Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K201009

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-21

    (460 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CHIRAVAC™ Blood Collection Needles are single use and are intended for venous blood collection in connection with blood collection tubes.

    Device Description

    The CHIRAVAC™ Blood Collection Needles consists of a double-ended hollow stainless steel thin-walled cannula, a threaded polypropylene hub, a sleeve that interrupts blood flow between filling multiple tubes, and polypropylene caps. The device is single use and supplied sterile (ethylene oxide sterilized). The stainless steel cannula and silicone cannula lubricants are classified as externally communicating with limited (≤24 hours) duration contact; all other device components have no contact with patient blood, tissue, or skin. Once the vein has been accessed by puncturing with the intravenous end of the cannula, the healthcare professional will begin the blood collection process by placing evacuated blood collection tubes on the cannula's non-patient end, where the sleeve acts as a non-return valve and allows for more than one tube to be used to collect blood from the vein puncture.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the CHIRAVAC™ Blood Collection Needles. It does not present a clinical study involving human readers or AI algorithms. Instead, it focuses on demonstrating substantial equivalence to a legally marketed predicate device through performance testing (bench testing, biocompatibility, sterilization, shipping, and shelf-life).

    Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, specifically concerning human reader performance, AI assistance, standalone algorithm performance, and sample sizes for training/test sets, cannot be fully provided as it is not present in the document.

    However, I can extract information related to the acceptance criteria for the device's physical and functional characteristics and the bench testing conducted to meet these criteria.

    Here's a summary of the available information:


    1. Table of Acceptance Criteria and Reported Device Performance (Summary of Bench Testing)

    The document lists several international standards (ISO) and internal procedures to which the CHIRAVAC™ Blood Collection Needles were tested. The "reported device performance" is generally implied as "passed" or "met requirements" for each test, as the overall conclusion is that the device is substantially equivalent to the predicate. The document does not provide specific quantitative results for each test (e.g., "Fragmented less than X%").

    Acceptance Criteria (Test Mode)Standard/ProcedureReported Device Performance (Implied)
    Performance Testing - Bench
    Color CodingISO 7864:2016, ISO 6009:2016Met standard requirements
    Size DesignationISO 7864:2016Met standard requirements
    CleanlinessISO 7864:2016, Internal Procedure SOP 4405Met standard requirements
    pH LimitsISO 7864:2016, ISO 9626:2016Met standard requirements
    Limits of Extractable MetalsISO 7864:2016Met standard requirements
    Inspection of Needle LengthISO 7864:2016Met specified lengths
    Inspection of DefectsISO 7864:2016No defects found
    Inspection of LubricationISO 7864:2016Met standard requirements
    Needle/Hub Bond StrengthISO 7864:2016Met standard requirements
    Lumen PatencyISO 7864:2016Patent
    FragmentationISO 7864:2016No fragmentation observed
    Inspection of Needle Sharpness/CleannessISO 7864:2016Met standard requirements
    Penetration Force TestISO 7864:2016Met standard requirements
    Inspection of Surface AppearanceISO 9626:2016Met standard requirements
    Inspection of Tubing DimensionsISO 9626:2016Met specified dimensions
    Stiffness TestISO 9626:2016Met standard requirements
    Breakage Resistance TestISO 9626:2016Met standard requirements
    Corrosion Resistance TestISO 9626:2016No corrosion observed
    Visual Inspection of LabelingInternal Procedure SOP 4113Conformed to labeling requirements
    Inspection of Blister/Label PerforationInternal Procedure SOP 4401No perforations/damage observed
    Vacuum Function EvaluationInternal Procedure SOP 4401Functional
    Test for Rubber Sleeve Returning to Original PositionInternal Procedure SOP 4401Returned to original position, no damage
    Inspection of Cannula TransparencyInternal Procedure SOP 4405Transparent
    Biocompatibility
    CytotoxicityISO 10993-5Non-cytotoxic
    SensitizationISO 10993-10Non-sensitizing
    IrritationISO 10993-10Non-irritating
    Acute Systemic ToxicityISO 10993-11Non-toxic
    HemocompatibilityASTM F756Hemocompatible
    Material Mediated PyrogenicityEuropean Pharmacopoeia 10.0 Article 2.6.8Non-pyrogenic
    Sterilization, Shipping & Shelf-life
    Packaging IntegrityInternal proceduresIntegrity maintained
    Sterile Barrier PackagingISO 7886-1Integrity maintained
    Shelf Life (5 years)ASTM F1980 (validated by accelerated aging)Established 5-year shelf life

    2. Sample Size Used for the Test Set and Data Provenance

    This document describes bench testing of a physical medical device (blood collection needles), not a diagnostic algorithm that relies on a "test set" of data in the common sense (e.g., medical images).

    • Sample Size: The document does not explicitly state the number of samples (e.g., individual needles or batches) used for each specific test. This level of detail is typically found in the full test reports, not the 510(k) summary.
    • Data Provenance: The tests are conducted according to international standards (ISO, ASTM, European Pharmacopoeia) and internal Standard Operating Procedures (SOPs). The testing was performed by CHIRANA T. Injecta or third-party labs on their behalf, based in the Slovak Republic (Submitter's address). This is a prospective evaluation of manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is not a study requiring expert interpretation of data to establish a "ground truth" for a diagnostic task. The ground truth for device performance is established by objective measurements against predefined acceptance criteria from the relevant standards (e.g., physical dimensions, strength, sterility).

    4. Adjudication Method for the Test Set

    Not applicable. No "adjudication" is described as this is not a diagnostic performance study involving multiple human readers or AI.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an MRMC comparative effectiveness study, nor does it involve AI. It's a submission for a physical medical device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This device does not involve an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on:

    • International Standards: e.g., ISO 7864, ISO 6009, ISO 9626, ISO 10993 series, ASTM F756, ASTM F1980, European Pharmacopoeia. These standards define the acceptable physical, chemical, and biological properties.
    • Predicate Device Characteristics: The device is compared to a legally marketed predicate (Sol-M Blood Collection Needles, K182146) to demonstrate substantial equivalence, implying the predicate's performance against relevant standards serves as a benchmark for safety and effectiveness.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device, so there is no training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set or associated ground truth as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1