Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113166

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-16

    (202 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearPath Upper Gl is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

    Device Description

    The ClearPath Upper GI includes one suction tube + tip, one irrigation tube + tip, and a sleeve to attach to the endoscope. The ClearPath Upper GI is connected to the ClearPath Controller and the ClearPath Tubing (K091305) and allows for irrigation and evacuation of debris from the upper gastrointenstinal tract during an endoscopic procedure. The device can accommodate several sizes and configurations of endoscopes with a change in the attachment ring.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the ClearPath Upper GI device, which is an irrigation/evacuation system for endoscopic procedures. The submission is for a modified design of a previously cleared device (K093779).

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics. Instead, it describes performance testing in qualitative terms and asserts equivalence to the predicate device.

    Acceptance Criteria Category (Implied)Reported Device Performance
    Bench Testing
    Dimensional compliancePerformed to specifications
    Strength compliancePerformed to specifications
    Functional compliancePerformed to specifications
    Mechanical compliancePerformed to specifications
    Compliance with endoscopesPerformed to specifications
    Animal Testing
    No dislodgement of tipPerformed as expected
    Ease of maneuverability/advancePerformed as expected
    Quality of visibilityPerformed as expected
    Quality of irrigationPerformed as expected
    No occlusion/interruption to suctionPerformed as expected
    Quality of evacuation (blood, bile, etc.)Performed as expected
    Overall PerformanceThe modified ClearPath Upper GI disposable performed to its specifications and/or was equivalent to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a numerical sample size for either the bench or animal testing. It refers to "animal testing," implying more than one animal, but no specific count is given.
    • Data Provenance: The testing appears to be conducted by the manufacturer, EasyGlide Ltd., which is based in Kfar Truman, Israel. The study type is prospective as it involves new testing performed for this 510(k) submission, referencing "We have performed bench and animal testing to verify..."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish a "ground truth" in the traditional sense (e.g., for diagnostic accuracy). The testing performed seems to be primarily engineering validation (bench tests) and observational assessment by presumably trained personnel during animal studies. No specific number or qualifications of experts for ground truthing are provided.

    4. Adjudication Method for the Test Set

    Not applicable. The reported studies are primarily performance verification against specifications and observational assessments during animal trials, not diagnostic accuracy studies requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is an irrigation/evacuation system, not a diagnostic imaging device typically requiring MRMC studies to assess human reader performance with and without AI assistance.

    6. Standalone Performance Study

    Yes, a standalone performance study was done in the sense that the device's performance was evaluated independently through bench and animal testing against its specifications and the performance of the predicate device. This is not "algorithm only" performance as the device is a physical medical device. The testing aimed to confirm that the modified device itself, when used as intended, meets its design and functional requirements.

    7. Type of Ground Truth Used

    The "ground truth" in this context is based on:

    • Pre-defined specifications for dimensional, strength, functional, and mechanical testing.
    • Observational assessment during animal testing for factors like dislodgement, maneuverability, visibility, irrigation quality, and evacuation effectiveness. This observational assessment likely relies on the expertise of veterinary or medical personnel conducting the animal procedures.
    • Equivalence to the predicate device (K093779), meaning the performance of the modified device was compared to the expected performance of the predicate.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical, mechanical system, not an AI/ML algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1