Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K230427

    Validate with FDA (Live)

    Date Cleared
    2023-07-11

    (144 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:
    Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:
    Neurosurgery - including removal of primary and secondary malignant and benign brain and spinal tumors, including but not limited to meningiomas and gliomas
    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
    Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
    General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The device within the scope of this premarket notification is the optional CUSA® Electrosurgery Module (CEM) accessory that is intended to be used with the 23 kHz components of the CUSA® Clarity Ultrasonic Surgical Aspirator System.

    The purpose of this submission is to modify the CEM nosecone accessory currently offered with CUSA Clarity to allow for connection with additional electrosurgical generators, to continue to provide electrosurgical capabilities to the user. The additional electrosurgical generators that the modified CEM Nosecone may be used with include the Medtronic FT10 (K191601), Medtronic FX8 (K181389), Erbe VIO 300D (K083452), and Erbe VIO 3 (K190823). Compatibility with the Medtronic Force FX (K143161) will be maintained as well.

    The CUSA Clarity 23kHz Expanded CEM Nosecone has the same intended use and technological characteristics as the predicate CUSA Clarity 23 kHz CEM Nosecone (K190180). The subject CEM nosecone will continue to allow the surgeon to apply immediate electrosurgical coagulation to bleeding tissue at the surgical site, with the same handpiece assembly that is removing unwanted tissue.

    AI/ML Overview

    The provided text describes the CUSA Clarity Ultrasonic Surgical Aspirator System, specifically focusing on a modified CEM nosecone accessory for the 23 kHz components. This submission is a 510(k) premarket notification claiming substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present acceptance criteria in a quantitative table format with corresponding device performance values. Instead, it describes various non-clinical tests undertaken to ensure the safety and efficacy of the device and its substantial equivalence to the predicate.

    The reported device performance is broadly stated as:

    • "Testing was determined successful and supports the conclusion that all product specifications and design inputs have been met."
    • "The results of the non-clinical testing indicate that the intended use of the device, fundamental scientific technology, and performance of the CUSA Clarity 23 kHz Expanded CEM Nosecone is substantially equivalent to the predicate device."

    Therefore, a table of quantitative acceptance criteria and specific reported device performance cannot be generated from the provided text. The information indicates that all tests were passed and specifications met, implying the device performed within acceptable limits.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical testing but does not specify a "test set" in terms of subject or patient data. The tests are focused on device characteristics rather than clinical outcomes with a patient population.

    • Sample size for test set: Not applicable in the context of device performance testing described. The tests are on the device itself (e.g., handpiece life, functionality, mechanical properties).
    • Data provenance: Not applicable in the context of patient data. The provenance is from internal testing conducted by Integra LifeSciences Corporation. There is no mention of country of origin for data related to clinical or patient studies, as none were performed. The tests are prospective as they were conducted as part of the submission process.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of experts: Not applicable. The "ground truth" for the non-clinical tests is established by engineering specifications, recognized standards (FDA guidance documents, ISO standards for biocompatibility, EMC, and electrical safety), and comparison to the predicate device's established performance.
    • Qualifications of experts: Not specified as a separate set of experts for ground truth. However, the development and testing would have been overseen by Integra LifeSciences Corporation's engineering and regulatory teams.

    4. Adjudication Method for the Test Set

    • Adjudication method: Not applicable. The testing described is objective device performance (e.g., sterilization, biocompatibility, electrical safety, mechanical, thermal effects). Success or failure is determined by meeting predefined engineering specifications and regulatory standards, not by an adjudication process between human experts on a specific outcome.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC study: No. The document explicitly states: "No clinical studies were performed or required as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate device."
    • Effect size of human readers with/without AI assistance: Not applicable, as no MRMC study or study involving human readers with or without AI assistance was conducted. The device is a surgical aspirator system, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone study: Not applicable. The CUSA Clarity Ultrasonic Surgical Aspirator System is a physical medical device, not an algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's acceptable performance is based on:

    • Established engineering specifications and design inputs: Ensuring the device functions as intended.
    • Compliance with FDA guidance documents and recognized standards: Such as those for sterilization, biocompatibility, EMC, electrical safety, thermal effects, and capacitive coupling.
    • Substantial equivalence to the legally marketed predicate device (CUSA® Clarity Ultrasonic Surgical Aspirator System K190180): This implies that the predicate's established safety and effectiveness profile serves as a benchmark for the new component.

    8. The Sample Size for the Training Set

    • Sample size for training set: Not applicable. This device is not an AI/ML algorithm that requires a training set. The performance is assessed through traditional engineering and regulatory compliance testing.

    9. How the Ground Truth for the Training Set Was Established

    • How ground truth for training set was established: Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221835

    Validate with FDA (Live)

    Date Cleared
    2022-08-22

    (60 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:
    Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:
    Neurosurgery - including removal of primary and secondary malignant and benign brain and spinal tumors, including but not limited to meningiomas and gliomas
    Gastrointestinal and Affiliated Organ Surgery – including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
    Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
    General Surgery – including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The device within the scope of this premarket notification is the CUSA® Clarity 23kHz Single-Sided Bone Tip Pack that is intended to be used with the 23 kHz Handpiece of the CUSA® Clarity Ultrasonic Surgical Aspirator System.

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is the newest device that was added to the Integra Lifesciences Corporation family of tissue ablation products. All CUSA systems are surgical aspirators that use ultrasonics and cavitation with irrigation and aspiration, to fragment, emulsify and remove unwanted tissue. It allows for the selective dissection of target tissue while preserving vessels, ducts, and other delicate structures.

    The CUSA Clarity 23kHz Single-Sided Bone Tip Pack has the same intended use and technological characteristics as the predicate CUSA Clarity system, including all predicate tip pack accessories. A bone tip is currently cleared with the predicate CUSA Clarity system (CUSA Clarity 23 kHz Bone Tip). The primary modification of the CUSA Clarity 23 kHz Single-Sided Bone Tip Pack when compared to the currently cleared CUSA Clarity bone tip is that the subject device has a limited abrasive fragmentation surface at the distal end of the tip when compared to the existing bone tip.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study that proves a device meets such criteria in the context of AI/ML performance. It appears to be a 510(k) summary for a medical device (CUSA® Clarity Ultrasonic Surgical Aspirator System) focusing on establishing substantial equivalence to a predicate device.

    The information provided describes:

    • Device Name: CUSA® Clarity Ultrasonic Surgical Aspirator System
    • Indication for Use: Surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue are desirable. This includes various surgical specialties like neurosurgery, gastrointestinal, urological, general, and laparoscopic surgery.
    • Predicate Device: CUSA® Clarity Ultrasonic Surgical Aspirator System K200774
    • Device Description: The subject device is a CUSA® Clarity 23kHz Single-Sided Bone Tip Pack. The primary modification compared to the predicate bone tip is a limited abrasive fragmentation surface at the distal end.
    • Non-clinical testing performed: Sterilization, shipping, stability, biocompatibility, EMC, electrical safety, and bench testing to verify requirements like tissue fragmentation rate, tip life, torque functionality, and functionality within environmental variations.
    • Conclusion: The non-clinical testing indicates substantial equivalence to the predicate device.
    • Clinical Studies: No clinical studies were performed or required, as performance tests were deemed sufficient to support substantial equivalence.

    Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC study results, or type of ground truth, as these details are not present in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221763

    Validate with FDA (Live)

    Date Cleared
    2022-07-15

    (28 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:

    Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:

    Neurosurgery - including removal of primary and secondary main and spinal tumors, including but not limited to meningiomas and gliomas

    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy

    Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy

    General Surgery - including removal of beingn or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures

    Laparoscopic Surgery - including removal of hepatic parenchyma in lapatic resection, lobectomy or trisegmentectony, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The devices within the scope of this premarket notification are the CUSA® Clarity 23kHz Laparoscopic Tip Packs that are intended to be used with the 23 kHz Handpiece of the CUSA® Clarity Ultrasonic Surgical Aspirator System.

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is the newest device that was added to the Integra Lifesciences Corporation family of tissue ablation products. There are three (3) systems currently marketed in the United States: CUSA Clarity Ultrasonic Surgical Aspirator System (CUSA Clarity), CUSA® Excel+ Ultrasonic Surgical Aspirator System (CUSA Excel+) and CUSA® NXT Ultrasonic Tissue Ablation System (CUSA NXT). All CUSA systems are surgical aspirators that use ultrasonics and cavitation, in combination with irrigation and aspiration, to fragment, emulsify and remove unwanted tissue. It allows for the selective dissection of target tissue while preserving vessels, ducts, and other delicate structures.

    The CUSA Clarity 23kHz Laparoscopic Tip Packs have the same intended use and technological characteristics as the predicate CUSA Clarity and CUSA Excel+ systems, including the predicate tip pack accessories. The primary modification of the subject tips compared to the predicate devices is that they are longer, in order to provide an alternate tip design suited for laparoscopic surgery to the CUSA Clarity portfolio. CUSA Clarity is already indicated for laparoscopic surgery. The purpose of the subject tips is to continue to fill out the CUSA Clarity Tip portfolio to align with that of the legacy CUSA Excel+ system and meet user needs; the CUSA Excel+ system has a tip design very similar to the subject tips. Additionally, testing confirmed that the modifications reduce the frictional force between a trocar and the flue as compared to the predicate CUSA Excel+ Laparoscopic Tip, making the device easier to insert and retract from the trocar during surgical use.

    AI/ML Overview

    The provided text K221763 is a 510(k) summary for the CUSA® Clarity Ultrasonic Surgical Aspirator System, specifically for new 23kHz Laparoscopic Tip Packs. The submission claims substantial equivalence to predicate devices, and therefore, does not present a typical study with acceptance criteria and device performance analysis as one would expect for a de novo device or a device requiring clinical trials.

    The document states: "No clinical studies were required or performed as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate devices." This means the FDA cleared the device based on non-clinical (bench) testing, demonstrating that it performs as intended and is as safe and effective as existing legally marketed predicate devices.

    Therefore, the requested information, which typically pertains to studies proving a device meets acceptance criteria, is not fully applicable in the context of this 510(k) summary claiming substantial equivalence. However, I can extract the relevant information regarding the non-clinical testing performed and the conclusions drawn.

    1. A table of acceptance criteria and the reported device performance

    The document lists types of bench testing performed but does not explicitly provide a table of quantitative acceptance criteria and corresponding device performance results for each criterion. It broadly states that "Bench testing to verify requirements" was conducted.

    Acceptance Criteria Category (Implied)Reported Device Performance (Implied from the summary)
    Sterilization, Shipping, and StabilityMet FDA guidance documents and recognized standards
    BiocompatibilityMet FDA guidance documents and recognized standards
    EMC and Electrical SafetyMet FDA guidance documents and recognized standards
    Tissue Fragmentation RateFunctionality within specification
    Tip Life with CEM (Continuous Emulsification Mode)Functionality within specification
    Torque Functionality During AssemblyFunctionality within specification
    Functionality During Environmental VariationsFunctionality within specification
    Mechanical and PerformanceFunctionality within specification
    Trocar-Flue FrictionReduced frictional force between a trocar and the flue compared to predicate. Verified to make device easier to insert and retract.
    Thermal Effects and Capacitive CouplingMet FDA guidance document requirements

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for each non-clinical test. Given these are bench tests for substantial equivalence, the "data provenance" would be laboratory testing conducted by the manufacturer, Integra LifeSciences Corporation, in the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable. For bench testing used in a 510(k) to demonstrate substantial equivalence, ground truth as established by human experts (e.g., radiologists) is not typically required. Performance is measured against engineering specifications and regulatory standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable, as it pertains to human interpretation/adjudication of results, which was not part of this non-clinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. No clinical or comparative effectiveness studies involving human readers or AI assistance were performed or required for this 510(k) submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is a surgical aspirator, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical testing, the "ground truth" refers to established engineering specifications, recognized industry standards (e.g., ISO for biocompatibility, IEC for electrical safety), and FDA guidance documents. Tissue fragmentation rate, tip life, and mechanical performance would be measured against internal design requirements and predicate device performance.

    8. The sample size for the training set

    This section is not applicable. This is not an AI/ML device, and therefore no "training set" of data was used.

    9. How the ground truth for the training set was established

    This section is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200774

    Validate with FDA (Live)

    Date Cleared
    2020-06-23

    (90 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA Clarity Ultrasonic System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:

    Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:

    Neurosurgery - including removal of primary and secondary main and spinal tumors, including but not limited to meningiomas and gliomas

    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy

    Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy

    General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures

    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The CUSA Clarity Ultrasonic Surgical Aspirator System (CUSA Clarity) is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective dissection of target tissue while preserving vessels, ducts and other delicate structures. The CUSA Clarity consists of a console that provides power and control of the ultrasonic, aspiration and irrigation functions, two surgical handpieces that provide ultrasonic mechanical energy (23 kHz and 36 kHz), a footswitch to allow user control over the ultrasonics, titanium surgical tips (variety of models), irrigation flues, suction/irrigation system (manifold tubing and vacuum canister) and accessories used for assembly/disassembly and reprocessing.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to prove the device meets these criteria:

    Device: CUSA Clarity Ultrasonic Surgical Aspirator System
    510(k) Number: K200774

    This submission appears to be an update to an existing device (CUSA Clarity K182809), specifically an expansion of its neurosurgical indications for use, rather than a new device entirely. Therefore, the "acceptance criteria" and "study" described are primarily focused on supporting this updated indication for neurosurgery.

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance for this device, based on the provided text, is a substantial equivalence determination. This means the device is being compared to a predicate device (CUSA Clarity Ultrasonic Surgical Aspirator; K182809) rather than requiring specific performance targets independent of a predicate. The "acceptance criteria" revolve around demonstrating the new indications for use are safe and effective, and that the device's technological characteristics remain the same.

    Acceptance Criteria CategoryReported Device Performance
    New Neurosurgery Indications Safety and EffectivenessThe device can be safely and effectively used in neurosurgery, including removal of primary and secondary malignant and benign brain and spinal tumors (e.g., meningiomas and gliomas). The literature showed CUSA to be useful for the resection of tumors ranging from soft to firm consistencies, with benefits including enhanced tissue selectivity, preservation of healthy tissue, and more efficient debulking and resection.
    Technological Equivalence to PredicateThe technological characteristics of the device are the same compared to the predicate device.
    General Indications Equivalence to PredicateNo changes to the general Indications for Use compared to the predicate, confirming continued equivalence.

    2. Sample size used for the test set and the data provenance

    • Test Set (Clinical Data): 72 peer-reviewed articles discussing neurosurgical cases.
    • Number of Cases: Approximately 1,706 cases reported across these 72 articles.
    • Data Provenance: Retrospective, derived from published peer-reviewed clinical literature. The country of origin is not specified but would be global as it's from peer-reviewed literature.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable in the traditional sense for this submission. The "ground truth" for this substantial equivalence determination is established by the existing body of peer-reviewed clinical literature. The "experts" are the authors and peer reviewers of the 72 articles themselves, who conducted the studies and produced the reported outcomes. Their qualifications would vary widely but generally include neurosurgeons, neurologists, and other medical professionals involved in the care and study of neurosurgical patients. The submission does not describe an independent panel of experts reviewing cases specifically for this 510(k) alongside the device.

    4. Adjudication method for the test set

    None in the context of an independent adjudication process for this 510(k) submission. The clinical data comes from already published and peer-reviewed articles, implying that any internal adjudication or consensus methods used by the authors of those individual studies would have occurred prior to publication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an ultrasonic surgical aspirator, not an AI-powered diagnostic or interpretive device. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to its function or clearance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a surgical tool, not a standalone diagnostic algorithm.

    7. The type of ground truth used

    The ground truth is established by clinical outcomes data reported in peer-reviewed medical literature. This includes observations on the safety and effectiveness of the CUSA device in various neurosurgical procedures, tumor debulking, tissue selectivity, and patient outcomes over approximately 40 years of use.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The clinical literature serves as the justification for the expanded indications.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190180

    Validate with FDA (Live)

    Date Cleared
    2019-04-01

    (59 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.
    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive Surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:
    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
    Urological Surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
    General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The device within the scope of this premarket notification is the optional CUSA® Electrosurgery Module (CEM) accessory that is intended to be used with the 23 kHz components of the CUSA® Clarity Ultrasonic Surgical Aspirator System. The CUSA® Clarity Ultrasonic Surgical Aspirator System is the newest device added to the Integra Lifesciences Corporation family of tissue ablation products. There are three (3) systems currently marketed in the United States: CUSA® Clarity Ultrasonic Surgical Aspirator System, CUSA® Excel+ Ultrasonic Surgical Aspirator System and CUSA® NXT Ultrasonic Tissue Ablation System. Of these currently marketed products, a CEM accessory is offered with the CUSA® Excel+ Ultrasonic Surgical Aspirator System that allows for coagulation capabilities at the surgical site. The CEM accessory that is offered with the CUSA® Clarity is nearly identical to its predecessor, modified for use with the CUSA® Clarity system. The CEM accessory will allow the surgeon to apply immediate electrosurgical coagulation to bleeding tissue at the surgical site, with the same handpiece assembly that is removing unwanted tissue. The CEM accessory consists of a nosecone, cable and plug which are molded together and packaged as one unit.

    AI/ML Overview

    The provided document is a 510(k) summary for the CUSA® Clarity Ultrasonic Surgical Aspirator System with the optional Electrosurgery Module (CEM) accessory. It describes the device, its intended use, and argues for its substantial equivalence to previously cleared predicate devices.

    Based on the document, here's an analysis of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of quantitative acceptance criteria directly tied to specific performance metrics for the CEM accessory. Instead, it describes categories of non-clinical tests conducted to ensure safety and efficacy, and states that these tests were "successful."

    Acceptance Criteria Category (Implicit)Reported Device Performance (Summary)
    SterilizationSuccessful per FDA guidance and recognized standards
    Shipping and StabilitySuccessful per FDA guidance and recognized standards
    BiocompatibilitySuccessful per FDA guidance and recognized standards
    EMC and Electrical SafetySuccessful per FDA guidance and recognized standards
    Bench Testing (General Device Requirements)Successful, "all 13 protocols within the 4 testing categories... met product specifications and design inputs."
    - Handpiece and Tip Life with CEMFunctionality verified
    - Functionality within environmental variationsFunctionality verified
    Mechanical and Performance TestingFunctionality verified
    Thermal Effects and Capacitive CouplingFunctionality verified per FDA guidance document

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state sample sizes for individual tests. It broadly mentions "non-clinical testing" and refers to "all 13 protocols within the 4 testing categories." This indicates various tests were performed, likely on multiple units or components, but specific numbers are not provided. The data provenance is implied to be from Integra LifeSciences Corporation's internal testing labs, likely in the USA given the company address, but no specific country of origin is mentioned for the data. All testing appears to be prospective as it was conducted specifically to support this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable and therefore not provided. The study conducted was non-clinical (bench, mechanical, electrical, etc.) and did not involve human experts establishing ground truth in a clinical sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used in clinical studies, particularly when establishing ground truth from multiple human readers/experts. The testing described is non-clinical.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The device is an ultrasonic surgical aspirator with an electrosurgery module, not an AI-based diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an AI algorithm, so the concept of standalone algorithm performance is not applicable. The device's performance was evaluated through non-clinical bench testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests, the "ground truth" would be defined by engineering specifications, recognized standards (e.g., for biocompatibility, EMC, electrical safety), and predefined functional requirements (e.g., handpiece life, functionality under environmental variations, thermal effects). These are objective, measurable criteria established through engineering principles and regulatory standards rather than subjective expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    This information is not applicable. As this is not an AI/machine learning device, there is no concept of a "training set."

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set mentioned in the context of this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182809

    Validate with FDA (Live)

    Date Cleared
    2018-12-07

    (65 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:

    Neurosurgery, Plastic and Reconstructive Surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:

    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy

    Urological Surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy

    General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures

    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The CUSA® Clarity Ultrasonic Surgical Aspirator System was originally cleared October 14, 2016 (K161882). The purpose of this submission is to obtain FDA clearance for expanding the current CUSA Clarity indication for use statement to add a hard tissue indication (bone) and to add a 23 kHz handpiece and related accessories for use with the previously cleared CUSA Clarity system. The submission includes a bone tip that facilitates the removal of hard tissue, and additional tips intended for the removal of soft tissue.

    The CUSA Clarity system shares the same intended use and principle of operation as previous CUSA systems. All CUSA systems are surgical aspirators that use ultrasonics and cavitation, in combination with irrigation and aspiration, to fragment, emulsify and remove unwanted tissue. It allows for the selective dissection of target tissue while preserving vessels, ducts, and other delicate structures.

    The CUSA Clarity consists of a console that provides power and control of the ultrasonic, aspiration and irrigation functions, two surgical handpieces that provides ultrasonic mechanical energy (23 kHz and 36 kHz), a footswitch to allow user control over the ultrasonics, titanium surgical tips (variety of models), irrigation flues, suction/irrigation system (manifold tubing and vacuum canister) and accessories used for assembly/disassembly and reprocessing. The CUSA Clarity system, including the 23 kHz components that are the subject of this submission, will be labelled MR Unsafe.

    AI/ML Overview

    The provided text is a 510(k) summary for the CUSA Clarity Ultrasonic Surgical Aspirator System. It details an FDA clearance for expanding the device's indications to include hard tissue (bone) and adding a 23 kHz handpiece and related accessories.

    Based on the information provided, here's a description of the acceptance criteria and study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document summarizes performance testing but does not explicitly present acceptance criteria in a formal table with pass/fail results. Instead, it lists categories of testing and a general statement of success.

    Acceptance Criteria CategoryReported Device Performance
    Sterilization, cleaning, shipping and stability"Testing was determined successful for all 42 protocols within the 5 testing categories above and supports the conclusion that all product specifications and design inputs have been met."
    Biocompatibility"Testing was determined successful for all 42 protocols within the 5 testing categories above and supports the conclusion that all product specifications and design inputs have been met."
    Software testing"Testing was determined successful for all 42 protocols within the 5 testing categories above and supports the conclusion that all product specifications and design inputs have been met."
    Electrical Safety and EMC (Electromagnetic Compatibility)"Testing was determined successful for all 42 protocols within the 5 testing categories above and supports the conclusion that all product specifications and design inputs have been met."
    Bench Testing (e.g., Tissue Fragmentation Rate, Handpiece and Tip Life, Torque functionality during assembly, Operating frequency, Functionality within specification during environmental variations, Aspiration and Irrigation flow)"Testing was determined successful for all 42 protocols within the 5 testing categories above and supports the conclusion that all product specifications and design inputs have been met."

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample sizes used for any of the performance tests. It also does not mention data provenance (e.g., country of origin, retrospective or prospective) as the studies were non-clinical bench tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the studies were non-clinical performance (bench) tests, not human-in-the-loop studies requiring expert ground truth establishment.

    4. Adjudication method for the test set:

    This information is not applicable as the studies were non-clinical performance (bench) tests, not human-in-the-loop studies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC comparative effectiveness study was done. The submission explicitly states: "No clinical studies were performed or required as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate devices." This device is a surgical aspirator, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a surgical tool, not an algorithm. The performance tests ("standalone" in the sense of device-only operation) were bench tests to verify system requirements, such as tissue fragmentation rate, handpiece life, and electrical safety.

    7. The type of ground truth used:

    For the non-clinical performance tests, the "ground truth" would be defined by the predefined technical specifications and standards (e.g., acceptable fragmentation rates, lifespan expectations, safety limits). The document does not explicitly state the specific ground truth metrics but implies they were based on established engineering and regulatory standards for medical device performance.

    8. The sample size for the training set:

    This information is not applicable. The CUSA Clarity Ultrasonic Surgical Aspirator System is a physical medical device, not a machine learning model that requires training data.

    9. How the ground truth for the training set was established:

    This information is not applicable as the device is not a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161882

    Validate with FDA (Live)

    Date Cleared
    2016-10-14

    (95 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CUSA® Clarity Ultrasonic System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable.

    The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:

    Neurosurgery. Plastic and Reconstructive surgery. Orthopedic Surgery. Gynecological Surgery and Thoracic Surgery and the following specific uses:

    Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy

    General Surgery - including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures

    Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy

    Device Description

    The CUSA® Clarity Ultrasonic Surgical Aspirator System is the newest device to be added to the Integra Lifesciences Corporation family of tissue ablation products. There are two (2) systems currently marketed in the United States: CUSA® Excel+ Ultrasonic Surgical Aspirator System and CUSA® NXT Ultrasonic Tissue Ablation System These two systems as well as all predecessor devices share(d) the same principle of operation. All CUSA systems are surgical aspirators which use ultrasonics and cavitation to fragment and emulsify tissue, and aspiration at the end of the surgical tip to remove unwanted tissue. Each model has unique features and attributes that may not be found in the others. Like the predicate devices. CUSA Clarity is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies, and removes unwanted tissue. It allows for the selective of target tissue while preserving vessels, ducts, and other delicate structures. The CUSA Clarity consists of a console that provides power and control of the ultrasonic, aspiration and irrigation functions, a surgical handpiece that provides ultrasonic mechanical energy, a footswitch to allow user control over the ultrasonics, a titanium surgical tip, irrigation flue, and a suction/irrigation system (manifold tubing and vacuum canister). The CUSA Clarity system, including all accessories and components, will be labeled MR Unsafe.

    AI/ML Overview

    This document refers to the CUSA Clarity Ultrasonic Surgical Aspirator System, which is an ultrasonic surgical aspirator. This type of device fragments, emulsifies, and aspirates soft tissue during surgical procedures. The document does not describe the specific acceptance criteria or a detailed study proving the device meets those criteria in the way one would expect for an AI/ML powered device, image analysis software, or a diagnostic tool.

    Instead, the provided text outlines the regulatory submission for a traditional medical device (an ultrasonic surgical aspirator). The focus is on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing performance against specific clinical efficacy metrics for a new diagnostic or predictive algorithm.

    Therefore, many of the requested categories for AI/ML device studies (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable or not explicitly detailed in this type of submission for a physical surgical device.

    Based on the provided text, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document states: "A suite of performance tests was executed to show substantial equivalence with the predicate devices." It then lists categories of testing. However, it does not provide a table with specific quantitative acceptance criteria (e.g., in terms of fragmentation rate, aspiration efficiency, or tissue selectivity) and corresponding performance metrics for the CUSA Clarity. Instead, the general "acceptance criterion" is successful completion of tests demonstrating substantial equivalence.

    Acceptance Criteria Category (Inferrred)Reported Device Performance (Summary)
    SterilizationMet per FDA Guidance documents and recognized standards
    CleaningMet per FDA Guidance documents and recognized standards
    Shipping StabilityMet per FDA Guidance documents and recognized standards
    BiocompatibilityMet per FDA Guidance documents and recognized standards
    SoftwareMet per FDA Guidance document and recognized standards
    EMC and Electrical SafetyMet per FDA recognized standards
    Bench Testing (Performance)Met per various internal protocols; all design inputs fulfilled
    Overall GoalSubstantial equivalence to predicate devices (CUSA Excel+ and CUSA NXT)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size used for the test set: Not explicitly stated. The document refers to "a suite of performance tests" and "50 protocols" completed successfully. These protocols would involve testing samples of the device and its components, but the specific number of units or test runs for each protocol is not provided.
    • Data provenance: Not explicitly stated. The testing appears to be internal to the manufacturer (Integra LifeSciences Corporation) and conducted in a lab/bench setting to verify engineering performance and compliance with standards, rather than clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This concept is not applicable in the context of this device submission. Ground truth, in the AI/ML sense, is usually associated with labeling or interpreting medical images or data. For a surgical aspirator, "ground truth" would relate to the physical performance characteristics of the device (e.g., its power output, aspiration flow, material compatibility), which are measured by instruments and verified against engineering specifications and standards, not by human experts interpreting clinical data for diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are typically used for establishing consensus on "ground truth" labels in clinical data for AI/ML models. For a physical device's performance testing, results are typically determined by measurement instruments and engineering analysis against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are specific to evaluating the clinical performance of diagnostic or prognostic tools, often involving human readers (e.g., radiologists) and sometimes comparing their performance with and without AI assistance. The CUSA Clarity is a surgical instrument, not a diagnostic imaging or AI-powered interpretation tool. No such study is mentioned or relevant here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The CUSA Clarity is a human-operated surgical instrument. There is no "algorithm only" performance to evaluate in this context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As discussed, the concept of "ground truth" in the AI/ML sense is not directly applicable. The "ground truth" for this device's performance is established by engineering specifications, recognized national and international standards (e.g., for sterilization, biocompatibility, electrical safety, EMC), and internal protocols that define acceptable performance ranges for physical characteristics. Compliance with these specifications and standards constitutes meeting the "ground truth" for a physical device.

    8. The sample size for the training set

    Not applicable. The CUSA Clarity is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1