Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K041700

    Validate with FDA (Live)

    Date Cleared
    2004-07-30

    (38 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cantor & Nissel Hand Painted Prosthetic (hioxifilcon A) Soft Contact Lens for Daily Wear is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens is disinfected using a hydrogen peroxide lens care system only.

    Device Description

    The Cantor & Nissel Hand Painted Prosthetic (hioxifilcon A) Soft Contact Lens for Daily Wear is a hand painted lens with an iris pattern to mask a disfiguring or unsightly eye condition. The lens may be hand painted in an iris or a non-sighted eye or clear in the center for a sighted eye. Lens opacity is obtained by partially of lenses with FDA "listed" color additives in amounts not to exceed the minimum reasonably required to accomplish the intended iris pattern. The colorants are permanent and are not leached from a lens. The lenses are painted by skilled artists to an eye care practitioners specifications.

    AI/ML Overview

    This document is a 510(k) summary of safety and effectiveness for a medical device called "Cantor & Nissel Hand Painted Prosthetic (hioxifilcon A) Soft Contact Lens for Daily Wear." It's a premarket notification to the FDA to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's the breakdown of the requested information, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The provided 510(k) summary does not contain a table of explicit acceptance criteria or reported device performance in terms of quantitative metrics (e.g., success rates, accuracy, sensitivity, specificity). Instead, it relies on demonstrating substantial equivalence to a predicate device.

    The "acceptance criteria" can be inferred from the comparison table, which highlights the similarities in intended use, indications for use, opacity, color additive characteristics, and opaquing agent. The "performance" is implicitly stated as being equivalent to the predicate device, thereby meeting the safety and efficacy standards.

    Here's a table summarizing the comparison, which serves as the basis for demonstrating equivalence rather than a set of performance metrics:

    CharacteristicCANTOR & NISSEL Hand Painted PROSTHETICPredicate Device: Prosthetic (hefilcon A) Soft Contact Lens ~ K992950
    Intended UseCosmetic Management of conditions such as corneal, iris, or lens abnormalities.Cosmetic Management of conditions such as corneal, iris, or lens abnormalities.
    Indications For Use StatementIndicated for daily wear to enhance or alter the apparent eye color, including ocular masking, in sighted or non-sighted eyes requiring a prosthetic contact lens for cosmetic management of conditions such as corneal, iris or lens abnormalities. Also prescribed for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons, or for occlusive therapy for conditions such as diplopia, amblyopia, or extreme photophobia. Disinfected using a hydrogen peroxide lens care system only.Indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, in sighted or non-sighted eyes requiring a prosthetic contact lens for cosmetic management of conditions such as corneal, iris or lens abnormalities. Also prescribed for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons, or for occlusive therapy for conditions such as diplopia, amblyopia, or extreme photophobia. Disinfected with a chemical disinfection system.
    OpacityOne or more FDA listed color additivesPartially or totally White
    Color Additive CharacteristicsThe color additives used are not removed by lens handling and approved cleaning/disinfecting procedures. The optical and performance characteristics are not altered by the lens coloring process.The color additives used are not removed by lens handling and approved cleaning/disinfecting procedures. The optical and performance characteristics are not altered by the lens coloring process.
    Opaquing Agent UsedUse of FDA listed reactive dyes in amounts to accomplish desired hand-painted iris patterns.Titanium Dioxide

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a test set or clinical study with a specific sample size. This 510(k) is based on demonstrating substantial equivalence to an existing device, implying that the safety and effectiveness are established through the predicate's history, not new clinical testing for this specific device model. Therefore, there is no mention of data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no new clinical study or test set requiring ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no new clinical study or test set is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a contact lens, not AI software.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a contact lens, not AI software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no new clinical study or test set is described. The "ground truth" for the device's acceptable performance is its substantial equivalence to a previously cleared device.

    8. The sample size for the training set

    Not applicable. This is a physical medical device (contact lens), not a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device (contact lens), not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992500

    Validate with FDA (Live)

    Date Cleared
    1999-08-18

    (23 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CANTOR & SILVER 5X PROSTHETIC Tinted Soft Contact Lens are indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted eyes that require a prosthetic contact lens for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.

    Device Description

    The CANTOR & SILVER 5X PROSTHETIC Tinted Soft Contact Lens are tinted with FDA "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lens containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed reactive color additive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The CANTOR & SILVER 5X PROSTHETIC Tinted Soft Contact Lens are tinted to the eyecare professional instructions.

    The prosthetic color additive effect is formed by reacting one or more of the reactive color additives listed in this paragraph with (poly hydroxyethyl methacrylate). The reactive color additives that may be used either alone or in combination are: reactive blue 21, reactive blue 19, reactive blue 19, reactive blue 4. reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15, reactive yellow 86, or reactive orange 78. The color additives used are not removed by lens handling or approved cleaning/disinfecting procedures. The prosthetic lens tinting process does not alter the optical and/or performance characteristics of the finished tinted soft contact lens.

    Tint Patterns Available:

    1. Clear lens with Black Pupil. Pupil sizes available in 2.0 mm to 12.5 mm.
    2. Black Occluder Lens. A Central Black area that occludes light. Available to full lens diameter in 0.5 mm increments.
    3. Black Annular with clear pupil. Black Annular diameter range 7.5 mm to full lens diameter in 0.5 mm increments. Clear Pupil diameter range 2.0 mm to 7.5 mm in 0.5 mm increments.
    4. Tinted lens with Black Pupil. Uses the CANTOR & SILVER 5X PROSTHETIC Tinted Soft Contact Lens with black pupil. Pupil sizes available in 2.0 mm to 12.5 mm diameter.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Cantor & Silver 5X Prosthetic Tinted Soft Contact Lens:

    Acceptance Criteria and Device Performance:

    The provided document, a 510(k) summary, primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel acceptance criteria for specific performance metrics. Therefore, explicit "acceptance criteria" and "reported device performance" in the traditional sense of a clinical trial with quantitative endpoints are not present. Instead, substantial equivalence is claimed based on comparable intended use, indications, and material/manufacturing characteristics.

    The table below summarizes the characteristics presented for substantial equivalence. The "Acceptance Criteria" column reflects the implicit criteria for being considered substantially equivalent (i.e., being similar to the predicate device), and the "Reported Device Performance" column reflects the characteristics of the subject device.

    Acceptance Criteria (Implied: Similar to Predicate)Reported Device Performance (Cantor & Silver 5X Prosthetic)
    Intended Use: Cosmetic Management of conditions such as corneal, iris, or lens abnormalities. (Matching predicate device)Intended Use: Cosmetic Management of conditions such as corneal, iris, or lens abnormalities.
    Indication: Comprehensive indications for daily wear, cosmetic management, refractive correction, and occlusive therapy. (Matching predicate device, except for disinfection system)Indication: Daily wear for enhancing/altering eye color, ocular masking, corneal/iris/lens abnormalities, refractive ametropia correction (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons, occlusive therapy (diplopia, amblyopia, extreme photophobia). Disinfected using hydrogen peroxide only. Frequent replacement program.
    FDA "listed" colored additives: Use of specified reactive colored additives. (Matching predicate device, but specific additives differ)FDA "listed" colored additives: Reactive black 5, reactive blue 4, reactive blue 19, reactive 21, reactive blue 163, reactive yellow 15, reactive yellow 86, reactive orange 78, reactive red 11, reactive red 180.
    Uses and restrictions: Color additives not to exceed minimum required for prosthetic effect. (Matching predicate device)Uses and restrictions: Color additives not to exceed minimum reasonably required to accomplish the intended prosthetic effect.
    Color Additive Characteristics: Not removed by handling/cleaning; optical/performance characteristics not altered by coloring. (Matching predicate device)Color Additive Characteristics: Not removed by lens handling and approved cleaning/disinfecting procedures. Optical and performance characteristics are not altered by the lens coloring process.

    Study Details:

    The provided document does not contain information about a specific clinical study with quantitative performance metrics to "prove the device meets acceptance criteria" in the way one might expect for a new device claiming superiority or unique performance. Instead, the submission is based on substantial equivalence to existing legally marketed devices.

    Therefore, many of the requested study-specific details are not applicable or explicitly stated in this 510(k) summary.

    1. Sample size used for the test set and the data provenance: Not applicable. No specific "test set" or clinical study with patient data detailed in this 510(k) summary. The comparison is against characteristics of predicate devices.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No "ground truth" derived from expert consensus on a test set is mentioned.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. No adjudication process for study data is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a contact lens, not an AI diagnostic tool, so an MRMC study is irrelevant.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm or AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of a clinical performance study. The "ground truth" for the substantial equivalence claim would be the established characteristics and regulatory approval of the predicate devices.
    7. The sample size for the training set: Not applicable. No "training set" is mentioned as this is not an AI/machine learning device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K990757

    Validate with FDA (Live)

    Date Cleared
    1999-04-21

    (44 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CANTOR & SILVER SX Tinted (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia. The leases may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.

    The CANTOR & SILVER 5X Tinted (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-disessed eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10.00 diopters. The lenses are disinfected using a hydrogen peroxide lens care system only, and are available in a frequent replacement program.

    Device Description

    The CANTOR & SILVER 5X Tinted (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact lens are fabricated from hioxifilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The color enhanced soft contact lenses are tinted with "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lenses containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed color reactive additive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The finted contact lenses are available in a light, medium or dark shade of the following enhance colors; Blue, Green, Aqua, Yellow, Lavender, Brown, Ultra Violet, and Amber.

    The color additive (timing) effect is formed by reacting one or more of the "listed" additives with HEMA based (poly hydroxyethy) methacrylate) soft contact lens. The color additives used are not removed by lens handling and cleaning/disinfecting procedures as recommended. Except for affecting the amount of light transmitted through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens.

    In the bydrated state, the lens conforms to the curvature of the eye covering the comea and extending slightly beyond the limbus forming a transparent optical surface. The (hioxifilcon A) soft hydrophilic contact lens a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

    AI/ML Overview

    This document is a 510(k) premarket notification for a contact lens, which means it's focused on demonstrating substantial equivalence to existing devices rather than proving performance against specific acceptance criteria through a clinical study with detailed statistical outcomes.

    Therefore, the information regarding acceptance criteria, specific study designs, sample sizes for training/test sets, expert adjudication methods, MRMC studies, or standalone algorithm performance, as typically sought for AI/ML device evaluations, is not present in this type of submission.

    Instead, the submission focuses on comparing the new device's characteristics to a predicate device (Durasoft 2 Colors) to demonstrate that it does not raise new questions of safety or effectiveness.

    Here's a breakdown of what is available, and what is not:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Implicit): The implicit acceptance criterion is "substantial equivalence" to the predicate device, meaning the new device's characteristics (materials, production method, lens function, color additives, and use restrictions) are similar enough to an already approved device that it's considered equally safe and effective.
    • Reported Device Performance: Instead of performance metrics, the document provides the physical properties and characteristics of the CANTOR & SILVER 5X Tinted contact lens, and then compares them to the predicate device to show similarity.

    Here's the table from the document, re-formatted to highlight the comparison for "performance" in this context:

    CharacteristicCANTOR & SILVER 5X TINTED (Reported)DURASOFT 2 COLORS (Predicate)Substantial Equivalence?
    PRODUCTION METHODLathe-CutLathe-CutYes
    LENS FUNCTIONRefractive medium that focuses light rays from near and distant objects on the retina, while compensating for refractive error, including (astigmatism)Refractive medium that focuses light rays from near and distant objects on the retina, while compensating for refractive error, including (astigmatism)Yes
    FDA "listed" colored additivesThe reactive colored additives consist of reactive black 5, reactive blue 4, reactive blue 19, reactive 21, reactive blue 163, reactive yellow 15, reactive yellow 86, reactive orange 78, reactive red 11 and reactive red 180The colored pigments consist of iron oxides, chromium oxide greens, titanium dioxide, [phthalo-cyaninato (2-)] copper, carbazole violet and phthalocyanine green.Note: Different additives, but the document claims equivalence in usage and restrictions.
    Uses and restrictionsThe color additives listed above may be used to color contact lenses in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect.The color additives listed above may be used to color contact lenses in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect.Yes
    Color Additive CharacteristicsThe color additives used are not removed by lens handling and cleansing/disinfecting procedures. The optical and performance characteristics are not altered by the lens coloring process.The color additives used are not removed by lens handling and cleansing/disinfecting procedures. The optical and performance characteristics are not altered by the lens coloring process.Yes
    Colors OfferedBlue, Green, Aqua, Yellow, Lavender, Brown, Ultra Violet and AmberSky blue, Jade green, Aquamarine and Violet blueNote: Different range of colors, but this is a styling difference, not a functional one impacting safety/effectiveness.
    Other Physical PropertiesRefractive Index: 1.515 (dry) 1.404 (hydrated), Light Transmission: >70%, Water Content: 58% ± 2%, Specific Gravity: 1.308 (dry) 1.136 (hydrated), Oxygen Permeability: $20 X 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @ 35°C)(Not explicitly listed for predicate, but assumed to be within acceptable ranges for soft contact lenses for equivalence claim)Yes

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable: This submission does not describe a clinical performance study with a test set in the way a modern AI/ML device would. The "study" here is essentially a technical comparison and declaration of equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable: There is no "ground truth" establishment in the context of expert consensus for a test set described. The ground truth for regulatory approval is the established safety and effectiveness of the predicate device and the new device's proven similarity to it through material science, manufacturing, and toxicology data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable: No test set adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No: This is a contact lens, not an AI-powered diagnostic device. No MRMC study was conducted or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable: This is a physical medical device (contact lens), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Implicit Ground Truth: The "ground truth" for this submission is the established safety and effectiveness profile of the predicate device (Durasoft 2 Colors) and general scientific understanding of contact lens materials and manufacturing. The manufacturer is demonstrating that their new device is functionally and materially similar enough to the predicate that it can rely on its established safety profile. This includes:
      • Pre-clinical toxicology and manufacturing/chemistry data (K952620, referenced for BENZ-G 5X).
      • Material properties (hioxifilcon A) being well-understood.
      • Manufacturing processes (lathe-cut) being standard and controlled.

    8. The sample size for the training set:

    • Not Applicable: There is no machine learning "training set" for a contact lens.

    9. How the ground truth for the training set was established:

    • Not Applicable: As above, no machine learning training set or ground truth establishment in that context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K981779

    Validate with FDA (Live)

    Date Cleared
    1998-06-24

    (35 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CANTOR & SILVER 5X (hioxifilcon A) Spherical Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity.

    The CANTOR & SILVER 5X (hioxifilcon A) Toric Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism up to 10 Diopters.

    Device Description

    The CANTOR & SILVER 5X (hioxifileon A) Soft (Spherical & Toric) Daily Wear Contact lenses are fabricated from hioxifilcon A, which in the dry (unhydrated) state may be machined. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

    In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (hioxifileon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

    The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 58% water by weight. The physical properties of the lens are:

    Refractive Index: 1.515 (dry) 1.404 (hydrated)
    Light Transmission (clear): greater than 95% T
    Light Transmission (tinted): greater than 95% T
    Color Pigment Name: Phthalocyanato (2) - (copper)
    Water Content: 58% ± 2%
    Specific Gravity: 1.308 (dry) 1.136 (hydrated)
    Oxygen Permeability: 20 X 10 to the power of -11 (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method)

    AI/ML Overview

    The provided text describes the CANTOR & SILVER 5X (hioxifilcon A) Soft (Spherical & Toric) Daily Wear Contact Lens. This submission focuses on establishing substantial equivalence to previously marketed predicate devices rather than proving the device meets specific acceptance criteria through a clinical study with detailed performance metrics.

    Here's an analysis based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a performance study with numerical targets. Instead, it demonstrates substantial equivalence to predicate devices by comparing manufacturing methods, lens function, and material properties. The "performance" here is an assertion of similarity.

    CharacteristicPredicate Device (Acceptance Criteria - Implicit)CANTOR & SILVER 5X (Reported Performance)
    Production MethodLathe-CutLathe-Cut
    Lens FunctionRefractive medium focusing light, correcting refractive error (including astigmatism)Refractive medium focusing light, correcting refractive error (including astigmatism)
    MaterialHydrophilic PolymerHydrophilic Polymer
    Water Content58%58%
    Polymer Content42%42%
    Polymerhioxifilcon Ahioxifilcon A
    Dk Value2020
    Refractive Index (hydrated)1.4041.404
    Specific Gravity (hydrated)1.1361.136
    Light TransmissionGreater than 95% TGreater than 95% T

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes a comparison against predicate devices to establish substantial equivalence, not a performance study that would typically involve a "test set" of patients or data. Therefore, information about a test set sample size or its provenance (country of origin, retrospective/prospective) is not applicable in this context. The study is a comparison of product specifications and characteristics.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since this is a submission for substantial equivalence based on material properties and manufacturing, there was no "test set" requiring ground truth established by experts. The claims rely on the established properties of the materials and the manufacturing process, which are compared to the predicate devices.

    4. Adjudication Method for the Test Set

    As there was no test set in the sense of clinical performance data requiring expert review, an adjudication method is not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size

    An MRMC study is relevant for evaluating the performance of AI systems or interventions where human readers are involved in assessing cases. This submission is for a contact lens, which is a medical device, not an AI or imaging diagnostic tool. Therefore, an MRMC comparative effectiveness study was not done, and there is no effect size related to human readers improving with AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study

    This concept is applicable to AI and diagnostic algorithms. The CANTOR & SILVER 5X contact lens is a physical medical device. Therefore, a standalone (algorithm only) performance study was not done.

    7. Type of Ground Truth Used

    The "ground truth" in this context is the established specifications and properties of the predicate devices and the materials used. The substantial equivalence argument rests on matching these characteristics.

    8. Sample Size for the Training Set

    There is no "training set" as this is not an AI or machine learning device. The product development would involve engineering and material science, not data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1