    Intended Use

    This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patients' body, fluids, waste or environment.

    Device Description

    Class 1 Nitrile Patient Examination Glove 80 LZA, powder free that meets all the requirements of ASTM Standard D3578-95 and FDA Water Leak Test.

    AI/ML Overview

    This document describes the acceptance criteria and performance of the Cashmere Non-Sterile, Powder Free Nitrile Examination Gloves (Blue and Green).

    1. Table of Acceptance Criteria and Reported Device Performance:

    TESTAcceptance Criteria (ASTM D3578-95)Reported Device Performance (Cashmere Powder Free Nitrile Exam Gloves)
    Watertight (1000 ml)GI AQL=4.0%Pass GI AQL=4.0%
    Length (mm)
    Size XSMin 230240 mm minimum for all sizes
    SMin 230
    MMin 230
    LMin 230
    XLMin 230
    Palm width (mm)
    Size XS-75 - 78
    S80 +/- 1082 - 88
    M95 +/- 1092 - 98
    L111 +/- 10102 - 108
    XL-111 - 115
    Thickness (mm) (Single Layer)
    FingerMin 0.080.10 minimum
    PalmMin 0.080.10 minimum
    Physical Properties - Before Aging
    Tensile Strength (Mpa)Min 1416.5
    Ultimate Elongation (%)Min 500550
    Physical Properties - After Aging
    Tensile Strength (Mpa)Min 1423.1
    Ultimate Elongation (%)Min 500520
    Powder Content- (FDA Requirement)Below 2 mg / glove

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each test. However, the performance data is based on the ASTM D3578-95 standard and the FDA 1000 ml watertight test, which would typically specify sampling plans. The data provenance is from SEAL POLYMER INDUSTRIES SDN. BHD. in Lahat, Perak, Malaysia. The study appears to be a retrospective analysis of product performance against established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For product performance testing against standard specifications like ASTM D3578-95, the "ground truth" is typically defined by the objective measurement procedures within the standard, not by expert consensus on individual cases.

    4. Adjudication method for the test set:

    Not applicable. The performance data is based on objective measurements and compliance with a standard (ASTM D3578-95) and FDA requirements, not on subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a submission for a medical glove, a physical device, and not an AI-powered diagnostic or decision support tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    The ground truth used for performance evaluation is based on established industry standards and regulatory requirements, specifically:

    • ASTM D3578-95 Standard for Rubber Examination Gloves.
    • FDA 1000 ml watertight test for pinhole requirements.
    • FDA requirements for minimum powder residual content.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. This is a physical medical device.