510(k) Data Aggregation

    K Number
    K961000

    Validate with FDA (Live)

    Date Cleared
    1996-10-02

    (204 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CAPIOX® SX18 Hollow Fiber Oxygenator is used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integral heat exchanger is used to warm or cool the blood or perfusion fluid flowing through the device.

    The CAPIOX SX18 Hardshell Reservoir (detachable) is used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal.

    Device Description

    CAPIOX® SX18 Hollow Fiber Oxygenator contains an integrated heat exchanger and a detachable Hardshell Reservoir. This design permits an integrated system for ease of use as well as independent use of the oxygenator and of the Hardshell reservoir to accommodate a variety of circuit configurations.

    The SX18 oxygenator is a membrane oxygenator consisting of microporous hollow fibers with an integrated heat exchanger, consisting of stainless steel tubes.

    The detachable Hardshell reservoir has a rotatable connection with the oxygenator which permits optimum positioning for connections in the circuit. The venous blood inlet port is also rotatable to permit minimizing tubinq lengths which could result in lower circuit priming volumes.

    The Hardshell reservoir contains a defoamer and a screen filter in the venous blood inlet section. The total capacity of the reservoir is 4,000 mL.

    The cardiotomy section of the hardshell reservoir contains a defoamer and a cardiotomy filter to facilitate qas bubble removal and the removal of particulates/emboli from suctioned blood entering the reservoir.

    A detachable sampling system is positioned at the top of the hardshell reservoir which contains 3 three-way stopcocks. These stopcocks can be used for sampling. The sampling system contains a one-way valve permitting withdrawal of liquid samples but prohibiting entry of air into the blood exiting the oxygenator.

    AI/ML Overview

    The provided text describes the CAPIOX® SX18 Hollow Fiber Oxygenator with detachable Hardshell Reservoir and its substantial equivalence to the Cobe CML Excel. However, the document does not contain a detailed study proving the device meets specific acceptance criteria with reported device performance statistics, sample sizes, ground truth establishment, or expert-based reviews.

    The information provided largely focuses on demonstrating "substantial equivalence" to a predicate device (Cobe CML Excel) rather than detailing a specific study with quantitative acceptance criteria and results.

    Here's a breakdown of what can be extracted and what is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    This specific table is not provided in the document. The text mentions "performance tests" for "Gas Transfer" and "Pressure Drop of Heat Exchanger and Oxygenator" but does not give specific numerical acceptance criteria or the measured performance values for the SX18. It only states, "some differences in performance were observed... however, these differences do not present clinically significant differences."

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided. There is no mention of a specific "test set" in the context of a clinical or ex-vivo study with a defined sample size.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. The document makes no mention of expert review or ground truth establishment for a test set.

    4. Adjudication Method for the Test Set

    This information is not provided. There is no mention of an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not provided. The document describes a comparison between the SX18 and CML Excel devices, but not a study involving human readers' performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This information is not applicable as the device is a medical oxygenator, not an algorithm or AI system. The "performance" discussed relates to the physical device's function.

    7. Type of Ground Truth Used

    This information is not explicitly stated in terms of a formal "ground truth" as typically understood in AI/algorithm studies. The performance comparison against the predicate device (Cobe CML Excel) serves as the basis for "substantial equivalence." The implicit ground truth is the established performance and safety profile of the Cobe CML Excel and generally accepted physiological requirements for oxygenation during cardiopulmonary bypass.

    8. Sample Size for the Training Set

    This information is not applicable as the device is not an AI/algorithm-based system requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above.


    Summary of Available Information Related to Performance and Equivalence:

    The document focuses on demonstrating Substantial Equivalence through comparison with a predicate device, the Cobe CML Excel, based on:

    • Intended Use: Same
    • Design and Materials: Similar, with some differences (e.g., integrated heat exchanger location, detachable reservoir, type of tubing, specific membrane technology - hollow fibers vs. plate membrane). The document asserts that these differences "do not present significant differences in the function and intended uses of the devices."
    • Technology and Principles of Operation: Substantially equivalent (e.g., membrane technology, gravity/vacuum collection, defoamers, filters).
    • Specifications: Some differences (e.g., priming volume, membrane surface area, maximum blood flow range). Claims these "do not affect the substantial equivalence... since both provide adequate gas exchange for clinical use."
    • Performance: "Comparison... was conducted relative to the extended acceptable blood flow rate specification to include 0.5 - 2 LPM."
      • Tests performed: Gas Transfer, Pressure Drop of Heat Exchanger and Oxygenator.
      • Result: "some differences in performance were observed... however, these differences do not present clinically significant differences."

    Additional Safety Information Provided:

    • Gas Transfer relative to time
    • Heat Exchanger Performance Factor
    • Sterilization validation (SAL of 10-5)
    • Ethylene oxide residual limits
    • Manufacturing control tests (100% performance and leak testing)
    • Blood contacting materials tested per FDA GPM #G95-1 (ISO-10993)

    In conclusion, while the document asserts the device's performance and safety through comparison with a predicate, it does not provide the specific-study details, quantitative acceptance criteria, or performance data in the format requested. The primary method of "proof" is the demonstration of substantial equivalence rather than a detailed report of a specific study meeting predefined, quantitative acceptance criteria.