Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K150262
    Date Cleared
    2015-04-03

    (58 days)

    Product Code
    Regulation Number
    870.3375
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Crux Vena Cava Filter (VCF) System is indicated for the prevention of recurrent pulmonary embolism via placement in the inferior vena cava (IVC) in the following situations:

    • pulmonary thromboembolism when anticoagulants are contraindicated:
    • failure of anticoagulant therapy in thromboembolic disease;
    • emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced;
    • chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.

    The Crux VCF may be removed according to the instructions contained in the section "Optional Retrieval of the Crux VCF" in patients who no longer require a vena cava filter. Retrieval of the filter can be performed by femoral or jugular approach.

    Device Description

    The Crux Vena Cava Filter (VCF) System is an endovascular medical device used in the prevention of recurrent pulmonary embolism (PE). The system is comprised of a self-expanding Nitinol filter delivered from a single-use, disposable 9Fr catheter, which can be used percutaneously to deploy the filter. The filter wireforms are composed of two opposing selfexpanding Nitinol spiral elements connected at each end with Nitinol crimps. One of each wireform is formed into a sinusoidal shaped retrieval tail to aid in retrieval of the filter using a snare. Each retrieval tail has an atraumatic plasma ball and a radiopaque tantalum marker band to facilitate visualization. There are five tissue anchors attached to the wireform elements with Nitinol tubing. The filter is designed to treat IVC diameters of 17 to 28mm.

    The delivery catheter for the Crux VCF System is a disposable, 9Fr introducer-sheath-compatible, single-use delivery catheter. The filter is provided loaded in the Crux VCF System for jugular or femoral approach delivery. The delivery catheter is an over-the-wire system, 0.035" guidewirecompatible, and is comprised of a polycarbonate inner shaft and a nylon outer shaft. The polyimide inner shaft is comprised of the guidewire lumen, and a flexible radiopaque tracking tip with a radiopaque marker band. The outer shaft has a radiopaque marker band, a Touhy-Borst hemostasis valve, and a one-way check-valve for flushing.

    The filter can be retrieved with commercially available snares and sheaths via either the jugular or femoral approach.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Crux Vena Cava Filter System. The document focuses on regulatory approval based on substantial equivalence to a predicate device and addresses labeling changes, not on a study proving the device meets specific acceptance criteria in terms of clinical performance as an AI/CAD/imaging device.

    Therefore, many of the requested details, such as a table of accuracy metrics, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types related to AI/CAD/imaging device performance, are not applicable to this submission.

    However, the document does mention performance data related to device functionality.

    Here's the information that can be extracted or deduced from the provided text:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (General Description)Reported Device Performance Statement
    Proper blood flow with renal vein inflow (evaluated by CFD Modeling)All tests met the pre-determined acceptance criteria.
    Radial forces at pararenal placement (evaluated by VCF Radial Force Test)All tests met the pre-determined acceptance criteria.
    Catheter pressure integrity for power injectionAll tests met the pre-determined acceptance criteria.
    Clinical workflow of VCF deployment at pararenal locationAll tests met the pre-determined acceptance criteria.
    Power injection through the delivery catheter during simulated useAll tests met the pre-determined acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated. The "tests" mentioned were bench and animal studies, not human clinical trials for endpoint efficacy.
    • Data Provenance: Not explicitly stated. The studies were described as "bench and animal device testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. The "tests" focused on device functionality (e.g., flow, radial force, pressure integrity, workflow) rather than clinical diagnostic or therapeutic outcomes requiring expert ground truth for interpretation.

    4. Adjudication method for the test set:

    • Not Applicable. This is not relevant for the type of device functionality tests described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/CAD/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI/CAD/imaging device.

    7. The type of ground truth used:

    • For the bench testing, the "ground truth" would be established by validated engineering principles, physical measurements, and computational models (e.g., CFD).
    • For the animal testing, the "ground truth" would be direct observation and measurement in the animal model.
    • Not pathology, outcomes data, or expert consensus in the context of diagnostic interpretation, as this is a device for mechanical intervention.

    8. The sample size for the training set:

    • Not Applicable. This is not a machine learning or AI device that requires "training sets."

    9. How the ground truth for the training set was established:

    • Not Applicable. This is not a machine learning or AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K122585
    Manufacturer
    Date Cleared
    2013-01-17

    (146 days)

    Product Code
    Regulation Number
    870.3375
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Crux VCF System is indicated for the prevention of recurrent pulmonary embolism via percutaneous placement in the inferior vena cava (IVC) in the following situations:

    • Pulmonary thromboembolism when anticoagulants are contraindicated .
    • . Failure of anticoagulant therapy in thromboembolic diseases
    • Emergency treatment following massive pulmonary embolism where anticipated benefits of . conventional therapy are reduced
    • Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is . contraindicated

    The Crux VCF may be removed according to the instructions contained in the Instructions for Use section "Optional Retrieval of the Crux VCF" in patients who no longer require a vena cava filter. Retrieval of the filter can be performed by femoral or jugular approach.

    Device Description

    The Crux VCF System is an endovascular medical device used in the prevention of recurrent pulmonary embolism (PE). The system consists of a self-expanding Nitinol filter delivered from a single-use, disposable 9Fr catheter, which can be used percutaneously to deploy the filter. The filter wireforms are composed of two opposing self-expanding Nitinol spiral elements connected at each end with Nitinol crimps. One end of each wireform is formed into a sinusoidal shaped retrieval tail to aid in retrieval of the filter using a snare. Each retrieval tail has an atraumatic plasma ball and a radiopaque tantalum marker band to facilitate visualization. There are five tissue anchors attached to the wireforms elements with Nitinol tubing. The filter is designed to treat IVC diameters of 17 to 28mm.

    The Delivery Catheter for the Crux VCF is a disposable, 9Fr introducer-sheath-compatible, single-use Delivery Catheter. The filter is provided loaded in the Crux VCF System for jugular or femoral approach delivery. The Delivery Catheter is an over-the-wire system, 0.035" guidewire-compatible, and consists of a polycarbonate inner shaft and a nylon outer shaft. The polyimide inner shaft consists of the guidewire lumen and a flexible, radiopaque tracking tip and a radiopaque marker band. The outer shaft has a radiopaque marker band, a Touhy-Borst hemostasis valve and a one-way check-valve for flushing.

    The filter can be retrieved with commercially available snares and sheaths via either the jugular or femoral approach.

    AI/ML Overview

    This document outlines the performance data for the Crux Vena Cava Filter System, as presented in its 510(k) submission.

    Device: Crux Vena Cava Filter System

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly met by demonstrating "substantial equivalence" to a predicate device through bench testing. The document states that "Bench test results demonstrate that the device meets the established specifications and is comparable to the predicate device supporting substantial equivalence." Specific performance metrics and criteria are not explicitly detailed in the provided text, but the tests conducted were:

    Acceptance Criteria (Implied)Reported Device Performance
    Device meets established specifications."Bench test results demonstrate that the device meets the established specifications."
    Device is comparable to the predicate device."Bench test results demonstrate that the... device is comparable to the predicate device supporting substantial equivalence." and "The results also are comparable to the established predicate device supporting substantial equivalence."
    No adverse impact on safety and effectiveness compared to the predicate device due to design updates."Based on the design verification results, the updates in Delivery Catheter do not have an adverse impact on the safety and effectiveness of the VCF when compared to the predicate device."
    Performance of the updated delivery catheter is not adversely affected compared to the predicate device."As demonstrated in the results of the bench testing, these differences in design do not adversely affect the performance when compared to the predicate device and the devices are substantially equivalent."

    2. Sample Size for Test Set and Data Provenance

    The provided text does not specify the sample sizes used for the in-vitro bench tests (Dimensional and Visual Inspection, Deployment Trackability, Bond Strength). The data provenance is assumed to be from Crux Biomedical, Inc.'s internal testing, as it's part of a 510(k) submission to the FDA. The testing is in-vitro, meaning it was conducted in a lab setting, not on human or animal subjects, so country of origin isn't relevant in the same way as clinical data.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. The study involved bench testing (in-vitro testing) of the device's physical properties and performance. There was no ground truth established by experts in a clinical context for a diagnostic output.

    4. Adjudication Method for the Test Set

    Not applicable, as this was bench testing, not a clinical study with subjective assessments requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes a 510(k) submission based on bench testing to demonstrate substantial equivalence of a medical device to a predicate device, not a comparative effectiveness study involving human readers or AI.

    6. Standalone Performance Study (Algorithm Only)

    No. This device is a physical medical device (vena cava filter system), not an algorithm or AI system. Therefore, standalone algorithm performance is not applicable.

    7. Type of Ground Truth Used

    The "ground truth" for this submission is based on established engineering specifications and performance benchmarks derived from the predicate device. The tests are designed to verify that the updated device components (specifically the delivery catheter) meet these physical and mechanical requirements and perform comparably to the predicate.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI or machine learning system that requires a "training set" of data.

    9. How Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120402
    Manufacturer
    Date Cleared
    2012-07-13

    (156 days)

    Product Code
    Regulation Number
    870.3375
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Crux VCF System is indicated for the prevention of recurrent pulmonary embolism via percutaneous placement in the inferior vena cava (IVC) in the following situations:

    • Pulmonary thromboembolism when anticoagulants are contraindicated .
    • Failure of anticoagulant therapy in thromboembolic diseases .
    • Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced
    • Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or . is contraindicated

    The Crux VCF may be removed according to the instructions contained in the Instructions for Use section "Optional Retrieval of the Crux VCF" in patients who no longer require a vena cava filter. Retrieval of the filter can be performed by femoral or jugular approach.

    Device Description

    The Crux VCF System is an endovascular medical device used in the prevention of recurrent pulmonary embolism (PE). The system consists of a self-expanding Nitinol filter delivered from a single-use, disposable 9Fr catheter, which can be used percutaneously to deploy the filter. The filter wireforms are composed of two opposing self-expanding Nitinol spiral elements connected at each end with Nitinol crimps. One end of each wireform is formed into a sinusoidal shaped retrieval tail to aid in retrieval of the filter using a snare. Each retrieval tail has an atraumatic plasma ball and a radiopaque tantalum marker band to facilitate visualization. There are five tissue anchors attached to the wireforms elements with Nitinol tubing. The filter is designed to treat IVC diameters of 17 to 28mm.

    The delivery catheter for the Crux VCF is a disposable, 9Fr introducer-sheathcompatible, single-use delivery catheter. The filter is provided loaded in the Crux VCF System for jugular or femoral approach delivery. The delivery catheter is an over-thewire system, 0.035" guidewire-compatible, and consists of a polycarbonate inner shaft and a nylon outer shaft. The inner shaft consists of the guidewire lumen and a flexible, radiopaque tracking tip. The outer shaft has a radiopaque distal marker band, a Touhy-Borst hemostasis valve and a one-way checkvalve for flushing.

    The filter can be retrieved with commercially available snares and sheaths via either the jugular or femoral approach.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Crux Vena Cava Filter System, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Defined as "Primary Endpoint of Clinical Success")Reported Device Performance
    Technical success of deployment98% (123/125)
    Freedom from PE (Pulmonary Embolism)2.4% recurrent PE
    Freedom from filter migration0 cases of migration
    Freedom from device related adverse events requiring interventionNot explicitly quantified, but overall Clinical Success was 96%
    Overall Clinical Success: Expected to exceed 80% (91.8% lower one-sided 95% CL)96% (91.8% lower one-sided 95% CL)
    Retrieval success (Secondary Endpoint)98% (53/54)

    Note: The document explicitly states, "Clinical Success was 96% (91.8% lower one-sided 95% CL exceeding the 80)." This indicates that the acceptance criterion for overall clinical success was likely a lower bound of 80% with a 95% confidence interval.


    Study Details

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: 125 subjects (for clinical trial)
    • Data provenance: Prospective, single-arm, multinational clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not explicitly provided in the given text. Clinical trials often involve multiple clinicians for assessment, but the number and qualifications of experts specifically for establishing "ground truth" (e.g., confirming PE diagnoses, migration, etc.) are not detailed.

    4. Adjudication method for the test set:

    • This information is not explicitly provided in the given text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving AI assistance for human readers was not performed. This document describes a clinical trial for a physical medical device (vena cava filter), not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this refers to a physical medical device, not an algorithm. Therefore, no standalone algorithm performance was assessed.

    7. The type of ground truth used:

    • The ground truth was established through clinical outcomes observed in the prospective clinical trial. This includes:
      • Technical success of deployment (observed during procedure)
      • Occurrence of recurrent PE (diagnosed clinically)
      • Filter migration (diagnosed clinically/radiographically)
      • Device-related adverse events requiring intervention (diagnosed clinically)
      • Retrieval success (observed during removal procedure)
      • Radiographic anomalies (e.g., IVC thrombus).

    8. The sample size for the training set:

    • This information is not applicable as this describes a physical medical device and its clinical trial, not an AI or machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • This information is not applicable for the same reasons as above.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1