Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K041653

    Validate with FDA (Live)

    Date Cleared
    2005-07-01

    (379 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ComPel 1-Ply Surgical gowns are medical devices that are intended to be worn by operating room personnel during surgical procedures to protect them by inhibiting the migration of liquids across critical areas of the gown.

    When processed according to instructions, ComPel 1-Ply Surgical gowns will function and perform as protective apparel. The gowns are reusable medical devices that are intended to be processed through 100 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterile products that are intended to be processed by health care facilities and/or contract laundry companies.

    Device Description

    All the components used in these ComPel® Gowns are made from 100% Dacron. All the ouripented acomPel fabrics are dyed and treated with a fluorocarbon finish.

    ComPel® Gowns will function as surgical attire when processed according to instructions through 100 complete cycles - wash, dry and sterilization. These products will be manufactured and distributed as non-sterile gowns that are intended to be processed by health care facilities and/or contract sterilization/laundry companies.

    AI/ML Overview

    This document describes the regulatory submission for the "ComPel 1-Ply Surgical Gowns." It focuses on demonstrating that the device is substantially equivalent to legally marketed predicate devices. The study conducted is a series of laboratory tests to confirm the physical properties and performance characteristics of the surgical gowns against established standards.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Test NameStandard / MethodReported Device PerformanceComments
    Barrier Properties
    Liquid Barrier (Hydrostatic) TestAATCC #127-1980Not explicitly quantified, but stated to meet the "Level 2 Barrier" classification.Per ANSI/AAMI PB70:2003 standard, the 1-ply surgical gowns with flat felled seam design are classified as a Level 2 Barrier after successful testing.
    Impact Penetration TestAATCC 42-2000Not explicitly quantified.Successfully completed.
    Durability
    Durability through ProcessingThrough 100 processings (wash, dry and sterilization)Gowns will function and perform as protective apparel through 100 complete cycles.This indicates the gowns maintained their functional integrity after 100 cycles of washing, drying, and sterilization.
    Material Strength
    Strength TestASTM D-1682-75Not explicitly quantified.Successfully completed.
    Other Properties
    FlammabilityCode of Federal Regulations 1610 (CS-191-53)Not explicitly quantified.Successfully completed.
    LintIST 160.0-83Not explicitly quantified.Successfully completed.
    Biocompatibility/Safety
    Cytotoxicity (MEM Elution)MG023Not explicitly quantified.Successfully completed.
    Primary Skin IrritationFHSANot explicitly quantified.Successfully completed.
    Sterilization Compatibility
    Compatibility with Steam SterilizationNot specifiedNot explicitly quantified.Successfully completed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each individual test. It refers to "the tests that have been successfully completed," implying that a sufficient number of samples were tested to demonstrate compliance.

    • Sample Size: Not specified in the provided text for individual tests.
    • Data Provenance: The data is generated from laboratory testing of the ComPel 1-Ply Surgical Gowns, conducted by the manufacturer (Standard Textile Company 1). This is prospective data generation for the purpose of regulatory submission. The country of origin for the data generation would be the United States, as the manufacturer is based in Ohio.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable for this device. The "ground truth" for a surgical gown is adherence to established industry standards and test methods for physical and biological properties. These standards themselves are developed by expert consensus, but the testing itself does not require expert adjudication of "ground truth" for individual samples in the way a diagnostic AI might. The results are objective measurements against predefined thresholds within the standards.

    4. Adjudication Method for the Test Set

    This is not applicable. The tests are objective laboratory measurements against predefined standards. There is no human "adjudication" in the sense of multiple reviewers making subjective decisions. The results of the tests are either within the acceptance criteria for the standard or they are not.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where human interpretation plays a significant role, often comparing AI-assisted human performance against unaided human performance. A surgical gown is a protective barrier, not a diagnostic tool requiring image interpretation or similar cognitive tasks.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical product (a surgical gown), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on performance against established industry standards and test methods.

    • For barrier properties: Adherence to AATCC #127-1980 and AATCC 42-2000, which then correlates to a classification like "Level 2 Barrier" as per ANSI/AAMI PB70:2003.
    • For physical properties (strength, durability, flammability, lint): Meeting thresholds defined in ASTM D-1682-75, CFR 1610, IST 160.0-83, and manufacturer's internal durability specifications (lasting 100 cycles).
    • For biocompatibility: Meeting criteria in MEM Elution (MG023) and Primary Skin Irritation (FHSA).
    • For sterilization: Compatibility with steam sterilization implies the material structure and performance are not negatively affected.

    These standards and methods represent a consensus of scientific and technical understanding for the performance of surgical attire.

    8. The Sample Size for the Training Set

    This is not applicable. This device is a physical product, not a machine learning model. Therefore, there is no "training set." The materials used in the gowns (100% Dacron, fluorocarbon finish) are known materials with established properties.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1