However, the document focuses on mechanical performance testing using engineering standards (like ASTM F2077, F2267-04, F2119, F2182, F2213, F2052) to demonstrate the device's strength and substantial equivalence to predicate devices. It does not present any clinical study data or involve acceptance criteria related to a device's diagnostic or therapeutic performance in the context of human-machine interaction or AI performance.

    Therefore, many of the requested elements for describing acceptance criteria and a study that proves the device meets them (such as human reader improvement with AI assistance, expert consensus for ground truth, sample sizes for training/test sets for an algorithm, etc.) are not applicable to this document.

    The document primarily demonstrates the structural and material integrity of the medical implant through non-clinical testing.

    Here's what can be extracted and what cannot:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Based on the provided text, the acceptance criteria are not in the form of specific performance metrics for an AI/diagnostic device, but rather compliance with established engineering standards for interbody fusion devices.

    Acceptance Criteria (from Performance Standards)Reported Device Performance
    Static and dynamic axial compression per ASTM F2077Demonstrated sufficient strength and substantial equivalence to predicate devices.
    Static and dynamic axial compression shear per ASTM F2077Demonstrated sufficient strength and substantial equivalence to predicate devices.
    Static and dynamic torsion per ASTM F2077Demonstrated sufficient strength and substantial equivalence to predicate devices.
    Subsidence Testing per ASTM F2267-04Demonstrated sufficient strength and substantial equivalence to predicate devices.
    Expulsion testingDemonstrated sufficient strength and substantial equivalence to predicate devices.
    MR image artifact per ASTM F2119Demonstrated substantial equivalence to predicate devices.
    RF induced heating per ASTM F2182Demonstrated substantial equivalence to predicate devices.
    MR induced torque per ASTM F2213Demonstrated substantial equivalence to predicate devices.
    MR induced displacement force per ASTM F2052Demonstrated substantial equivalence to predicate devices.
    No new worst-case compared to previously cleared devices for mechanical performance (via validated finite element analysis)Achieved; design changes do not introduce a new worst-case.

    2. Sample size used for the test set and the data provenance: Not applicable. This involves mechanical testing of physical devices or simulations, not a test set of data like in an AI/diagnostic study. The "test set" here refers to the physical devices manufactured for testing. Data provenance is not relevant in the context of mechanical engineering tests on manufactured components.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts: Not applicable. Ground truth for mechanical properties is established by the rigorous application of testing standards and physical measurements, not by human experts adjudicating images or clinical data.

    4. Adjudication method for the test set: Not applicable for mechanical or simulation testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical medical device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device.

    7. The type of ground truth used: For the mechanical properties, the "ground truth" is derived from the physical laws governing material science and mechanics, validated through standardized testing methods (e.g., ASTM standards) and finite element analysis.

    8. The sample size for the training set: Not applicable. There is no "training set" in the context of validating a mechanical implant through performance testing. Finite element analysis (FEA) models are "trained" or developed based on material properties and engineering principles, but not in the same sense as an AI algorithm.

    9. How the ground truth for the training set was established: Not applicable. See point 8.