Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K052135

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-20

    (137 days)

    Product Code
    Regulation Number
    876.4480
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cyberwand Dual Probe Lithotripter is intended to be used for the fragmentation of urinary tract calculi in the kidney, ureter and bladder.

    Device Description

    The Cybersonics, Cyberwand Dual Probe Ultrasonic Lithotripsy system is an electromechanical device capable of fragmenting and aspirating calculi. The hand piece consists of an ultrasonic transducer containing the piezo-electronic elements, which are driven by a generator operating at 21 ± 1 kHz. The resulting longitudinal waves are propagated along the ultrasonic dual probe to the target stone. The ultrasonic transducer probes are hollow, permitting simultaneous suction.

    AI/ML Overview

    The provided K052135 document describes the Cybersonics Cyberwand Dual Probe Lithotripter, which is intended for the percutaneous fragmentation and removal of kidney stones. The basis for substantial equivalence is primarily against other legally marketed ultrasonic lithotripters, specifically the EMS Swiss Lithoclast Ultra (K012445) and the Karl Storz Calcuson (K973251).

    Acceptance Criteria and Study Details

    The document does not explicitly state "acceptance criteria" as a set of quantified thresholds. Instead, it presents a comparison of technological characteristics and performance metrics, particularly drill rate, against predicate devices to demonstrate substantial equivalence. The implication is that if the Cyberwand performs "as well or better" than the predicates, it meets the unstated acceptance criteria for safety and effectiveness.

    Here’s a breakdown of the information requested based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Implied)Reported Device Performance (Cybersonics Cyberwand)Predicate Performance (e.g., LithoClast Ultra)
    Technological Characteristics:
    Operating Frequency20–22 kHz24–26 kHz (LithoClast Ultra); 25–26kHz (Storz Calcuson)
    Utilities120–240 Vac120–240 Vac
    Autoclavable probes and hand pieceYesYes
    Probes threaded for assemblyYesYes
    Probe length (standard)400 mm400 mm
    Probe OD ≤ 3.75mmYesYes
    Ultrasonic Probe tip excursion ≤100 µmYesYes
    Detachable hand pieceYesYes
    Handpiece with rear suction connectionYesYes
    Piezo stack with Titanium hornYesYes
    Detachable footswitchYesYes
    Table Top GeneratorYesYes
    Applied Part BFYesYes
    Performance (Drill Rate):
    Mean Drill Rate (seconds)8.7 +/- 0.910.9 +/- 1.0 (LithoClast Ultra)

    Notes on Acceptance Criteria: The document primarily focuses on demonstrating that the Cyberwand's technological characteristics are comparable to predicate devices. For performance, the "acceptance criterion" appears to be that its drill rate is comparable to or better than predicate devices. The study results show the Cyberwand had a faster mean drill rate than all listed predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Drill Rate Study):
      • Cybersonics Cyberwand: 10 stones
      • LithoClast Ultra: 10 stones
      • Olympus LUS-2: 25 stones
      • Circon-ACMI USL-2000: 25 stones
      • Karl Storz Calcuson: 15 stones
      • Olympus LUS-1: 12 stones
      • Richard Wolf Model 2271-004: 10 stones
    • Data Provenance: The drill rate data was supplied by Indiana University School of Medicine. This suggests a prospective study specifically designed for this comparison.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The "ground truth" for the drill rate study appears to be the objective measurement of time taken to drill artificial stones. There is no mention of experts establishing a subjective ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. The drill rate study involves objective measurements of time, not subjective assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a medical instrument (lithotripter), not an AI-powered diagnostic or assistive tool for human readers. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance study was done in the form of the drill rate comparison. The "algorithm" here is the device itself, and its performance (drill rate) was measured independently without human intervention influencing the fragmentation process, beyond operating the device.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    • The ground truth for the drill rate study was objective performance measurement based on artificial stone fragmentation time. This is a direct physical measurement, not dependent on expert consensus, pathology, or clinical outcomes data.

    8. The Sample Size for the Training Set

    • Not applicable/provided. This submission does not describe a machine learning or AI algorithm that would require a training set. The device is a physical electromechanical lithotripter.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable/provided for the reasons stated above (no AI/ML training set).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1