Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K180238

    Validate with FDA (Live)

    Date Cleared
    2018-06-01

    (123 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.

    Device Description

    The CARTO® 3 EP Navigation System, Version 6.0 is a catheter based atrial and ventricular mapping system designed to acquire and analyze data points, and use this information to display 3D anatomical and electroanatomical maps of the human heart. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheters and reference devices. The system allows electrograms and cardiac maps display based on the received intracardiac signals from the catheters. The CARTO® 3 System V6.0 uses two distinct types of location technology – magnetic sensor technology and Advanced Catheter Location (ACL) technology.

    The CARTO® 3 System V6.0 consists of the following components:
    Patient Interface Unit (PIU) Graphic User Interface (GUI) Wide-Screen monitors, keyboard, and mouse Intracardiac In Port Intracardiac Out Port Power Supply Patches Connection Box and Cables Pedals Location Pad

    All hardware components of the CARTO® 3 System V6.0with the VISITAG SURPOINT™ Module are identical to those described for the predicate device.

    AI/ML Overview

    This FDA 510(k) summary (K180238) for the CARTO 3 EP Navigation System Version 6.0 and Accessories with the VISITAG SURPOINT™ Module focuses on demonstrating substantial equivalence to a predicate device (K170600), rather than establishing new acceptance criteria or proving performance of the fundamental technology. Therefore, much of the requested information regarding specific acceptance criteria, detailed study results, sample sizes for test/training sets, and expert involvement is not explicitly provided in this document.

    However, based on the information available, here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes the performance data in general terms, stating that "All testing passed in accordance with appropriate test criteria and standards." It does not provide specific quantitative acceptance criteria or detailed performance metrics. The focus is on demonstrating that modifications did not negatively affect existing features.

    Acceptance CriteriaReported Device Performance
    New and modified features verified (functional testing)All testing passed in accordance with appropriate test criteria and standards.
    Existing features not negatively affected (regression testing)All testing passed in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness.
    Magnetic location sensor accuracy (stated to be identical to predicate)(Implied to be same as predicate due to "identical magnetic location sensor accuracy" statement, but no specific values provided)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states, "The CARTO® 3 EP Navigation System, Version 6.0 with the VISITAG SURPOINT™ Module underwent extensive bench testing..." This indicates the testing was conducted on the device itself (likely multiple units or configurations) rather than on patient data. No sample size for a "test set" in terms of patient data or data provenance (country of origin, retrospective/prospective) is provided, as the study appears to be bench-based and not a clinical study involving patient data with ground truth.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided as the testing described is bench testing, not a clinical study involving expert-established ground truth.

    4. Adjudication Method for the Test Set:

    This information is not provided as the testing described is bench testing, not a clinical study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study is mentioned. The document focuses on device modifications and bench testing, not on comparing human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only) Performance:

    The document describes "extensive bench testing" to verify new and modified features. While this is likely focused on the algorithm's performance within the system, it's not explicitly labeled as an "algorithm-only standalone performance study" in the context of comparative clinical evaluations. The device itself is a "Programmable Diagnostic Computer," so its functionality is inherently algorithmic.

    7. Type of Ground Truth Used:

    For the bench testing, the "ground truth" would likely be derived from:

    • Engineering specifications and design documents: To verify that the new features function as intended.
    • Reference measurements/simulations: For aspects like magnetic location sensor accuracy, comparing device output to known physical parameters.
    • Predicate device behavior: For regression testing, verifying that existing features still perform identically or within expected tolerances compared to the predicate.

    No "expert consensus, pathology, or outcomes data" ground truth is relevant or mentioned for this type of testing.

    8. Sample Size for the Training Set:

    This information is not provided. The document describes enhancements to an existing system, implying refinement of existing algorithms rather than the development of entirely new ones requiring large-scale training sets for machine learning. If machine learning was used in the development of the "VISITAG SURPOINT™ Module" or "HD Coloring" feature, the details of its training are not disclosed in this summary.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided as there is no mention of a dedicated training set or machine learning model development in detail.

    In summary, this 510(k) pertains to modifications of an existing device and relies on bench testing to demonstrate substantial equivalence, rather than new clinical effectiveness studies with explicit acceptance criteria, patient data, and expert interpretation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170600

    Validate with FDA (Live)

    Date Cleared
    2017-07-25

    (146 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications

    Device Description

    The CARTO® 3 EP Navigation System, Version 6.0 is a catheterbased atrial and ventricular mapping system designed to acquire and analyze data points, and use this information to display 3D anatomical and electroanatomical maps of the human heart. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheters and reference devices. The system allows electrograms and cardiac maps display based on the received intracardiac signals from the catheters. The CARTO® 3 System V6.0 uses two distinct types of location technology - magnetic sensor technology and Advanced Catheter Location (ACL) technology.

    The CARTO® 3 System V6.0 consists of the following components:

    • . Patient Interface Unit (PIU)
    • Graphic User Interface (GUI)
    • Wide-Screen monitors, keyboard, and mouse
    • Intracardiac In Port ●
    • Intracardiac Out Port
    • Power Supply ●
    • Patches Connection Box and Cables ●
    • Pedals
    • Location Pad ●

    All hardware components of the CARTO® 3 System V6.0 are identical to those described for the predicate CARTO® 3 System V4.2.

    AI/ML Overview

    This document is a 510(k) Summary for the CARTO® 3 EP Navigation System, Version 6.0 and Accessories. It focuses on demonstrating substantial equivalence to a predicate device (CARTO® 3 EP Navigation System Version 4.2) rather than presenting a study demonstrating acceptance criteria met for a novel device performance.

    Therefore, much of the requested information (acceptance criteria, specific study design details like sample size for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment for a training set) is not explicitly provided in this document in the format typically seen for a new device's performance evaluation.

    However, based on the provided text, we can infer some information relevant to the performance evaluation that was conducted to support the updated device's safety and effectiveness.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of explicit acceptance criteria with corresponding performance metrics. Instead, it states that the device "passed all tests in accordance with appropriate test criteria and standards." The performance data section describes the type of testing conducted.

    Implied Acceptance Criteria (based on predicate equivalence and enhancements):

    • Safety and Effectiveness: No new questions of safety or effectiveness are raised compared to the predicate device.
    • Functionality of New Features: The new and modified features (ConfiDENSE Module annotations, CARTO VIZIGO™ Software, TOSHIBA X-ray system compatibility, CARTOSEG updates, GUI modifications) perform as intended.
    • Preservation of Existing Functionality: Existing features are not negatively affected by modifications. This implies that the performance of core features (like magnetic location mapping technology and sensor accuracy) is maintained at the level of the predicate device (Version 4.2).

    Reported Device Performance:

    • "The CARTO® 3 EP Navigation System, Version 6.0 passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness."
    • Specifically noted is that the device "Have identical magnetic location sensor accuracy" as the predicate.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated. The document mentions "extensive bench and pre-clinical testing under simulated clinical conditions." This suggests that a test set was used, but the size is not quantified.
    • Data Provenance: "simulated clinical conditions." This implies a retrospective or de-identified prospective approach if clinical data was used in simulation, or purely simulated data for bench testing. The country of origin for the data is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not explicitly stated. The evaluation was primarily engineering and functional testing rather than a clinical study requiring expert ground truth in the traditional sense. The "acceptance criteria and standards" would likely refer to engineering specifications and regulatory standards.

    4. Adjudication method for the test set

    Not explicitly stated. Given the nature of bench and pre-clinical testing, it's unlikely a formal clinical adjudication method (like 2+1 or 3+1) was used. Testing would involve verifying output against known inputs or engineering specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. This device is a navigation system, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images. The enhancements are primarily software features and compatibility updates.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone Performance: Yes, the described "extensive bench and pre-clinical testing" would represent a form of standalone performance evaluation for the system's functionalities, particularly the new modules and integrability, without direct human-in-the-loop performance being a primary assessment as a "diagnostic" device. For instance, the "magnetic location sensor accuracy" being identical to the predicate would be a standalone technical measurement.

    7. The type of ground truth used

    • Type of Ground Truth: For the "bench and pre-clinical testing," the ground truth would be based on:
      • Engineering specifications: For quantitative measurements like magnetic location sensor accuracy.
      • Predicate device's established performance: For verifying that existing features were not negatively impacted.
      • Expected functionality: For new features, verifying they operate as designed (e.g., specific annotations are produced, software integrates correctly).

    8. The sample size for the training set

    Not applicable in the context of this 510(k) summary. This document describes an update to an existing medical device, primarily involving software enhancements and hardware compatibility, not a de novo AI/ML algorithm development that would typically require a training set in the machine learning sense.

    9. How the ground truth for the training set was established

    Not applicable for the reasons stated above (no training set mentioned for an AI/ML algorithm in this context).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1