Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K102033

    Validate with FDA (Live)

    Date Cleared
    2011-04-07

    (262 days)

    Product Code
    Regulation Number
    870.2850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B. Braun Combitrans Monitoring Sets are intended for the extravascular measurement of venous, arterial, or pulmonary arterial blood pressure by converting hemodynamic waveforms into electrical signals.

    The Combitrans flush device provides a continuous flush rate from approximately 3 ml/hr and allows a switch to approximately 300 ml/hr. When using the Combitrans device in neonates, infants or small children. A syringe pump should be used to control the flow rate, which should not exceed 3 ml/hr.

    Device Description

    The Combitrans Monitoring Set is a pre-calibrated blood pressure transducer designed to transform physiological pressure waveforms into electrical signals using a piezoresistive sensor. The set is supplied in 6 configurations:

    1. Combitrans Monitoring Set, Venous (blue striped tubing).

    2. Combitrans Monitoring Set, Arterial (red striped tubing).

    3. Combitrans Monitoring Set, Pulmonary-Arterial (yellow striped tubing).

    4. Exadyn Combitrans Monitoring Set (blue and red striped tubing with additional stopcock for the measurement of venous and arterial blood pressure with one transducer).

    5. Combitrans Add-On Set (consists only of the transducer, flush device, three-way stopcock and pressure tubing for customized set design and extension.

    6. Combitrans Fixation Tape (includes a hook and loop band/Velcro fixation tape used to fix the Combitrans transducer to the patient forearm or a stand).

    Each set has the same basic intended use and transducer components, but varies according to the features indicated above. With each of the sets, a fluid (saline) filled pressure transmitting catheter can be connected to a venous, arterial or pulmonary-arterial patient catheter that is in indirect contact with the patient's bloodstream.

    Each of the six Combitrans Monitoring Sets listed above are composed of the following components: a disposable extra-vascular blood pressure transducer, a sensor that converts hemodynamic pressure waveforms into electrical signals, a transducer cable which is connected to a pressure transmitting catheter system, a flush device and a three-way stopcock. The Combitrans transducer and the sensor function together such that physiologic waveforms are transmitted from a fluid filled cavity inside the transducer housing onto the sensor via a silicone gel cushion.

    The Combitrans Monitoring Set finger-activated flush device provides a continuous flush from approximately 3 mL/hr and allows a switch to approximately 300 ml/hr. The Combitrans Monitoring Set interfaces with a standard infusion set and a transducer cable, which transmits electrical signals onto a monitor or into a computer for further processing, graphic presentation or storage. When using the Combitrans device in neonates, infants or small children, the standard gravity infusion set is not used but is exchanged for a syringe pump to control the flow rate. which should not exceed 3 ml/hr.

    Combitrans disposable transducers, as well as catheters and infusion systems, are indirectly in contact with the blood stream. They are supplied sterile, individually packaged and for single use only.

    Accessories available for use with the Combitrans Monitoring Sets include attachment plates for the attachment of the transducer to an infusion stand. A wide range of interface cables using different plug design and configurations are also available to facilitate storage or visualization of blood pressure measurements on patient monitors.

    AI/ML Overview

    The provided text describes the Combitrans Monitoring Set, a blood pressure transducer, and its 510(k) premarket notification (K102033). It outlines the device's intended use, components, and confirms its substantial equivalence to a predicate device.

    The acceptance criteria and study information provided mainly focus on the device's compliance with established standards for performance, electromagnetic compatibility, and biocompatibility, rather than a clinical study establishing specific accuracy metrics against a ground truth.

    Here's an analysis of the provided information based on your requested points:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Performance Standard: ANSI/AAMI BP22-1994/(R)2006: "Blood pressure transducers"Performance Met: "Results of performance testing demonstrate that the Combitrans Monitoring Sets meet all requirements of the standard."
    Electromagnetic Compatibility (EMC) Standard: IEC 60601-2-34 (2000-10) "Medical Electrical Equipment Part 2-34: Particular Requirements for the Safety, including Essential performance, of Invasive Blood Pressure Monitoring Equipment."EMC Met: "The Combitrans Monitoring Set fulfills all requirements for electromagnetic compatibility as defined by this standard."
    Conical Fittings Standard: ISO 594-2:1998 "Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock fittings."Fittings Met: "All conical fittings of the Combitrans Monitoring Sets were tested in accordance with the requirements identified in ISO 594-2:1998... The Combitrans Monitoring Set fulfilled all requirements for conical luers as defined by this standard."
    Biocompatibility Standard: ISO 10993-1:2003 (Biological evaluation of medical devices according to applicable patient contact categories)Biocompatibility Met: "All materials used for the Combitrans disposable pressure transducer sets met requirements for the biological evaluation of medical devices according to the applicable patient contact categories identified in ISO 10993-1:2003."
    Flush Rate (intended use): Continuous flush from approx. 3 mL/hr, switch to approx. 300 ml/hr. For neonates/infants/small children, syringe pump to control flow rate not exceeding 3 ml/hr.(This is an intended functional specification, and the document states the device provides this functionality as part of its description and intended use, implying it meets this functionality but does not present specific test results demonstrating the precision of these flow rates in a performance summary. It's assumed to be met as part of the overall design and performance.) "The Combitrans Monitoring Set finger-activated flush device provides a continuous flush from approximately 3 mL/hr and allows a switch to approximately 300 ml/hr." and "The Combitrans flush device provides a continuous flush rate from approximately 3 ml/hr and allows a switch to approximately 300 ml/hr. When using the Combitrans device in neonates, infants or small children a syringe pump should be used to control the flow rate, as the flow rate should not exceed 3 ml/hr."

    2. Sample size used for the test set and the data provenance
    The document does not specify sample sizes for the performance testing. The provenance of the data is implicit: it's from internal testing performed by B. Braun Melsungen AG to support their 510(k) application. There is no indication of country of origin for the data or whether it was retrospective or prospective. The studies mentioned are engineering and laboratory tests against standards, not clinical trials with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable or provided. The "ground truth" for these tests are the requirements and specifications defined by the referenced international and national standards (e.g., ANSI/AAMI BP22, IEC 60601-2-34, ISO 594-2, ISO 10993-1). Compliance is determined by the device's physical and functional performance against these objective standards, not through expert consensus on medical data.

    4. Adjudication method for the test set
    Not applicable. Testing performed was against established engineering and performance standards, not against a human-interpreted ground truth requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a blood pressure transducer, a medical device for measuring physiological parameters. It does not involve "readers," "cases," or Artificial Intelligence (AI) in the context of diagnostic interpretation, therefore no MRMC or AI effectiveness study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. There is no "algorithm" in the sense of AI or image analysis involved with this device. It is a physical transducer that converts pressure into electrical signals. Its performance is inherent to its design and manufacturing, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The "ground truth" for the device's performance is compliance with the objective requirements outlined in consensus standards such as ANSI/AAMI BP22-1994/(R)2006 (for blood pressure transducers), IEC 60601-2-34 (for electromagnetic compatibility), ISO 594-2:1998 (for conical fittings), and ISO 10993-1:2003 (for biocompatibility). These standards define measurable parameters and acceptable ranges.

    8. The sample size for the training set
    Not applicable. This device does not use machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established
    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1