Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K212071

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-13

    (529 days)

    Product Code
    Regulation Number
    882.5891
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cefaly Dual Connected - OTC and Cefaly Dual Enhanced with RFID - OTC are indicated for:

    1. Acute treatment of migraine with or without aura in patients 18 years of age or older
    2. Preventative treatment of migraine in patients 18 years of age or older

    Cefaly Dual Connected - Rx and Cefaly Dual Enhanced With RFID - Rx are indicated for:

    1. Acute treatment of migraine with or without aura in patients 18 years of age or older
    2. Prophylactic treatment of migraine in patients 18 years of age or older
    Device Description

    The Cefaly® Dual Series of devices are small, non-invasive, and portable devices meant to be worn on the forehead using a self-adhesive electrode to treat migraine similar to Cefaly® Dual (K201895 OTC and K173006 R.). The devices are similar in construction and materials and use the same components (electronics and electrical components, and firmware), including the component that comes into patient contact which is the CEFALY® Electrode. The notable differences between the Subject and Predicate devices are Bluetooth technology and mobile application for the purpose of displaying device data, the new docking station, RFID for electrode detection, audio/visual indicators, and the dimensions and the weight of the electrode.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the Cefaly Dual Series devices. It details the device's characteristics and compares them to predicate devices to demonstrate substantial equivalence. Crucially, this document does not contain information about a study that proves the device meets specific acceptance criteria in the format typically used for demonstrating the performance of a novel AI/ML-based medical device.

    The document states:

    • "No animal testing was required to justify the differences with the corresponding Predicate Devices." (page 40)
    • "No clinical testing was required to justify the differences with the corresponding Predicate Devices." (page 40)

    This indicates that the submission relies on demonstrating substantial equivalence to existing devices through bench testing and comparison of technological characteristics, rather than new clinical performance data to prove efficacy. The changes made to the device (e.g., Bluetooth, RFID, audio/visual indicators, charging system, weight, dimensions, electrode thickness, packaging) are assessed against safety and electrical standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10) and adhesion performance.

    Therefore, I cannot provide the requested information for a study proving acceptance criteria for an AI/ML device because:

    • No acceptance criteria for an AI/ML device are listed. The document pertains to a Transcutaneous Electrical Nerve Stimulator, not a device incorporating AI/ML for diagnostic or therapeutic decision-making.
    • No study proving device performance against such criteria is presented. The "Performance Testing - Bench" section (Table 5.7.1) lists general electrical and EMC safety standards, and adhesion testing for electrodes, but not performance metrics related to migraine treatment outcomes that would typically be evaluated in a clinical trial.

    Without a study described as such in the provided text, I cannot fill in the requested table or additional details. The information provided heavily focuses on hardware changes and their compliance with general safety standards, and substantiates that the functionality of the device (e.g., electrical stimulation parameters) remains equivalent to the cleared predicate devices, thus not requiring new clinical evidence of effectiveness for migraine treatment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201895

    Validate with FDA (Live)

    Device Name
    Cefaly Dual
    Manufacturer
    Date Cleared
    2020-09-29

    (83 days)

    Product Code
    Regulation Number
    882.5891
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cefaly® Dual is indicated for

    • The acute treatment of migraine with or without aura in patients 18 years of age or older.
    • The prophylactic treatment of episodic migraine in patients 18 years of age or older.
    Device Description

    The Cefaly® Dual device indicated for an OTC (also referred to as the Subject Device) is a supraorbital transcutaneous electrical nerve stimulator device to be applied on the forehead. A self-adhesive electrode with 2 conductive zones is placed on the forehead. This double electrode is directly connected to the device. The Subject Device is operated by a rechargeable battery. Pressure on the single button allows selecting and starting a stimulation program, which runs automatically. The electrical impulses generated by the Subject Device are transmitted transcutaneously via the supraorbital electrode to excite (trigger action potentials on) the supratrochlearis and supraorbitalis nerves. Supratrochlearis and supraorbitalis (or supratrochlear and supraorbital) nerves belong to the upper branch of the trigeminal nerve (V1). Therefore, the supraorbital neurostimulation is also known as external trigeminal nerve stimulation. The supraorbital neurostimulation generates an analgesic effect and is intended to treat migraine headaches.

    AI/ML Overview

    This document is a 510(k) premarket notification approval letter for the Cefaly® Dual device, indicating its substantial equivalence to a previously cleared predicate device. It primarily focuses on the device's technological characteristics, indications for use, and a non-clinical human factors study to support a change from prescription (Rx) to over-the-counter (OTC) use.

    Based on the provided text, the device itself does not involve an AI/ML algorithm that requires specific acceptance criteria related to its performance in tasks like image classification or predictive modeling. Instead, it's a Transcutaneous Electrical Nerve Stimulator (TENS) device. Therefore, many of the requested points regarding AI/ML acceptance criteria, performance metrics (like sensitivity, specificity, AUC), ground truth establishment, training sets, and MRMC studies are not applicable to this submission.

    The "study that proves the device meets the acceptance criteria" in this context refers to demonstrating the device's equivalent performance and safety to its predicate, particularly for its proposed OTC use.

    Here's a breakdown based on the document:


    Acceptance Criteria for Cefaly® Dual (for OTC use):

    The acceptance criteria are implicitly tied to demonstrating the substantial equivalence to the predicate device (Cefaly® Dual K173006) and showing that the device is safe and effective for over-the-counter use without requiring a prescription.

    The document highlights the following as crucial for demonstrating equivalence/acceptance:

    • Identical Technological Characteristics: The device must be technologically identical to the predicate device in design, materials, energy source, electrical impulse parameters (pulse width, frequency, current, waveform), and software.
    • Identical Intended Use/Clinical Application: The clinical application and intended users (patients 18 years or older for acute treatment of migraine with or without aura, and prophylactic treatment of episodic migraine) must remain the same as the predicate.
    • Safety and Effectiveness for OTC Use: The primary additional "acceptance criterion" for the OTC claim is demonstrating that the device can be used safely and effectively by lay users without professional guidance. This is addressed through a Human Factors validation study.

    Study Proving Device Meets Acceptance Criteria:

    The study that proves the device meets the (implicit) acceptance criteria for OTC marketing is primarily a Non-Clinical Human Factors Validation Study.

    1. Table of Acceptance Criteria (Implicit) and Reported Device Performance:

    Acceptance Criterion (Implicit)Reported Device Performance and Evidence from Submission
    Technological Equivalence to Predicate DeviceMet: The Subject Device (Cefaly® Dual for OTC) is stated to be "identical to the legally marketed Predicate Cefaly® Dual (K173006) in terms of technological characteristics (design, material, and energy source)." Specific parameters like power source, channels, software, maximum output current, pulse width, frequency, session duration, waveform, shape, net charge per pulse, current density, and power density are listed as "Equivalent" to the Predicate.
    Identical Clinical Application and Intended UsersMet: The indications for use for acute and prophylactic treatment of migraine in patients 18 years or older are explicitly stated to be the same as the predicate device. The document explicitly says, "While the clinical application, the intended users and use remains identical..."
    Safety and Effectiveness for Over-The-Counter (OTC) UseMet: A "Human Factors validation study was conducted with the Cefaly® Dual (Subject Device)" to support the change in indications for use from Rx to OTC. The successful completion of this study implied the device is safe and effective for use by lay users. The FDA's clearance (K201895) further confirms this.
    Compliance with International StandardsMet: The Subject Device is stated to be "compliant to the same international standards as the legally marketed Predicate (K173006)."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document only mentions a "Human Factors validation study" but does not specify the sample size for this study.
    • Data Provenance: Not explicitly stated, but Human Factors studies are typically prospective tests conducted in a simulated or real-world use environment. The manufacturer is CEFALY Technology, based in Belgium, so participants could be from Europe or the US, but this is not detailed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This question is not applicable in the context of this device. A TENS device does not generate data that requires expert labeling or ground truth establishment in the way an AI/ML algorithm for medical image analysis would.
    • For Human Factors studies, "experts" are typically human factors specialists, usability engineers, and potentially clinicians who understand the device's use environment, rather than experts establishing "ground truth" in clinical data. The document does not detail the expertise of those involved in the human factors study design or evaluation.

    4. Adjudication Method for the Test Set:

    • Not applicable for this type of device and study. Human Factors studies involve observing user interactions, identifying use errors, and assessing user comprehension, rather than adjudicating clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, this was not done. The device cleared via a 510(k) pathway by demonstrating substantial equivalence to a predicate. The document explicitly states: "Clinical performance testing was not performed with the Subject Device to support equivalence, as there are no differences between the Subject Device and the Predicate Device."
    • MRMC studies are typically for AI/ML algorithms where human performance with and without AI assistance is being compared. This device is a direct-to-patient TENS.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:

    • Not applicable. This device is hardware with simple embedded software. It is not an AI/ML algorithm that operates in a "standalone" analytical capacity. Its performance is tied directly to its hardware function and user interaction.

    7. The Type of Ground Truth Used:

    • Not applicable for clinical efficacy. As noted, clinical performance testing was not performed for this submission. The "ground truth" for the Human Factors study would be the identification of all potential use errors and the successful demonstration that lay users can safely and effectively operate the device according to its instructions for use. This isn't "ground truth" in the diagnostic or prognostic sense.

    8. The Sample Size for the Training Set:

    • Not applicable. This device does not use an AI/ML algorithm that is "trained" on a dataset in the conventional sense. The software functions are fixed programs (Program 1 for acute treatment, Program 2 for prophylactic treatment) identical to the predicate device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173006

    Validate with FDA (Live)

    Device Name
    Cefaly Dual
    Manufacturer
    Date Cleared
    2017-11-28

    (62 days)

    Product Code
    Regulation Number
    882.5891
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cefaly® Dual is indicated for:

    • The acute treatment of migraine with or without aura in patients 18 years of age or older.

    • The prophylactic treatment of episodic migraine in patients 18 years of age or older.

    Device Description

    The Cefaly® Dual device is a supraorbital transcutaneous electrical nerve stimulator device to be applied on the forehead. A self-adhesive electrode with 2 conductive zones is placed on the forehead. This double electrode is directly connected to the device.

    The Cefaly® Dual is operated by a rechargeable battery. A pressure on the single button allows selecting and starting a stimulation program, which runs automatically.

    The electrical impulses generated by the Cefaly® Dual device are transmitted transcutaneously via the supraorbital electrode to excite (trigger action potentials on) the supratrochlearis and supraorbitalis nerves. Supratrochlearis and supraorbitalis (or supratrochlear and supraorbital) nerves belong to the upper branch of the trigeminal nerve (VI). Therefore, the supraorbital neurostimulation is also known as external trigeminal nerve stimulation. The supraorbital neurostimulation generates an analgesic effect and is intended to treat migraine headaches.

    AI/ML Overview

    This document does not contain the detailed information necessary to complete the requested table and provide answers for all points. The provided text is a 510(k) summary for the Cefaly® Dual device, making a claim of substantial equivalence to two predicate devices (Cefaly® Acute and Cefaly®). While it describes the device's technical specifications and indications for use, it does not include the results of a clinical study or performance testing against specific acceptance criteria for efficacy or safety.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. Table of acceptance criteria and reported device performance:

    This information is not available in the provided text. The document focuses on showing the Cefaly® Dual is technically equivalent to its predicates. It does not present clinical performance data or acceptance criteria for that data in this summary.

    2. Sample size used for the test set and the data provenance:

    This information is not available as no clinical test set or study data is presented.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not available as no clinical test set or ground truth establishment is described.

    4. Adjudication method for the test set:

    This information is not available.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not available. The Cefaly® Dual is a transcutaneous electrical nerve stimulator, not an AI-assisted diagnostic device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The Cefaly® Dual is a medical device, not an algorithm.

    7. The type of ground truth used:

    This information is not available.

    8. The sample size for the training set:

    This information is not available as no training set for a machine learning algorithm is mentioned.

    9. How the ground truth for the training set was established:

    This information is not available.


    Summary of what the document does provide:

    • Device: Cefaly® Dual, a supraorbital transcutaneous electrical nerve stimulator.
    • Indications for Use:
      • Acute treatment of migraine with or without aura in patients 18 years of age or older.
      • Prophylactic treatment of episodic migraine in patients 18 years of age or older.
    • Mechanism of Action: Transmits electrical impulses via a supraorbital electrode to excite supratrochlearis and supraorbitalis nerves (external trigeminal nerve stimulation) to generate an analgesic effect.
    • Substantial Equivalence: Claimed against Cefaly® Acute (K171446) and Cefaly® (K160237). The Cefaly® Dual combines the stimulation programs and indications of these two predicate devices.
    • Technical Characteristics: Detailed comparison of Cefaly® Dual with its predicates, showing identical physical dimensions, weight, housing materials, electrodes, power source, pulse width, waveform, and net charge per pulse. The key difference is that Cefaly® Dual includes two fixed programs (one for acute treatment at 100 Hz, one for prophylactic at 60 Hz), whereas each predicate only had one.
    • Non-Clinical Performance Testing: Cefaly® Dual is compliant with the same international standards (IEC 60601-1, -1-2, -1-6, -1-11, -2-10, IEC 62366) as the predicate devices. This indicates electrical safety, electromagnetic compatibility, usability, and specific requirements for nerve stimulators.

    The provided text is a regulatory submission for substantial equivalence based on technical and performance similarities to previously cleared devices, rather than a report on a clinical trial with specific acceptance criteria and outcome measurements for the new device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1