Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112036

    Validate with FDA (Live)

    Date Cleared
    2011-11-08

    (113 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    When used as a vertebral body replacement, the X-spine Calix PC System is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to tumor or trauma (i.e., fracture). The Calix device, when used as a vertebral body replacement, can be packed with either allograft or autograft.

    When used as an intervertebral body fusion device, the X-spine Calix PC System is intended for spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implanted via an open, anterior approach and packed with autogenous bone. Patients should receive 6 weeks non-operative treatment prior to treatment with the Calix intervertebral body fusion device.

    For all indications, this device is intended to be used with supplemental spinal fixation systems that have been cleared for use in the cervical, thoracic or lumbar spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems).

    Device Description

    The X-spine Calix PC System is a generally box-shaped device with various holes located throughout its geometry and teeth on the superior and inferior surfaces. The device is made from polyetheretherketone (Invibio PEEK-Optima LT1). The devices contain radiographic markers made from tantalum per ASTM F560. The technological difference between this system and the current Calix Spinal Implant System is the addition of a commercially pure titanium (CP Ti) coating per ASTM F1580 applied to the superior and inferior surfaces of each device.

    AI/ML Overview

    The provided text describes a medical device, the Calix™ PC Spinal Implant System, and non-clinical testing performed to establish its substantial equivalence to predicate devices. However, the document does not contain information regarding a study involving acceptance criteria, device performance, or human-in-the-loop performance using AI.

    Therefore, I cannot fulfill the request to provide information based on the provided input for acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or training set details.

    The document focuses on non-clinical biomechanical and coating characterization testing to demonstrate substantial equivalence to previously approved predicate devices, which is a common pathway for 510(k) clearances. It is not an efficacy or diagnostic performance study in the context of AI or human interpretation.

    Summary of what the document does provide regarding testing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not present acceptance criteria in a quantitative table with specific performance metrics and their corresponding values. It lists the types of tests conducted:

    Test TypeDescription
    Biomechanical Testing (Full Device Constructs)
    ASTM F2077 – Test Methods for Intervertebral Body Fusion Devices- Static and dynamic axial compression
    - Static torsion
    Expulsion testing (FDA Guidance)- As suggested by FDA Guidance for Intervertebral Body Fusion Devices
    Coating Characterization (Test Coupons)
    ASTM F1044 - Standard Test Method for Shear Testing- Static and dynamic shear
    ASTM F1147 - Standard Test Method for Tension Testing- Static tension
    Abrasion testing (FDA Guidance)- As suggested by FDA Guidance for Modified Metallic Surfaces Apposing Bone or Bone Cement

    The "reported device performance" is not quantified in the document beyond stating that "biomechanical testing results indicate that the Calix PC Spinal Implant System is substantially equivalent to predicate device performance and is as effective, and performs at least as safely and effectively as the cited predicate devices." Specific numerical results or pass/fail criteria are not detailed.

    Further information regarding your specific questions, as it relates to the provided text:

    1. Sample sizes used for the test set and the data provenance: Not applicable. The tests are non-clinical (biomechanical and material characterization) performed on device constructs and test coupons, not on human data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for non-clinical engineering tests is based on established ASTM standards and FDA guidance, not expert human interpretation.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a medical device for spinal implant, not an AI-powered diagnostic or interpretive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the non-clinical tests, the "ground truth" would be defined by the parameters and acceptance criteria set forth in the ASTM standards and FDA guidance documents cited for each test (e.g., specific load endurance, displacement limits, or bond strength thresholds).
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1