Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K040579

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-06-18

    (106 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cholestech LDX high sensitivity C-Reactive Protein (hs-CRP) is an in vitro diagnostic test for the quantitative determination of CRP in whole blood or serum. Measurement of CRP is useful as an aid in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.

    Device Description

    The Cholestech LDX System combines immunoassay and solid-phase technology to measure CRP. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin coated capillary tube), venous whole blood or serum. The sample is applied to a Cholestech LDX hs-CRP cassette. The cassette is then placed into the Cholestech LDX Analyzer where a unique system on the cassette separates the plasma from the blood cells. The resultant color in the reaction is measured by reflectance photometry. A brown magnetic stripe on each cassette contains the calibration information required for the Cholestech LDX Analyzer to convert the reflectance reading to the CRP concentration in mg/L.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific MetricAcceptance CriteriaReported Device Performance (Cholestech LDX hs-CRP)
    Accuracy (vs. Predicate)Slope (Serum)Not explicitly stated, implied to be close to 11.01
    Y-intercept (Serum)Not explicitly stated, implied to be close to 00.22
    "r" (Serum)Not explicitly stated, implied to be close to 10.975
    Slope (Whole Blood)Not explicitly stated, implied to be close to 11.06
    Y-intercept (Whole Blood)Not explicitly stated, implied to be close to 00.07
    "r" (Whole Blood)Not explicitly stated, implied to be close to 10.976
    Slope (Fingersticks)Not explicitly stated, implied to be close to 11.08
    Y-intercept (Fingersticks)Not explicitly stated, implied to be close to 0-0.02
    "r" (Fingersticks)Not explicitly stated, implied to be close to 10.981
    Assay RangeRange0.2 to 10 mg/L0.2 to 10 mg/L
    Hematocrit ToleranceToleranceN/A (implied to be within acceptable limits)30-55%
    InterferenceInterference %Less than 10%Less than 10% interference
    Precision%CV (Low Control)N/A (implied clinical acceptability)14.3%
    %CV (High Control)N/A (implied clinical acceptability)11.5%
    %CV (Serum Sample @ 6.5 mg/L)N/A (implied clinical acceptability)11.4%

    Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria for slope, y-intercept, and "r" values. However, for a device claiming substantial equivalence to a predicate, the expectation is that these metrics demonstrate a strong correlation and agreement with the predicate. The reported values of r (correlation coefficient) being close to 1 (0.975, 0.976, 0.981) and slopes being close to 1 (1.01, 1.06, 1.08) with small y-intercepts (-0.02, 0.07, 0.22) indicate good agreement and likely meet the implicit acceptance criteria for substantial equivalence. The assay range and interference criteria are explicitly stated and met. Precision values are reported but no explicit acceptance thresholds are provided.

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:
      • Accuracy (Serum vs. Predicate): 70 matched serum samples
      • Accuracy (Whole Blood vs. Predicate): 76 whole blood samples (both venous and fingerstick)
      • Precision: 2 levels of controls (Low and High) and one serum sample (6.5 mg/L). Each control was tested in duplicate, twice a day, over 20 days (total 80 replicates per level).
      • Hematocrit Tolerance & Interference: No specific sample sizes provided, but tested with "evaluated levels" of endogenous substances.
    • Data Provenance: Not explicitly stated, but clinical studies for such devices typically involve samples from a geographically diverse patient population. The document does not specify country of origin or whether the data was retrospective or prospective. Given the nature of a 510(k) submission for an in vitro diagnostic, it is generally assumed to be prospective clinical study data collected for the purpose of demonstrating device performance.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Number of Experts: Not applicable. This study does not involve expert readers assessing images or clinical cases to establish ground truth.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method (Test Set)

    • Adjudication Method: Not applicable. This study does not involve expert adjudication as it is a quantitative diagnostic test compared to a predicate device. The "ground truth" is established by the predicate device's measurement.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was it done? No. This is not an imaging AI device that relies on human interpretation. It's an in vitro diagnostic test that provides a quantitative measurement.

    6. Standalone (Algorithm Only) Performance Study

    • Was it done? Yes. The entire accuracy study comparing the Cholestech LDX hs-CRP device to the Dade Behring N high sensitivity CRP test is a standalone performance study. The device's output (CRP concentration) is directly compared to the predicate's output. There is no human-in-the-loop component in the measurement itself.

    7. Type of Ground Truth Used

    • Ground Truth Type: Predicate device measurement. The Dade Behring N High Sensitivity CRP test on the BN100 (K991385) is used as the reference standard to establish the "ground truth" for the accuracy study. This is a common approach for demonstrating substantial equivalence for new in vitro diagnostic devices.

    8. Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is an in vitro diagnostic immunoassay, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The device's performance is determined by its inherent chemical and optical properties, and its "calibration information" is encoded on the magnetic stripe of each cassette, not learned from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment (Training Set): Not applicable, as there is no "training set" in the context of this device. The device's calibration is factory-set and encoded on each cassette.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1