510(k) Data Aggregation

    K Number
    K050029

    Validate with FDA (Live)

    Device Name
    CARDIOMPO TEST
    Manufacturer
    Date Cleared
    2005-05-10

    (124 days)

    Product Code
    Regulation Number
    866.5600
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioMPO™ Test is comprised of the CardioMPO Reagent Kit, the CardioMPO Calibrator Kit, and the CardioMPO Control Kit.

    The CardioMPO Reagent Kit is an enzyme immunoassay intended for the quantitative determination of myeloperoxidase in human plasma, to be used in conjunction with clinical history, ECG and cardiac biomarkers to evaluate patients presenting with chest pain that are at risk for major adverse cardiac events, including myocardial infarction, need for revascularization, or death.

    The PrognostiX CardioMPO Calibrator Kit is intended for use with the CardioMPO Reagent Kit to establish a calibration curve that is used to determine MPO concentration.

    The PrognostiX CardioMPO Control Kit is intended for use with the CardioMPO Reagent Kit as an assayed quality control sample to monitor and evaluate the precision and accuracy of the CardioMPO Test.

    Device Description

    The CardioMPO Test is a sandwich enzyme immunoassay that uses two highly specific antibodies, one monoclonal and one polyclonal, and an enzyme labeled anti-rabbit IgG antibody for the measurement of MPO concentration in human plasma. The CardioMPO Test is comprised of the CardioMPO Reagent Kit, the CardioMPO Calibrator Kit, and the CardioMPO Control Kit.

    The CardioMPO Reagent Kit contains the following: a microtiter plate coated with a monoclonal anti-MPO antibody; a solution of primary polyclonal rabbit anti-MPO antibody; a solution of secondary goat anti-rabbit IgG antibody labeled with horseradish peroxidase solution; TMB substrate solution; Stop Solution; Wash Buffer Concentrate; Assay Buffer; and plate sealers.

    The CardioMPO Calibrator Kit contains six Calibrators comprised of human MPO in a phosphatebuffered matrix containing proteins, detergents, and stabilizers. Calibrators are intended to establish a calibration curve that is used to determine MPO concentration. Calibrator values are provided on individual Calibrator labels. Calibrators are provided ready-to-use.

    The CardioMPO Control Kit contains three Controls comprised of human MPO in a human plasma matrix. Controls are intended to monitor and evaluate the precision and accuracy of the CardioMPO Test. Ranges are provided on individual Control labels. Controls are freated in the same manner as patient samples.

    Calibrators are added directly to the appropriate wells of the Microtiter Plate. Assay Buffer is added to all wells that are intended for Control or sample analysis. Aliquots of Controls or samples are added to the appropriate wells and the plate is incubated for 60 minutes at 20-26°C. The wells are then washed with Wash Buffer to remove antigens not specifically bound to the immobilized antibody.

    A yellow primary polyclonal rabbit anti-MPO antibody is added to each well and incubated for 60 minutes at 20-26°C. This antibody binds to the MPO captured on the plate. The plate is again washed with Wash Buffer to remove unbound primary antibody.

    A blue secondary goat anti-rabbit IgG antibody, labeled with the enzyme horseradish peroxidase (HRP), is then added to each well and incubated for 30 minutes at 20-26°C. This antibody binds to the primary antibody, and the MPO in the sample is "sandwiched" between the monoclonal antibody on the solid phase and a complex of the rabbit anti-MPO and the HRP-goat anti-rabbit IgG antibody. The wells are washed with Wash Buffer to remove unbound HRP-labeled antibody.

    The substrate, tetramethylbenzidine (TMB), is then added to each well and incubated for 10 minutes at 20-26°C resulting in the development of a blue color. Color development is stopped with the addition of Stop Solution, changing the color to yellow.

    The enzymatic turnover of the substrate is determined spectrophotometrically at 450 nm, preferably with correction at 630 nm, and is directly proportional to the concentration of MPO. The absorbances of the Calibrators are used to plot a standard curve of absorbance versus MPO concentration from which the MPO concentration in the Controls or samples can be determined.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the PrognostiX CardioMPO™ Test:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" for all performance characteristics in a separate section with numerical targets. Instead, it presents performance characteristic results and implies their acceptability by stating "The PrognostiX CardioMPO Test is acceptably precise, linear, accurate, and is not subject to appreciable cross-reactivity or interference." and "The detection limit of 13 pM is acceptable for the intended use of this test."

    However, we can infer acceptance criteria based on the comparison to the predicate device and from general good practices for in vitro diagnostic devices, and report the observed performance:

    Acceptance Criteria (Inferred/Stated)Reported Device Performance (CardioMPO™ Test)Notes
    Analytical Performance
    Minimum Detection Limit13 pM MPOCalculated by interpolation of the mean + 2 SD of 24 replicates of 0 pM MPO Calibrator (F). Stated as "acceptable for the intended use."
    Total Precision in Plasma8.2%Compares favorably with the predicate device's <8% total precision. The main summary table lists "<9%", so 8.2% meets this. NCCLS guidelines EP5-A followed.
    Within-run Precision5.5%Compares favorably with the predicate device's <6% within-run precision. The main summary table lists "<6%", so 5.5% meets this. NCCLS guidelines EP5-A followed.
    Linearity Range13 pM to 5223 pM MPOWithin 60 pM or 9% in this interval. NCCLS guidelines EP6-A followed.
    Interfering Substances Bias≤10% for listed substancesNo interference observed at tested concentrations (or <1500 mg/L hemoglobin). NCCLS guidelines EP7-A followed. Hemoglobin levels >1500 mg/L caused >10% bias.
    Interfering Antibodies Recovery89% to 118% (mean 108%)No significant interference observed due to heterophilic antibodies.
    Cross-reactivityNo significant cross-reactionFor a list of tested substances (e.g., α-1 antrypsin, C-reactive protein, etc.) up to indicated concentrations. NCCLS guidelines EP7-A followed.
    Spiking Recovery89% to 105% (mean 97%)Fourteen lithium heparin plasma samples spiked with various MPO levels. Acceptable range for recovery in assays is typically 80-120%.
    Dilution Recovery1:2 dilution: 102% to 131% (mean 116%)Recoveries at dilutions 1:4 through 1:16 rose to 132%-145% due to decreased plasma matrix effects. PrognostiX does not recommend further dilution for samples exceeding the highest calibrator.
    Hook EffectNo hook effect up to 800,000 pM MPORoughly 150x greater than the upper end of the reportable range and 500x greater than the median clinical level.
    Clinical Performance
    Odds Ratio for MACE (All Patients)Increases from 1.0 to 3.3 across 4 quartilesStatistically significant (p < 0.001 for highest quartile). Demonstrates association with risk of MACE.
    Odds Ratio for MACE (TnT negative)Increases from 1.0 to 6.9 across 4 quartilesStatistically significant (p < 0.001 for highest quartile). Stronger association in this subgroup, highlighting potential utility for early risk stratification.
    Association with MACEDemonstrated association"The clinical evaluation... demonstrate[s] that plasma levels of MPO are associated with risk of myocardial infarction, revascularization, and death."

    2. Sample size used for the test set and the data provenance

    • Test Set (Clinical Study):

      • Sample Size: 560 banked plasma samples.
      • Data Provenance: Retrospective study of patients presenting to the Emergency Department within 24 hours of the onset of chest pain. These samples were obtained from a prior study (Brennan, et al, N Engl J Med 2003; 349:1595-1604).
      • Country of Origin: Not explicitly stated, but the cited paper by Brennan et al. is from the New England Journal of Medicine, suggesting a US or Western origin for the prior study.
    • Analytical Performance Studies (e.g., Precision, Linearity, Interference, etc.):

      • Sample Size:
        • Sensitivity: 24 replicates of 0 pM MPO Calibrator.
        • Precision: 9 samples (4 Matrix Controls, 5 Plasma Controls) assayed in duplicate, for 20 days.
        • Linearity: 9 MPO concentrations created by intermixing spiked and neat plasma samples.
        • Interfering Substances: Plasma containing 3000 pM and 1000 pM MPO.
        • Interfering Antibodies: Twenty-one plasma samples.
        • Cross-reactivity: Plasma pools containing MPO and Assay Buffer.
        • Spiking Recovery: Fourteen lithium heparin plasma samples.
        • Dilution Recovery: Thirteen lithium heparin plasma samples.
        • Reference Interval: "a population of apparently healthy blood donors" with 145 males and 149 females after outlier adjustment (initial sample size not explicitly given but >294).
      • Data Provenance: These are laboratory-based analytical studies, likely conducted internally by PrognostiX, Inc. and are prospective in nature for the purpose of this submission (i.e., substances were specifically "tested" and samples "analyzed" for the submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for the clinical study was the occurrence of "major adverse cardiac events (MACE), including myocardial infarction, need for revascularization, or death" over a 30-day and 6-month interval. This outcome was assessed by follow-up phone calls. The document does not specify:

    • The number of experts used to establish this ground truth.
    • The qualifications of those experts (e.g., cardiologists, emergency physicians).
    • The methods used to adjudicate MACE events (e.g., review of medical records, independent committee).
      Given it was "assessed by follow up phone calls," it's highly likely that trained personnel (possibly nurses or research coordinators) collected this information, potentially with physician review for confirmation, but the details are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not specify any formal adjudication method (like 2+1, 3+1) for the clinical outcomes (MACE) in the test set. The assessment relied on "follow up phone calls."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.
    • The CardioMPO™ Test is an in vitro diagnostic (IVD) assay for measuring a biomarker (myeloperoxidase) in plasma. It is not an imaging AI device that would involve human "readers" interpreting images with or without AI assistance. The test provides a quantitative result which is then used by clinicians in conjunction with other clinical information.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the analytical performance and clinical utility data presented are for the device operating in a standalone manner. The device measures MPO concentration; there is no human-in-the-loop performance component for the measurement itself. The interpretation of the MPO result (in conjunction with clinical history, ECG, and cardiac biomarkers) is where human expertise is applied. The clinical study demonstrated the MPO test's predictive power as a standalone biomarker.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the clinical study was outcomes data, specifically the occurrence of "major adverse cardiac events (MACE), including myocardial infarction, need for revascularization, or death" within 30 days and 6 months after presentation with chest pain.

    8. The sample size for the training set

    The document does not explicitly describe a separate "training set" for the clinical study. The 560 banked plasma samples were used to evaluate the device's performance in predicting MACE. The study used multivariate logistic regression models for analysis, which are typically applied to a dataset to assess relationships, but there's no mention of a separate training phase to develop the MPO test itself or a decision rule based on MPO levels from this dataset, beyond assessing its association with outcomes. It's more of a validation of the biomarker's utility.

    9. How the ground truth for the training set was established

    As there's no explicitly defined separate "training set" for the clinical study from the perspective of developing an AI algorithm, this question isn't directly applicable in the conventional sense of AI/ML model development. However, if we interpret "training set" as the data used to understand the reference ranges or analytical characteristics, the ground truth was established as follows:

    • Reference Interval: Plasma samples from "a population of apparently healthy blood donors" were screened, and outliers were removed using a two-stage method (Box-Cox transformation followed by robust outlier detection). The "ground truth" for normality was based on this statistical analysis of healthy individuals.
    • Analytical Studies (Precision, Linearity, etc.): Ground truth for these was established by standard laboratory practices (e.g., using known concentrations of MPO, spiked samples, or reference materials) according to NCCLS guidelines. These studies aim to characterize the assay's performance against expected analytical values.