510(k) Data Aggregation

    K Number
    K250921

    Validate with FDA (Live)

    Date Cleared
    2025-06-25

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use

    CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide preoperative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for generating the plan are identifiable on patient imaging scans.

    Indications for Use

    The CORIOGRAPH Pre-op Planning and Modeling Services are indicated for use for the following procedures:

    • Unicondylar Knee Replacement (UKR)
    • Total Knee Arthroplasty (TKA)
    • Primary Total Hip Arthroplasty (THA)
    • Primary Anatomic Total Shoulder Arthroplasty (aTSA)
    • Primary Reverse Total Shoulder Arthroplasty (rTSA)
    Device Description

    Coriograph Pre-Op Planning and Modeling Services (CORIOGRAPH) are Software as a Medical Device (SaMD) that provide pre-operative planning for orthopedic surgical procedures based on patient imaging, surgeon preferences and implant geometry. CORIOGRAPH is comprised of several medical software systems (modules), and these modules share a set of non-medical function software applications called the Case Processing System.

    CORIOGRAPH is a Software Medical Device System. The medical function modules share a graphic user interface where the surgeon provides their case preferences and patient imaging, retrieves PDF plans generated by the Pre-Op Plan modules, and launches the Modeler modules.

    CORIOGRAPH Pre-Op Plan Modules

    Using patient-specific information, patient imaging, and surgeon inputs, a pre-operative plan is generated by Smith+Nephew personnel. The plan provides initial alignment recommendations to the surgeon on implant placement based on the geometries of the implant and of the generated bone models.

    CORIOGRAPH Hip Modeler and CORIOGRAPH Shoulder Modeler Modules

    CORIOGRAPH Hip Modeler and CORIOGRAPH Shoulder Modeler are web-based software applications intended for surgeons to modify the patient-specific pre-operative plan and to evaluate the impact of the modifications through the impingement analysis tools such as Impingement-free Range-of Motion (ROM) and Activities of Daily Living (ADL) simulations.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for a medical device called CORIOGRAPH Pre-Op Planning and Modeling Services. It details the device's intended use, indications for use, and a comparison to predicate devices, focusing on the addition of shoulder arthroplasty procedures.

    However, the document does not contain the specific information requested regarding acceptance criteria and the detailed study proving the device meets these criteria. It mentions non-clinical bench testing, software verification testing (IEC 62304), credibility evaluation of kinematic models and ADL simulations, and summative usability validation testing (IEC 62366). It states that "Design verification and validation testing demonstrated that CORIOGRAPH Pre-Op Planning and Modeling Services (v 3.0) meets all design requirements and is as safe and effective as its primary and secondary predicate devices." and "The credibility evaluation demonstrated that the kinematic models and Activities of Daily Living (ADLs) simulations utilized in the subject device are clinically relevant."

    Crucially, it does not provide:

    • Specific acceptance criteria for accuracy, precision, or performance metrics.
    • Quantifiable reported device performance against those criteria.
    • Sample sizes for test sets used for performance evaluation (other than mentioning "representative users" for usability).
    • Data provenance (country of origin, retrospective/prospective) for performance data.
    • Details on expert ground truth establishment (number of experts, qualifications, adjudication method).
    • Information on MRMC comparative effectiveness studies.
    • Standalone (algorithm-only) performance.
    • Type of ground truth used (e.g., pathology, outcomes data) for clinical performance, if any.
    • Sample size for the training set or how its ground truth was established.

    The document is a high-level summary confirming regulatory clearance based on substantial equivalence, implying that the detailed testing data would have been part of the full 510(k) submission, but is not included in this public-facing summary.

    Therefore, based only on the provided text, I cannot complete the requested tables and details about acceptance criteria and study proof. The document confirms that testing was done and results met requirements, but does not provide the specifics of those results or the criteria themselves.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242272

    Validate with FDA (Live)

    Date Cleared
    2024-11-15

    (106 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use
    CORIOGRAPH Pre-Op Planning and Modeling Services are intended to provide preoperative planning for surgical procedures based on patient imaging, provided that anatomic landmarks necessary for generating the plan are identifiable on patient imaging scans.

    Indications for Use
    The CORIOGRAPH Pre-op Planning and Modeling Services are indicated for use for the following procedures:

    • · Unicondylar Knee Replacement (UKR)
    • · Total Knee Arthroplasty (TKA)
    • · Primary Total Hip Arthroplasty (THA)
    Device Description

    The subject CORIOGRAPH Hip Pre-Op Plan and CORIOGRAPH Modeler are medical function modules within the CORIOGRAPH Pre-Op Planning and Modeling Services are additional offerings being introduced by Blue Belt Technologies, Inc. to allow for pre-operative planning for surgical procedures based on patient imaging for primary total hip arthroplasty (THA). The CORIOGRAPH Hip Pre-Op Plan system will utilize Smith and Nephew personnel to generate patient specific bone models and preoperative plans for primary THA which will be viewable and editable on CORIOGRAPH Modeler 1.0. Together, CORIOGRAPH Hip Pre-Op Plan and CORIOGRAPH Modeler are the subject of this submission.

    AI/ML Overview

    The acceptance criteria for the CORIOGRAPH Pre-Op Planning and Modeling Services V2.0 were demonstrated through verification and validation testing, and summative usability testing. The provided document does not explicitly list numerical acceptance criteria values for metrics like accuracy, sensitivity, or specificity. Instead, it broadly states that testing demonstrated the safety and effectiveness of the software applications and that all design inputs were met.

    Acceptance Criteria and Reported Device Performance

    Note: The document does not provide specific quantitative acceptance criteria or detailed performance metrics. It indicates that the device met its design inputs and was found to be safe and effective.

    Acceptance Criteria CategoryReported Device Performance (as stated in document)
    Verification and Validation TestingDemonstrated the safety and effectiveness of the software applications used in CORIOGRAPH Pre-Op Planning & Modeling Services V2.0. All design inputs were met.
    Summative Usability TestingDemonstrated that participating surgeons were able to use the subject device safely and effectively in a simulated use environment.
    Credibility EvaluationDemonstrated that the kinematic models and Activities of Daily Living (ADLs) utilized in the subject device are clinically relevant.

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document mentions "sumative usability testing" was performed, indicating a test set was used, but does not specify the sample size for the test set.
      • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective) for the data used in testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document implies that "participating surgeons" were involved in the summative usability testing, but does not state the number of experts used to establish ground truth or their specific qualifications (e.g., years of experience, subspecialty).
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • The document does not describe any specific adjudication method used for the test set.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study or report any effect size for human reader improvement with AI assistance. The study focuses on the device's ability to generate pre-operative plans and its usability.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The document states that the software applications underwent "verification and validation testing," implying a standalone component to ensure functional correctness. However, it also highlights "summative usability testing" with "participating surgeons," which indicates human-in-the-loop performance was also evaluated, particularly for the overall service. It is not explicitly stated whether fully standalone performance was evaluated as a separate metric without any human involvement.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the "credibility evaluation," the document states it "demonstrated that the kinematic models and Activities of Daily Living (ADLs) utilized in the subject device are clinically relevant." This suggests clinical relevance or expert opinion/consensus as a form of ground truth for these aspects. For the outputs of the pre-operative planning, the implied ground truth is agreement with surgical principles and objectives, likely assessed by participating surgeons during usability testing.
    7. The sample size for the training set:

      • The document does not specify the sample size for the training set. It describes the device as providing "Pre-Op Planning and Modeling Services" based on patient imaging and does not detail a machine learning model's training process or associated dataset sizes.
    8. How the ground truth for the training set was established:

      • Since the document does not explicitly mention a training set or machine learning components in terms of specific algorithms that require labeled training data (beyond general software development and functionality), it does not describe how ground truth for a training set was established. The "services" aspect implies that the system is used by human personnel (Smith and Nephew personnel, as stated in device description) to generate plans, meaning the "training" might refer to the development and refinement of these human-led processes and software functionalities under established surgical guidelines.