510(k) Data Aggregation

    K Number
    DEN130033

    Validate with FDA (Live)

    Device Name
    COGNIVUE
    Date Cleared
    2015-06-05

    (723 days)

    Product Code
    Regulation Number
    882.1470
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cognivue testing is indicated as an adjunctive tool for evaluating perceptual and memory function in individuals aged 55-95 years old.

    Device Description

    Cognivue provides clinicians in a healthcare setting with objective measurements of cognitive function as a screening aid in the assessment of adults 55 years of age and older. This is done for the purpose of identifying a potential decline in cognitive function relative to baseline test performance of other age-normal adults, referring those adults for further testing where warranted, and monitoring changes in cognitive function over time.

    Cognivue presents a series of visual stimuli and a wedge-shaped cursor. The display presents stimuli at varying signal strengths with a moving domain-specific target. The patient identifies the target location by moving the cursor using the rotatory mouse manipulandum. The patient is scored based on the timing and accuracy of the responses. Repeated trials of varying difficulty characterize the subject's performance in each of the tested functional domains.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the CogniVue device meets those criteria, based on the provided text:


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryAcceptance Criteria (from study objective/endpoints)Reported Device Performance (from study results)
    Agreement with Reference Standard (SLUMS)Cognivue agrees with the non-reference standard, SLUMS, in identifying unimpaired and impaired test subjects.Agreement Analysis (SLUMS and Cognivue):- Overall relationship: Deming regression showed Cognivue = -29.98 + 4.465 SLUMS with r = 0.71 (p<.0001).- Categorical Agreement (including intermediate): - Implies SLUMS scores grouped with unimpaired: PPA = 56% (95% CI: .47 to .65), NPA = 95% (95% CI: .91 to .97). - Quadratic weighted kappa score = 0.57 (95% CI: .50 to .63), suggesting a clear categorical relationship.- Categorical Agreement (omitting intermediate): - PPA = 82% (95% CI: .72 to .89), NPA = 98% (95% CI: .93 to .99).- The study concludes "Cognivue can inform an impression that a patient is unimpaired or is impaired." and "Disagreement between Cognivue and SLUMS is attributable to subjects whom Cognivue test result classifies a subject as unimpaired and SLUMS scores who score in the intermediate classification of mildly impaired."
    Test-Retest ReliabilityCognivue shows test-retest reliability comparable to that of the SLUMS.Cognivue Test-Retest Regression Analysis (continuous scores):- Deming regression: Cognivue 2 = 9.06 + 0.936 Cognivue 1 with r = 0.90 (p=.0264).- Substantial and significant agreement.Cognivue Test-Retest Reliability Analysis (categorical classification):- PPA = 89% (95% CI: .77 to .95)- NPA = 93% (95% CI: .90 to .96)- Fisher-Freeman-Halton exact test yields p = 0.161, indicating Cognivue classifications (impaired, mildly impaired, unimpaired) do not differ significantly by time of test.Comparison to SLUMS Test-Retest Reliability:- SLUMS Deming regression: SLUMS 2 = 4.15 + 0.90 SLUMS 1 with r = 0.82 (p=.039).- SLUMS categorical classification: PPA = 87%, NPA = 87%.- Fisher-Freeman-Halton exact test for SLUMS yields p < 0.001, indicating SLUMS classifications do differ significantly by time of test. Cognivue's reliability appears better than SLUMS in terms of classification stability.
    Psychometric Validation (Construct Validity)(Implied: CogniVue scores correlate meaningfully with established neuropsychological tests, demonstrating construct validity)Psychometric Validation of the Cognivue Test Battery:- Pearson correlations between Cognivue Average Score and various neuropsychological tests (e.g., PegBrdLeft: 0.597, Animal_Name: 0.655, TrailsBTime: -0.663) averaged slightly higher than SLUMS correlations in many cases. Correlations ranged approximately from 0.40 to 0.66, which is stated as "within the acceptable ranges for determining construct validity for psychometric tests."
    Cut-off Values Establishment(Implicit: Robust determination of cut-off values for impaired, intermediate, and unimpaired classifications)Cognivue Cut-Offs Analysis:- Determined through a preliminary study of 92 subjects comparing Cognivue performance against SLUMS.- Two optimization methods were used (minimizing objective functions related to PPA/NPA and inaccuracy/error bias).- Established Cognivue scores: ≤50 (impaired), ≥75 (unimpaired), 51-74 (intermediate).- Example: For SLUMS < 21 or >=27 (impairment cut-off values), Cognivue related PPA = 86%, NPA = 50%, Accuracy = 0.86.- For SLUMS ≥ 27 and Cognivue ≥ 75 (unimpaired cut-off values): PPA = 66%, NPA = 67%, Accuracy = 0.66.

    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Main Clinical Validation Test Set: 401 subjects.
      • Test-Retest Reliability Test Set: 358 subjects (a subset of the 401, those available for repeat testing).
      • Cut-off Determination Test Set (Training Set for Cut-offs): 92 subjects.
      • Data Provenance: The subjects were enrolled at 13 older adult independent living communities and adult programs for independent living seniors who are living at home. The study was prospective in its design, involving specific recruitment and protocol execution for the study. No explicit country of origin is stated, but given the FDA De Novo filing, it's highly likely the data is from the USA.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The primary ground truth for classification was the St. Louis University Mental Status (SLUMS) Examination. While the SLUMS is a clinical tool administered by a "Tester," the text doesn't specify if this "Tester" is explicitly an "expert" (e.g., a neurologist or psychologist) or a trained administrator. Given that SLUMS scores have validated cut-off scores, it relies on the established validity of the SLUMS itself, rather than real-time expert consensus for each case. The study states "Tester #2 administration of the neuropsychological test battery," implying trained personnel administered the SLUMS.
      • The study used "licensed practitioners evaluating brain function" as the general user group for the device, implying that interpretation of SLUMS scores would typically fall under their purview.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • There was no adjudication method mentioned for establishing the ground truth (SLUMS scores). The SLUMS is a standardized psychometric test with established scoring rules and cut-offs.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No such MRMC comparative effectiveness study was conducted. The study design was a comparative study between the Cognivue device and a battery of non-reference standard tests (including SLUMS), not a study comparing human readers with and without AI assistance. The Cognivue is intended as an "adjunctive tool," meaning a clinician would use its output alongside other information, but the study did not quantify improvement of human performance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance assessment was done. The entire clinical validation focuses on the Cognivue device's output (an average score classified as impaired, intermediate, or unimpaired) and its agreement with the SLUMS examination and other neuropsychological tests. The device itself generates the score and classification automatically without real-time human interpretation or modification of its core output.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The primary ground truth used was a validated clinical psychometric scale: the St. Louis University Mental Status (SLUMS) Examination. This is a well-established cognitive screening tool with predefined cut-off scores for unimpaired, intermediate (mildly impaired), and impaired cognitive function. Secondary ground truth involved other established neuropsychological tests for construct validity.
    7. The sample size for the training set:

      • The cut-off values for Cognivue (impaired, intermediate, unimpaired) were determined using a preliminary study of 92 subjects. This set of 92 subjects essentially served as a training set for calibrating the device's classification thresholds against the SLUMS.
    8. How the ground truth for the training set was established:

      • For the 92 subjects used to establish the Cognivue cut-offs, the ground truth was established by administering the St. Louis University Mental Status (SLUMS) Examination. The SLUMS scores were then used to classify these 92 subjects into impaired, intermediate, and unimpaired categories based on their established cut-offs (e.g., >26 unimpaired, <21 impaired for high school graduates). The Cognivue scores were then optimized to align with these SLUMS classifications.