Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121183

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-07-26

    (99 days)

    Product Code
    Regulation Number
    888.3660
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Comprehensive® Reverse Shoulder is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants and a functional deltoid muscle is necessary.

    The Comprehensive® Reverse Shoulder is indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency.

    Glenoid components with Hydroxyapatite (HA) coating applied over the porous coating are indicated only for uncemented biological fixation applications. The Glenoid Baseplate components are intended for cementless application with the addition of screw fixation.

    Interlok® finish humeral stems are intended for cemented use and the MacroBond® coated humeral stems are intended for press-fit or cemented applications. Humeral components with porous coated surface coating are indicated for either cemented or uncemented biological fixation applications.

    Device Description

    The Comprehensive® Reverse Shoulder is intended for total shoulder replacement in a reverse shoulder configuration. Unlike traditional total shoulder replacement, a reverse shoulder employs a ball for articulation on the glenoid side of the joint and a polyethylene bearing surface on the humeral side of the joint. This device configuration increases the lever arm of the deltoid muscle bundle to provide stability and the ability to raise the arm. This is especially useful in cases where a patient has a non-functioning rotator cuff which severely limits traditional joint replacement options.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "Comprehensive® Reverse Shoulder - E1® Polyethylene Claims". It details the device's indications for use, regulatory classification, and claims related to its E1® Antioxidant Infused Technology.

    Here's an analysis of the acceptance criteria and supporting studies, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria ClaimReported Device Performance and Test Details
    Claim 1: E1® Antioxidant Infused Technology prevents oxidative degradation of polyethylene.Test: Environmental Stress Cracking Testing. Cyclically loaded GUR1020 and GUR1050 E1® test specimens in an air atmosphere at 80°C for 5 weeks. Testing per Nabar, Sean, et al. Transactions of the 54th Annual Meeting of the ORS, Poster No. 1684. Performance: E1® specimens showed no evidence of environmental stress cracking. Infrared spectroscopy showed no detectable oxidation in loaded or unloaded samples (oxidation indices <0.1).
    Claim 2: E1® Antioxidant Infused Technology protects polyethylene from oxidation and cracking during environmental stress crack testing. (Comparative)Test: Comparative Environmental Stress Cracking Testing. Cyclically loaded test specimens in an air atmosphere at 80°C for 5 weeks per Nabar, Sean, et al. - GUR1050 E1® against: GUR1050 gamma sterilized (25-40kGy in argon) polyethylene AND sequentially crosslinked and annealed polyethylene (GUR 1050 barstock, 33kGy gamma irradiated in air, annealed at 130°C in air, repeated for total 99kGy, then machined). - GUR1020 E1® against: GUR1050 direct compression molded polyethylene gamma sterilized (25-40kGy) in argon. Performance: E1® material was the only material tested that showed no evidence of environmental stress cracking or fracture and no detectable oxidation (indices <0.1). Both gamma sterilized and sequentially crosslinked and annealed polyethylene showed evidence of increased oxidation and cracking or fracture.
    Claim 3: E1® Antioxidant Infused Technology maintains the mechanical strength of conventional UHMWPE under small punch testing. (Comparative)Test: Small Punch Testing per ASTM F2183. - E1® GUR1050 material compared to GUR1050 gamma sterilized in argon isostatic compression molded (ICM) UHMWPE. - E1® GUR1020 material compared to GUR1050 gamma sterilized (25-40kGy) in argon direct compression molded (DCM) UHMWPE. Performance: - E1® GUR1050 ultimate load: 105 ± 5.5N. GUR1050 ICM control ultimate load: 75.4 ± 5.3N. (p<0.001) - E1® GUR1050 ultimate load: 97.2 ± 6.4N. DCM control ultimate load: 86.6 ± 7.5N. (p<0.001) E1® materials had statistically significantly greater ultimate loads than the respective controls.
    Claim 4: E1® Antioxidant Infused Technology maintains mechanical strength after accelerated aging.Test: Accelerated aging per ASTM F2003 (70°C and 5 atm of oxygen for 14 days). Followed by small punch testing per ASTM F2183 (for ultimate load) and tensile testing per ASTM D638 (for ultimate tensile strength and yield strength). Performance: No significant decrease (P>0.05) in ultimate load, ultimate tensile strength, or yield strength after accelerated aging for either E1® GUR1050 or E1® GUR1020 material. - E1® GUR1020: Ultimate load (before/after aging) 97.2±6.4N / 100.0±5.0N. Ultimate tensile strength 45.8±1.6 MPa / 46.1±2.9 MPa. Yield strength 22.6±0.2 MPa / 22.8±0.3 MPa. - E1® GUR1050: Ultimate load (before/after aging) 105.0±5.5N / 115.0±3.2N. Ultimate tensile strength 43±3 MPa / 43±2 MPa. Yield strength 24.2±0.2 MPa / 24.4±0.2 MPa.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document describes material testing, not patient-level data. The "test sets" refer to the numbers of material specimens or samples used in the mechanical and chemical tests.

    • Sample Size: Not explicitly stated as a single number. For Claim 3, ultimate loads are given with standard deviations (e.g., 105±5.5N), implying multiple samples were tested for each group, but the exact number of replicates is not provided. Similarly for Claim 4, multiple samples are implied. For Claims 1 and 2, it refers to "specimens" without numerical counts.
    • Data Provenance: Not applicable in the context of patient data. The tests are bench tests conducted on materials. The "provenance" refers to the specific materials (GUR1020 and GUR1050 UHMWPE) and their processing methods (isostatically compression molded, crosslinked, gamma irradiation, vitamin E infusion, etc.). It's a laboratory-based study.
    • Retrospective/Prospective: Not applicable. These are in-vitro, bench-top material tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This document details material properties and performance through standardized mechanical and chemical testing, not expert interpretation of clinical data. The "ground truth" is established by the results of these physical tests, which are objective measurements.

    4. Adjudication Method for the Test Set

    Not applicable. This refers to consensus-building among experts for ambiguous cases in clinical studies. For material performance tests, observations (like presence/absence of cracks, oxidation indices, ultimate load measurements) are objective and do not require expert adjudication to establish truth.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a material science study to evaluate the properties of polyethylene used in a shoulder prosthesis, not an AI-based diagnostic or assistive device. There are no human readers or AI involved in interpreting the results.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a material science study, not an algorithm or AI performance evaluation.

    7. The Type of Ground Truth Used

    The ground truth is established by objective measurements and observations from validated laboratory tests on material samples. Examples include:

    • Absence of environmental stress cracking (visual inspection/microscopy).
    • Oxidation indices from infrared spectroscopy (quantitative chemical analysis).
    • Ultimate load measurements from small punch testing (quantitative mechanical test).
    • Ultimate tensile strength and yield strength from tensile testing (quantitative mechanical test).

    These are direct physical and chemical properties of the material, measured using established scientific methods, rather than expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    Not applicable. There is no concept of a "training set" in these material performance studies. The tests are conducted on representative samples of the manufactured material to evaluate its inherent properties.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1