Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K234015

    Validate with FDA (Live)

    Date Cleared
    2024-03-11

    (83 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Caphosol® is indicated for dryness of the mouth or throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the conditions are temporary or permanent. Caphosol® is also indicated as an adjunct to standard oral care in treating mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with an amelioration of pain.

    Caphosol® may be used for relief of dryness of the oral mucosa when hyposalivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy, infection or dysfunction of the salivary glands; inflammation of the mouth or throat; fever; emotional factors such as fear or anxiety; obstruction of the salivary ducts; Sjogren's syndrome; and Bell's Palsy.

    Caphosol® is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atropine or other anticholinergic agents that suppress salivary secretion.

    It may be used as part of an oral hygiene program for patients with dry mouth. Caphosol® provides intensive hygiene of the oral cavity, and may be used to help relieve bad taste and to relieve offensive nasal discharge and crusting.

    Device Description

    Caphosol® is an electrolyte solution resembling human saliva, designed in part to replace the normal ionic and pH balance in the oral cavity. It is intended as a mouth rinse to moisten, lubricate, and clean the oral cavity including the mucosa of the mouth, tongue, and throat. Caphosol® maintains moistness in the oral cavity. It relieves diffuse dryness and fissuring of the: oral mucosa, as well as painful tongue conditions due to hyposalivation. Patients having this condition are also prone to dental caries and candidal infections.

    Caphosol® is a partial substitute for natural saliva. Caphosol® facilitates chewing and speaking; loosens tough mucus; prevent mucous membranes from sticking together, helps remove nasal crust and relieve nasal soreness; improves adherence of dentures, and also relieves bad taste. Caphosol® is an adjunct to standard oral care for treating mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with an amelioration · of pain.

    Caphosol® is a preparation comprising two separately packaged aqueous solutions, a phosphate solution (Caphosol® A) and a calcium solution (Caphosol® B) which when both sachet solutions are combined in equal volumes form a solution supersaturated with respect to both calcium and phosphate ions.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding Caphosol® Artificial Saliva. It primarily focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than presenting a performance study for new acceptance criteria.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets them, specifically in the context of an AI/human comparative effectiveness study, standalone algorithm performance, or training set details, is not present in this document.

    The document states: "No performance testing was conducted." This explicitly indicates that a study demonstrating device performance against acceptance criteria, as typically understood for medical device efficacy or AI performance, was not performed or submitted for this specific 510(k).

    The "Substantial Equivalence" section details how the new device (Caphosol® in sachets) is compared to the predicate device (Caphosol® in vials). The core of the submission is that the new device has not changed its composition or indications for use, and the only difference is the packaging (sachets versus vials).

    Here's the information that is available in the document, largely pertaining to the comparison with the predicate:

    1. Table of Acceptance Criteria and Reported Device Performance: This document does not present a table of new acceptance criteria and corresponding device performance data from a new study. Instead, it compares the characteristics of the subject device (Caphosol® in sachets) to the predicate device (Caphosol® in vials) to establish substantial equivalence.

      FeatureCaphosol® Artificial Saliva in sachets (Subject Device)Caphosol® Artificial Saliva in vials (Predicate Device K030802)
      Product CodeLFDLFD
      Indications for UseSame as PredicateSame as Subject
      Device DescriptionSame as Predicate (electrolyte solution resembling human saliva)Same as Subject
      Composition (% w/w)Dibasic Sodium Phosphate 0.032Dibasic Sodium Phosphate 0.032
      Monobasic Sodium Phosphate 0.009Monobasic Sodium Phosphate 0.009
      Calcium Chloride 0.052Calcium Chloride 0.052
      Sodium Chloride 0.569Sodium Chloride 0.569
      Purified Water 99.338Purified Water 99.338
      Packagesachetsvials
      Shelf Life3 years3 years
      Shelf Life TestingYes (includes liquid and package testing)Yes (includes liquid and package testing)
    2. Sample size used for the test set and the data provenance: Not applicable. The document states "No performance testing was conducted." The "testing" mentioned refers to accelerated shelf-life testing, not clinical or performance efficacy trials with a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No test set requiring expert ground truth was used for this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No test set requiring adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an artificial saliva solution, not an AI-powered diagnostic or assistance tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. No ground truth was established as "No performance testing was conducted."

    8. The sample size for the training set: Not applicable. This is not an AI/machine learning device; therefore, no training set was used.

    9. How the ground truth for the training set was established: Not applicable. (See answer to #8).

    In essence, this 510(k) submission leverages the substantial equivalence pathway, asserting that because the active ingredients, concentration, and intended use of the Caphosol® product remain identical to the previously cleared predicate, and only the packaging format has changed (from vials to sachets), no new performance data is required to establish safety and effectiveness. Shelf-life testing was conducted to ensure the integrity of the new packaging and the stability of the liquid over time, which is considered "similar" to the predicate's testing, with minor adjustments for the flexible package type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162167

    Validate with FDA (Live)

    Date Cleared
    2017-03-31

    (240 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Caphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Caphosol Dispersible is also indicated as an adjunct to standard oral care in the treatment of the mucositis that may be caused by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with amelioration of pain.

    Device Description

    Caphosol is a supersaturated calcium phosphate, electrolyte mouth rinse used as artificial saliva, designed in part to replace ionic and pH balance in the oral cavity, and formulated as a tablet to be dissolved in water prior to use.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "Caphosol Artificial Saliva Tablets." It primarily focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed information about acceptance criteria or a study proving that the device meets such criteria in the context of an AI/ML device.

    However, I can extract the information that is present and indicate what is not available for your specific request on AI/ML device acceptance criteria and study details.

    Here's a breakdown based on the provided document and your request:

    Device: Caphosol Artificial Saliva Tablets.
    Device Type (relevance to AI/ML): This is a pharmaceutical-like product (artificial saliva) and not an AI/ML device. Therefore, the questions related to AI/ML specific performance, such as human reader improvement with AI assistance or standalone algorithm performance, are not applicable.


    1. Table of acceptance criteria and the reported device performance

    The document does not specify formal "acceptance criteria" for performance in the way an AI/ML device would. Instead, it focuses on demonstrating substantial equivalence to a predicate device (Caphosol liquid artificial saliva) based on:

    Acceptance Criterion (Implicit for Substantial Equivalence)Reported Device Performance (Caphosol Artificial Saliva Tablets)
    Intended UseCaphosol Dispersible is indicated for dryness of the mouth and throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent. Also indicated as an adjunct to standard oral care in the treatment of mucositis caused by radiation or high dose chemotherapy. Relief of dryness is associated with amelioration of pain. (Identical to predicate in core function, slight difference in specificity not raising concerns).
    Technology/Operating PrincipleSupersaturated calcium phosphate, electrolyte mouth rinse designed to replace ionic and pH balance in the oral cavity. Active chemical components (Calcium and Phosphate) are identical to the predicate. The final solution prepared after dissolution of a tablet or after mixing 2 ampules (predicate) are identical in quantity of the active ingredients. Substantially equivalent to predicate regarding operating principle and function.
    Formulation (Active Ingredients)Contains: Sodium phosphate dibasic anhydrous, Sodium phosphate monobasic anhydrous, Calcium chloride. (Identical active components to predicate).
    BiocompatibilityAssessed in accordance with ISO - 10993-1. (Implies compliance with standard).
    Safety (Adverse Effects)"A literature search has not shown any clinically significant adverse effects being reported."

    Regarding the study that proves the device meets the acceptance criteria (as requested for an AI/ML device):

    The document explicitly states: "Clinical performance testing was not conducted for the subject device."

    Therefore, the following points which are highly relevant for AI/ML device studies are not applicable or not present in this document:

    1. Sample size used for the test set and the data provenance: Not applicable as no clinical performance testing was done.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI/ML device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.

    Summary for this specific document:

    This document describes a 510(k) submission for a non-AI/ML medical device. The "acceptance criteria" are implicitly met by demonstrating substantial equivalence to a legally marketed predicate device, focusing on intended use, technological characteristics, active ingredients, and biocompatibility. No clinical performance study, in the typical sense of showing efficacy or safety for novel AI/ML devices, was reported to have been conducted for this submission, as it relied on the established safety and efficacy of the predicate and the similar nature of the product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030802

    Validate with FDA (Live)

    Device Name
    CAPHOSOL
    Manufacturer
    Date Cleared
    2003-11-25

    (257 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Caph sol is indicated for dryness of the mouth or throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the conditions are temporary or permanent. Caphosol is also indicated as an adjunct to standard oral care in treating the mucositis that may be caused by and increating or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is assoc iated with an amelioration of pain.

    Capt osol may be used for relief of dryness of the oral mucosa when hyposalivation results from the f )llowing: surgery, radiotherapy near the salivary glands, chemotherapy; infection of dysf inction of the salivary glands; inflammation of the mouth or throat; fever; emotional fact rs such as fear or anxiety; obstruction of the salivary ducts; Sjögren's syndrome; and Bell's Pals /.

    Cap 10sol is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atro ine or other anticholinergic agents that suppress salivary secretion.

    It may be used as part of an oral hygiene program for patients with dry mouth. Caphosol provides intensive hygiene of the oral cavity, and may be used to help relieve bad taste and to relieve offensive nasal discharge and crusting.

    Device Description

    Caphosol is an clectrolyte solution resembling human saliva, designed in part to replace the normal ionic and pH balance in the oral cavity. It is intended as a mouth rinse to moisten, lubricate, and clean the oral cavity including the mucosa of the mouth, tongue and throat. Caphosol maintains moistness in the oral cavity. It relieves diffuse dryness and fissuring of the oral mucosa, as well as painful tongue conditions due to hyposalivation. Patients having this condition are also prome to dental caries and candidal infections.

    Caphosol is a partial substitute for natural saliva. Caphosol facilitates chewing and speaking; loosens tough mucus; prevent mucous membrancs from sticking together; helps remove nasal crust and relieve nasal soreness; improves adherence of dentures, and also relieves bad taste. The purpose of this premarket notification is to expand the claims for Caphosol to allow Caphosol to be marketed as an adjunct to standard oral care for treating the mucositis that may be mansed by radiation or high dose chemotherapy. Relief of dryness of the oral mucosa in these conditions is associated with an amelioration of pain.

    Caphosol is a preparation comprising two separately packaged aqueous solutions, a phosphate solution (Caphosol A) and a calcium solution (Caphosol B) which when both ampoule solutions are combined in equal volumes form a solution supersaturated with respect to both calcium and phosphate ions. The solution after mixing the contents of solutions A and B contains:

    Ingredient:% w/w
    Dibasic Sodium Phosphate0.032
    Monobasic Sodium Phosphate0.009
    Calcium Chloride0.052
    Sodium Chloride0.569
    Purified Waterqs ad
    AI/ML Overview

    The provided text is for a 510(k) premarket notification for a medical device (Caphosol Artificial Saliva) and describes its intended use, comparison to a predicate device, and the FDA's substantial equivalence determination. It does not contain any information about acceptance criteria or a study proving that the device meets such criteria.

    The document is a regulatory submission for a device, which typically focuses on demonstrating substantial equivalence to a previously cleared device, rather than presenting new clinical study data with performance criteria. The FDA letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This indicates that the clearance is based on similarity to an existing device, not on new clinical performance data against specific acceptance criteria.

    Therefore, I cannot provide the requested information from the given text. All the fields regarding acceptance criteria, study details, sample sizes, expert qualifications, and ground truth are not present in the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991938

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-08-10

    (62 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Caphosol is indicated for dryness of the mouth or throat (hyposalivation, xerostomia), regardless of the cause and regardless of whether the condition is temporary or permanent.

    Caphosol is indicated for relief of dryness of the oral mucosa when hyposalivation results from the following: surgery, radiotherapy near the salivary glands, chemotherapy, infection or dysfunction of the salivary glands; inflammation of the mouth or throat; fever; emotional factors such as fear or anxiety, obstruction of the salivary ducts, Sjögren's syndrome; and Bell's Palsy.

    Caphosol is also indicated for dryness of the oral mucosa due to drugs such as antihistamines or atropine or other anticholinergic agents that suppress salivary secretion.

    It may be used as part of an oral hygiene program for patients with dry mouth. Caphosol provides intensive hygiene of the oral cavity, and may be used to help relieve bad taste and to relieve offensive nasal discharge and crusting.

    Device Description

    Caphosol™ Artificial Saliva Consisting of a mixture of two solutions Caphosol™ A -- 15 mL Phosphate Solution and Caphosol™ B - 15 mL Calcium Solution

    AI/ML Overview

    Based on the provided text, there is no information available to answer the specific questions about the acceptance criteria and the study that proves the device meets them.

    The document is a 510(k) summary for "Caphosol™ Artificial Saliva," which primarily focuses on establishing substantial equivalence to a predicate device (GLANDOSANE (SALIVART®)). It outlines the device's name, classification, predicate device, and indications for use.

    The document does not contain:

    • Any specific acceptance criteria for the device's performance.
    • Details of a study conducted to demonstrate that the device meets such criteria.
    • Information regarding sample sizes, data provenance, expert involvement, adjudication methods, or MRMC studies.
    • Results of a standalone algorithm performance study (as this is an artificial saliva, not an AI/algorithm-driven device).
    • Details about the ground truth for training or testing sets.

    Therefore, I cannot provide the requested table or answer the specific questions based on the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1