Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K213262

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-07

    (99 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ HP Chemical Indicator is a vaporized hydrogen peroxide multivariable chemical indicator. It is designed for routine monitoring of the following cycles:

    · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    • Standard, Advanced, Express, Flex or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    Device Description

    The Celerity™ HP Chemical Indicator is a chemical indicator strip consisting of indicator ink containing the reactive chemicals printed on one end of a polypropylene strip. If the critical variables are achieved, the color of the indicator ink changes from magenta to orange/yellow or lighter yellow when exposed to the V-PRO® Low Temperature Sterilization System cycles or ASP STERRAD® System cycles. The indicator is validated to function as a multiple variable indicator with increased resistance characteristics similar to ISO 11140-1:2014 end points for a Type 4 Vaporized Hydrogen Peroxide (VHP) Chemical Indicator (CI). No changes have been made to the device other than additional testing and updating the labeling for multivariable indicator claims.

    AI/ML Overview

    The provided text describes the acceptance criteria and study results for the CELERITY HP Chemical Indicator, a medical device used to monitor sterilization cycles. Here's a breakdown of the information requested:

    Acceptance Criteria and Device Performance

    1. Table of acceptance criteria and the reported device performance:

    TestingPurposeAcceptance CriteriaStudy Result
    Multivariable CI Performance TestingDemonstrates that the indicator reacts to all critical variables (time, temperature, and sterilant concentration) of the sterilization process.Pass testing per Table 8 of ISO 11140-1:2014 (complete color change)Pass. All CI show a complete color change.
    Fail testing per Table 8 of ISO 11140-1:2014 (incomplete color change)Pass. All CI show an incomplete color change.
    Endpoint color stabilityDemonstrates that the processed indicator color does not significantly change over the labeled time period.Color remains stablePass. Color stable up to 15 months.
    Simulated Use Testing (Leveraged from K192020)Demonstrates that the indicator performs as expected in a commercial sterilizer with worst-case load.Complete color change in a full cycle. Incomplete color change in an abbreviated cycle.Pass. Complete color change in a full cycle. Incomplete color change in an abbreviated cycle.

    Regarding the other information, the provided text does NOT contain details about the following aspects, which are typically relevant for AI/ML-based medical devices or studies involving human readers:

    2. Sample size used for the test set and the data provenance: This information is not explicitly stated. The study refers to "All CI" for the multivariable performance, implying a batch of Chemical Indicators were tested. The provenance (country of origin, retrospective/prospective) is also not detailed, though the manufacturing facility is in the United Kingdom.

    3. Number of experts used to establish the ground truth for the test set and their qualifications: This device is a chemical indicator that shows a color change, not an AI/ML system requiring expert interpretation of images. Therefore, expert consensus on ground truth in that context is not applicable. The "ground truth" here is the physical reaction of the chemical indicator to specific sterilization parameters as defined by ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for a chemical indicator.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical chemical indicator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is a physical chemical indicator.

    7. The type of ground truth used: The ground truth for this device's performance is established by ISO 11140-1:2014 standards for chemical indicators and verified by subjecting the indicators to controlled sterilization conditions with specified time, temperature, and sterilant concentration. The "complete color change" and "incomplete color change" are the physical manifestations defining the ground truth.

    8. The sample size for the training set: Not applicable. This is an engineered chemical product, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192020

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-05

    (38 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ HP Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when placed within packs to be sterilized, through a visible change from red to orange/yellow, when the device has been exposed to the:

    · Lumen. Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    The Celerity™ Vaporized VH2O2 Process Indicator Adhesive Label is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed units, when affixed to packs to be sterilized, through a visible change from magenta to orange/yellow or lighter, when the pack has been exposed to the:

    · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including STERRAD NX and 100NX with ALLClear Technology.

    The VERIFY V24 Self-Contained Biological Indicator Vial Label is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed SCBIs, whilst affixed to SCBI vials to be sterilized, through a visible change from red to orange/yellow, when the SCBI has been exposed to the:

    · Lumen. Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or

    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including STERRAD NX and 100NX with ALLClear Technology.

    Device Description

    The Celerity HP Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units through a visible change from magenta to orange/yellow or lighter, when the device has been exposed to the:

    • . Lumen, Non-Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    • . Standard. Advanced, Express, Flex or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    The Celerity "10 HP Chemical indicator is provided in three formats:

    • Version 1: Celerity™ Chemical Indicator (CI) .
    • . Version 2: Celerity™ Vaporized |VH2O2 Process Indicator Adhesive Label (PI)
    • Version 3: VERIFY V24 Self-Contained Biological Indicator Vial Label (PI)

    The three formats differ in their intended location inside the load to be sterilized. Version 1 is to be placed inside a pack to be sterilized. Version 2 is to be affixed to the outside of a pack by means of the adhesive back. Version 3 is to be affixed to the outside of a Self-Contained Biological Indicator by means of the adhesive back.

    The proposed devices are manufactured by application of the indicator ink by printing onto a substrate in a variety of shapes in three constructions:

    • Construction C Polypropylene ●
    • Construction B - Spun-bonded polyolefin with an adhesive and a glassine backing paper
    • Construction D - polypropylene with an adhesive and a glassine backing paper.

    The proposed devices will consist of one of the above substrates with an indicator ink printed thereon.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Testing CategoryMethodologyAcceptance CriteriaStudy Result
    Type 1 Performance TestingANSI/AAMI/ISO 11140-1:2014- When exposed to VHP Resistometer cycles at the Fail parameters (Test Points 2 and 3): 100% Celerity™ HP Chemical Indicator stripes must show a FAIL result.- When exposed to VHP Resistometer cycles at the Pass condition (Test Point 1): Minimum 90% Celerity™ HP Chemical Indicator stripes must show a PASS result.Pass
    Simulated Use TestingInternal Method Used- When exposed to V-PRO maX Non Lumen 1/84 cycles with ≥50 lb load (Fail condition): 100% Celerity™ HP Chemical Indicator stripes must show a FAIL result.- When exposed to V-PRO and STERRAD cycles with the worst case loads: Minimum 90% Celerity™ HP Chemical Indicator stripes must show a PASS result.Pass
    Fluorescent Light Stability TestingInternal Method & ANSI/AAMI/ISO 11140-1:2014- Indicator start color to remain unchanged after exposure to fluorescent lighting for ≥ 30 days before processing.- When exposed to VHP Resistometer cycles at the Fail parameters following exposure to fluorescent lighting for ≥ 30 days: 100% of Indicators tested must show a FAIL results (non-visible color change or color change to another color except orange/yellow).- When exposed to VHP Resistometer cycles at the PASS parameters following exposure to fluorescent lighting for ≥ 30 days: Minimum 90% Indicators tested must show a PASS result (orange/yellow or lighter).Pass
    Temperature Extremes Exposure (Freeze/Thaw) TestingInternal Method & ANSI/AAMI/ISO 11140-1:2014- Indicator start color to remain unchanged after exposure to extreme temperatures before processing.- When exposed to VHP Resistometer cycles at the Fail parameters following exposure to extreme temperatures: 100% of Indicators tested must show a FAIL results (non-visible color change or color change to another color except orange/yellow).- When exposed to VHP Resistometer cycles at the PASS parameters following exposure to extreme temperatures: Minimum 90% Indicators tested must show a PASS result (orange/yellow or lighter).Pass
    Transference TestingANSI/AAMI/ISO 11140-1:2014- Indicators placed between a piece of the same substrate and held together with tape, then processed in a double V-PRO maX 2 Lumen cycle (no load) and evaluated for off-set, bleed and migration.Pass
    - Samples with reference and text ink processed in a double V-PRO maX 2 Lumen cycle (no load) and evaluated for smearing, discoloration and fading.Pass
    Adhesion Stability (Vials) TestingInternal Method Used- When exposed to two consecutive V-PRO maX 2 Lumen cycles: • 0% samples show bleeding of the adhesive• 100% must be firmly affixed• 0% must show signs of peeling• Minimum 90% must show a PASS resultPass
    Shelf Life TestingANSI/AAMI/ISO 11140-1:2014- After the process indicator has been aged: Minimum of 90% of indicators must PASS Type 1 PASS conditions.- After the process indicator has been aged: 100% of indicators must FAIL.Pass
    Post-Processing Stability TestingInternal Method Used- Indicator processed ink color remains unchanged for a minimum of 6 months following exposure to a PASS or FAIL cycle.Pass

    2. Sample Size and Data Provenance for Test Set:

    The document does not explicitly state the specific sample sizes (number of indicators) used for each individual test in the performance studies. However, for "Type 1 Performance Testing", "Simulated Use Testing", "Fluorescent Light Stability Testing", "Temperature Extremes Exposure Testing", "Shelf Life Testing", and "Adhesion Stability (Vials) Testing", the acceptance criteria mention percentages (e.g., "100% Celerity™ HP Chemical Indicator stripes", "Minimum 90% Celerity™ HP Chemical Indicator stripes"). This implies that multiple indicators were tested for each condition.

    The data provenance is not explicitly stated in terms of country of origin. The manufacturing facility is in the United Kingdom, but the testing may have occurred elsewhere. The studies are described in the context of a 510(k) submission, which typically involves prospective testing designed to meet regulatory requirements.

    3. Number of Experts and Qualifications for Ground Truth of Test Set:

    This information is not provided. The performance testing for these chemical indicators relies on objective color changes and adherence to established standards (ANSI/AAMI/ISO 11140-1:2014), rather than expert interpretation of the results. The 'ground truth' is the objective physical change of the chemical indicator based on exposure to sterilization conditions.

    4. Adjudication Method for the Test Set:

    Not applicable. The performance testing for chemical indicators is based on objective observable color changes and compliance with quantitative standards, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices where human readers interpret images or other data, and the comparison is how AI assistance affects their diagnostic accuracy. Chemical indicators, like the Celerity™ HP, are designed for direct visual interpretation of a color change, not complex interpretation by multiple human readers.

    6. Standalone (Algorithm Only) Performance Study:

    Yes, the studies described are standalone performance studies of the chemical indicator itself. The device is a "Type 1 vaporized hydrogen peroxide sterilization process indicator" that provides a direct visual cue (color change) to indicate exposure to sterilization conditions. Its performance is evaluated based on its intrinsic ability to change color accurately under specific conditions, without human interpretation beyond basic color recognition.

    7. Type of Ground Truth Used:

    The ground truth used is primarily based on physical exposure to defined sterilization conditions (VHP Resistometer cycles at specific "Fail parameters" and "Pass condition", and various V-PRO and STERRAD cycles with worst-case loads). The observable color change (or lack thereof) in the chemical indicator is then compared against these known exposure conditions to determine if the indicator provides an accurate visual signal according to the standard.

    8. Sample Size for Training Set:

    Not applicable. Chemical indicators are passive devices that rely on a chemical reaction for their function, not on machine learning algorithms that require a "training set" of data.

    9. How Ground Truth for Training Set was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183295

    Validate with FDA (Live)

    Date Cleared
    2019-01-04

    (38 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ HP Chemical Indicator (CI) is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when placed within packs to be sterilized, through a visible change from red to orange/yellow, when the device has been exposed to the:
    · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    · Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including those systems with ALLClear Technology.

    A version of the Celerity™ HP Chemical Indicator (CI) may be utilized on the V24 Self-Contained Biological Indicator Vial Label and is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed SCBIs, whilst affixed to SCBI vials to be sterilized, through a visible change from red to orange/yellow, when the SCBI has been exposed to the:
    · Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP STERRAD® System, including STERRAD NX and 100NX with ALLClear Technology.

    Device Description

    The Celerity™ HP Chemical Indicator (CI) is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when placed within packs to be sterilized, through a visible change from red to orange/yellow, when the device has been exposed to the:

    • . Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle of a V-PRO® Low Temperature Sterilization System, or
    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP ● STERRAD® System, including those systems with ALLClear Technology.

    A version of the Celerity™ HP Chemical Indicator may be utilized on the V24 Self-Contained Biological Indicator Vial Label and is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed SCBIs, whilst affixed to SCBI vials to be sterilized, through a visible change from red to orange/yellow, when the SCBI has been exposed to the:

    • Lumen, Non Lumen, Flexible, Fast Non Lumen or Fast sterilization cycle . of a V-PRO® Low Temperature Sterilization System, or
    • Standard, Advanced, Express, Flex Scope or Duo cycles of an ASP ● STERRAD® System, including those systems with ALLClear Technology

    The CI is composed of the following components:

    • Substrate ●
    • Indicator Ink

    The PI is composed of the following components

    • Substrate
    • Indicator Ink
    • Adhesive ●
    • Backing Paper
    AI/ML Overview

    This document describes the Celerity™ HP Chemical Indicator (CI), a Type 1 vaporized hydrogen peroxide sterilization process indicator. The device is designed to distinguish between processed and unprocessed units through a visible color change from red to orange/yellow when exposed to specific sterilization cycles of V-PRO® Low Temperature Sterilization Systems and ASP® STERRAD® Systems.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Test NameDescription / StandardReported Device Performance
    Type 1 Performance TestingConformance to ANSI/AAMI/ISO 11140-1:2014 requirements for a Type 1 Hydrogen Peroxide Chemical Indicator.Pass
    Simulated Use TestingSimulation of typical in-use applications to verify performance.Pass
    Fluorescent Light Stability TestingAssessment of indicator stability under fluorescent light exposure.Pass
    Temperature Extremes Exposure (Freeze/Thaw) TestingEvaluation of indicator performance after exposure to extreme temperatures.Pass
    Transference TestingEvaluation to ensure ink does not transfer from the indicator to other materials.Pass
    Adhesion Stability (Vials) TestingEvaluation of the adhesive's stability for the version used on Self-Contained Biological Indicator Vial Labels.Pass
    Shelf Life TestingDetermination of the product's stability and performance over its intended shelf life.Pass
    Post-Processing Stability TestingEvaluation of the indicator's stability and color integrity after being processed through sterilization cycles.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test or the data provenance (e.g., country of origin, retrospective/prospective). However, it implies that the testing was conducted by STERIS Corporation or its subsidiary Albert Browne Ltd., located in the United Kingdom. Given the nature of chemical indicator testing, it is likely that the testing involved multiple samples for each stress/condition investigated. The studies are non-clinical performance data, implying these were laboratory-based experiments.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to a chemical indicator device. The "ground truth" for chemical indicators is based on whether the indicator changes color in response to specific sterilization parameters (e.g., presence of hydrogen peroxide) and according to established international standards (ANSI/AAMI/ISO 11140-1:2014). This is a physical or chemical reaction, not a subjective interpretation requiring expert consensus.

    4. Adjudication Method for the Test Set

    Not applicable. The color change of a chemical indicator is a direct, observable outcome, not subject to human adjudication in the same way as, for example, image interpretation. The "adjudication" is essentially a pass/fail determination based on the observable color change as per the device's design and specified standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for devices that rely on human interpretation, such as imaging AI algorithms, to assess the impact of the device on reader performance. Chemical indicators do not involve human interpretation in this manner.

    6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance tests of the chemical indicator itself. The device is a "Type 1 process indicator," meaning it provides a visual confirmation of exposure to the sterilization process. Its function is entirely automated in terms of its chemical reaction to sterilization conditions; there is no "human-in-the-loop" performance aspect in the mechanism of the indicator itself. Its performance is assessed purely on its ability to change color accurately under specified conditions.

    7. The Type of Ground Truth Used

    The ground truth used for this chemical indicator's performance evaluation is based on controlled exposure to validated sterilization cycles and conformance to established international standards (ANSI/AAMI/ISO 11140-1:2014). The expected color change (red to orange/yellow) when exposed to these conditions, and the lack of change when not exposed, constitutes the ground truth.

    8. The Sample Size for the Training Set

    No training set is mentioned or applicable. Chemical indicators operate based on chemical reactions; they do not involve machine learning algorithms that require training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1