510(k) Data Aggregation

    K Number
    K012416

    Validate with FDA (Live)

    Date Cleared
    2001-10-26

    (88 days)

    Product Code
    Regulation Number
    886.1850
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ConfoScan 2 Confocal Microscope is intended for use in the precise display and digital storage of images of the corneal layers (endothelium, descemet's membrane, stroma, bowman's membrane, and epithelium) and corneal components (endothelial cells, stromal keratocytes, nerve fibers, basalar and superficial cells). The microscope is also used for optical pachymetry, post-surgery interface evaluation, endothelial cell analysis, haze detection, foreign bodies, corneal dystrophies, keratoconus, keratitis, and trauma.

    The ConfoScan 2 Confocal Microscope is indicated for use as a diagnostic tool for observation of the cell layers of the anterior parts of the eye.

    Device Description

    The Nidek Technologies Srl ConfoScan 2 Confocal Microscope is a fully digital, corneal confocal microscope that allows a user to view and image human corneal layers in vivo. The confocal arrangement consists of one slit that illuminates the tissue and a second slit that filters the reflected light from unfocused layers. The user can view, magnify, measure, and photograph scparate layers of the transparent structures and tissues of the cornea. One examination records 350 different images.

    The ConfoScan 2 microscope is supplied with a Quartz Halogen light source and a scauning slit, suitable for performing ophthalmic examination. The system includes a 15-inch LCD S-VGA Monitor and a Personal Computer, which incorporates the dedicated NAVIS software.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ConfoScan 2 Confocal Microscope:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state quantitative acceptance criteria or detailed device performance metrics in the way one might see for a diagnostic accuracy study. Instead, the performance evaluation focuses on compliance with established standards and equivalence to a predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Electrical Safety: Meets all requirements of EN 60601-1.ConfoScan 2 Confocal Microscope found to meet all requirements of EN 60601-1.
    Electromagnetic Compatibility: Meets all requirements of EN 60601-1-2.ConfoScan 2 Confocal Microscope found to meet all requirements of EN 60601-1-2.
    Programmable Electrical Medical Systems: Meets all requirements of EN 60601-1-4.ConfoScan 2 Confocal Microscope found to meet all requirements of EN 60601-1-4.
    Ophthalmic Instruments Fundamental Test Requirements & Methods: Meets all requirements of ISO 15004.ConfoScan 2 Confocal Microscope found to meet all requirements of ISO 15004.
    Light Energy Levels: Negligible amount of light energy transferred to the patient during normal examination (1-2 minutes).Amount of light energy transferred to the patient during 1-2 minute examination found to be negligible.
    Endothelial Cell Analysis Software Validation: Automatic cell density values comparable to results from at least two other commercially available specular microscopes.134 samples analyzed, and resulting automatic cell density values validated by comparison to two other commercially available specular microscopes. (The exact level of agreement/equivalence is not quantified in the summary, only that it was "validated").
    Substantial Equivalence: Same intended use and fundamental scientific technology as the predicate device (Tomey ConfoScan Confocal Microscope K972953)."The results of this comparison demonstrate that the ConfoScan 2 Confocal Microscope device has the same basic technological characteristics as the predicate device and is equivalent to the marketed predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: For the "Endothelial Cell Analysis Software Validation," 134 samples (extracted from corneal images) were used.
    • Data Provenance: The document does not specify the country of origin of the data or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts used to establish ground truth for the "Endothelial Cell Analysis Software Validation," nor their qualifications. The validation involved "comparison to the results of two other commercially available specular microscopes," implying these microscopes (and their inherent cell analysis methods) served as a reference or "ground truth surrogate," rather than a distinct panel of human experts.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method. The validation was a comparison against other commercially available devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly described. The performance evaluation focuses on device-to-standard compliance and device-to-device comparison for the endothelial cell analysis software. There is no mention of human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone evaluation of the algorithm's performance was done for the "Endothelial Cell Analysis Software Validation." The ConfoScan 2 Confocal Microscope's software analyzed 134 samples and produced automatic cell density values, which were then compared to other commercial devices. This is a standalone assessment of the algorithm's output.

    7. The Type of Ground Truth Used

    For the "Endothelial Cell Analysis Software Validation," the ground truth was based on the results obtained from two other commercially available specular microscopes. This suggests a comparative reference standard, rather than a definitive "true" value derived from pathology or long-term outcomes. For the other tests (electrical safety, etc.), the "ground truth" was compliance with harmonized standards.

    8. The Sample Size for the Training Set

    The document does not provide information regarding the sample size used for training the ConfoScan 2 Confocal Microscope's software or any algorithms.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information regarding how the ground truth for any potential training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K971953

    Validate with FDA (Live)

    Device Name
    CONFOSCAN
    Date Cleared
    1997-08-14

    (78 days)

    Product Code
    Regulation Number
    886.1850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tomey "ConfoScan" Confocal Microscope is intended for use in carrying out a visualization of the cell layers of the anterior parts of the eye.
    The Tomey "ConfoScan" Confocal Microscope is intended for use as a diagnostic tool for looking at the cell layers of the anterior parts of the eye.

    Device Description

    The Tomey "ConfoScan" Confocal Microscope is a scanning slit microscope system that is used to carry out a visualization of the cell layers of the anterior parts of the eye. The prime attributes of a Confocal Microscope include both improved contrast and spatial resolution over more conventional microscopy. The improved image rendition is accomplished in the Tomey "ConfoScan" Confocal Microscope by the introduction conjugate scanning slits into the viewing and illuminating light paths of the microscope. By illuminating and viewing an object through restricted apertures, it is possible to eliminate the contributions of scattered light from other regions in the object plane that can degrade contrast and spatial resolution.

    AI/ML Overview

    The provided text describes the Tomey ConfoScan Confocal Microscope, but it does not contain explicit acceptance criteria or a detailed study proving the device meets specific performance criteria. Instead, it focuses on demonstrating substantial equivalence to predicate devices.

    Here's a breakdown based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    No explicit acceptance criteria are provided in terms of performance metrics (e.g., sensitivity, specificity, accuracy, resolution targets). The document highlights characteristics of the ConfoScan and compares them to predicate devices, implying these characteristics are accepted by virtue of being similar or equivalent to already approved devices.

    CharacteristicTomey ConfoScanPredicate Tandem Scanning Confocal Microscope (for performance comparison)Predicate Tomey Specular MicroscopePredicate Tomey ViVA
    Ophthalmic MicroscopeYesYesYesNo
    Cornea ContactYesYesYesNo
    Floating Objective LensYesUnknownNoNA
    Working distance0-10 mm0-8 mm0-1.3 mm100+ cm
    Type Scanning ApertureSlitNipkow DiskNANA
    Aperture Light Budget3%0.25-1%NANA
    Light Source100 W Halogen100 W Hg ArcIRIR
    Objective LensAchroplan with Floating MountConvex Applanating Dipping ConeDipping Cone with Floating MountNA
    Depth Resolution1 micron< 1 micronNANA
    Horizontal Resolution10 microns6 micronNANA
    Pre SterilizedNoNoNoNA

    2. Sample Size for Test Set and Data Provenance

    • No specific test set or clinical study data is provided in this submission. The "Non-Clinical Testing" section describes tests like elution, sterilization, biocompatibility, and electrical/thermal safety, but these are not performance studies on a "test set" of patient data.
    • The entire submission relies on comparison to legally marketed predicate devices to demonstrate substantial equivalence, rather than a clinical performance study with a distinct test set.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable. Since there's no clinical performance study presented that uses a "test set" requiring ground truth establishment by experts, this information is not present. The submission focuses on device characteristics and safety/functionality tests.

    4. Adjudication Method for Test Set

    • Not applicable. No test set requiring expert adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. The document does not describe an MRMC study comparing human readers with and without AI assistance. The device is a microscope for visualization, not an AI diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    • No. The Tomey ConfoScan is a physical ophthalmic microscope. It is not an algorithm, and therefore, a standalone algorithm-only performance study is not relevant or included.

    7. Type of Ground Truth Used

    • Not applicable. As no clinical performance study is described, there is no ground truth, expert consensus, pathology, or outcomes data used for this purpose. The "ground" for approval is the demonstration of substantial equivalence to existing devices.

    8. Sample Size for Training Set

    • Not applicable. The Tomey ConfoScan is a hardware device for visualization, not an AI/machine learning system that requires a "training set."

    9. How Ground Truth for Training Set Was Established

    • Not applicable. As the device does not use a training set, the establishment of ground truth for it is irrelevant.

    Summary of the Study Proving Device Meets Acceptance Criteria (as per the provided text):

    The "study" presented is primarily a non-clinical performance testing and a comparison of characteristics to predicate devices to demonstrate substantial equivalence.

    • Non-Clinical Testing:

      • Elution testing: The ConfoScan "passed the Elution test performed according to the requirements of the Ministry of Health and Welfare in Japan." This indicates it meets safety standards related to leaching of substances.
      • Sterilization: The objective lens is intended for disinfection by the end-user following recommended procedures. This implies the device design allows for user-level sterilization.
      • Biocompatibility: The biocompatibility issues are stated to be "identical to those of the many other ophthalmic microscopes employing short focal length immersion lenses." This indicates the materials are considered safe based on prior experience with similar devices.
      • Electrical and Thermal Safety: The ConfoScan system was "certified for Electrical and Thermal safety by the JQA (Japan Quality Assurance Organization)." This demonstrates compliance with relevant safety standards.
    • Substantial Equivalence Comparison: The core of the evidence is the "Side-by-Side Comparison with Predicate Devices" (Table 1). This table maps characteristic features, working parameters (e.g., working distance, depth/horizontal resolution), and light sources of the ConfoScan against three predicate devices. The claim is that the device "has the same intended use and target population as the predicate devices, and has equivalent effectiveness for its intended use."

    Conclusion: The provided document establishes the device's acceptability not through a traditional clinical study with defined acceptance criteria and performance metrics, but by demonstrating safety (non-clinical testing) and showing its fundamental characteristics and intended use are "substantially equivalent" to devices already legally marketed, thus implying equivalent safety and effectiveness.