510(k) Data Aggregation

    K Number
    K963097

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-10-18

    (70 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BMW's CVC-PICC is designed to establish peripheral access to the central venous system for administration of fluids including, but not limited to, hydration agents, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal.

    Device Description

    BMW Medical, Inc.'s peripherally inserted central venous catheters are configured as long-term single lumen catheters. They are composed of extruded silicone rubber tubing which is homogeneously mixed with barium sulfate (BaSO ) prior to extrusion to impart radiopacity. The catheters vary in length and diameter to accommodate specific patient populations. A 2-piece plastic adapter is attached to the proximal end of each single lumen catheter. The catheter is mechanically locked onto the adapter with a silicone rubber sleeve. The other end of the adapter terminates in a female luer which allows attachment of the catheter to a male luer fitting. The adapter houses a slitted disc of silicone rubber which serves as a three-way valve. The valve remains closed when the catheter is not in use. The valve opens inward toward the distal tip of the catheter when fluids are infused into the patient via the catheter, and it opens outward toward the proximal hub during aspiration of blood samples.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a medical device and describes its equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in the way a clinical trial or performance study for a novel device would. Therefore, much of the requested information regarding sample size, expert ground truth, adjudication, MRMC studies, and training sets is not applicable to this type of submission.

    However, I can extract the information provided regarding performance and equivalence.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Reduce potential for air embolism"Performance data indicate that the BMW CVC-PIC catheter is at least as effective in minimizing these potentials as are the predicate devices..."
    Reduce bleedback"Performance data indicate that the BMW CVC-PIC catheter is at least as effective in minimizing these potentials as are the predicate devices..."
    Valve function maintained after guide wire/stylet passage"...even when a guide wire or stylet stiffener is passed through the valve located in the CVC-PICC hub. Studies have shown that repeated insertion and removal of the guide wires or stylet through the valve does not compromise its function."
    No hemolysis of blood during aspiration"No hemolysis of blood is observed when aspirating blood through BMW's valve..."
    Low aspiration pressures"...aspiration pressures are low."
    Biocompatibility of hub material"Published biocompatibility flow chart requirements indicate that the BMW's hub material meets biocompatibility requirements, and is therefore acceptable for use in this catheter application."
    Catheter stiffness comparable to predicate"Further, test data collected on the stiffness... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Elongation at break comparable to predicate"Further, test data collected on the... elongation at break... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Radiopacity comparable to predicate"Further, test data collected on the... radiopacity... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Flow rate comparable to predicate"Further, test data collected on the... flow rate... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Tensile strength comparable to predicate"Further, test data collected on the... tensile... strength... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Burst strength comparable to predicate"Further, test data collected on the... burst strength... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Valve function comparable to predicate"Further, test data collected on the... valve function... demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."
    Leak test results comparable to predicate"Further, test data collected on the... leak test data demonstrate that the BMW CVC-PICC is substantially equivalent to the predicate devices."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for any tests, nor does it detail the provenance of the data (country, retrospective/prospective). The tests mentioned appear to be laboratory or bench tests ("performance data," "test data collected on") rather than clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a 510(k) submission based on engineering and performance testing for substantial equivalence, not a study requiring expert-established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a catheter, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a catheter, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance appears to be based on established engineering standards and comparative performance against the predicate devices in various physical and functional tests. For biocompatibility, "published biocompatibility flow chart requirements" served as the standard.

    8. The sample size for the training set

    Not applicable. This is a medical device, not a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable.