510(k) Data Aggregation

    K Number
    K063042

    Validate with FDA (Live)

    Date Cleared
    2007-05-30

    (239 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioMem® CM 3000-12BT is a Hotter recorder which is indicated for patients who may benefit from a long-term continuous electrocardiographic (ECG) recording, including, but not limited to, those with complaints of palpitations, syncope, chest pain, shortness of breath, or those that need to be monitored to judge their current cardiac functionality.

    Device Description

    The CM 3000-12BT is a Holler recorder designed to be used in conjunction with the evaluation software GardioDay® (K051471). This recorder is not capable of any diagnosis nor can it provide any interpretation of the clata. The CM 3000-12BT acquires, digitizes and stores data to be analyzed by CardioDay®. The CM 3000-12BT uitlizes a 10-lead electrode hookup and placement to provide CardioDay" with 12 channels of full disclosure for Holler analysis. The cardiac data provided by CardioDay® is used by trained medical personnel to assist in the diagnosis of patients with various rhythm patterns. The CM 3000-12BT Holter recorder stores 12 ECG channels continuously for up to 48-hours including the detection of pacemaker pulses. A keyped is available to set up system contiguration, to enter patient's ID and name, to check lead quality during hook-up, and to start the recording. During the recording, the keypad can be used to enter event markers. The CM 3000-12BT has a LCD screen to allow ECG display during the hook-up, lead quality check, system configuration and various messages for the hock-up technician. The CM 3000-12BT uses one or two AA batterles, and a removable memory card for data storage.

    ECG data and palient data can be Iransmitted via a Bluetooth connection (BT module from Amber wireless GmbH) from the CM 3000-12BT to CardidDay" and also from CardioDay® to the CM 3000-12BT

    AI/ML Overview

    Here's an analysis of the provided text regarding the CardioMem CM 3000-12BT, focusing on acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly list specific quantitative acceptance criteria or detailed performance metrics from a test study. Instead, it relies on a comparison to predicate devices and adherence to relevant standards. The "New/Modified Device" column serves as the "reported device performance" in this context, compared against the "Legally Marketed Device" columns.

    SpecificationAcceptance Criteria (Implied by Predicate Device)Reported Device Performance (CardioMem CM 3000-12BT)
    Online dataNoNo
    Patient hookup10 ECG electrodes10 ECG electrodes
    Number of ECG channels desired1212
    ECG lead namesRA, LA, LL, RL, V1, V2, V3, V4, V5, V6RA, LA, LL, RL, V1, V2, V3, V4, V5, V6
    A to D sample rate1024 samples/sec (from K051686) / 720 samples/sec (from K052556)1024 samples/sec
    A to D resolution12 bit12 bit
    Pacemaker detectionNo (from K051686) / Not specified (from K052556)Yes
    Open-Lead detectionYesYes
    Memory typeCompactFlash™ Memory CardCompactFlash™ Memory Card
    Data transfer methodVia removable memory cardVia removable memory card
    Online data transferVia OptoLink cable OL1000 (from K051686) / Via Bluetooth connection (from K052556)Via Bluetooth connection or USB connection
    Memory card data formatStandard file systemStandard file system
    Liquid crystal display (LCD)YesYes
    Display purposeDisplay ECG, check lead quality, input patient ID, display messagesDisplay ECG, check lead quality, input patient ID, display messages
    KeyboardProtected touch keys (membrane)Protected touch keys (membrane)
    System configurationCheck lead quality, input patient ID and name, start recordingCheck lead quality, input patient ID and name, start recording
    System configuration methodPer keyboard and LCD displayPer keyboard and LCD display
    Marker buttonYesYes
    Size108 * 66 * 22 mm108 * 66 * 22 mm
    Weight<160 g (without battery)<160 g (without battery)
    Carrier pouchCarrier bag (Pouch)Carrier bag (Pouch)
    Battery1 or 2 x 1.5 V AA alkaline1 or 2 x 1.5 V AA alkaline
    Battery check prior to recordingYesYes
    External patient cableYesYes
    Record identification procedureYesYes
    ECG channelYesYes

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Clinical testing was not required to demonstrate substantial equivalence of safety and effectiveness." This implies that there was no specific "test set" of patient data used in a clinical study as would be typical for many medical devices. The evaluation for this 510(k) focused on non-clinical testing and comparison to predicate devices. The data provenance would therefore be from the technical specifications and performance characteristics of the device itself and the predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since no clinical testing with a "test set" was performed, this information is not applicable and not provided in the document.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used for expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states "Clinical testing was not required."

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The device is an ECG recorder, not an algorithm that performs analysis. It acquires, digitizes, and stores data to be analyzed by separate evaluation software (CardioDay®) and "trained medical personnel." Therefore, the concept of a "standalone" algorithm performance study is not directly applicable to this device, which is a data acquisition unit. Its performance is related to its ability to accurately record physiological signals. Non-clinical testing focused on this aspect.

    7. Type of Ground Truth Used

    The ground truth for the device's technical specifications and electrical performance would be established through engineering measurements, laboratory testing, and adherence to established industry standards (e.g., IEC 60601 series, ANSVAAMI EC 38). For example, the A to D sample rate and resolution would be verified against internal specifications and industry benchmarks for Holter recorders.

    8. Sample Size for the Training Set

    Not applicable. This device is a hardware recorder, not an AI/ML algorithm that requires a training set. The "training" in this context would refer to the device's design, engineering, and manufacturing process.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm for this hardware device. The device's functionality is based on its engineering design and adherence to technical requirements, which are verified through non-clinical testing against established industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051686

    Validate with FDA (Live)

    Date Cleared
    2005-07-14

    (21 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioMem® CM 3000-12 digital Holter recorder is intended to continuously record up to 48 hours of ECG data on a digital flash memory card. The CardioMem® CM 3000-12 performs no cardiac analysis by itself and is intended to be used with the analysis evaluation software CardioDay®. The recorded data are downloaded to a PC for analysis and following evaluation by a trained physician or health care professional.

    Federal law restricts CardioMem® CM 3000-12 to use on order of a physician.

    This device is available only upon the order of a physician or other licensed medical professional.

    The CardioMem® CM 3000-12 is a Holter recorder which is indicated for patients who may benefit from a long-term continuous electrocardiographic (ECG) recording, including, but not limited to, those with complaints of palpitations, syncope, chest pain, shortness of breath, or those that need to be monitored to judge their current cardiac functionality.

    Device Description

    The CM 3000-12 is a Holter recorder designed to be used in conjunction with the evaluation software CardioDay® (not included in this 510(k)). This recorder is not capable of any diagnosis nor can it provide any interpretation of the data. The CM 3000-12 acquires, digitizes and stores data to be analyzed by CardioDay". The CM 3000-12 uiliizes a 10-lead electrode hookup and placement to provide CardioDay® with 12 channels of full disclosure for H analysis. The cardiac data provided by CardioDay® is used by trained medical personnel to assist in the diagnosis of patients with various rhythm patterns. The CM 3000-12 Holter recorder stores 12 ECG channels continuously for up to 48-hours. A keypad is available to set up system configuration, to enter patient's ID and name, to check lead quality during hook-up. and to start the recording. During the recording, the keypad can be used to enter event markers. The CM 3000-12 has a LCD screen to allow ECG display during the hook-up, lead quality check, system configuration and various messages for the hook-up technician. The CM 3000-12 uses one or two AA batteries, and a removable memory card for data storage.

    AI/ML Overview

    The provided document is a 510(k) summary for the CardioMem® CM 3000-12 Holter recorder. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study.

    The device, CardioMem® CM 3000-12, is a Holter recorder that acquires, digitizes, and stores ECG data. Crucially, the document states: "This recorder is not capable of any diagnosis nor can it provide any interpretation of the data." and "The CardioMem® CM 3000-12 performs no cardiac analysis by itself and is intended to be used with the analysis evaluation software CardioDay®."

    Therefore, the "acceptance criteria" and "device performance" in this context do not refer to diagnostic accuracy or clinical effectiveness metrics, but rather to the recorder's technical specifications and compliance with regulatory standards for device safety and electrical performance. The study described is not a clinical performance study but a series of tests to confirm safety and efficacy based on technical standards.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document implicitly defines "acceptance criteria" through comparison with predicate devices and adherence to regulatory standards (IEC and AAMI). The "reported device performance" is essentially that the device meets these technical specifications and standards.

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Equivalence to Predicate Devices:
    Continuous recording of ECG dataRecords 12 ECG channels continuously for up to 48-hours
    No online data monitoring & alarmNo
    Patient Hookup (10 ECG electrodes)10 ECG electrodes
    Number of ECG channels derived (12)12
    A to D sample rate (1024 samples/sec - matching one predicate)1024 samples/sec
    A to D resolution (12 bit - matching one predicate)12 bit
    Uses CompactFlash™ Memory CardCompactFlash™ Memory Card
    Data transfer via removable memory cardVia removable memory card
    Standard file system for memory card dataStandard file system
    Liquid Crystal Display (LCD) presentYes
    Display purposes (ECG, lead quality, patient ID, messages)Display ECG, check lead quality, input patient ID, display messages
    Keyboard type (Protected touch keys)Protected touch keys (membrane)
    System configuration method (keyboard/LCD)Per keyboard and LCD display
    Marker button presentYes
    External patient cable presentYes
    Record identification procedureYes
    ECG channel previewYes
    Safety and Efficacy Standards Compliance:
    Compliance with Quality System Regulation 21 CFR 820Basis for development
    Environmental and safety tests (including EMC)Tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-2-47 and ANSI/AAMI EC 38:1998 standards.
    (Lack of Pacemaker Detection - deviation from some predicates)No (Pacemaker detection)

    2. Sample size used for the test set and the data provenance

    The document does not describe a clinical "test set" in the traditional sense involving patient data for diagnostic evaluation. The "tests" mentioned are primarily engineering and electrical safety tests, not performance on a clinical dataset. Therefore, there is no sample size of patients or data provenance (country, retrospective/prospective) for a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As this is not a clinical performance study involving diagnostic outputs, there was no ground truth for a test set established by medical experts. The device explicitly states it performs no analysis or diagnosis.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a Holter recorder, not an AI analysis tool. It does not perform analysis itself, nor does it involve human readers interpreting AI-assisted outputs.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a data acquisition and storage tool, not an algorithm performing standalone analysis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for this device's acceptance is its adherence to technical specifications and safety standards, rather than diagnostic accuracy against clinical ground truth.

    8. The sample size for the training set

    Not applicable. This document is about a hardware device, a Holter recorder, not an AI or analytical algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.