Search Results
Found 1 results
510(k) Data Aggregation
(67 days)
Monitoring combined pressure and electrical impedance (volume) in the cardiovascular system
The Cath-Cath-Catheter is a disposable/single use, sterile pressure-volume catheter transducer that combines one or two pressure transducers and a series of electrodes mounted at various locations along the distal segment of the catheter body. It terminates in one electrical connection at the proximal end.
The medical device in question is the Millar Instruments Cardia-Cath, a Pressure-Volume (PV) Catheter Transducer. The provided text outlines the acceptance criteria and corresponding study results for this device, primarily focusing on verification testing, biocompatibility, sterility, and accelerated aging, demonstrating its substantial equivalence to a legally marketed predicate device (Millar model SPC-550).
Here's an organized breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Verification Testing (BP-22) | - Catheters must meet the same specifications as the legally marketed predicate device as defined in its IFU. - Compliance with BP-22 standard requirements. - Temperature Error Band at Zero Pressure: ≤ 3.24 mmHg (for predicate device). - Sensitivity Error Band: ≤ 3.16% (for predicate device). | - All tested catheters met the testing requirements and complied with BP-22 standard requirements. - Deviation 1: Temperature Error Band at Zero Pressure relaxed to 3.5 mmHg. - Deviation 2: Sensitivity Error Band relaxed to 3.5%. - The modified device IFU includes the new (relaxed) values to ensure end-users obtain the same level of data accuracy. |
| Transportation Testing | - Compliance with ASTM D4169-08 standards for fully packaged finished product. - Samples must pass all tests including bubble leak and packaging integrity. - No impact on device functionality (safety and effectiveness) after transportation testing. | - Samples passed all tests including bubble leak test and packaging integrity, as per DDL report No: 1009057 Rev. A. - Functionality testing (Millar reference VRP-2010-0423-01 VAL-180) confirmed no impact on device functionality, safety, and effectiveness. |
| Biocompatibility (ISO 10993-1) | - Requirements for External communicating device, circulating blood, contact duration A-Limited (<24hrs): Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Systemic toxicity (acute), and Haemocompatibility. - Statistical similarity to the predicate device (Model SPC-550). | - Nelson Labs analysis determined that the Cardia-Cath (modified device) was statistically similar, and therefore substantially equivalent, to the predicate (legally marketed) device, Model SPC-550 for all tests: Modified ASTM Hemolysis (Direct Contact Method), Partial Thromboplastin time (PTT) test, MEM Elution, Complement Activation Test, ISO Acute Systemic Injection Test, Rabbit Pyrogen Test (Material Mediated)- ISO, ISO Intracutaneous Reactivity Test, Thrombogenicity Study in Dogs ISO, ISO Guinea Pig Maximization Sensitization Test. |
| Sterility | - Method: EtO-Overkill, validated per ANSI/AAMI ISO 11135. - Sterility Assurance Level (SAL): ≥ 10⁻⁶. - Pyrogen Free: Limulus Amebocyte Lysate test (LAL) should not show significant interference with lysate gel formation. - Residuals: Ethylene Oxide (EO) < 10mg, Ethylene Chlorohydrin < 5mg (per ANSI/AAMI/ISO 10993-7). - EtO breathable materials (Tyvek) for sterilant penetration and maintenance of sterility. | - Sterility method EtO-Overkill validated per ANSI/AAMI ISO 11135. - Actual SAL: 3.33 times more than 10⁻⁶ (meeting the requirement). - LAL test showed no significant interference with lysate gel formation, indicating pyrogen-free status. - Maximum EO residuals: 0.3096mg (<10mg). - Maximum Ethylene Chlorohydrin residuals: 0.1079mg (<5mg). - Packaging uses EtO Breathable materials (Tyvek). |
| Accelerated Aging | - 2-year accelerated aging test per ASTM F1980-07. - Device must pass test criteria after accelerated aging. - Substantial equivalence to the predicate device (Model SPC-550) after aging. | - A 2-year accelerated aging test was performed by DDL. - The tested samples were inspected by Millar, and the Cardia-Cath passed the test criteria and maintained substantial equivalence to the predicate device, Model SPC-550. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the exact sample sizes for each test set (e.g., number of catheters tested for BP-22, transportation, or accelerated aging, or the number of samples for biocompatibility). It mentions:
- "Selected samples of the modified catheters were tested to the same specification..." for verification testing.
- "Samples" were tested for transportation testing.
- "Samples of the legally marketed device and the modified device were sent to Nelson Laboratories" for biocompatibility.
- The accelerated aging test was performed on "the actual modified device. Cardia-Cath."
Data Provenance: The studies appear to be prospective verification and validation tests conducted by Millar Instruments, DDL (for transportation and accelerated aging), and Nelson Labs (for biocompatibility). The testing was for a modified device (Cardia-Cath) to establish its substantial equivalence to a legally marketed predicate device (Millar model SPC-550). The country of origin of the data is implied to be primarily the United States, given the involved organizations (Millar Instruments in Houston, TX; Nelson Labs; DDL; FDA submission).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of information is generally not applicable to the non-clinical bench and lab testing described in the document. The "ground truth" here is defined by established international standards (ISO 14971, ISO 10993-1, ISO 11135, ASTM D4169-08, ASTM F1980-07) and internal specifications of the manufacturer, rather than expert consensus on clinical findings.
However, the document states:
- "The final report [for verification testing] has been reviewed and approved by qualified individuals."
- Biocompatibility testing was performed "following GLP requirements" by Nelson Labs, an accredited laboratory.
This indicates that experts with relevant qualifications (e.g., engineers for verification testing, toxicologists/biologists for biocompatibility) would have been involved in the design, execution, and review of these tests. Specific numbers and detailed qualifications are not provided.
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or patient data by multiple human readers. This document describes non-clinical performance testing (bench testing, lab analysis, physical testing) against predefined technical specifications and standards. Therefore, an adjudication method for a test set in the conventional sense is not applicable. The "adjudication" is inherent in the objective measurements and pass/fail criteria of the specified tests.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document describes non-clinical performance and safety testing (bench, lab, and physical tests) of a medical device against established standards and a predicate device. It does not involve human readers interpreting cases, nor does it assess the effect of AI assistance on human reader performance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
The device described is a physical catheter transducer, not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" is not applicable here. The performance described is inherent to the physical device itself.
7. The Type of Ground Truth Used
The ground truth for the various tests is defined by:
- Industry Standards: ISO 14971:2007 (Risk Management), ISO 10993-1 (Biocompatibility), ISO 11135 (Sterilization), ASTM D4169-08 (Transportation), ASTM F1980-07 (Accelerated Aging).
- Predicate Device Specifications: The "same specification as the legally marketed device as defined in its IFU" for general performance, and specific values for parameters like Temperature Error Band and Sensitivity Error Band of the predicate device (SPC-550).
- Regulatory Limits: Maximum permissible levels for EO and Ethylene Chlorohydrin residuals per ANSI/AAMI/ISO 10993-7.
- Objective Measurements: Directly measured parameters against these standards and specifications (e.g., pressure readings, percentage errors, biological responses, SAL values).
8. The Sample Size for the Training Set
The concept of a "training set" and "training data" is applicable to machine learning or AI models. Since this document describes a physical medical device and its non-clinical verification and validation testing, there is no training set in the context of AI/ML.
9. How the Ground Truth for the Training Set Was Established
As there is no training set for an AI/ML model, this question is not applicable.
Ask a specific question about this device
Page 1 of 1