510(k) Data Aggregation

    K Number
    DEN130018

    Validate with FDA (Live)

    Device Name
    CYTOSCAN(R) DX
    Manufacturer
    Date Cleared
    2014-01-17

    (30 days)

    Product Code
    Regulation Number
    866.5920
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CytoScan® Dx Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) in genomic DNA obtained from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. CytoScan® Dx Assay is intended for the detection of CNVs associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features. Assay results are intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, including confirmation by alternative methods, parental evaluation, clinical genetic evaluation, and counseling, as appropriate. Interpretation of assay results is intended to be performed only by healthcare professionals, board certified in clinical cytogenetics or molecular genetics. The assay is intended to be used on the GeneChip® System 3000Dx and analyzed by Chromosome Analysis Suite Dx Software (ChAS Dx Software).

    This device is not intended to be used for standalone diagnostic purposes, preimplantation or prenatal testing or screening, population screening, or for the detection of, or screening for, acquired or somatic genetic aberrations.

    Device Description

    The CytoScan Dx consists of five reagent modules, a microarray kit, and analysis software. The five reagent modules are:

      1. MOD R L A, CytoScan® Dx Pre-PCR contains buffers, nucleotides, enzyme, and primers and adaptors for amplification;
      1. MOD T E W, CytoScan® Dx Pre-PCR contains buffer and nuclease free water for amplification;
      1. MOD F L H, CytoScan® Dx Post-PCR contains buffers, nucleotides, and enzyme for fragmentation, labeling and hybridization;
    • MOD S AH W PB, CytoScan® Dx Post-PCR contains buffers, nuclease free water, and purification beads for stain and array hold;
      1. MOD E PW CytoScan® Dx Post-PCR contains buffer for elution and purification wash.

    The CytoScan® Dx Post-PCR CytoScan® Dx Array kit, 6-pack is designed for 6 runs. The microarray contains approximately 2,696,550 functional markers, each of which is approximately 25 bases long.

    ChAS Dx Analysis Software and Browser v1.0.0 analyzes CEL file microarray data.

    AI/ML Overview

    The Affymetrix® CytoScan® Dx Assay is a qualitative assay intended for the postnatal detection of copy number variations (CNV) in genomic DNA from peripheral whole blood in patients referred for chromosomal testing based on clinical presentation. It detects CNVs associated with developmental delay, intellectual disability, congenital anomalies, or dysmorphic features.

    Here's an analysis of the acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly listed in a single table within the provided text. However, they can be inferred from the "Required Mitigations" and the detailed performance study descriptions. The reported device performance is extracted from the study results.

    Inferred Acceptance Criteria and Reported Device Performance for Affymetrix® CytoScan® Dx Assay

    Acceptance Criteria CategorySpecific Performance Metric / CriterionReported Device Performance
    Reproducibility (Site-to-Site)Overall pairwise replicate agreement for CNVs (>=50% overlap, same state)79.4% for all CNVs (range 65.7-100%). 75.5% for gains, 82.4% for losses (kb-based, 50% overlap). 82.2% for all CNVs (marker-based, 50% overlap).
    Overall pairwise replicate agreement for CNVs (>=80% overlap, same state)71.9% for all CNVs (kb-based, 80% overlap).
    Reproducibility of Copy Number (0, 1, 2, 3, 4+)82.7% agreement across all CNVs (regardless of location/size/marker number).
    Median % Absolute Endpoint Deviation (Overall, excluding HR)0.03 (mean, 0.00 min, 0.00 median, 0.77 max) for both markers and kb.
    Reproducibility of LOH calls (3-4 Mb)91.9% call rate; 92.4% pairwise replicate agreement (50% overlap); 86.0% (80% overlap).
    Detection Limit (DNA Input)Reliable detection of CNV markers10 ng genomic DNA is a conservative lower limit; assay functions appropriately with input up to 500 ng.
    Detection Limit (Mosaicism)Reliable detection of mosaic segmentsReliably detects mosaicism greater than 20% (mean AUC=0.96 at 15% mosaicism, but still states >20% being reliable).
    Analytical Specificity (Interfering Substances)Array QC and assay performance to meet acceptance criteria in presence of bilirubin, triglycerides, hemoglobin.All results met sponsor's acceptance criteria; no interference observed.
    Analytical Specificity (Carryover)No significant difference in copy number state determination or array QC metrics between first and fifth runs in sequential processing.No significant difference in copy number state (Wilcoxon signed rank statistic (S) = 0.5, p-value = 1.00) or array QC metrics (p > 0.05).
    Analytical Specificity (Cross-Contamination)SNPQC ≥ 12 at approximately 20% contamination.SNPQC values ≥12 indicate absence of substantial cross-contamination.
    Accuracy (Overall CNVs vs. Sequencing)Agreement between CytoScan Dx and sequencing (>50% overlap or size, same state)78.8% agreement (all CNVs).
    Accuracy (Overall CNVs vs. Composite Method)Agreement between CytoScan Dx and composite method (>50% overlap or size, same state)88.7% agreement (all CNVs, excluding hypervariable regions).
    Accuracy (LOH Regions vs. Sequencing)PPA for LOH regions >=3 MB100.0% (159/159) for LOH regions 3 MB and greater.
    Accuracy (Syndromic LOH vs. Published)Detection of known LOH regions (e.g., from Gibson et al.)Identified all 20 (100.0%) specifically described regions (>=5 MB) by Gibson et al.
    Endpoint AccuracyEndpoint agreement (<=12 markers for losses, <=25 markers for gains) for CNVs with state agreement (excluding HR).93.4% overall (94.8% for gains, 92.4% for losses).
    Clinical Sensitivity (Sample Level)PPA between RPC (pathogenic) and CytoScan Dx (pathogenic)82.0% (105/128).
    Clinical Sensitivity (CNV Level)Analytical accuracy: percentage of RPC-identified CNVs identified by CytoScan Dx.94.0% (639/680) (analytical agreement).
    Clinical Sensitivity (Syndrome Agreement)Overall interpretation positive agreement.80.7%.
    Expected Values (Normal Population)Interpretation of CytoScan Dx Assay results for phenotypically normal individuals.3 out of 108 phenotypically normal individuals were interpreted as pathogenic.
    Minimum CNV Size DetectedSet at 25 kb and 25 markers for losses, 50 kb and 50 markers for gains.Reproducibility for gains (50-75kb) 78.8% (50% overlap); losses (25-50kb) 81.0% (50% overlap).Accuracy for gains (50-75kb) 93.5%; losses (25-50kb) 91.6% (using composite method).
    Minimum LOH Size DetectedSet at 3 Mb for LOH.Reproducibility of 3-4 Mb LOH 91.9% (call rate); accuracy with sequencing 100.0% (3-5 Mb).

    2. Sample Sizes and Data Provenance

    Reproducibility Studies (Site-to-Site and Lot-to-Lot):

    • Test Set Sample Size: 93 genomic DNA samples (48 purified from blood, 44 from cell lines from Coriell and WHO, 1 control from ATCC) for site-to-site study. 47 genomic DNA samples (including 1 control) for lot-to-lot study (12 blood gDNA, 11 cell-line gDNA, 1 cell-line control gDNA).
    • Data Provenance: The general nature of the samples (purified blood, cell lines from Coriell, WHO, ATCC) suggests a mix of sources, likely from the United States or international repositories. The site-to-site study implies data collection across three sites (details on geographic location not specified beyond "3 sites"). The lot-to-lot study was conducted at "1 investigative site - Affymetrix, Santa Clara, CA."
    • Retrospective/Prospective: These appear to be retrospective studies using pre-existing genomic DNA samples.

    Accuracy Study (Comparison to Sequencing & Composite Method):

    • Test Set Sample Size: 137 gDNA samples (48 purified from blood, 86 from Coriell Cell Repository cell lines, 3 from ATCC cell lines). 5 samples were excluded, leaving 132 evaluable samples.
    • Data Provenance: Similar to reproducibility studies, a mix of blood samples and cell lines from public repositories.
    • Retrospective/Prospective: Retrospective.

    Clinical Sensitivity Study:

    • Test Set Sample Size: 960 gDNA samples.
    • Data Provenance: Retrospective clinical study using samples from previously tested DD/ID (developmental delay/intellectual disability) patients referred to three clinical laboratories for routine chromosomal testing. The origin of the patients is not explicitly stated but implies clinical settings, likely within the US, given the FDA context.
    • Retrospective/Prospective: Retrospective, but samples were "collected prospectively" for the original routine testing, then retrospectively analyzed with CytoScan Dx.

    Expected Values/Reference Range Study:

    • Syndromic Samples: 149 samples representing known syndromes.
    • Normal Samples: 108 phenotypically normal subjects.
    • Data Provenance: Syndromic samples collected at 8 sites, normal samples at 1 site. No specific country mentioned.
    • Retrospective/Prospective: Retrospective.

    3. Number of Experts and Qualifications for Ground Truth

    Clinical Sensitivity Study (Establishing Clinical Interpretation Ground Truth):

    • Number of Experts: Two independent cytogeneticists for interpreting CytoScan Dx results. The historical "routine patient care (RPC)" interpretations were made by various cytogeneticists at the original clinical laboratories.
    • Qualifications: "Healthcare professionals, board certified in clinical cytogenetics or molecular genetics." This is stated in the intended use and product classification.

    Accuracy Study (Analytical Ground Truth):

    • Number of Experts: Not applicable in the same way as clinical interpretation. The ground truth was established by "validated high throughput sequencing" and a "composite analytical method" based on molecular methods. These methods themselves would have been developed and validated by experts, but the determination of CNVs from these methods is a technical analysis rather than expert consensus on clinical significance.
    • Qualifications: Not specified beyond the methods themselves being "validated."

    4. Adjudication Method

    • Clinical Sensitivity Study: For CytoScan Dx interpretations, results were provided to one of two independent cytogeneticists who then generated an overall clinical laboratory interpretation. It does not explicitly state that both cytogeneticists interpreted every case, nor does it describe a formal adjudication process if their interpretations differed. However, the study involved comparing these interpretations to historical RPC interpretations.
    • Analytical Ground Truth (e.g., Sequencing, Composite Method): Adjudication methods for the ground truth itself (sequencing results, composite method results) are not detailed, implying standard analytical pipelines were used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A formal MRMC comparative effectiveness study where human readers interpret cases with and without AI assistance (or device assistance in this context) to measure an "effect size of how much human readers improve with AI vs without AI assistance" was not explicitly described in the provided text.
    • The clinical sensitivity study involved human cytogeneticists interpreting CytoScan Dx results and comparing them to historical RPC interpretations. This demonstrates the performance of the device in a clinical context but doesn't directly quantify the improvement of human readers with vs. without the device's assistance in real-time. The "diagnostic yield" comparison (13.8% for CytoScan Dx vs 13.3% for RPC, stratified by RPC methods) suggests comparable performance to existing complex workflows, but not necessarily an "improvement effect size" for a single reader.

    6. Standalone Performance Study (Algorithm Only)

    • The CytoScan Dx Assay includes the Chromosome Analysis Suite Dx Software (ChAS Dx Software) for analysis. The "Performance Characteristics" section describes the analytical performance (precision, accuracy, detection limits) of the device as a whole, which inherently includes the algorithm's performance in detecting CNVs.
    • However, the device's intended use specifically states: "Interpretation of assay results is intended to be performed only by healthcare professionals, board certified in clinical cytogenetics or molecular genetics. This device is not intended to be used for standalone diagnostic purposes".
    • Therefore, while the algorithm's performance is integral to the device's overall performance, a study focusing solely on the algorithm's "standalone" performance without any human-in-the-loop interpretation (e.g., direct output of diagnosis, bypassing expert review) was not presented as this falls outside its intended use as an interpretative tool for experts. The analytical accuracy studies effectively evaluate the algorithm's ability to detect CNVs compared to reference methods.

    7. Type of Ground Truth Used

    • Analytical Performance Studies (Reproducibility, Accuracy, Detection Limits):

      • Sequencing: For overall CNV accuracy and LOH accuracy. This is a high-resolution molecular method.
      • Validated High-Throughput Method: Also used for CNV accuracy.
      • Composite Analytical Method: Combining results from various molecular methods (e.g., sequencing, qPCR) was used when sequencing alone didn't confirm CNVs. This aims to provide a more robust ground truth.
      • Published LOH Regions: For LOH accuracy, comparison to a published table from Gibson et al., derived from HapMap sample genotypes, which can be considered highly curated and expert-validated data.
      • Known Chromosomal Aberrations: For detection limit and matrix comparison studies, cell lines with "known chromosomal aberrations" were used.
    • Clinical Sensitivity Study:

      • Clinical Interpretation by Board-Certified Cytogeneticists: This formed the ground truth for potential diagnostic yield and sample/syndrome agreement. The comparison was against "routine patient care (RPC)" interpretations, which themselves were based on various methods (karyotype, FISH, microarray, etc.) and interpreted by cytogeneticists.

    8. Sample Size for the Training Set

    • The provided document describes validation studies, not product development or machine learning model training. It does not specify a separate "training set" for the algorithms within the ChAS Dx software.
    • The signal intensity of the patient sample is compared to a "reference DNA, which is based on an average of over 400 samples." This "reference DNA" could be considered analogous to a training or calibration dataset used to establish baseline copy number profiles, but it's not described as a "training set" in the context of supervised machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • As mentioned above, there is no explicit "training set" described for a machine learning algorithm. The "reference DNA, which is based on an average of over 400 samples" is key for baseline comparison.
    • The method for establishing the ground truth for these 400+ reference samples is not detailed in the provided text. It is implied these are normal, healthy samples used to establish a baseline for copy number variations.