    K Number
    K182830

    Validate with FDA (Live)

    Date Cleared
    2019-04-26

    (203 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown is intended to protect health care personnel and patients from the transfer of microorganisms, body fluids, and particulate material. Not intended for use in the operating room.

    The proposed Cardinal Health™ Poly-Coated Open-Back Protective Gown meets the barrier protection requirements of AAMI Level 4 per ANSI/AAMI PB70:2012. Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, but has an open back which is non-protective.

    The Cardinal Health™ Poly-Coated Open-Back Protective Gowns is a single use, disposable medical device provided non-sterile.

    Device Description

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown is a Class II medical device under the FDA product code of FYC, General & Plastic Surgery Panel, and Regulation 21 CFR 878.4040. The device description of the Cardinal Health™ Poly-Coated Open-Back Protective Gown is in accordance with the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and Guidance for Industry and FDA Staff: Premarket Notification Requirements Concerning Gowns Intended for Use in Healthcare Settings, issued on December 9, 2015.

    The Cardinal Health™ Poly-Coated Open-Back Protective Gown consists of a laminate (blue spunbond polypropylene coated with polyethylene) throughout the body of the gown (including sleeves). The sleeves have a thumbhook cuff formed into the sleeve or have a knit cuff sewn onto sleeve for keeping the sleeves in place on the wearer. The collar has a neck removal feature and the belt ties are integrated into the body. These gowns are non-sterile, single use, disposable gowns that are packaged in pastic bags and/or dispenser boxes. These gowns are offered in blue and come in both Universal and XX-Large sizes.

    AI/ML Overview

    The document describes the non-clinical performance testing of the Cardinal Health™ Poly-Coated Open-Back Protective Gown. It evaluates the device against specific acceptance criteria and details the study methods used.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Element of ComparisonSpecification (Acceptance Criteria)Reported Device Performance (Mean (min/max))
    Basis Weight (ASTM D3776/D3776M-17)39.0 gsm min (average) 42.0 gsm target 45.0 gsm max (average)43.3 (39.9 / 46.1) gsm (Sleeve and Body material)
    Tensile Strength, CD (ASTM D5034-17)5.8 lb min (individual) 3.5 lb min (average)14.2 (12.6 / 16.1) lb
    Tear Strength, MD (ASTM D5587-15)6.7 lb min (individual) 2.8 lb min (average)6.0 (4.4 / 7.5) lb
    Flammability (CPSC, Part 1610)Class 1Class 1
    Hydrostatic Head (AATCC 127:2017)140 cmH20 target 100 cmH20 min (individual)>130 (130 / >130) cmH20 >51 (51 / >51) inH20
    Viral barrier (ASTM F1671-13)PassPass
    Liquid Barrier Performance Classification Properties (ANSI/AAMI PB70:2012 / ASTM F1671)Meets AAMI Level 4 performance requirementsAll areas tested meet Level 4 performance requirements
    BiocompatibilityNon-cytotoxic, non-irritating, non-sensitizing per ISO-10993-1Non-cytotoxic, non-irritating, non-sensitizing and met requirements per ISO-10993-1

    2. Sample size used for the test set and the data provenance

    • Liquid Barrier Performance Classification Properties: Testing was performed using 3 lots and 32 samples per lot in each critical zone area (body or sleeve, and heat sealed seam(s)).
    • Other tests (Basis Weight, Tensile Strength, Tear Strength, Hydrostatic Head): The document provides mean and min/max values, but a specific sample size for these individual tests beyond the "Finished Good Test Results" is not explicitly stated.
    • Data Provenance: The document implies that the testing was conducted internally by the manufacturer, Cardinal Health, or by a contracted lab. The data is presented as "Finished Good Test Results" for the "Subject Device: Cardinal Health™ Poly-Coated Open-Back Protective Gown." This data would be considered prospective as it's generated specifically for the device submission. The country of origin for the data is not specified, but the submission is to the U.S. FDA by a company based in Illinois, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The tests described are non-clinical performance tests that use standardized methodologies (e.g., ASTM, AATCC, ISO standards) to measure physical and barrier properties. These types of tests do not typically involve human expert interpretation or "ground truth" in the same way clinical studies or diagnostic device evaluations would. The "ground truth" is established by the test method itself and the specified performance criteria. Therefore, no human experts were used to establish ground truth for this test set in the context of diagnostic interpretation.

    4. Adjudication method for the test set

    Not applicable. As noted above, these are non-clinical, objective performance tests following standardized methods, not subjective clinical assessments requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a protective gown, which is a physical medical device, not an AI-powered diagnostic tool. Therefore, an MRMC study related to human readers and AI assistance is not relevant or presented.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm or AI system.

    7. The type of ground truth used

    The ground truth for the non-clinical performance tests is based on predefined objective standards and specifications (e.g., AAMI Level 4 for barrier protection, specific tensile strength values, flammability classes). These standards define what constitutes a "pass" or "fail" for each characteristic.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model, and therefore there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.