Found 2 results

510(k) Data Aggregation

    K Number
    K191301

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-11

    (120 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ceribell Pocket EEG Device is intended to record and to present the EEG signals in visual and audible formats in real time. The visual and audible signals assist trained medical staff to make neurological diagnoses. The Pocket EEG Device is intended to be used in a professional healthcare facility environment.

    Additionally, the EEG Recording Viewer Software component of the Pocket EEG Device incorporates a Seizure Detection component that is intended to mark previously acquired sections of EEG recordings in patients greater than or equal to 18 years of age that may correspond to electrographic seizures in order to assist qualified clinical practitioners in the assessment of EEG traces. The Seizure Detection component provides notifications to the user when detected seizure prevalence is "Frequent," "Abundant," or "Continuous," per the definitions of the American Clinical Neurophysiology Society Guideline 14. Notifications include an on-screen display on the Pocket EEG Device and the optional sending of an e-mail message to a clinician. Delays of up to several minutes can occur between the beginning of a seizure and when the Seizure Detection notifications will be shown to a user.

    The Pocket EEG Device does not provide any diagnostic conclusion about the subject's condition and Seizure Detection notifications cannot be used as a substitute for real time monitoring of the underlying EEG by a trained expert.

    Device Description

    The Ceribell Pocket EEG Device is a previously cleared EEG monitoring system which includes a portable 8-channel system that is being upgraded in this 510(k) notification K191301 to include a seizure detection module. The device connects to 10 patient electrodes (5 left, 5 right), which are used to form the 8 channels. The device may be used with any scalp EEG electrodes, and the system includes the following components:

    • Pocket EEG Device: a portable, battery powered, 8-channel EEG monitoring system.
    • Power adapter: 100-240 V AC power adapter used to charge the Pocket EEG Device.
    • Micro-USB cable: cable used to connect Pocket EEG Device to power adapter for charging and to connect to a computer to transfer EEG recording files. When the Pocket EEG Device is connected to a power adapter of a computer, all EEG acquisition functions are automatically disabled.
    • EEG Recording Viewer Software: EEG review software for viewing EEG recordings on a computer. The EEG Recording Viewer Software includes a Seizure Detection software module that assists qualified users in reviewing and annotating EEG by marking previously acquired sections of EEG that may correspond to electrographic seizures.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test DescriptionAcceptance CriteriaReported Device Performance (PPA, 95% CI Lower Bound)Conformance to PPA Acceptance CriteriaReported Device Performance (FP/hr, 95% CI Upper Bound)Conformance to FP/hr Acceptance Criteria
    "Frequent" seizure activity notifications (≥10% seizure burden), PPA and FP/hrCeribell non-inferior to Persyst 13 with 10% non-inferiority marginNot directly reported in tableN/A (non-inferiority not in table)Not directly reported in tableN/A (non-inferiority not in table)
    "Frequent" seizure activity notifications (≥10% seizure burden), PPA and FP/hrPPA: 95% CI lower bound ≥ 70%90.00%PASS0.351PASS
    "Abundant" seizure activity notifications (≥50% seizure burden), PPA and FP/hrFP/hr: 95% CI upper bound ≤ 0.446
    "Abundant" seizure activity notifications (≥50% seizure burden), PPA and FP/hrPPA: 95% CI lower bound ≥ 70%81.39%PASS0.381PASS
    "Continuous" seizure activity notifications (≥90% seizure burden), PPA and FP/hrFP/hr: 95% CI upper bound ≤ 0.446
    "Continuous" seizure activity notifications (≥90% seizure burden), PPA and FP/hrPPA: 95% CI lower bound ≥ 70%78.26%PASS0.272PASS
    FP/hr: 95% CI upper bound ≤ 0.446
    Seizure Burden output, L1-distance vs. reference standardCeribell non-inferior to Persyst 13 with 10% non-inferiority marginNot directly reported in tableN/ANot directly reported in tableN/A

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 60 subjects (30 from ICU dataset and 30 from EMU dataset).
    • Data Provenance: Retrospective. Data was obtained from previous studies that retrospectively reviewed all adult EEGs from patients who underwent inpatient EEG monitoring over a period of time. Data from one hospital was used for ICU EEG data, and data from a second hospital was used for EMU EEG data. The study states the validation dataset was "from previous studies" and "retrospectively obtained from pre-existing EEG databases." The specific country of origin is not explicitly stated, but the submission is to the U.S. FDA, implying the data is likely from the U.S.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: An "expert panel" was used, with an agreement between "at least two neurologists required." The exact number of individual experts beyond "at least two" is not specified.
    • Qualifications of Experts: EEG trained neurologists (physicians who have obtained fellowship training in epilepsy or neurophysiology). They were affiliated with "multiple different institutions."

    4. Adjudication Method for the Test Set

    • The ground truth was established by an "expert panel, with agreement between at least two neurologists required to determine the reference standard." This indicates an adjudication method where agreement between at least two experts was necessary. It suggests a form of consensus-based ground truth.

    5. Was a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Done? If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

    • No, a multi-reader, multi-case (MRMC) comparative effectiveness study with human readers assisting the AI or vice-versa was not explicitly described in this document. The study compared the Ceribell device's performance to a predicate device (Persyst 13) and a consensus-based ground truth established by experts. The device's role is to "assist qualified clinical practitioners" but the study format does not evaluate the improvement of human readers with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done?

    • Yes, a standalone performance evaluation of the Ceribell Seizure Detection module was done. The module was run on the validation dataset and its performance metrics (PPA and FP/hr) were calculated against the established reference standard. While the device's intended use is to "assist trained medical staff," the validation primarily assessed the algorithm's ability to detect seizures independently.

    7. The Type of Ground Truth Used

    • The ground truth used was expert consensus. It was established by "EEG trained neurologists who reviewed and annotated the EEG recordings for seizure episodes." Specifically, "Each seizure episode was manually annotated by an expert panel, with agreement between at least two neurologists required to determine the reference standard."

    8. The Sample Size for the Training Set

    • The sample size for the training set is not specified in the provided document. The document explicitly states that the validation dataset was "completely separate and independent from the data used to design and train the algorithm," and "No patients from the validation dataset were used for algorithm training or development," but it does not provide details on the training data itself.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not provide details on how the ground truth for the training set was established. It only emphasizes that the validation dataset was separate from the training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K170363

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-05-08

    (91 days)

    Product Code
    Regulation Number
    882.1400
    Age Range
    0.5 - 12
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ceribell Pocket EEG Device is intended to record and to present the EEG signals in visual and audible formats in real time. The visual and audible signals assist trained medical staff to make neurological diagnoses. The Pocket EEG Device does not provide any diagnostic conclusion about the subject's condition and does not provide any automated alerts of an adverse clinical event. The Pocket EEG Device is intended to be used in a professional healthcare facility environment.

    Device Description

    The Ceribell Pocket EEG Device (subject) is a portable 8-channel EEG monitoring system. The device connects to 10 patient electrodes (5 left, 5 right), which are used to form the 8 channels, and may be used with any scalp EEG electrodes. The Device includes the following items:

    • Pocket EEG Device: a portable, battery powered, 8-channel EEG monitoring system. .
    • Power adapter: 100-240 V AC power adapter used to charge the Pocket EEG Device. ●
    • . Micro-USB cable: cable used to connect Pocket EEG Device to power adapter for charging and to connect to a computer to transfer EEG recording files. When the Pocket EEG Device is connected to a power adapter of a computer, all EEG acquisition functions are automatically disabled.
    • EEG Recording Viewer Software: EEG review software for viewing EEG recordings on a . computer.
    AI/ML Overview

    I am sorry, but based on the provided text, there is no acceptance criteria or study information related to device performance in terms of diagnostic effectiveness or clinical outcomes.

    The provided 510(k) summary focuses entirely on:

    • Technical specifications and safety testing (e.g., electrical safety, electromagnetic compatibility, battery safety, mechanical robustness).
    • Comparison to a predicate device based on intended use and technological characteristics (e.g., number of channels, digital vs. analog technology).

    The device is explicitly stated to not provide any diagnostic conclusion or automated alerts. Its function is to "record and to present the EEG signals in visual and audible formats in real time" to "assist trained medical staff to make neurological diagnoses." Therefore, there would not be an "acceptance criteria" related to diagnostic accuracy for this particular device as it does not perform automated diagnosis.

    To directly answer your specific questions based only on the provided text:

    1. A table of acceptance criteria and the reported device performance:

      • No acceptance criteria or reported device performance for diagnostic accuracy or clinical outcomes are mentioned. The performance data listed are for safety and technical standards compliance.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable as no clinical performance study involving a test set for diagnostic accuracy is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable as no clinical performance study involving ground truth establishment is described.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as no clinical performance study involving a test set is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study or AI assistance is mentioned. The device presents raw EEG signals to trained medical staff.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device does not provide "diagnostic conclusions" or "automated alerts." It is explicitly a human-in-the-loop device where trained medical staff make diagnoses based on the presented signals.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable as no clinical performance study requiring ground truth is described.
    8. The sample size for the training set:

      • Not applicable as no machine learning algorithm development for diagnostic purposes is described. The device is a data acquisition and presentation tool.
    9. How the ground truth for the training set was established:

      • Not applicable as no machine learning algorithm development for diagnostic purposes is described.

    In summary, the provided document describes a 510(k) clearance for a medical device that records and presents EEG signals for trained medical staff to interpret, and it focuses on demonstrating the device's technical and safety performance against established standards, not on its diagnostic efficacy or any AI-driven capabilities.