510(k) Data Aggregation

    K Number
    K983211

    Validate with FDA (Live)

    Device Name
    CENTAURI
    Date Cleared
    1998-10-09

    (25 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The laser is indicated for Incision, Excision, Hemostasis, Ablation of Tissue, and Vaporization of Tissue in the oral cavity. The additional representative indications include Removal of Caries, Cavity Preparation, Modification or Etching of Enamel prior to acidentify, and Modification or Etching of Dentin prior to acid etching.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text to extract the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria & Performance MetricStated Acceptance CriteriaReported Device Performance
    Pulpal Temperature RiseMax 5.5°C rise (to avoid irreversible damage)- 20 mJ: 0.08°C - 50 mJ: 1.07°C - 100 mJ: 0.84°C - 150 mJ: 1.12°C - 200 mJ: 2.3°C - Drill (Air): 0.25°C - Drill (H2O): 0.05°C
    Pulpal Healing (H&E Histology)No deleterious effectNo deleterious effect for laser or control treatment
    Pulp Vitality (Long-term)No compromise in pulp vitality (1.5 year follow-up)No compromise in pulp vitality over 1.5 year follow-up
    Pulp Vitality (Patient-matched, pre/post)No difference in pulpal vitality post-surgery over 3 months compared to pre-surgery/drillNo difference in pulpal vitality pre-surgery, post-surgery, and over 3 months
    Pulp Vitality (Randomized, pre/post)No significant difference in pulp vitality between laser and control after 3 monthsNo significant difference between laser and control in pulp vitality pre-surgery and after a 3 month period
    Surface Morphology (non-treatment site)No changes in surface morphologyNo changes in surface morphology except at the treatment site
    Surface Morphology (treatment site)Equivalent surface changes to drillDrill and laser show equivalent surface changes at the treatment site
    Structural Morphology (Etched Tooth)Roughened dentin/enamel, no cracking/fissuring/charring, open dentinal tubules, vaporized organic material (leaving inorganic components), no remaining caries, crater below marginEffective removal of tooth structure, no fissures/cracks, scaled/roughened surface without thermal damage. At 430 mJ, hydroxyapatite vaporized, pores developed, no visible cracks.
    Adjacent Structures DamageDamage less than drill when defocusedDamage due to inadvertent lasing of adjacent structures usually less than the drill, as laser does not cut effectively when defocused.
    Caries Removal EffectivenessMore effective than the drillInvestigators rate caries removal more effective than the drill
    Cavity Preparation EffectivenessEqual to the drillInvestigators rate cavity preparation with the laser equal to the drill
    Laser Etching EffectivenessMuch more effective than acid etchingInvestigators rate laser etching much more effective than acid etching
    Overall Equivalence to Standard TreatmentEquivalent to standard treatmentAnimal and human studies demonstrate that the Er:YAG laser is equivalent to standard treatment
    Caries Removal Completion RateNot explicitly stated but 100% is implied for "completely removed"100% of teeth treated had all classes of caries (I-V) completely removed
    Cavity Preparation & Restoration AdequacyNot explicitly stated but 100% after 3 months is implied100% of treated teeth had adequate cavity preparation and restoration after 3 months
    Pediatric Use Complications (Primary Teeth)No reported complications- Phase I (51 procedures): 12% on primary teeth, no complications. - Phase II (130 procedures): 15% on primary teeth, no complications. - Additional data (219 procedures): 22% on primary teeth, no complications. - Total (400 procedures): 19% on primary teeth, no complications.
    Pediatric Use Anesthesia RequestNo anesthesia requested by patientNo anesthesia requested by patient in any pediatric primary tooth procedures (Phase I). (Implied for others as well as "no reported complications")

    2. Sample Size Used for the Test Set and Data Provenance

    • Pulpal Temperature Study:

      • Sample Size: Not explicitly stated, but the data is from "teeth in dogs" (Zach and Cohen) and other substantiating studies (Powell, et. Al.). The presented table shows experimental conditions, not a specific number of teeth in a test set. This appears to be a benchmark/foundational study.
      • Data Provenance: Implied to be retrospective from existing literature (Zach and Cohen, Powell, et. Al.) for the benchmark, and experimental for the device's specific temperature measurements. Country of origin not specified.
    • Pulp Vitality Studies:

      • Study 1 (Histological): "H&E histological evidence over pulpal healing time". Sample size not specified. Data provenance: "Animal and human studies" in surface morphology suggests this could be both, but not explicitly stated here.
      • Study 2 (1.5 year follow-up): Sample size not specified.
      • Study 3 (Patient-matched): 33 patients, two teeth each (total 66 teeth). Data provenance: Not specified (but likely prospective clinical study).
      • Study 4 (Randomized): 125 randomized adult teeth. Data provenance: Not specified (but likely prospective clinical study).
    • Surface Morphology Studies:

      • Sample Size: Not specified for animal or human studies.
      • Data Provenance: "Animal and human studies" (likely retrospective or previously published experimental for establishing general findings, then performed with the device).
    • Structural Morphology Studies:

      • Sample Size: Not specified for Hibst and Keller or Paghdiwala. Laser dosimetries ranged from 50-350 mJ (Hibst and Keller) and 430 mJ (Paghdiwala).
      • Data Provenance: Reported by "Hibst and Keller" and "Paghdiwala," suggesting these are from published literature used as reference points, or studies performed by them.
    • Caries Removal / Cavity Prep / Equivalence Studies:

      • Study 1,2,3 (Investigator Ratings): Sample size of investigators not specified.
      • Study 4 (Overall Equivalence): Sample size not specified. Data provenance: "Animal and human studies."
      • Study 5 (Caries Removal Rate): 125 randomized human adult teeth. Data provenance: "Multi-site study," implies prospective clinical data.
      • Study 6 (Cavity Prep & Restoration Adequacy): 125 randomized human adult teeth. Data provenance: "Multi-site study," implies a prospective clinical data.
    • Pediatric Use Studies (Phase I, II, Additional Patient Data):

      • Phase I: 51 laser procedures (6 on primary teeth).
      • Phase II: 130 laser procedures (18 on primary teeth).
      • Additional Patient Data: 219 laser procedures (48 on primary teeth).
      • Total: 400 laser procedures (68 on primary teeth).
      • Data Provenance: Clinical data described as "Phase I," "Phase II," and "additional patient data," suggesting prospective clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Pulp Vitality (Histological): Not specified.
    • Pulp Vitality (Clinical): "Pulp vitality measurements" (Studies 3 & 4). The method of establishing ground truth (e.g., specific tests like electric pulp testing, thermal tests) is implied by "measurements," but the number and qualifications of experts interpreting these are not specified.
    • Surface Morphology: "Animal and human studies using SEM". Ground truth likely established by expert interpretation of SEM images. Number and qualifications of experts not specified.
    • Structural Morphology: "Hibst and Keller reported," "Paghdiwala showed." These are likely based on expert interpretation of micrographs. Number and qualifications not specified for these external references.
    • Caries Removal / Cavity Prep / Etching Ratings: "Investigators rate." Number and qualifications of these "investigators" are not specified.
    • Caries Removal Completion (125 teeth study): "completely removed in 100%". The method for determining "completely removed" (e.g., visual inspection, excavation to hard dentin, radiographs, pathological examination) and the number/qualifications of experts making this determination are not specified.
    • Cavity Preparation & Restoration Adequacy (125 teeth study): "still adequate after 3 months in 100%". The criteria for "adequate" and the number/qualifications of experts making this determination are not specified.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). The "all blinded" statements for Pulp Vitality and Surface Morphology suggest efforts to reduce bias, but not a specific consensus or adjudication process. For investigator ratings, it's unclear if there was a consensus process or if individual investigator opinions were aggregated.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an AI component in this device. Therefore, no MRMC study involving AI assistance for human readers was conducted or reported.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is a physical laser system, not an algorithm. Therefore, no standalone algorithm performance study was reported.

    7. The Type of Ground Truth Used

    • Pulpal Temperature: Objective thermal measurements (°C rise) compared against a physiological benchmark (irreversible damage threshold).
    • Pulp Vitality:
      • H&E histological evidence (pathology).
      • Pulp vitality measurements (clinical tests, likely electric, thermal, or visual assessment of health status).
    • Surface Morphology: SEM (Scanning Electron Microscopy) observations and interpretations.
    • Structural Morphology: Detailed visual/microscopic observations (etched surface characteristics, cracks, charring, open tubules) and reported findings from scientific literature (Hibst, Keller, Paghdiwala).
    • Caries Removal / Cavity Prep / Etching Efficacy: "Investigators rate" (expert opinion/clinical judgment).
    • Caries Removal Completion / Cavity Prep & Restoration Adequacy (125 teeth study): Clinical assessment of presence/absence of caries and adequacy of preparation/restoration. The specific methodology for these assessments and the experts performing them are not detailed beyond "completely removed" and "adequate."
    • Pediatric Complications: Clinical observation/reporting of complications by healthcare professionals.

    8. The Sample Size for the Training Set

    The concept of a "training set" is not applicable as this is a physical medical device, not an AI/machine learning model. The various studies and experiments described serve as validation/testing.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable (see point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K933841

    Validate with FDA (Live)

    Device Name
    CENTAURI
    Date Cleared
    1997-05-05

    (1370 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The laser is indicated for Incision, Excision, Hemostasis, Ablation of Tissue, and Vaporization of Tissue in the oral cavity. The additional representative indications include Removal of Caries, Cavity Preparation, Modification or Etching of Enamel prior to acid etching, and Modification or Etching of Dentin prior to acid etching.

    Device Description

    Centauri Er:YAG Laser System

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and the study data that demonstrate the device meets those criteria:

    Device Name: Centauri Er:YAG Laser System

    Indications For Use: Incision, Excision, Hemostasis, Ablation of Tissue, and Vaporization of Tissue in the oral cavity. Additional representative indications include Removal of Caries, Cavity Preparation, Modification or Etching of Enamel prior to acid etching, and Modification or Etching of Dentin prior to acid etching.

    Contraindications: Children under the age of 18 years.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for each performance metric with numerical thresholds. Instead, it presents various performance measures and concludes that the device is "equivalent to standard treatment" or "more effective" than drilling/acid etching for certain outcomes. Therefore, I will infer the acceptance criteria from the reported findings, primarily focusing on equivalence to established methods or a lack of deleterious effects.

    Acceptance Criteria (Inferred)Reported Device Performance
    Pulpal Temperature Rise: Max pulpal temperature rise of ≤ 5.5°C (to avoid irreversible damage).Observed Pulpal Temperature Rise: - 20 mJ, 10 Hz, 2 sec H2O cooling: 0.08°C - 50 mJ, 10 Hz, 2 sec H2O cooling: 1.07°C - 100 mJ, 10 Hz, 2 sec H2O cooling: 0.84°C - 150 mJ, 10 Hz, 2 sec H2O cooling: 1.12°C - 200 mJ, 10 Hz, 2 sec H2O cooling: 2.3°C Conclusion: All observed temperature rises are significantly below the 5.5°C threshold.
    Pulp Vitality: No deleterious effect on pulp vitality; equivalent to control/drill methods over time.Observed Pulp Vitality: - H&E histological evidence over pulpal healing time: No deleterious effect for laser or control treatment. - Pulp vitality measurements over 1.5 year follow-up: No compromise in pulp vitality. - In 33 patients, two teeth (laser vs. drill) showed no difference in pulpal vitality pre-surgery, post-surgery, and over three months. - In 125 randomly treated adult teeth, no significant difference in pulp vitality pre-surgery and after three months compared to drill.
    Surface Morphology: No changes in surface morphology except at the treatment site; equivalent to drill at treatment site.Observed Surface Morphology: - Animal and human studies using SEM: No changes in surface morphology except at the treatment site. - Animal and human studies using SEM: Drill and laser show equivalent surface changes at the treatment site.
    Structural Morphology: Roughened dentin/enamel, no cracking, fissuring, or charring; open dentinal tubules; no remaining caries.Observed Structural Morphology: - Ideal etched tooth (description given): "roughened dentin or enamel surface and no evidence of cracking, fissuring or charring. The dentin demonstrates open dentinal tubules. Organic material has been vaporized leaving the inorganic components of the tooth - leaving greater tooth surface area. In addition, a cavity preparation should show no remaining evidence of caries and a crater created by removal of tooth structure below the margin of the preparation." - Hibst and Keller: Effective removal of tooth structure with ultrastructural changes; no fissures or cracks; surface scaled and roughened without signs of thermal damage. - Paghdiwala: No visible cracks radiated from craters.
    Adjacent Structures: Damage due to inadvertent lasing of adjacent structures is less than or equivalent to the drill.Observed Adjacent Structures Effects: - Damage due to inadvertent lasing of adjacent structures is usually less than the drill, as the laser does not cut effectively when defocused.
    Efficacy (Caries Removal): Effective removal of caries.Observed Efficacy (Caries Removal): - Investigators rate caries removal more effective than the drill. - In a multi-site study of 125 randomized human adult teeth, 100% of Class I-V caries were completely removed.
    Efficacy (Cavity Preparation): Equivalent to drill.Observed Efficacy (Cavity Preparation): - Investigators rate cavity preparation with the laser equal to the drill. - In a multi-site study of 125 randomized human adult teeth, 100% of cavity preparations and restorations were still adequate after 3 months.
    Efficacy (Etching): More effective than acid etching.Observed Efficacy (Etching): - Investigators rate laser etching much more effective than acid etching.
    Overall Equivalence: Equivalent to standard treatment.Observed Overall Equivalence: - Animal and human studies demonstrate that the Er: YAG laser is equivalent to standard treatment.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Pulpal Temperature: The table (TABLE 1) shows measurements at various energy outputs and repetition rates for "Time with H2O Cooling". No explicit sample size for this specific test is provided beyond listing the parameters. A control "Drill in Air" and "Drill in H2O" are also mentioned, suggesting a comparison study. The study references Zach and Cohen (1965) and Powell et al. (1990) for the temperature threshold, implying the context for these measurements is based on established dental research, likely involving animal models initially.
    • Pulp Vitality:
      • 33 patients (comparison of two teeth per patient: laser vs. drill).
      • 125 randomly treated adult teeth (comparison of laser vs. drill, source not explicitly stated but implies human).
      • One and a half-year follow-up for general pulp vitality measurements (sample size not specified, but context suggests human patients).
    • Surface Morphology: Animal and human studies (specific sample sizes not provided).
    • Efficacy:
      • 125 randomized human adult teeth for caries removal.
      • 125 randomized human adult teeth for cavity preparation and restoration adequacy.

    Data Provenance: The document references investigations from various researchers (e.g., Zach and Cohen, Powell, Hibst and Keller, Paghdiwala, Arcoria, Kumazaki). This suggests a compilation of findings from multiple studies, both animal and human. The "multi-site study" for efficacy suggests data from different clinical locations. The general impression is that this is a retrospective summary of existing research and possibly some newly compiled data to support the 510(k) submission. No specific country of origin is mentioned for the sample data beyond the general context of medical device regulatory submissions in the USA.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document directly mentions "Investigators" rating efficacy for caries removal, cavity preparation, and laser etching. It also mentions "blinded" assessments for Pulp Vitality and Surface Morphology.

    • Pulpal Temperature: The ground truth for the acceptable temperature rise (5.5°C) is established by publications from Zach and Cohen (1965) and Powell et al. (1990). These would be established dental researchers or pathologists.
    • Pulp Vitality: "H&E histological evidence" implies assessment by pathologists. "Pulp vitality measurements" would likely be performed and interpreted by dental professionals. The term "blinded" suggests that the assessors were unaware of whether the treatment was laser or control. The number of experts is not specified.
    • Surface Morphology: "SEM demonstrated" implies interpretation by experts trained in electron microscopy, likely with dental or material science backgrounds. The term "blinded" is used. The number of experts is not specified.
    • Structural Morphology: This section relies on findings reported by Hibst and Keller (dental researchers) and Paghdiwala (who specifically researched the Er:YAG laser in hard dental tissues). Their qualifications would be in dentistry, laser physics, and material science as applied to dentistry.
    • Efficacy: "Investigators rate..." suggests multiple dental professionals or clinicians were involved in evaluating the outcomes of caries removal, cavity preparation, and etching effectiveness. Their specific qualifications (e.g., years of experience as a dentist or oral surgeon) are not detailed, nor is their exact number. Given the context of a 510(k), these "investigators" would typically be qualified dentists or oral surgeons.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). For the "blinded" evaluations (Pulp Vitality, Surface Morphology), it implies individual expert assessments, but doesn't describe a formal consensus or tie-breaking process. For the "Investigators rate..." efficacy statements, it suggests a collective judgment rather than individual adjudication results for each case. The "multi-site study" implies that data might have been collected and analyzed collectively rather than through a strict adjudication panel for each individual finding reported.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not describe a formal MRMC comparative effectiveness study in the way this term is typically used for AI-assisted image interpretation studies. However, it does present data that can be considered a comparative effectiveness assessment:

    • Comparison to Drill: Multiple sections compare the laser's performance to the conventional drill for pulpal temperature, pulp vitality, surface morphology, and efficacy in caries removal (laser more effective), cavity preparation (laser equal), and adjacent tissue damage (laser usually less).
    • Comparison to Acid Etching: The laser is rated "much more effective" than acid etching.
    • "Investigators rate..." and "Animal and human studies demonstrate..." suggest a collective assessment over multiple cases by multiple readers (investigators/experts) comparing the laser to standard methods.

    Effect Size: The document provides qualitative rather than quantitative effect sizes for "human readers improve with AI vs without AI assistance." Instead, it focuses on the device's standalone performance or direct comparison to current standards:

    • Caries removal: laser "more effective" than drill.
    • Cavity preparation: laser "equal" to drill.
    • Laser etching: "much more effective" than acid etching.
    • Adjacent structures: damage "usually less" than drill when defocused.

    There is no mention of human-in-the-loop performance augmentation by the device. The device is a direct treatment tool, not an AI diagnostic aid for human readers.


    6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study

    Yes, the studies described are essentially standalone performance studies of the Centauri Er:YAG Laser System as a treatment device. The device performs the action (caries removal, cavity preparation, etching, etc.) directly. The "efficacy" and "safety" evaluations are assessments of the laser's direct effects, not its assistance to a human operator in a diagnostic or interpretive task. The quantitative measurements (e.g., pulpal temperature rise) and qualitative observations (e.g., histological evidence, SEM changes) are direct outputs of the laser's application.


    7. Type of Ground Truth Used

    The ground truth used is a combination of:

    • Expert Consensus/Clinical Judgment: For efficacy ratings (caries removal, cavity preparation, etching), "Investigators rate..." suggests expert consensus or clinical judgment.
    • Histopathology: "H&E histological evidence" for pulpal healing.
    • Physiological Measurements: "Pulp vitality measurements" (likely using cold/electric pulp tests or similar clinical methods). "Temperature in °C" for pulpal temperature rise.
    • Imaging/Microscopy: "SEM demonstrated" for surface morphology.
    • Established Scientific Literature: The 5.5°C pulpal temperature threshold is derived from benchmark publications (Zach and Cohen, Powell et al.). Structural morphology descriptions also rely on existing research findings.
    • Outcome Data: "100% of Class I-V caries were completely removed" and "100% of cavity preparation and restoration was still adequate after 3 months" are direct outcome measures.

    8. Sample Size for the Training Set

    The document does not describe a "training set" in the context of an algorithm or AI system. This 510(k) submission is for a physical medical device (laser system) and describes its performance validation, not the development and training of an AI algorithm. Therefore, this question is not applicable to the provided document. The studies referenced are for validating the device's safety and efficacy, not for training a model.


    9. How the Ground Truth for the Training Set Was Established

    As there is no "training set" described for an algorithm, this question is not applicable. The information provided relates to the validation of a physical laser device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K932683

    Validate with FDA (Live)

    Device Name
    CENTAURI
    Date Cleared
    1997-05-05

    (1432 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The laser is indicated for Incision, Excision, Hemostasis, Ablation of Tissue, and Vaporization of Tissue in the oral cavity. The additional representative indications include Removal of Caries, Cavity Preparation, Modification or Etching of Enamel prior to acid etching, and Modification or Etching of Dentin prior to acid etching.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification for the Centauri Er:YAG Laser System, seeking FDA clearance for marketing. While it contains performance data and claims regarding the device's efficacy, it does not outline specific, pre-defined acceptance criteria for regulatory approval in the way modern AI device submissions do. Instead, the document presents study results to demonstrate substantial equivalence to legally marketed predicate devices.

    Therefore, I cannot provide a table of "acceptance criteria" and "reported device performance" in the strict sense requested, because explicit, quantitative acceptance criteria for this type of medical device in 1997 are not detailed in this document. The document describes studies aiming to show the device performs safely and effectively for its intended use, comparable to existing methods.

    However, I can extract the reported device performance and infer the implicit "acceptance criteria" based on the presented results and the claims made for "substantial equivalence."

    Here's an analysis of the provided information, framed as closely as possible to your request:


    Analysis of Centauri Er:YAG Laser System Performance and Supporting Studies

    1. Table of (Inferred) Acceptance Criteria and Reported Device Performance:

    Feature/Metric(Inferred) Acceptance CriteriaReported Device Performance
    Pulpal Temperature Increase≤ 5.5 °C (to avoid irreversible damage)- 20 mJ: 0.08°C - 50 mJ: 1.07°C - 100 mJ: 0.84°C - 150 mJ: 1.12°C - 200 mJ: 2.3°C - Drill in Air: 0.25°C - Drill in H2O: 0.05°C (All well below the 5.5°C threshold)
    Pulpal Vitality (Long-term)No deleterious effect / no compromise in pulp vitality / no significant difference compared to control (drill)- H&E histological evidence showed no deleterious effect for laser or control treatment. - Pulp vitality measurements over 1.5 year follow-up showed no compromise. - No difference in pulp vitality pre-surgery, post-surgery, and over 3 months in 33 patients (laser vs. drill in same patient). - No significant difference between laser and control over 3 months in 125 randomized teeth.
    Caries Removal EffectivenessComplete removal of all classes of caries100% of caries (all classes I-V) completely removed in 125 randomized human adult teeth.
    Cavity Preparation & RestorationAdequate preparation and restoration after 3 months100% of cavity preparations and restorations still adequate after 3 months in 125 randomized human adult teeth.
    Surface Morphology (SEM)No changes except at the treatment site; equivalent to drill at treatment site; no fissures/cracks, no thermal damage- Animal and human studies demonstrated no changes in surface morphology except at the treatment site. - Equivalent surface changes to drill at treatment site. - Ideal etched tooth presented roughened dentin/enamel, no cracking/fissuring/charring, open dentinal tubules, vaporized organic material. - Hibst and Keller reported no fissures/cracks, scaled/roughened surface without thermal damage.
    Adjacent Structures DamageLess damage than drill when inadvertent lasing occursDamage due to inadvertent lasing of adjacent structures is "usually less than the drill," as the laser does not cut effectively when defocused.
    Overall Efficacy (Expert Opinion)More effective than drill for caries removal; equal to drill for cavity preparation; more effective than acid etching for etching; equivalent to standard treatment overall.- Investigators rated caries removal more effective than the drill. - Investigators rated cavity preparation with the laser equal to the drill. - Investigators rated laser etching much more effective than acid etching. - Animal and human studies demonstrated equivalence to standard treatment.

    2. Sample Sizes and Data Provenance:

    • Pulpal Temperature Study: Not explicitly stated as a "test set" with human or animal subjects, but rather an experimental measurement of the device's physical effect. The data provenance is not specified beyond being "a benchmark publication" and subsequent substantiation.
    • Pulp Vitality Studies:
      • Study 1 (H&E): Not specified.
      • Study 2 (1.5 yr follow-up): Not specified.
      • Study 3 (33 patients): 33 human patients ("two teeth in each of 33 patients"). Prospective. Country of origin not specified.
      • Study 4 (125 teeth): 125 randomized human adult teeth. Prospective. Country of origin not specified.
    • Caries Removal & Cavity Preparation Studies: 125 randomized human adult teeth for each study. Multi-site study. Prospective. Country of origin not specified.
    • Surface Morphology (SEM) Studies: "Animal and human studies" - sample sizes not specified.
    • Structural Morphology Studies: Refers to published literature by Hibst and Keller, and Paghdiwala. These are not explicitly device-specific studies for this submission but rather foundational science.
    • Efficacy (Investigator Ratings): "Investigators" - number not specified.

    3. Number of Experts and Qualifications for Ground Truth for Test Set:

    The studies described do not involve a "test set" of images for diagnostic purposes requiring expert consensus in the way AI/ML device submissions typically do. Instead, they refer to clinical outcomes, histological analysis, and direct measurements.

    • Pulp Vitality (H&E histological evidence): Implies a pathologist's evaluation. Qualifications not specified.
    • Pulp Vitality (measurements): Objective clinical measurements.
    • Caries Removal & Cavity Prep: Implies clinical assessment by dental professionals (investigators). Qualifications not specified.
    • Surface Morphology (SEM): Interpretation by experts in electron microscopy. Qualifications not specified.
    • Efficacy (Investigator ratings): The "investigators" are implicitly dental/medical professionals performing the procedures and making subjective assessments. Qualifications not specified.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this device submission. The studies involve direct clinical outcomes, measurements, or expert assessments rather than a diagnostic adjudication process. "All blinded" is mentioned for pulp vitality and surface morphology studies, indicating blinding of outcome assessors to treatment arm, which is a form of bias mitigation, but not an adjudication method for ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This is a laser system for dental procedures, not an AI diagnostic device. There is no mention of human readers or AI assistance in the context of diagnostic interpretation. The comparisons are between the laser and a traditional drill.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. This is a physical laser device, not an algorithm. Its performance is evaluated by its physical effects and clinical outcomes, not as a standalone software algorithm.

    7. Type of Ground Truth Used:

    • Pulpal Temperature: Direct physical measurement of temperature increase in °C.
    • Pulpal Vitality:
      • Histological evidence (pathology).
      • Objective clinical measurements of pulp vitality.
    • Caries Removal: Clinical assessment of complete removal by treating dentists/investigators.
    • Cavity Preparation & Restoration: Clinical assessment of adequacy by treating dentists/investigators.
    • Surface Morphology: SEM imaging and expert interpretation (morphological assessment).
    • Efficacy: Subjective ratings by clinical investigators.

    8. Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device requiring a training set.

    9. How Ground Truth for the Training Set Was Established:

    Not applicable.


    Summary Interpretation:

    This 510(k) submission for the Centauri Er:YAG Laser System provides data from a series of studies (some device-specific, some referencing existing literature) to demonstrate that its performance aligns with established safety thresholds (e.g., pulpal temperature increase) and clinical effectiveness when compared to the traditional dental drill for carving/ablation. The "acceptance criteria" are implied by the safe and effective performance demonstrated, particularly the lack of harm and comparable or superior efficacy to existing methods.