    K Number
    K040821

    Validate with FDA (Live)

    Date Cleared
    2004-04-23

    (24 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Coalescent Surgical PG-CLIP™ and accessories are intended for endoscopic and non-endoscopic peripheral vascular and general surgical soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomoses in peripheral vascular blood vessels, grafts and other tubular structures. The PG-CLIP is not intended for use in cardiac and/or thoracic procedures; including, but not limited to intra-cardiac and coronary artery bypass grafting (CABG) procedures.

    Device Description

    The Coalescent Surgical PG-CLIP™ device is a self-closing clip for peripheral vascular anastomosis and tissue approximation/attachment applications. The PG-CLIP consists of specially designed vascular clips optionally connected to needles via flexible members. These clips may or may not include coils surrounding a Nitinol wire core. This design allows precise placement of clips prior to closure and facilitation of an interrupted "suture" technique through elimination of knot tying. The device is fabricated from standard medical and implantable grade materials. Accessories include Clip delivery and removal devices as well as vessel cutting and retention tools.

    AI/ML Overview

    The provided text is a 510(k) summary for the Coalescent Surgical PG-CLIP™ device. It's a premarket notification to demonstrate that the device is substantially equivalent to legally marketed predicate devices. This type of submission focuses on comparing a new device to an existing one, rather than proving the new device's performance against specific acceptance criteria through extensive new studies.

    Therefore, the document does not contain the kind of information requested regarding acceptance criteria, a study proving the device meets those criteria, or specific performance metrics in the way a clinical trial or algorithm validation study would.

    Here's why the requested information cannot be extracted and what the document does state:

    • No new performance data was acquired for this specific submission. The crucial sentence is: "No test data has been acquired related to this restricted indication for use." This means the submission relies on the existing performance data and substantial equivalence demonstrated for the predicate devices.
    • The submission is based on "substantial equivalence." This regulatory pathway avoids the need for a full clinical trial to establish new safety and effectiveness criteria if the device is sufficiently similar to an already cleared device.

    Given this, I cannot fill out the requested table and answer the specific questions about performance studies. However, I can explain why based on the provided text.


    Reason For Inability to Fulfill Request:

    The provided 510(k) summary explicitly states: "No test data has been acquired related to this restricted indication for use." (Section: Test Data). This means that for this particular submission (K040821), no new studies, performance data, or acceptance criteria beyond those established for the predicate devices were used or provided to the FDA. The submission relies on establishing "substantial equivalence" to previously cleared devices (Coalescent Surgical U-CLIP™ K031623, K024366, K023125, K021407, K013664, K012317, K994160, and K971588).

    Therefore, specific information regarding:

    1. A table of acceptance criteria and reported device performance
    2. Sample sizes used for the test set and data provenance
    3. Number and qualifications of experts for ground truth
    4. Adjudication method
    5. MRMC comparative effectiveness study results
    6. Standalone performance
    7. Type of ground truth used
    8. Sample size for the training set
    9. How training set ground truth was established

    is not available within this document because such studies were not conducted or reported for this specific 510(k) submission. The FDA clearance for K040821 was based on the substantial equivalence to predicate devices, leveraging their existing pre-market notification data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K024366

    Validate with FDA (Live)

    Date Cleared
    2003-01-17

    (17 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Coalescent Surgical U-CLIP™ is intended for endoscopic and non-endoscopic general soft tissue and prosthetic material approximation/attachment and/or ligation and the creation of anastomoses in blood vessels, grafts and other tubular structures; including use in cardiovascular and coronary artery bypass grafting procedures.

    Device Description

    The Coalescent Surgical U-CLIP™ device is a self-closing clip for vascular anastomosis and tissue approximation/attachment applications. The U-CLIP consists of specially designed vascular clips optionally connected to needles via flexible members. These clips may not include coils surrounding a Nitinol wire core. This design allows precise placement of clips prior to closure and facilitation of an interrupted "suture" technique through elimination of knot tying. The device is fabricated from standard medical and implantable grade materials. Accessories include Clip delivery and removal devices as well as vessel retention tools.

    AI/ML Overview

    Due to the nature of the provided text, which is an FDA 510(k) summary for a medical device and not a study report, it does not contain the specific acceptance criteria or the detailed results of a study as typically presented for AI/ML device evaluations. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against quantified acceptance criteria for a novel AI/ML application.

    Therefore, many of the requested items (e.g., sample sizes for test/training, number of experts, adjudication methods, MRMC study results, ground truth types for AI/ML) are not applicable or extractable from this document.

    However, I can extract information related to the device characteristics and the claim of equivalence.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't define explicit numerical acceptance criteria for AI/ML performance metrics. Instead, it states:

    CriterionReported Device Performance
    Functional Characteristics"In vitro and in vivo data have confirmed that functional characteristics are substantially equivalent to the predicate device cited as well as applicable USP suture testing requirements. All data fell well within, both, internal specification requirements, as well as external standard requirements and device performance expectations."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable. This document describes a medical device, not an AI/ML algorithm. The "test set" would refer to the in vitro and in vivo studies performed, but specific sample sizes are not provided in this summary.
    • Data Provenance: The document does not specify the country of origin for the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not Applicable. As this is not an AI/ML study, there is no mention of experts establishing a "ground truth" in the context of an algorithmic evaluation. The evaluation is based on standard biomedical testing.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No information on adjudication methods for an AI/ML test set is present.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • Not Applicable. This is not an AI/ML device, so an MRMC comparative effectiveness study in the context of human readers analyzing AI output is not relevant nor mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance relies on established in vitro and in vivo biological and mechanical testing methods, and adherence to applicable USP suture testing requirements and internal/external standard requirements. It's not a "ground truth" in the sense of expert-annotated data for an AI/ML system.

    8. The Sample Size for the Training Set:

    • Not Applicable. This document does not describe the development or training of an AI/ML algorithm; therefore, a "training set" in that context is not relevant.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K021407

    Validate with FDA (Live)

    Date Cleared
    2002-07-03

    (61 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The modified Coalescent Surgical U-CLIP™ is intended for endoscopic and non-endoscopic tissue and prosthetic material approximation/attachment and the creation of anastomoses in blood vessels, grafts and other tubular structures; including use in cardiovascular and coronary artery bypass grafting procedures.

    Device Description

    The modified Coalescent Surgical U-CLIP™ is a single self-closing clip for vascular anastomosis and tissue approximation/attachment applications. The U-CLIP consists of a specially designed vascular clip connected to needles via flexible members. This design allows precise placement of clips prior to closure and facilitation of an interrupted "suture" technique. The device is fabricated from standard medical and implantable grade materials. Accessories include Clip removal devices and vessel retention tools.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Coalescent Surgical U-CLIP™ and accessories, focusing on its substantial equivalence to a predicate device. This type of regulatory submission primarily relies on demonstration of equivalence rather than extensive de novo clinical trials with predefined acceptance criteria and performance metrics in the way a novel AI algorithm would.

    Therefore, the requested information about acceptance criteria, detailed study design, sample sizes, expert ground truth, MRMC studies, or standalone algorithm performance is not present in the provided document, as it is not typically required for a 510(k) submission based on substantial equivalence to a predicate device.

    However, I can interpret the available information to answer what data was presented and how the device met its 'acceptance criteria' in the context of a 510(k).

    Here's how the information relates to your request:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Characteristics Equivalence (to predicate device)"In vitro data confirm that functional characteristics are substantially equivalent to the predicate device cited. All data fell well within, both, internal specification requirements, as well as external standard requirements and device performance expectations."
    Substantial Equivalence (to predicate device K013664)The submission states: "The modified Coalescent Surgical U-CLIP™ and accessories is substantially equivalent in intended use, fabrication, design and method of operation to the following predicate device: Coalescent Surgical U-CLIP™ 510(k) K013664." And "Based upon the product technical information, intended use and clinical performance information provided in this pre-market notification, the modified Coalescent Surgical U-CLIP™ and accessories has been shown to be substantially equivalent to a currently marketed predicate device." The FDA concurred with this determination, as indicated by the clearance letter.
    Adherence to Internal and External Standards"All data fell well within, both, internal specification requirements, as well as external standard requirements and device performance expectations."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified. The document only mentions "In vitro data" and "product technical information." Clinical data with specific sample sizes for a "test set" in the context of an AI device are not present.
    • Data provenance: Not specified. The data is described as "In vitro data." No information about country of origin or whether it was retrospective/prospective is provided, as this is not relevant for an in-vitro performance test of a mechanical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not specified. This information is relevant for AI algorithms relying on expert annotations for ground truth. This document pertains to a medical device (surgical clip), where "ground truth" would be established through engineering specifications, material testing, and performance against defined physical standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified. Adjudication methods are typically used when human experts review and categorize data, often in the context of AI model evaluation or clinical trials with subjective endpoints. This document focuses on a physical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study is completely irrelevant for this type of medical device (surgical clip). MRMC studies are designed to evaluate the impact of medical imaging devices or AI aids on human reader performance, which isn't the function of the U-CLIP.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. The U-CLIP is a physical surgical device, not an algorithm. Therefore, "standalone algorithm performance" doesn't apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device would be based on engineering specifications, material properties, and performance benchmarks for mechanical strength, biocompatibility, and functional operation (e.g., clip closure force, material integrity under stress, sealing capability). The document states "internal specification requirements, as well as external standard requirements and device performance expectations." This implies that the device was tested against established physical and material standards.

    8. The sample size for the training set

    • Not applicable. This device is not an AI algorithm, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As above, there is no "training set" for this physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013664

    Validate with FDA (Live)

    Date Cleared
    2002-01-24

    (79 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Coalescent Surgical U-CLIP™ is intended for endoscopic and non-endoscopic use in tissue and prosthetic material approximation and the creation of anastomoses in blood vessels, grafts and other tubular structures; including cardiovascular and coronary artery bypass grafting procedures.

    Device Description

    The Coalescent Surgical U-CLIP™ is a single self-closing clip for vascular anastomosis and tissue approximation applications. The U-CLIP consists of a specially designed vascular clip with a needle connected to one end via a flexible member. This design allows precise placement of clips prior to closure and facilitation of an interrupted technique. The device is fabricated from standard medical and implantable grade materials.

    AI/ML Overview

    I am sorry, but the provided text does not contain the specific information required to complete the table and answer the questions asked about acceptance criteria and study details. The document is a 510(k) summary for a medical device (Coalescent Surgical U-CLIP™), which primarily focuses on establishing substantial equivalence to predicate devices, rather than detailing specific performance acceptance criteria and study results in the manner requested.

    The document mentions "Clinical data confirm that functional characteristics are substantially equivalent to the predicate, and fall within, both, internal specification requirements, as well as device performance expectations," but it does not provide:

    • A table of specific acceptance criteria.
    • Reported device performance against those criteria.
    • Sample sizes for test or training sets.
    • Data provenance.
    • Details about expert involvement, adjudication, or ground truth establishment.
    • Information on MRMC studies or standalone algorithmic performance.

    Therefore, I cannot fulfill your request based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K012317

    Validate with FDA (Live)

    Date Cleared
    2001-08-31

    (39 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Coalescent Surgical U-Clip™ is intended for endoscopic and non-endoscopic use in tissue and prosthetic material approximation and the creation of anastomoses in blood vessels, grafts and other tubular structures.

    Device Description

    The Coalescent Surgical U-Clip™ is a single self-closing clip for vascular anastomosis and tissue approximation applications. The U-Clip consists of a specially designed vascular clip with a needle connected to one end via a flexible member. This design allows precise placement of clips prior to closure. The device is fabricated from standard medical and implantable grade materials.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Coalescent Surgical U-Clip™. However, it does not contain the detailed information required to fill out all sections of your request regarding acceptance criteria and the specific study proving the device meets them. The document states that "in vitro data confirm that functional characteristics are substantially equivalent to the predicate device cited. Testing included in vitro studies. All data fell well within, both, internal specification requirements, as well as external standard requirements and predicate performance expectations." This is a high-level summary and lacks the specifics you're asking for.

    Here's an attempt to answer based on the available text, with explicit notes where information is missing:


    Acceptance Criteria and Device Performance

    The document does not explicitly state quantitative acceptance criteria or detailed reported device performance in a table format. It generally claims that "All data fell well within, both, internal specification requirements, as well as external standard requirements and predicate performance expectations."

    Acceptance Criteria (Not explicitly stated, inferred requirements for equivalence)Reported Device Performance (Summary statement)
    Functional characteristics substantially equivalent to predicate deviceData confirmed substantial equivalence
    Met internal specification requirementsAll data fell well within
    Met external standard requirementsAll data fell well within
    Met predicate performance expectationsAll data fell well within

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      The document mentions "in vitro studies" but does not specify the sample size for any test set or the data provenance. It only states these were "in vitro test data."

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      This information is not provided in the document. The testing described is "in vitro data," implying laboratory or bench testing rather than human subject testing requiring expert ground truth establishment in a clinical context.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      This information is not provided in the document. Given the "in vitro" nature, clinical adjudication methods are unlikely to be relevant here.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No MRMC study was mentioned. This device is an implantable surgical clip, not an AI/software diagnostic tool, so such a study would not be applicable.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      No standalone algorithm performance study was mentioned. This device is a physical surgical implant, not an algorithm. The "in vitro" testing implies standalone device performance, but not in the context of an algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For "in vitro data," the "ground truth" would likely be established engineering specifications, material properties, mechanical test standards, and comparison to the predicate device's known performance characteristics. The document does not explicitly state the type of ground truth beyond fulfilling "internal specification requirements" and "external standard requirements."

    7. The sample size for the training set:
      This information is not applicable as the device is a physical medical device, not a machine learning algorithm that requires a training set.

    8. How the ground truth for the training set was established:
      This information is not applicable as the device is a physical medical device, not a machine learning algorithm.