    Intended Use

    BMW's multi-lumen central venous catheters have the same intended use as Bard Access Systems' Groshong long-term central venous catheters, and Bard's Hickman preamendment long-term central venous catheters.

    BMW's Clampless Valved Catheter-Tunneled is designed for patients who require long-term access to the central venous system for administration of fluids including, but not limited to, hydration fluids, antibiotics, chemotherapy, analgesics, nutritional therapy, and blood products. It is also indicated for blood specimen withdrawal. This product is effective for venous access in adults, children, and infants who require intravenous therapy.

    Device Description

    BMW Medical has developed a catheter that provides the benefits of the valve incorporated at the distal tip of the Groshong catheter on an open-ended catheter. The BMW valve is external to the central venous system and is protected inside the rigid catheter adapter.

    BMW Medical, Inc.'s multi-lumen catheters are composed of extruded silicone rubber tubing which is homogeneously mixed with barium sulfate (BaSO,) prior to extrusion to impart radiopacity. The catheters vary in diameter and number of lumens to accommodate specific applications.

    A silicone rubber joint permanently attaches the multi-lumen catheter tubing to the 2 or 3 legs required to access each lumen of a dual or a triple lumen catheter. A 2-piece rigid plastic adapter is attached to the proximal end of each single lumen leg. Each leg is mechanically locked onto the barbed end of the adapter with a silicone rubber sleeve. The other end of the adapter terminates in a female luer which allows for attachment of the catheter to a male luer fitting.

    The adapter houses a slitted disc of silicone rubber which serves as a three-way valve. The valve remains closed when the catheter is not in use. The valve opens toward the distal tip of the catheter when fluids are infused into the patient via the catheter, and it opens toward the female luer of the proximal hub during aspiration of blood samples.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (BMW Medical, Inc.'s Multi-Lumen Clampless Valved Catheter-Tunneled) and does not contain the information requested in your bullet points regarding acceptance criteria and a study proving a device meets them in the context of an AI/ML device.

    The document describes a physical medical device (a catheter) and its comparison to predicate devices, focusing on physical properties, materials, and basic functional performance like flow rate and leak tests. It's a premarket notification for a traditional medical device, not an AI/ML diagnostic or assistive tool.

    Therefore, I cannot extract the following information from the provided text:

    • A table of acceptance criteria and the reported device performance (for AI/ML performance): The text discusses "performance data" for the catheter's valve (reducing air embolism, bleedback, no hemolysis, low aspiration pressures) and physical properties (stiffness, elongation, radiopacity, flow rate, tensile/burst strength). However, these are for a physical device, not an AI algorithm.
    • Sample sized used for the test set and the data provenance: Not applicable.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.
    • If a standalone performance was done: Not applicable (for AI).
    • The type of ground truth used: For the physical device, it's about physical measurements and functional characteristics against established standards or predicate devices, not "ground truth" as understood in AI/ML.
    • The sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes the regulatory submission for a physical medical device (a catheter) and therefore does not contain the information you requested about acceptance criteria and studies related to an AI/ML device's performance.