510(k) Data Aggregation

    K Number
    K031636

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-08-11

    (76 days)

    Product Code
    Regulation Number
    870.3535
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The balloon pump is an electromechanical system used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle.

    Device Description

    The CS100 Intra-Aortic Balloon Pump ( IABP) is a cardiac assist device. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. The CS100 is equipped with the flexibility to aid medical staff in supporting a broad range of patients. It is also designed to work in conditions which are unique to the operating room, catheterization laboratory, critical care unit and transport. The CS100 IABP has three operation modes; auto, semi auto and manual. The auto operation mode provides simplicity and minimizes operator intervention. The semi-auto and manual operation modes provide operators with flexibility for difficult clinical cases.

    AI/ML Overview

    The given text is a 510(k) summary for the Datascope CS100 Intra-Aortic Balloon Pump. This document describes the device, its intended use, and claims substantial equivalence to predicate devices. However, it does not contain the specific details required to fully address your request regarding acceptance criteria and a study proving those criteria are met.

    Here's a breakdown of why the requested information cannot be extracted from the provided text and what can be inferred:

    Information Not Present in the Document:

    • Table of Acceptance Criteria and Reported Device Performance: The document states that the device "complies with the voluntary standards identified in section six of this submission" and that "The results of all measurements demonstrated that the CS 100 Intra-Aortic Balloon Pump is as safe, as effective and performs as well as the predicate devices." However, it does not provide a table outlining specific performance acceptance criteria (e.g., timing accuracy, pressure limits, inflation/deflation rates) or quantitative reported device performance metrics against those criteria.
    • Sample Size for the Test Set and Data Provenance: The document mentions "Performance testing" but does not specify the sample size of any test set used or where the data came from (e.g., country of origin, retrospective/prospective).
    • Number of Experts Used to Establish Ground Truth and Qualifications: This information is not provided. The testing described focuses on "Hardware and software testing," "Environmental testing," "Safety testing," and "Performance testing," which are typically engineering and functional tests, not clinical evaluations requiring expert ground truth in the traditional sense.
    • Adjudication Method: Not applicable or not described for the type of testing outlined.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: This type of study is typically done for diagnostic imaging devices where human readers interpret results. The CS100 is an electromechanical pump, so an MRMC study is not relevant to its type of performance evaluation.
    • Standalone (algorithm only) Performance: The device is a physical electromechanical system, not solely an algorithm. Its performance is always "standalone" in the sense that it operates as a device. However, if "standalone" refers to an AI component, the provided text doesn't suggest an AI algorithm separate from the device's operational modes. The "auto operational mode" is mentioned as an automated feature, but details about its specific algorithm and standalone performance metrics are not given.
    • Type of Ground Truth Used: Given the nature of the device (an intra-aortic balloon pump), the "ground truth" would likely relate to objective physiological measurements and adherence to engineering specifications rather than clinical "ground truth" like pathology for a diagnostic device. The document does not specify.
    • Sample Size for the Training Set: Not applicable in the context of traditional clinical or image-based AI training. The device has an "auto operation mode," but details about how this mode was "trained" (if it involves machine learning in the modern sense) are not provided. The phrase "Code design and code reviews" suggests traditional software development.
    • How Ground Truth for the Training Set Was Established: See above.

    What can be inferred/extracted from the provided text:

    While a direct table describing acceptance criteria and a detailed study proving they are met is absent, here's what the document implies about the testing process:

    1. Acceptance Criteria and Device Performance (Inferred/General)

    • Acceptance Criteria (General): The device's performance must demonstrate that it is "as safe, as effective and performs as well as the predicate devices," specifically the Datascope System 96 IABP (K965236) and the Arrow AutoCAT Intra-Aortic Balloon Pump (K983866). This implies that its functional performance (e.g., inflation/deflation timing, pressure delivery, synchronization with ECG/arterial pressure) must be comparable to these established devices.
    • Reported Device Performance (General): The conclusion states: "The results of all measurements demonstrated that the CS 100 Intra-Aortic Balloon Pump is as safe, as effective and performs as well as the predicate devices." This is a summary statement, not specific performance metrics.

    2. Type of Study Done

    The document describes a comprehensive product development and verification testing process rather than a single, specific clinical study with detailed performance metrics. The listed activities include:

    • Requirements specification review
    • Hardware and software testing
    • Code design and code reviews
    • Environmental testing
    • Safety testing
    • Performance testing
    • Hardware and Software validation
    • Auto Operation Mode (specific evaluation of this new feature)

    Summary of Non-Applicable/Missing Information:

    SectionInformation from Text (K031636)
    Acceptance Criteria & Reported PerformanceAcceptance Criteria: Implicitly, the device must be "as safe, as effective and performs as well as the predicate devices" (Datascope System 96 IABP and Arrow AutoCAT IABP). Reported Performance: "The results of all measurements demonstrated that the CS 100 Intra-Aortic Balloon Pump is as safe, as effective and performs as well as the predicate devices." (Specific quantitative criteria and performance values are NOT provided.)
    Sample Size (Test Set) & Data ProvenanceNot specified. "Performance testing" is mentioned, but no details on sample size or data origin (e.g., country, retrospective/prospective clinical data) are given. The testing appears to be primarily engineering/validation based.
    Number & Qualifications of Experts (Ground Truth)Not applicable or not specified. The testing described is functional/engineering validation rather than clinical assessment requiring expert interpretation for ground truth.
    Adjudication MethodNot applicable.
    Multi-Reader Multi-Case (MRMC) Comparative Effectiveness StudyNo. This type of study is not relevant for an electromechanical pump.
    Standalone (Algorithm Only) Performance StudyNot a purely algorithmic device. The "auto operation mode" is a feature of the device; its standalone performance metrics (if any were specifically isolated from overall device performance) are not detailed. It implies testing was done for this mode as part of the overall "Performance testing" and "Hardware and Software validation."
    Type of Ground Truth Used (Test Set)Not explicitly stated, but for an electromechanical device, ground truth would typically be objective engineering specifications, validated physical and electrical measurements, and possibly animal or human physiological data from controlled environments to ensure synchrony and therapeutic effect. Not "expert consensus," "pathology," or "outcomes data" in the typical AI/diagnostic sense.
    Sample Size (Training Set)Not applicable. The device relies on traditional software (code design, code reviews) and hardware; there's no indication of a "training set" in the machine learning sense for its "auto operation mode."
    How Ground Truth for Training Set Was EstablishedNot applicable. The "auto operation mode" is described as providing simplicity and minimizing operator intervention, likely developed through engineering design, control algorithms, and testing against desired physiological responses and synchronization with ECG/arterial pressure. If machine learning was involved (unlikely for this era/device type), no details are given for training or ground truth establishment.

    In conclusion, the 510(k) summary for the Datascope CS100 Intra-Aortic Balloon Pump affirms that extensive testing and validation were performed to demonstrate substantial equivalence to predicate devices, focusing on safety, effectiveness, and performance. However, it does not provide the specific quantitative acceptance criteria or detailed study results that you are requesting. Such details would typically be found in the actual test reports and validation documents submitted to the FDA, which are generally not part of the publicly available 510(k) summary.