Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K172603

    Validate with FDA (Live)

    Date Cleared
    2017-11-27

    (89 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cerafix Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 4.9 in2 (31.7cm2) or less in area. For example, 4.0 in x 1.2 in (10.0 cm x 3.1 cm) would be an acceptable defect size.

    Device Description

    Cerafix® Dura Substitute is a resorbable implant for repair of dural defects. The device can be applied as an onlay matrix or sutured in place. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.

    AI/ML Overview

    The provided text describes the Cerafix® Dura Substitute, a medical device for dura mater repair, and its equivalence to predicate devices, particularly focusing on expanding its indicated defect size and application method. The document, however, does not contain a detailed study proving acceptance criteria for an AI/ML device. Instead, it discusses the substantial equivalence of the Cerafix® Dura Substitute based on technological characteristics and animal studies, which is typical for a 510(k) submission for a non-AI medical device.

    Therefore, many of the requested points regarding AI/ML device acceptance criteria and studies cannot be answered from this document. I will focus on the information that is present regarding the non-AI device.


    Acceptance Criteria and Device Performance (Based on the non-AI device context):

    The document describes "Indications for Use" and "Technological Characteristics" which serve as the de facto acceptance criteria for the expanded use of the Cerafix® Dura Substitute via a 510(k) submission. The "reported device performance" is primarily demonstrated through equivalence to a predicate device and side-by-side animal studies.

    Acceptance Criteria (from Proposed Indications for Use)Reported Device Performance
    Indicated as a dura substitute for the repair of dura mater.Subject device (Cerafix® Dura Substitute) has identical technological characteristics, principles of operation, material performance, and biocompatibility to the reference device (previously approved Cerafix® Dura Substitute, K153613, K161278). The reference device was indicated for dura mater repair. Side-by-side animal studies in a canine duraplasty model (dural defects 18 mm x 25 mm) showed "equivalent safety and performance between the subject and predicate device."
    Indicated for defects of 4.9 in² (31.7 cm²) or less in area.Previously, the device was indicated for defects of 4.4 in² (28.3 cm²). The submission includes "Data included in this submission to justify defect increase" (though the data itself is not presented in this summary). The side-by-side animal study utilized dural defects of 18 mm x 25 mm (equal to 4.5 cm² or ~0.7 in²), which is consistent with the general purpose of evaluating a dura substitute but is smaller than the maximum defect size being justified. The equivalence argument for the larger defect size relies on "Data included in this submission to justify defect increase" in comparison to the reference device, which had a slightly smaller indicated defect size.
    May be applied as an onlay matrix or sutured in place.The previous version (reference device) was applied with "tensionless suture application." The primary predicate device (DuraMatrix™ Collagen Dura Substitute) "can be cut by surgeon and placed on dural defect and used as an onlay membrane or sutured in place." The subject device has "Equivalent" principles of operation to the predicate regarding application methods. Side-by-side animal implantation studies were performed with the subject and predicate device "utilizing dural defects (18 mm x 25 mm) without the use of suture (onlay)." Results showed "equivalent safety and performance."
    Biocompatibility: BiocompatibleBiocompatibility testing was previously submitted for K153613 and K161278 and confirmed. "Biocompatible" is listed as a common characteristic with both the reference and predicate devices.
    Sterility: Sterile, SAL 10⁻⁶"Sterile, SAL 10⁻⁶" is listed as a common characteristic with the reference device. "Sterile" is listed for the predicate.
    Pyrogenicity: Non-pyrogenic"Non-pyrogenic" is listed as a common characteristic with the reference and predicate devices.
    Resorbable: Yes"Yes" is listed as a common characteristic with the reference device. The predicate is listed as "Not Applicable," but the subject device's resorbable nature is considered "Equivalent to reference device."

    Regarding AI/ML Specific Information (points 2-9):

    The provided document describes a 510(k) submission for a physical medical device (Cerafix® Dura Substitute), not an AI/ML driven software device. Therefore, the following points are not applicable to this submission based on the provided text:

    1. Sample size used for the test set and the data provenance: Not applicable, as this is a physical device, not an AI/ML algorithm requiring a test set for performance evaluation in the described manner.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML algorithm's predictions is not relevant here.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable in the context of AI/ML ground truth. The "ground truth" for this physical device's performance is derived from biological/physiological responses in animal models and comparison to known predicate device performance.
    7. The sample size for the training set: Not applicable. This is not an AI/ML algorithm that requires a training set.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of Relevant "Study" for the physical device:

    • Study Type: Side-by-side animal implantation studies.
    • Model: Canine duraplasty model.
    • Procedure: Dural defects (18 mm x 25 mm) created. Both the subject device (Cerafix® Dura Substitute) and the predicate device were implanted without the use of suture (onlay).
    • Outcome Measured: "Equivalent safety and performance" between the subject and predicate devices.
    • Conclusion: This animal study supported the expanded indication for onlay application and reinforced the safety and performance for dura repair. The justification for the increased defect size (from 4.4 in² to 4.9 in²) is stated as "Data included in this submission to justify defect increase," but the specific data from that justification is not detailed in this summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161278

    Validate with FDA (Live)

    Date Cleared
    2016-08-08

    (94 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cerafix® Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 4.4 in2 (28.3 cm2) or less in area. For example, 4.0 in x 1.1 in (10.1 cm x 2.8 cm) would be an acceptable defect size.

    Device Description

    Cerafix® Dura Substitute is a resorbable implant for repair of dural defects and is to be used with tensionless sutures. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.

    AI/ML Overview

    The provided document is a 510(k) summary for the Cerafix Dura Substitute. It describes the device, its indications for use, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's a breakdown of the acceptance criteria and study information provided:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" with numerical targets and reported performance values for each criterion in the way one might expect for a diagnostic or AI device. Instead, it focuses on demonstrating "substantial equivalence" to a predicate device.

    The primary "acceptance criteria" implicitly revolve around demonstrating equivalent performance to the predicate device in terms of:

    • Safety and Efficacy: No significant differences in adverse events or repair outcomes.
    • Biocompatibility: No new biocompatibility concerns.
    • Mechanical Properties: Equivalent per unit area to the predicate.
    • Neoduralization and Resorption: Similar mechanisms and timelines to the predicate.
    • Absence of Complications: No CSF leaks, hydrocephalus, hemorrhage, or infection.

    The reported device performance is that the subject device (new Cerafix Dura Substitute) was found to be equivalent to the predicate device (previously cleared Cerafix Dura Substitute - K153613) in all these aspects.

    Key Comparison Points and Performance (Implicit Acceptance Criteria and Reported Performance):

    Acceptance Criteria (Implicit)Reported Device Performance (Subject Device)
    Principles of Operation: Cut by surgeon, placed with tensionless suture, 2-3mm suture line, 1cm overlap.Equivalent to predicate device.
    Material of Construction: Porous polymer matrix, Porous PGLA / PDO matrix.Equivalent to predicate device.
    Surgical Application Restrictions: No specific orientation requirement.Equivalent to predicate device.
    Sterility: Sterile, SAL 10-6.Equivalent to predicate device.
    Packaging: Double sterile pack, nested pouch in chipboard envelope.Equivalent to predicate device.
    Pyrogenicity: Non-pyrogenic.Equivalent to predicate device.
    Resorbable: Yes.Equivalent to predicate device.
    Biocompatibility: Biocompatible.Equivalent to predicate device (no new biocompatibility testing was conducted; relied on previous submission for predicate).
    Mechanical Properties: Equivalent per unit area.Equivalent to predicate device (no new mechanical testing was conducted; relied on previous submission for predicate).
    Clinical Performance (Animal Study): Absence of CSF leaks, hydrocephalus, hemorrhage, infection.No CSF leaks observed in either group throughout the duration of both studies. All animals appeared healthy with normal neurological evaluations.
    Clinical Performance (Animal Study): Neoduralization and absorption mechanism.Similar in the mechanism of neoduralization and absorption, independent of the size of the induced dural defect. Demonstrated signs of resorption with infiltration of fibrovascular connective tissue and successful neoduralization.
    Clinical Performance (Animal Study): Overall safety and efficacy.Equivalent clinical performance at each time point, successfully repaired induced dural defects independent of defect size. Demonstrates equivalent safety and efficacy when compared to the predicate device.
    Indications for Use: Repair of dura mater for defects up to 4.4 in² (28.3 cm²).The subject device supports repair of dura mater for defects up to 4.4 in² (28.3 cm²), which is a larger indicated defect size than the predicate device (1.9 in² (12.5cm²)). The side-by-side animal study concluded equivalency independent of defect size, supporting this larger indication.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: The "test set" was the animal study.
      • Canine Bilateral Duraplasty Model: "Each time point evaluated 3 test and 3 control animals, each with 2 defects."
      • This means a total of 6 animals per time point (3 for subject device, 3 for predicate device).
      • Since it states "at both time points" (implying at least two), it would be a minimum of 12 animals in total (6 animals x 2 time points), generating 24 defects.
    • Data Provenance: The study was a "canine bilateral duraplasty model." This indicates it was a prospective animal study. The country of origin of the data is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Ground Truth Establishment: For the animal study, tissue samples were processed by histopathology techniques and analyzed for dural integrity/neoduralization and local tissue reactions according to ISO 10993-6.
    • Number and Qualifications of Experts: The document does not specify the exact number or qualifications of the individuals who performed the histopathology analysis or interpreted the results. It only mentions "tissue samples from each defect site were processed by histopathology techniques and analyzed."

    4. Adjudication method for the test set

    • The document does not describe an adjudication method for the animal study results. The histopathology analysis would likely have been performed by a qualified pathologist, but no multi-reader review or consensus method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This document pertains to a medical device (dura substitute), not an AI algorithm. Therefore, there is no discussion of human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm only) performance study was not done. This is a hardware medical device, not an algorithm.

    7. The type of ground truth used

    • For the animal study, the ground truth was established through histopathology analysis (morphological and cellular evaluation of tissue samples) and clinical observations (e.g., absence of CSF leaks, neurological evaluations). These are essentially expert observations/assessments based on established scientific methods.

    8. The sample size for the training set

    • This device is not an AI algorithm; therefore, there is no training set in the context of machine learning. The "predicate device" study (K153613) serves as a baseline/reference, from which the current subject device draws its "equivalent" conclusions regarding mechanical and biocompatibility data.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for an AI algorithm. The performance of the predicate device (K153613) was established through its own set of non-clinical and potentially animal/clinical studies, which presumably used similar ground truth methods (e.g., pathology, clinical assessment).
    Ask a Question

    Ask a specific question about this device

    K Number
    K153613

    Validate with FDA (Live)

    Date Cleared
    2016-03-16

    (90 days)

    Product Code
    Regulation Number
    882.5910
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cerafix® Dura Substitute is indicated as a dura substitute for the repair of dura mater. This device is indicated for defects of 1.9 in² (12.5cm²) or less in area. For example, 1.2 in x 1.6 in (3 cm x 4 cm) would be an acceptable defect size.

    Device Description

    Cerafix® Dura Substitute is a resorbable implant for repair of dural defects and is to be used with tensionless sutures. Cerafix® Dura Substitute is a soft, white, pliable, nonfriable, porous polymer matrix. Cerafix® Dura Substitute is available in a variety of sizes and is supplied sterile and nonpyrogenic in a single-use nested pouch configuration, which is enclosed within a protective chipboard envelope.

    AI/ML Overview

    The provided text describes the Cerafix® Dura Substitute, a medical device intended for the repair of dura mater, and its journey through FDA 510(k) clearance. The document details the device's characteristics, indications for use, and the non-clinical testing performed to establish its substantial equivalence to predicate devices.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for many of the mechanical and biological tests were framed as "Equivalent to Predicate or Reference Device" or "Meets Final Device Specification." For some, specific thresholds were mentioned.

    TestAcceptance CriteriaReported Device Performance
    Mechanical Testing
    ThicknessEquivalent to Predicate or Reference DevicePASS
    Mass per AreaEquivalent to Predicate or Reference DevicePASS
    Tensile StrengthEquivalent to Predicate or Reference DevicePASS
    Suture Pull-Out StrengthEquivalent to Predicate or Reference DevicePASS
    Burst StrengthEquivalent to Predicate or Reference Device; and burst strength greater than anticipated intracranial pressuresPASS (burst strength greater than anticipated intracranial pressures)
    Shrink TemperatureShow stability at applicable temperaturesPASS (showed stability)
    Fiber DiameterMeets Final Device SpecificationPASS (meets specification)
    Pore SizeMeets Final Device SpecificationPASS (meets specification)
    Biocompatibility Testing
    ISO Cytotoxicity MEM ElutionNon-cytotoxicCell culture exhibited no reactivity; non-cytotoxic.
    Guinea Pig Maximization - SensitizationNon-irritating, no sensitization responseDid not elicit a sensitization response; non-irritant.
    Intracutaneous Irritation ReactivityNon-irritatingNon-irritating.
    Hemolysis AssayNon-hemolyticFound to be non-hemolytic.
    Genotoxicity (Mouse Lymphoma Assay)Non-genotoxicEquivalent to negative control; non-genotoxic.
    Genotoxicity (Mouse Micronucleus Assay)Non-mutagenicConsidered non-mutagenic.
    Genotoxicity (Bacterial Mutagenicity)Non-mutagenicConsidered non-mutagenic.
    Pyrogenicity (Rabbit Pyrogen Test)Non-pyrogenicExhibited a negative response; non-pyrogenic.
    Acute Systemic ToxicityNon-toxicConsidered non-toxic.
    Endotoxin TestingLess than 2.15 EU/deviceLess than 2.15 EU/device; non-pyrogenic.
    Subchronic Toxicity (90-day animal study)Non-toxicShowed the device to be non-toxic.
    Chronic Toxicity (180-day animal study)Non-toxicShowed the device to be non-toxic.
    Side-by-Side Animal StudyEquivalent safety and performance to predicate deviceShowed equivalent safety and performance.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each mechanical test (e.g., number of samples tested for tensile strength or burst strength). It mentions "side-by-side bench testing versus the predicate or commercially available reference device" for mechanical tests, and for biocompatibility, it refers to standard ISO/ASTM tests using animals (e.g., guinea pigs, rabbits, mice) and cell cultures. The data provenance is pre-clinical testing, likely conducted in a laboratory setting. There is no mention of country of origin for the data or whether it was retrospective or prospective in the context of human data, as this is a pre-market clearance based on non-clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for this device's clearance is based on established scientific principles and comparison to legally marketed predicate devices through defined acceptance criteria in mechanical and biocompatibility testing, not on expert consensus of clinical data.

    4. Adjudication method for the test set

    Not applicable. This device clearance relies on objective laboratory and animal testing, not human-based adjudication of clinical outcomes or images.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to the pre-market clearance of a physical medical implant (dura substitute), not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness is discussed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an algorithm or AI device.

    7. The type of ground truth used

    The "ground truth" for this regulatory submission is a combination of:

    • Predicate Device Equivalence: The primary ground truth is established by demonstrating that the Cerafix® Dura Substitute's technological characteristics, performance, and safety are substantially equivalent to a legally marketed predicate device (Ethisorb™ Dura Patch) and a reference device (DuraGen Plus™ Dural Regeneration Matrix).
    • Established Scientific Standards: Compliance with ISO and ASTM standards for biocompatibility and mechanical properties (e.g., non-cytotoxic, non-pyrogenic, appropriate burst strength).
    • Animal Study Outcomes: Equivalence in safety and performance based on side-by-side animal implantation studies compared to the predicate device.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of this device's pre-market clearance, as it's not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1